GB/T 223.71-1997Methods for chemical analysis of iron, steel and alloy--The gravimetric method after combustion in the pipe furnace for the determination of carbon content (English PDF)
钢铁及合金化学分析方法 管式炉内燃烧后重量法测定碳含量
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 223.71-1997 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
March 17, 1997
Implementation date
September 1, 1997
Scope
GB/T 223.71-1997 is the English-translated version of 钢铁及合金化学分析方法 管式炉内燃烧后重量法测定碳含量.
GB/T 223.71-1997 is the Chinese standard "Methods for chemical analysis of iron, steel and alloy--The gravimetric method after combustion in the pipe furnace for the determination of carbon content".
Its scope clause reads: This Standard specifies the gravimetric method after combustion in the pipe furnace for the determination of sulfur content. This Standard applies to the determination of carbon content of 0.10% (m/m) ~ 5.00% (m/m) in iron, steel, high temperature alloy and precision alloy.
Its clauses include application scope; method summary; reagents and materials; analytical result and its expression; precision.
Found what you need?
Document preview — GB/T 223.71-1997
National Standard of the People's Republic of China
- ICS
- 77.080
- Classification
- H 11
- Replacing
- GB 223.71-1991
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Application Scope4
- 2 Normative References4
- 3 Method Summary4
- 4 Reagents and Materials4
- 5 Apparatus6
- 6 Sampling and Sample Preparation8
- 7 Analytical Procedures8
- 8 Analytical Result and Its Expression11
- 9 Precision11
- 10 Test Report12
- after Combustion in the Pipe Furnace13
1 Application Scope
This Standard specifies the gravimetric method after combustion in the pipe furnace for the determination of sulfur content.
This Standard applies to the determination of carbon content of 0.10% (m/m) ~ 5.00% (m/m) in iron, steel, high temperature alloy and precision alloy.
2 Normative References
The provisions in following documents become the provisions of this Standard through reference in this Standard. All editions shown are effective when this Standard is issued. All standards will be revised, however, parties who reach an agreement based on this Standard are encouraged to study if the latest versions of the following standards are applicable.
GB 222-84, Method of Sampling Steel for Determination of Chemical Composition and Permissible Variations for Product Analysis
GB 6379-86, Precision of Test Methods - Determination of Repeatability and
Reproducibility for a Standard Test Method by Interlaboratory Tests
3 Method Summary
Test sample and fluxing agent burn in a pipe furnace at high temperature (1 250°C ~
1 350°C) by pumping in oxygen; carbon is fully oxidized into carbon dioxide. After removing sulphur dioxide, use the soda asbestos absorption flask of a known mass to absorb the carbon dioxide in the mixed gas. Weigh the increment of the absorption flask and calculated the carbon content of test sample based on the increment.
4 Reagents and Materials
4.1
Oxygen of purity not lower than 99.5% (m/m).
7.1.2
Pump in oxygen, check whether the pipeline and valve of the whole equipment
leaks, place the porcelain boat loaded with test sample in the porcelain tube, use the silicone rubber stopper to seal the porcelain tube, observe the chromic acid saturated sulphuric acid solution in the quadruple ball and if there is any backflow, it indicates the equipment does not leak.
7.1.3
Connect the water absorption flask and carbon dioxide absorption flask, unwind the valves in succession, pump in oxygen (at the flow rate of 500 mL/min ~ 800 mL/min), conduct gas washing for 3 min ~ 5 min, and close the valves of oxygen and two absorption flasks. Use the long hook to pull back the porcelain boat. After weighing, repeat the operation of pumping in oxygen until the difference between two weighed values is less than 0.3 mg; and analysis may be started after putting the carbon dioxide absorption flask back.
