GB/T 36406-2018Test method of tin penetrated on float glass surface (English PDF)
平板玻璃表面渗锡量的测试方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 36406-2018 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
June 7, 2018
Implementation date
May 1, 2019
Scope
GB/T 36406-2018 is the English-translated version of 平板玻璃表面渗锡量的测试方法.
GB/T 36406-2018 sets out how to measure the tin that penetrates the surface of float glass, expressed as micrograms of tin per square centimetre of glass surface.
It specifies the apparatus, general rules on reagents, water and parallel testing, the test methods, how results are expressed, precision requirements and the test report.
The first method etches the tin side of the sample with hydrofluoric and nitric acid, oxidizes the tin and measures the red phenylfluorone-CTMAB complex by spectrophotometry at 510 nm against a standard curve.
The second method uses graphite furnace atomic absorption with a zirconium-coated graphite tube and a tartaric acid-zirconium matrix modifier. It applies to flat glass produced by the float process and is intended for glass testing laboratories. It was issued on 7 June 2018 and has applied since 1 May 2019.
Found what you need?
Document preview — GB/T 36406-2018
National Standard of the People's Republic of China
- ICS
- 81.040.10
- Classification
- Q 33
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Apparatus5
- 4 General Rules5
- 5 Test Methods6
- 6 Result Expression14
- 7 Precision Requirements14
- 8 Test Report14
1 Scope
This Standard specifies the apparatus, general rules, test methods, result expression, precision requirements, test report for the test method of tin penetrated on float glass surface.
This Standard is applicable to the flat glass produced by float process.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.
GB/T 6682 Water for Analytical Laboratory Use - Specification and Test Methods
GB/T 8170 Rules of Rounding off for Numerical Values & Expression and Judgement of Limiting Values
GB/T 9721 Chemical Reagent - General Rules for the Molecular Absorption Spectrophotometry (Ultraviolet and Visible)
GB/T 12806 Laboratory Glassware - One-Mark Volumetric Flasks
GB/T 12808 Laboratory Glassware - One Mark Pipettes
GB/T 15337 General Rules for Atomic Absorption Spectrometric Analysis
GB/T 21389 Vernier Dial and Digital Display Calipers
4 General Rules
4.2 Reagent
The density of commercially available solution reagents listed in this Standard refers to the density (rho) at 20°C ambient conditions. The unindicated concentration of the used acid refers to the concentration of commercially available acid.
Use the volume ratio to express the dilution degree of the reagent, for instance, hydrochloric acid (1+3) means that the 1 part by volume of concentrated hydrochloric acid is mixed with 3 parts by volume of water.
4.3 Parallel test
Perform two sets of tests in parallel under repetitive conditions.
5 Test Methods
5.1 Phenylfluorone-cetyltrimethylammonium bromide spectrophotometry
5.1.1 Principle
The surface of the glass sample was etched with a mixed acid solution of hydrofluoric acid-nitric acid; transfer all the penetrated tin into the mixed acid solution. Use potassium permanganate to oxide all the divalent tin into tetravalent tin. Tin(Iv)- phenylfluorone-cetyltrimethylammonium bromide forms a red ternary complex in a sulfuric acid medium. Measure its absorbance at a wavelength of 510nm. In a certain centration range, the absorbance is proportional to the concentration of tin. From this, the amount of penetrated tin per unit surface area (µg/cm2) of the glass sample can be calculated.
5.1.2 Reagent
5.1.2.1 Nitric acid
The density is about 1.42g/mL; the mass fraction is 65%~68%.
5.1.2.2 Hydrofluoric acid
Warning. Hydrofluoric acid is toxic and corrosive. Wear rubber gloves during operation to prevent skin contact!
The density is about 1.14g/mL; the mass fraction is 40% above.
5.1.2.15 Phenylfluorone (0.3g/L)
Take 0.15g of phenylfluorone; dissolve into 5mL of sulfuric acid solution (5.1.2.7); dilute with anhydrous ethanol to 500mL.
5.1.2.16 Tin standard stock solution (100µg/mL)
Accurately take 0.1000g of high-purity tin metal (more than 99.9% tin content; wash the surface oxide with dilute hydrochloric acid before weighing; then wash thoroughly with water and anhydrous ethanol; dry it); place into 10mL of sulfuric acid (5.1.2.3); evaporate till there is sulfur dioxide white smoke; take off to cool off. Use sulfuric acid solution (5.1.2.8) to dissolve; then transfer into 1000mL volumetric flask; dilute with sulfuric acid solution (5.1.2.8) to the scale; shake up; or use the nationally recognized standard solution to prepare.
5.1.2.17 Tin standard use solution (10.0µg/mL)
Accurately pipette 10.00mL of tin standard stock solution (5.1.2.16) into 100mL volumetric flask; dilute with sulfuric acid solution (5.1.2.9) to the scale; shake up.
