GB/T 26792-2019High performance liquid chromatography (English PDF)
高效液相色谱仪
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for GB/T 26792-2019 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
October 18, 2019
Implementation date
May 1, 2020
Scope
GB/T 26792-2019 is the English-translated version of 高效液相色谱仪.
GB/T 26792-2019 is the Chinese national standard for high performance liquid chromatographs. It sets the requirements, test methods, inspection rules and the marking, packaging, transport and storage of the instrument, and does not apply to special-purpose instruments built on the same principle.
The normal working conditions it fixes are 5 °C to 35 °C, 20 % to 80 % relative humidity, 75 kPa to 106 kPa, an AC supply of 220 V ± 22 V at 50 Hz ± 0,5 Hz, grounding resistance not above 4 ohm and a well ventilated room.
It then covers appearance, the infusion pump - whose pressure drop must stay within 3 MPa in the sealing test - and the detectors, diode array and evaporative light scattering included.
It replaces GB/T 26792-2011, was issued on 18 October 2019 and has applied since 1 May 2020.
Found what you need?
Document preview — GB/T 26792-2019
National Standard of the People's Republic of China
- ICS
- 17.180.99
- Classification
- N 52
- Replacing
- GB/T 26792-2011
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009. This standard replaces GB/T 26792-2011 "High Performance Liquid Chromatograph".
Compared with GB/T 26792-2011, the main technical contents of this standard are as follows.
--- Revised the scope of non-applicability (see Chapter 1, Chapter 1 of the 2011 edition);
--- Added "good ventilation in the room" [see 3.1g)];
--- Revised the requirements for sealing indicators (see 3.3.1, 3.3.1 of the 2011 edition);
--- Modified Table 1 flow set value error and flow stability requirements, increased the flow rate of 5.0mL/min, 10.0mL/min Test points (see Table 1, Table 1 of the 2011 edition);
--- Modified the main technical indicators of the detector, and summarized the articles as shown in Table 2 (see Table 2, Table 3.5, 3.5 of the 2011 edition);
--- Added diode array detector indicators and test methods, evaporative light scattering detector indicators and test methods (see Table 2, 4.5.1) 4.5.4);
---Modified security requirements (see 3.7, 3.7 of the 2011 edition);
---Modified the column oven test procedure (see 4.4.1.2, 2011 edition of 4.4.1.2);
--- Modified the dynamic baseline noise and drift test of the refractive index detector (see 4.5.2.1, 4.5.2.1 of the 2011 edition);
--- Modified the solution for the minimum detection concentration of the refractive index detector (see 4.5.2.2, 2011 edition 4.5.2.2);
--- Modified the solution for the minimum detection concentration of the fluorescence detector (see 4.5.3.3, 2011 edition 4.5.3.3).
1 Scope
This standard specifies the requirements, test methods, inspection rules and marking, packaging, transportation and storage of high performance liquid chromatography.
This standard applies to high performance liquid chromatography (hereinafter referred to as the instrument).
This standard does not apply to special instruments based on the principle of high performance liquid chromatography.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 191-2008 packaging storage and transportation icon
GB/T 2829-2002 Periodic inspection count sampling procedures and tables (applicable to the inspection of process stability)
GB/T 6682-2008 Analytical laboratory water specifications and test methods
GB/T 11606-2007 Analytical instrument environmental test method
GB/T 12519-2010 General technical conditions for analytical instruments
GB/T 13384-2008 General technical conditions for mechanical and electrical product packaging
Standard column for GB/T 30433 liquid chromatograph test
GB/T 34065-2017 Safety requirements for analytical instruments
3 Requirements
3.1 Normal working conditions
The normal working conditions of the instrument are as follows.
a) Ambient temperature. 5 ° C ~ 35 ° C;
b) Relative humidity. 20%~80%;
c) atmospheric pressure. 75kPa~106kPa;
d) Power supply. AC voltage 220V±22V, frequency 50Hz±0.5Hz;
e) Grounding resistance <= 4ohm;
f) indoors should avoid flammable, explosive and corrosive gases and strong vibration, electromagnetic interference and air convection;
g) The room should be well ventilated.
3.2 Appearance
3.2.1 The appearance of the instrument should be smooth and clear, the writing should be clear, the surface should be coated with uniform color, and there should be no obvious scratches, exposed bottoms, cracks, blistering, burrs, etc. phenomenon.
3.2.2 Each adjustment knob, button, switch, etc. should work normally and without looseness; the indication and display should be clear and complete.
3.3 Infusion pump
3.3.1 Sealing
The infusion pump flow path is cut off, the pressure reaches 90% of the upper limit value, the infusion pump stops running for 10 minutes, and the pressure drop should be no more than 3 MPa.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191-2008Graphical symbols marking for handling and storage of packages
- GB/T 2829-2002Sampling procedures and tables for periodic inspection by attributes
- GB/T 6682-2008Water for analytical laboratory use - Specification and test methods
- GB/T 11606-2007Methods of environmental test for analytical instruments
- GB/T 12519-2010General specification of analytical instruments
- GB/T 13384-2008General specifications for packing of mechanical and electrical product
Cited by
- GB/T 14491-2025Propylene oxide for industrial use
- GB/T 46135.3-2025Test methods for organic luminescence electronic and hole function material - Part 3: Purity
- GB/T 43724-2024Test methods for liquid crystal monomers
- GB/T 41726-2022Identification and detection methods of monomer ginsenoside
- GB/T 42241-2022Nanotechnology—Determination of [60]/[70] fullerene purity—High performance liquid chromatography
- GB/T 22533-2018Grade quality of cultivated ginseng
Similar standards
Editions of GB/T 26792
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 26792-2019 | High performance liquid chromatography | current edition | Current |
| GB/T 26792-2011 | High performance liquid chromatography | previous edition | Superseded |
This page sells the current edition, GB/T 26792-2019. Earlier editions are listed for reference only.
How to Download GB/T 26792-2019
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 11606-2007 — Methods of environmental test for analytical instruments
GB/T 13384-2008 — General specifications for packing of mechanical and electrical product
GB/T 30433-2021 — Standard column used for evaluating liquid chromatograph