Scope

GB/T 9721-2006 is the English-translated version of 化学试剂 分子吸收分光光度法通则(紫外和可见光部分).

GB/T 9721-2006 sets the general rules Chinese laboratories follow when a chemical reagent is examined by molecular absorption spectrophotometry in the ultraviolet and visible range.

It defines the terms used - absorption peak, absorption spectrum, absorption pool thickness, amount-of-substance concentration and molar absorptivity - explains the principle the method rests on, and states which water, organic solvents, buffer solutions and standard sample solutions have to be prepared, with the water grades taken from GB/T 6682-1992.

It calls for a single-wavelength single-beam or dual-beam spectrophotometer meeting JJG 178-1996, JJG 682-1990 or JJG 689-1990, and it fixes the determination conditions, from light source to sample solution, and the determination methods themselves.

It covers impurity determination at absorption wavelengths between 200 nm and 850 nm. It replaces GB/T 9721-1988, was issued on 1 September 2006 and has applied since 1 April 2007.

Found what you need?

Document preview — GB/T 9721-2006

National Standard of the People's Republic of China

ICS
71.040.40
Classification
G 60
Replacing
GB/T 9721-1988

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions4
  • 4 Principles of method6
  • 5 Reagent7
  • 6 Instruments8
  • 7 Determination8

Foreword

This standard replaces GB/T 9721-1988 "Chemical reagent - General rules for the molecular absorption spectrophotometry (ultraviolet and visible)". As compared with GB/T 9721-1988, the main changes are as follows.

- ADD the references (2 of this version);

- ADD the terms and symbols (3.4, 3.16, and 3.18 of this version);

- MODIFY the relevant contents of the "Instruments" part (6.1, 6.2 and 6.3 of 1988 version; 6 of this version);

- SUPPLEMENT the specification for the technical requirements of the "absorption pool" (7.1.4 of 1988 version; 7.1.4 of this version);

- SUPPLEMENT the application conditions as required by the "absorbance reading range" (7.1.6 of 1988 version; 7.1.6 of this version);

- ADD the "Determination of molar absorptivity" (7.2.4 of this version) and the "Determination of mass absorptivity" (7.2.6 of this version);

- ELIMINATE the "precision" (8 of 1988 version).

1 Scope

This standard specifies the instrument requirements and determination methods of chemical reagent molecular absorption spectrophotometry (ultraviolet and visible).

This standard applies to the determination of the impurity contents in the chemical reagent of the absorption wavelength between 200 nm ~ 850 nm AND the qualitative and quantitative analysis of the organic chemical reagents, indicators, and special reagents.

2 Normative references

The provisions in following documents become the provisions of this Standard through reference in this Standard. For the dated references, the subsequent amendments (excluding corrections) or revisions do not apply to this Standard; however, parties who reach an agreement based on this Standard are encouraged to study if the latest versions of these documents are applicable.

For undated references, the latest edition of the referenced document applies.

GB/T 6682-1992 Water for laboratory use; Specifications (eqv ISO 3696:1987)

GB/T 14666 Terms for analytical chemistry

JJG 178-1996 Visible spectrophotometer

JJG 682-1990 Dual-beam ultraviolet visible spectrophotometer

JJG 689-1990 Ultraviolet, visible, near-infrared spectrophotometer

3 Terms and definitions

The terms and definitions as defined in GB/T 14666 AND the following terms and definitions apply to this standard.

3.1 Absorption peak

It refers to the wavelength of the absorption spectrum at the maximum absorbance value, in the unit of nm.

3.2 Absorption spectrum

It refers to, when the concentration of the substance to be determined and the thickness of the absorption pool are constant, the curve of the absorbance (or any function of the absorbance) corresponding to the wavelength (or any function of the wavelength).

3.3 Thickness (L)

It refers to the distance between the two parallel and transparent internal surface planes of the absorption pool, in the unit of mm or cm.

3.4 Amount-of-substance concentration (c)

It refers to the ratio of the amount of substance to the volume of solution, in the unit of mol/L.

3.5 Molar absorptivity (epsilon)

It refers to the absorptivity epsilon with the thickness expressed in cm AND the concentration expressed in mol/L, in the unit of L/(cm·mol).

4 Principles of method

The valence electrons in the molecule of the solution can selectively absorb ultraviolet or visible light, AND transit from the ground state to the excited state, forming an ultraviolet-visible absorption spectrum. Qualitative analysis is carried out based on the absorption peak and the molar absorptivity in the UV-visible absorption spectrum.

5 Reagent

5.1 Water

The water used to prepare solution for the instrument calibration shall comply with the level II water requirements in GB/T 6682-1992. The water used for sample test shall be of the level II or level III water in GB/T 6682-1992 in accordance with the requirements of the product standard.

5.2 Organic solvents

SELECT appropriate solvent based on the requirements of the product standards, and CHECK whether the solvent used for the test complies with the requirements in the vicinity of the determination wavelength; AND it shall be […]

5.3 Buffer solution

It shall be prepared in accordance with the requirements of the product standard; AND the reagents used shall be free from interference absorption in case of determination at the UV zone.

5.4 Standard sample solution

It shall be prepared together with the sample solution to be determined in accordance with the requirements of the product standard.

6 Instruments

In accordance with the sample determination requirements, it shall select the single-wavelength single-beam or single-wavelength dual-beam UV visible spectrophotometer which complies with the requirements of JJG 178-1996, JJG 682-1990 or JJG 689-1990.

7 Determination

7.1 Selection of determination conditions

7.1.1 Light source

SELECT the light source based on the determined wavelength; USE the hydrogen lamp (or tritium lamp) when the determined wavelength is 200 nm ~ 350 nm, AND the tungsten lamp (or iodine tungsten lamp) when the determined wavelength is 350 nm ~ 850 nm.

7.1.3 Determined wavelength

In the development of product standards, the determined wavelength is determined in accordance with the position of the substance to be determined, which is generally selected from the wavelength range of the maximum absorbance.

7.1.4 Absorption pool

In accordance with the requirements of use, it may, in accordance with the provisions of clause 2.2 "Matching of absorption pool" in JJG 178-1996 OR the provisions of 12 "Specification of standard substance quartz absorption pool" in JJG 682-1990, select the absorption pool; […]

7.1.5 Sample solution

It shall be prepared in accordance with the provisions of product standards.

The solution shall be uniform and non-scatter, that is, there shall not be bubbles and suspended substances that affect the absorption of light.

7.2 Determination method

7.2.1 Absorption curve method

When using the automatic recording instrument, it can automatically scan and draw the absorption curve.

7.2.2 Working curve method

PREPARE more than 4 standard solutions with the concentrations in appropriate proportions in accordance with the provisions of the product standards; USE the blank solution (or solvent) as the reference solution, and meanwhile USE the absorbance of the blank solution (or solvent) to conduct calibration.

7.2.5 Sensitivity determination

CONDUCT determination in accordance with the specified conditions of the product standards. Sensitivity can be expressed by the molar absorptivity.

7.2.6 Determination of mass absorptivity

CONDUCT determination in accordance with the specified conditions of the product standards.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Cited by

Similar standards

Editions of GB/T 9721

EditionTitleRevisionStatus
GB/T 9721-2006Chemical reagent - General rules for the molecular absorption spectrophotometry (ultraviolet and visible)current editionCurrent
GB/T 9721-1988Chemical reagent - General rules for the molecular absorption spectrophotometry (ultraviolet and visible)previous editionSuperseded

This page sells the current edition, GB/T 9721-2006. Earlier editions are listed for reference only.

How to Download GB/T 9721-2006

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 9721-2006

$75.00