Scope
GB/T 9722-2023 is the English-translated version of 化学试剂 气相色谱法通则.
GB/T 9722-2023 is the Chinese national set of general rules for analysing chemical reagents by gas chromatography.
It fixes the materials, the instrument and the arithmetic: standard samples whose main component is not lower than 99.9%, carrier and combustion gases of not lower than 99.99% purity, a chromatograph meeting JJG 700 with capillary or packed columns chosen by stationary phase, bore, length and film thickness, and the measurement of peak height, width at half height and area, with the vertical, valley-to-valley and tangent ways of splitting overlapping peaks.
Quantification is by correction factor, normalization or external standard, the last with a curve of no fewer than five standard solutions. Clause 1 applies it to main components and impurities in organic reagents containing volatile components.
Issued on 6 August 2023, it replaces GB/T 9722-2006 and has applied since 1 March 2024.
Found what you need?
Document preview — GB/T 9722-2023
National Standard of the People's Republic of China
- ICS
- 71.040.30
- Classification
- G 60
- Replacing
- GB/T 9722-2006
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope5
- 2 Normative References5
- 3 Terms and Definitions5
- 4 Method and Principle6
- 5 Reagents and Materials6
- 6 Instruments6
- 7 Test Conditions7
- 8 Operating Method8
- 9 Qualitative Analysis11
- 10 Quantitative Analysis12
- 11 Method Error17
- 12 Data Quality Assurance18
- 13 Environmental Requirements, Safety Precautions and Waste Disposal19
- Appendix A (informative) Chromatographic Columns21
- Appendix B (normative) Illustrations and Calculation Formulas of Related Terms24
- Appendix C (informative) Acceptable Ranges of Precision and Trueness of the Method26
- Bibliography27
1 Scope
This document specifies the instrument requirements and analytical methods for the gas chromatography for chemical reagent.
This document is applicable to the determination of main components and impurities of organic chemical reagents containing volatile components.
2 Normative References
The contents of the following documents constitute indispensable clauses of this document through the normative references in the text. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB/T 4946 Terms of Gas Chromatography
GB 4962 Technical Safety Regulation for Gaseous Hydrogen Use
GB/T 8170 Rules of Rounding off for Numerical Values & Expression and Judgement of Limiting Values
JJG 700 Gas Chromatographs
TSG 23-2021 Regulation on Safety Technology for Gas Cylinder
3 Terms and Definitions
What is defined in GB/T 4946, and the following terms and definitions are applicable to this document.
3.1 Asymmetric factor
A parameter that describes the degree of asymmetry of a chromatographic peak.
3.2 Height of an effective plate
The length of unit effective plate.
4 Method and Principle
After the sample and its components being determined are vaporized, they enter the chromatographic column at the same time with the carrier gas. The difference in physical and chemical properties, such as. adsorption or dissolution, desorption or analysis of each component being determined between the gas-solid or gas-liquid phases is used to form a difference in the migration speed of the components in the column for separation.
5 Reagents and Materials
5.1 Standard Sample
The mass fraction of the main body content of the standard sample shall not be lower than 99.9%. When high-purity standard samples cannot be obtained for special substances, standard samples with clarified main body content shall be used.
5.2 Reference Substance
The reference substance shall be traceable to International System of Units (SI) or certified reference substance.
5.3 Carrier Gas
The purity is not lower than 99.99%. Before use, dehydration devices (silica gel and molecular sieve), activated carbon and deoxidizer, etc. shall be used for purification.
5.4 Combustion Gas
The purity is not lower than 99.99%. Before use, dehydration devices (silica gel and molecular sieve), activated carbon and deoxidizer, etc. shall be used for purification.
5.5 Air
It does not contain dust, hydrocarbons, moisture and corrosive substances that may affect the normal operation of the instrument. Before use, dehydration devices (silica gel and molecular sieve) and activated carbon, etc. shall be used for purification.
6 Instruments
6.1 Composition of Gas Chromatograph
The gas chromatograph consists of gas path system, sample injection system, separation system, detection system, temperature control system and data processing system, as shown in Figure 1.
6.2 Performance of Gas Chromatograph
The performance requirements for the gas chromatograph shall comply with the requirements of JJG 700.
6.3 Chromatographic Columns
The chromatographic columns are divided into capillary columns and packed columns. The important parameters of chromatographic columns are stationary phase, inner diameter, column length and liquid film thickness. The users shall select appropriate chromatographic columns based on sample properties. See Appendix A for the parameters of capillary chromatographic columns and the preparation and handling of packed columns.
7 Test Conditions
In accordance with the characteristics and specifications requirements of the product and the components to be determined, select the optimum conditions as specified below.
8 Operating Method
8.1 Peak Height Measurement
Draw a vertical line from the top of the peak to the bottom of the peak. The distance from the point where the vertical line intersects with the upper edge of the chromatographic peak baseline to the top is the peak height (h). Or calculate the difference between the signal value at the peak apex and the baseline signal value at the same retention time as the peak apex (see Figure 2).
8.2 Measurement of Peak Width at Half Height
Draw a straight line parallel to the peak bottom from the midpoint of the peak height and intersect with both sides of the peak. The distance between two points on the same side of the peak height line is the peak width at half height (Wh/2) (see Figure 2).