5.1.3 Drawing of standard curve
Accurately pipette 0.00mL, 1.00mL, 2.00mL, 3.00mL, 4.00mL, 5.00mL of tin standard use solution (5.1.2.17) into a group of 100mL volumetric flasks; separately add 10mL of sulfuric acid solution (5.1.2.8); titrate potassium manganate solution (5.1.2.11) till red color appears; add water to about 30mL. Titrate ascorbic acid solution (5.1.2.12) till the solution is colorless; titrate another excessive 2 drops. Successively add 6mL of oxalic acid aqueous solution (5.1.2.13), 10mL of ascorbic acid solution (5.1.2.12), 10mL of CTMAB solution (5.1.2.14), 8mL of penylfluorone solution (5.1.2.15). Shake well after each addition of a solution. Finally, dilute with water to the scale; shake up; prepare the standard solution series with tin contents of 0.00µg, 10.0µg, 20.0µg, 30.0µg, 40.0µg, 50.0µg. After 20min, take blank reagent as a reference on the spectrophotometer at the wavelength of 510nm; use 0.5cm cuvette to measure the absorbance of the solution; draw the standard curve of tin content in tin standard solution corresponding to the absorbance.
5.1.4 Analysis steps
5.1.4.1 Preparation of sample solution
Cut the glass piece into a rectangular piece with size of 5cm×5cm; use the tin surface tester to distinguish the to-be-tested surface. Use the Vernier caliper (3.4) to accurately measure the length and width of the to-be-tested surface of the glass sample; calculate its surface area (S1). Clean it, and place into plastic erosion plate (3.3.1); and maintain […]
5.2.2.7 Sulfuric acid solution (1+1)
Warning. Improper dilution is dangerous!
Slowly inject 1 part by volume of sulfuric acid into 1 part by volume of water, mix evenly.
5.2.2.9 Zirconium oxychloride solution (Zr10mg/mL)
Take 0.36g of zirconium oxychloride (ZrOCl2·8H2O); dissolve into 10mL of water.
5.2.2.10 Tartaric acid-zirconium matrix improver (Zr 1mg/mL; tartaric acid 100mg/mL)
Pipette 1mL of zirconium oxychloride solution (5.2.2.9) into a small beaker containing 1mL of nitric acid solution (5.2.2.6); successively add 8mL of water, 1g of tartaric acid (5.2.2.8); dissolve completely.
5.2.2.11 Tin standard stock solution (100µg/mL)
Accurately take 0.1000g of high-purity tin metal (more than 99.9% tin content; wash the surface oxide with dilute hydrochloric acid before weighing; then wash thoroughly with water and anhydrous ethanol; dry it); place into 100mL beaker; add 10mL of nitric acid (5.2.2.1); after dissolving, transfer into 1000mL volumetric flask; dilute with nitric acid solution (5.2.2.6) to the scale; shake up; use the nationally recognized standard solution to prepare.
5.2.2.12 Tin standard intermediate solution (10µg/mL)
Accurately pipette 10.00mL of tin standard stock solution (5.2.2.11) into 100mL volumetric flask; dilute with the nitric acid solution (5.2.2.6) to the scale; shake up.
5.2.2.13 Tin standard use solution (1µg/mL)
Accurately pipette 10.00mL of tin standard intermediate solution (5.2.2.12) into 100mL volumetric flask; dilute with nitric acid solution (5.2.2.6) to the scale; shake up.
5.2.3 Drawing of standard curve
5.2.3.1 Use a micropipette to transfer about 150µL of zirconium oxychloride solution (5.2.2.9) into the pyrolytic graphite tube; select the instrument operating parameters and conditions as per the instrument operating instructions; perform a full process atomization in the graphite furnace; prepare the zirconium-coated graphite tube.
5.2.3.2 Separately and accurately pipette 0.00mL, 1.00mL, 2.00mL, 4.00mL, 6.00mL, 8.00mL, 10.00mL of tin standard use solution (5.2.2.13) into a group of 100mL […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 6682Water for analytical laboratory use - Specification and test methods
- GB/T 8170Rules of rounding off for numerical values & expression and judgement of limiting values
- GB/T 9721Chemical reagent - General rules for the molecular absorption spectrophotometry (ultraviolet and visible)
- GB/T 12806Laboratory glassware -- One-mark volumetric flasks
- GB/T 12808Laboratory glassware - One mark pipettes
- GB/T 15337General rules for atomic absorption spectrometric analysis
Similar standards
How to Download GB/T 36406-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 12806-2011 — Laboratory glassware -- One-mark volumetric flasks
GB/T 12808-2015 — Laboratory glassware - One mark pipettes
GB/T 15337-2008 — General rules for atomic absorption spectrometric analysis