GB/T 6920-1986Water quality--Determination of pH value--Glass electrode method (English PDF)

水质 pH值的测定 玻璃电极法

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for GB/T 6920-1986 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

October 10, 1986

Implementation date

March 1, 1987

Scope

GB/T 6920-1986 is the English-translated version of 水质 pH值的测定 玻璃电极法.

GB/T 6920-1986 is the Chinese standard "Water quality. Determination of pH value. Glass electrode method". Its scope clause reads: 1.1 This method is applicable to the determination of the pH value of drinking water, surface water and industrial wastewater.

1.2 The color, turbidity, colloidal substance, oxidizing agent, reducing agent and high salinity of water do not interfere with the determination; however, in the strong acidic solution with pH < 1, there is so-called "acid error", which can be determined according to the acidity; in the alkaline solution with pH >10, due to the existence of a large amount of sodium ions, the error is generated; therefore, the reading is relatively low, which is commonly referred to as "sodium error".

For the methods of eliminating the "sodium error", in addition to use the specially-made electrode of "low sodium error", it can also adopt the standard buffer solution which is similar to the pH value of the solution under test, in order to calibrate the instrument.

The temperature can affect the electric potential of the electrodes and the ionization. Its clauses include scope of application; definitions; principle; reagents; instruments; sample preservation; steps; degree of precision (see table 3); annotations; test report. It was issued on OCTOBER 10, 1986 and took effect on MARCH 1, 1987.

Found what you need?

Document preview — GB/T 6920-1986

National Standard of the People's Republic of China

ICS
13.060
Classification
Z 50

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • 1 Scope of application3
  • 2 Definitions3
  • 3 Principle4
  • 4 Reagents4
  • 5 Instruments7
  • 6 Sample preservation7
  • 7 Steps7
  • 8 Degree of precision (see Table 3)8
  • 9 Annotations8
  • 10 Test Report9

1 Scope of application

1.1 This method is applicable to the determination of the pH value of drinking water, surface water and industrial wastewater.

1.2 The color, turbidity, colloidal substance, oxidizing agent, reducing agent and high salinity of water do not interfere with the determination; however, in the strong acidic solution with pH < 1, there is so-called "acid error", which can be determined according to the acidity; in the alkaline solution with pH >10, due to the existence of a large amount of sodium ions, the error is generated; therefore, the reading is relatively low, which is commonly referred to as "sodium error". For the methods of eliminating the "sodium error", in addition to use the specially-made electrode of "low sodium error", it can also adopt the standard buffer solution which is similar to the pH value of the solution under test, in order to calibrate the instrument.

The temperature can affect the electric potential of the electrodes and the ionization balance of the water. It shall be noted that the compensation device of the adjusting instrument should be consistent with the temperature of the solution, and the temperature error of the standard buffer solution between the sample to be measured and the calibrating instrument should be within ± 1°C.

The pH is defined from the operation. For solution X, it should measure the galvanic cell's. reference electrode|KCl concentrated solution|| solution X |H2| Pt electromotive force EX. Replace the solution X, whose pH(X) is unknown, into the standard 1 This definition is cited from p.151 of GB 3100-3102-82 Quantities and units. pH solution S, and measure the battery's electromotive force ES, then, pH(X)=pH (S) + (Es - Ex) F / (RTln10)

Therefore, the defined pH is the dimensionless quantity. pH has no theoretical significance. the definition is the practical definition. However, within the limited range of the dilute aqueous solution whose amount-of-substance concentration is less than 0.1mol·dm3, when it is neither strong acidity nor the strong alkalinity (2 < pH < 12), then it has the followings according to the definition. pH=-log10[c(H+)y / (mol·dm3)] ± 0.02 In the formula, c(H+) represents the amount-of-substance concentration of the hydrogen ion (H+). y represents the activity coefficient of the typical 1-1-valence electrolyte in the solution.

2 Definitions 1

3 Principle

The pH value is obtained by measuring the electromotive force of the battery. This battery is usually composed of that the saturated calomel electrode is acted as the reference electrode AND the glass electrode is acted as the indicator electrode. At the temperature of 25°C, if the solution changes 1 pH unit, the change of the potential difference will be […]

4 Reagents

4.1 Preparation method of the standard buffer solution (abbreviated as standard solution)

4.1.2 When measuring the pH, according to the three possible conditions -- the water sample can present to acidity, neutrality and alkalinity -- the following three standard 2 The unit of electrical conductance is Siemens which is denoted by the symbol of "S". 1S = 1 ohm-1. solutions are often prepared.

4.4 The pH value of the standard solution can vary slightly with the changes of the temperature. The pH (S) values of certain common standard solutions are shown in Table 2.

4.1.1 Mass of the reagent and distilled water
4.1.2.1 pH standard solution A (pH 4.008, 25°C)

Weigh and take 10.12g of potassium acid phthalate (KHC8H4O4) which has been dried for 2-3 hours at 110~130°C. Dissolve it in the water AND dilute it to 1L in volumetric flask.

4.3 Preservation of the standard solution

4.3.1 Standard solution shall be sealed in the polyethylene bottle or hard glass bottle for preservation.

4.3.2 Under the condition of room temperature, the standard solution is generally stored for 1~2 months. When finding the phenomenon of turbidity, moldiness and sedimentation, it should not be used anymore.

5 Instruments

5.1 Acidity meter or ion concentration meter. The instrument for the routine test should accurate to 0.1pH unit. The pH range is from 0 to 14.If there is special requirement, the instrument with higher precision shall be used.

6 Sample preservation

It is best to conduct the site measurement. Otherwise, the samples should be kept at the 0~4°C after the sampling, and they should be measured within 6 hours after the sampling.

7 Steps

7.1 Instrument calibration. The operating procedure shall be carried out according to the operation instruction of the instrument. Firstly, adjust the water sample and the standard solution to the same temperature; record the measurement temperature; and adjust the temperature-compensation knob of the instrument to this temperature.

7.2 Sample determination

When determine the sample, it should carefully rinse the electrode with distilled water, and then rinse it with the water sample, then immerse the electrode in the sample. Shake it carefully or stir it to make it uniform. Keep it in still standing. Record the pH value when the reading is stable.

8 Degree of precision (see Table 3)

Repeatability 5 Reproducibility 6

9 Annotations

9.1 The glass electrode should be immersed in the distilled water for more than 24 hours before use.

9.2 When measuring the pH, the bulb of the glass electrode shall be completely immersed in the solution, and it should allow it to be slightly higher than the ceramic core end of the calomel electrode, in order not to bump the cup when stirring.

9.4 The liquid level of the saturated potassium chloride solution in the calomel electrode must be higher than the mercury body. It should have little potassium chloride crystal at the room temperature, in order to ensure the saturation of the potassium chloride solution, however, it is important to note that the potassium chloride crystal should not be too much, in order to prevent from blocking the access of the solution to be tested.

9.5 When measuring the pH, in order to reduce the dissolution or volatilization of the carbon dioxide from the air and the water sample, it should not open the bottle of water sample in advance before measuring the water sample.

10 Test Report

The test report shall include the followings. […]

59.16 mV

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Cited by

Similar standards

How to Download GB/T 6920-1986

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 6920-1986

$35.00