8.3 Calculation of Peak Area
Calculate the integrated value of the difference between the signal value and the baseline signal value from the start of the peak to the end of the peak; or the product of the peak height and the peak width at half height.
8.4 Overlapping Peak Area Segmentation Method
8.4.1 Vertical method
As shown in Figure 3, when the areas of two peaks are similar, draw a vertical line from the peak valley toward the baseline to divide the peak on the baseline into two parts, so as to obtain the respective peak areas.
8.4.2 Peak-valley - peak-valley method
As shown in Figure 4, in accordance with the adjacent peak valley and peak valley linear distribution, obtain the corresponding peak area. This method is applicable to background overlapping peaks.
8.4.3 Tangent method
As shown in Figure 4, when a small peak overlaps the tail of a large peak, draw a tangent line from the peak valley to the foot of the large peak, and use the part above the tangent line as the area of the small peak. The exponential function curve method can also be used instead of the tangent method for peak segmentation.
9 Qualitative Analysis
9.1 Comparative Qualitative Analysis of Known Substance
9.1.1 Adding known substance to increase peak height
Under the same chromatographic conditions, add a known substance to the sample to be tested, and compare the changes in the chromatographic peak height of the sample to be tested before and after the addition. If the peak of the component to be determined in the sample to be tested increases, it can be considered that the component and the known substance are the same substance.
9.1.2 Qualitative analysis with double-column or multi-column
On two or more chromatographic columns of different polarities, respectively analyze the known substance and the sample to be tested under the same chromatographic conditions. If the retention values of the known substance and unknown component in the sample to be tested are the same on different chromatographic columns, the component and the known substance can be considered to be the same substance.
9.2 Qualitative Analysis of Relative Retention Value
9.2.1 Under the same chromatographic conditions, respectively perform chromatographic analysis on the standard sample and the sample to be tested and calculate the relative retention Baseline
9.2.2 If the standard sample cannot be obtained, the relative retention value of the known substance can be obtained by searching the literature, then, under the test conditions (column temperature, stationary phase and reference substance) provided by the literature value, determine the relative retention value of the sample to be tested; if it is consistent with the literature value, it can be identified as the same substance.
10 Quantitative Analysis
10.1 Correction Factor
10.1.1 General requirements
This document adopts a mass correction factor for component i relative to the main body.
For individual components listed in technical indicators, the mass correction factor will be used regardless of the mass fraction. Among the components to be determined, for homologues with relatively close carbon numbers or substances with small thermal conductivity differences, a correction factor may be added depending on the specific situation.
10.2 Normalization Method
When adopting the normalization method for quantitative analysis, the following requirements shall be satisfied.
10.4 External Standard Method
10.4.1 General requirements
When adopting the external standard method for quantitative analysis, the following requirements shall be satisfied.
10.4.2 Single-point external standard method
Prepare a standard solution with a mass fraction close to that of the component to be determined.
In accordance with the same determination conditions, respectively determine the standard solution and the sample to be tested. Through the peak area ratio of the component to be determined and the external standard component, calculate the mass fraction of the component to be determined.
10.4.3 Standard curve method
Prepare no less than 5 standard solutions with proportional mass fractions and quantitatively inject the sample. Take the mass fraction of the component to be determined as the x-coordinate and the measured peak area as the y-coordinate to draw a standard working curve. Under the same determination conditions, inject the same amount of the sample to be tested for chromatographic analysis, determine the peak area of the sample, and in accordance with the standard working curve, calculate the mass fraction of the component to be determined (see Figure 6).
10.6 Result Expression
10.6.1 The arithmetic mean of two repeated determination results shall be used as the test result, which shall be rounded in accordance with the stipulations of GB/T 8170.
10.6.2 When the mass fraction of the component to be tested is not less than 1%, the test result shall be accurate to 0.01% (mass fraction).
10.6.3 When the mass fraction of the component to be tested is not less than 0.01% and less than 1%, the test result shall be accurate to 0.001% (mass fraction). […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 27 pages — is available in the English PDF.
Referenced standards
Normative references
Cited by
- GB/T 47119-2026Testing method for the thermal stability of organic heat transfer fluids - Gas chromatography with internal standard method
- GB 16663-2025Alcohol-based liquid fuel
- GB 45944-2025Perfluorohexanone extinguishing agent
- GB/T 14491-2025Propylene oxide for industrial use
- GB/T 20436-2025Dimethylsiloxane cyclics mixture
- GB/T 23953-2025Dimethyldichlorosilane for industrial use
Similar standards
Editions of GB/T 9722
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 9722-2023 | Chemical reagent—General rules for the gas chromatography | current edition | Current |
| GB/T 9722-2006 | Chemical reagent—General rules for the gas chromatography | previous edition | Superseded |
This page sells the current edition, GB/T 9722-2023. Earlier editions are listed for reference only.
How to Download GB/T 9722-2023
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4962-2008 — Technical safety regulation for gaseous hydrogen use
GB/T 4946-2008 — Terms of gas chromatography
GB/T 8170-2008 — Rules of rounding off for numerical values & expression and judgement of limiting values
Secure payment via Stripe
Payments accepted
GB/T 9722-2023
$300.00