NOTE 1 Pump in oxygen for the newly-loaded carbon dioxide absorption flasks for 20 min ~ 30 min, analyze the test samples of high secondary carbon content and weigh. Pump in oxygen once again for 3 min ~ 5 min, weigh and repeat 1 ~ 2 times, and the difference between two weighed values shall be less than 0.3 mg (when the weight of loss is much more than the value, it indicates that there is water loss, so the value is unusable).
2 When the height of soda asbestos in the carbon dioxide absorption flask becoming white is close to one third of the height of all soda asbestos, the soda asbestos must be replaced. During the mounting of the carbon dioxide absorption flask, operate quickly; or else, the soda asbestos will absorb the carbon dioxide and water in the air soon. The mounting of the carbon dioxide absorption flask can affect the analytical results. If the glass wool is spread uniformly on the bottom, carbon dioxide will rise uniformly during absorption and the absorption is satisfactory, and otherwise, the absorption is unsatisfactory.
7.2 Blank test
Spread fluxing agent (see Table 1) uniformly in the porcelain boat, carry out determination as specified in 7.3.3 and 7.3.4 and determine carefully the increment of the carbon dioxide absorption flask (m1). The blank value is normally not greater than
0.3 mg; and if the blank value is abnormal, find the cause before continuing operation and carry out blank test once again.
7.3.5
Check the porcelain boat and if the slag is uneven and the section of slag has
pores, it indicates the combustion is not complete, so test samples shall be weighed and determined once again.
NOTE 1 The time for preheating high-alloy test samples shall be increased; and oxygen may be pumped in by observing the rise of the liquid level of chromic acid saturated sulphuric acid in the quadruple ball.
2 During the continuous analysis of test samples, the content may be calculated as the difference between the weight measured this time minus the weight measured last time. If the analysis is interrupted for more than 20 min, oxygen shall be pumped in once again for gas washing for 3 min ~ 5 min and the weight of the carbon dioxide absorption flask shall be measured once again.
8 Analytical Result and Its Expression
The content content expressed in mass percentage is calculated in accordance with
Formula (1).
where, m1 - the mass of carbon dioxide measured in blank test, g;
m2 - the mass of carbon dioxide measured in the analysis of test sample, g;
m - the mass of test sample, g;
0.272 9 - the coefficient for conversion of carbon dioxide into carbon.
9 Precision
The precision of this Standard is the 10 carbon levels selected from 8 laboratories in
1988; and each sulphur level is determined by each laboratory through three joints tests as specified in GB 6379. See Annex A for the test samples. See Annex B for the raw data (measured values) given by all laboratories.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Cited by
- GB/T 32219-2026Cylinder mills - Cast grinding cylpebs
- GB/T 8602-2026Rough-rolled tyres for railway use
- GB/T 20932-2025Pig-irons - Definition and classification
- GB/T 42669-2023Corrosion resistant bulb flats for cargo oil tanks
- GB/T 17445-2022Cast grinding balls
- GB/T 8601-2021Forged and rolled solid wheels for railway
Similar standards
Editions of GB/T 223.71
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 223.71-1997 | Methods for chemical analysis of iron, steel and alloy--The gravimetric method after combustion in the pipe furnace for the determination of carbon content | current edition | Current |
| GB 223.71-1991 | Methods for chemical analysis of iron, steel and alloy--The gravimetric method after combustion in the pipe furnace for the determination of carbon content | previous edition | Superseded |
This page sells the current edition, GB/T 223.71-1997. Earlier editions are listed for reference only.
How to Download GB/T 223.71-1997
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 223.11-2008 — Iron, steel and alloy -- Determination of chromium content -- Visual titration or potentiometric titration method
GB/T 223.11-2025 — Iron, steel and alloy - Determination of chromium content - Titrimetric method and spectrophotometric method
GB/T 223.12-1991 — Methods for chemical analysis of iron, steel and alloy. The sodium carbonate separation-diphenyl carbazide photometric method for the determination of chromium content
Secure payment via Stripe
Payments accepted
GB/T 223.71-1997
$65.00