Scope
GB/T 17040-2019 is the English-translated version of 石油和石油产品中硫含量的测定 能量色散X射线荧光光谱法.
GB/T 17040-2019 gives the Chinese method for determining sulfur in petroleum and petroleum products by energy dispersive X-ray fluorescence spectrometry. The specimen is placed in the X-ray beam and the intensity of the excited sulfur Kalpha line, at about 2.3 keV, is compared with that of prepared standard samples.
It covers sulfur mass fractions from 0.0017% to 4.6% in single-phase liquid products or ones soluble in hydrocarbon solvents, among them gasoline, diesel, biodiesel, jet fuel, kerosene, naphtha, residual fuel, crude oil and lubricating base oils.
It specifies the spectrometer, the reagents, the sample cups and window films, calibration, sampling, the procedure, calculation, reporting, quality control and precision.
It cites GB/T 4756, GB/T 8170, GB/T 27867, NB/SH/T 0843, ASTM D6259 and ASTM D7343. It replaces GB/T 17040-2008, was issued on 25 March 2019 and has applied since 1 October 2019.
Found what you need?
Document preview — GB/T 17040-2019
National Standard of the People's Republic of China
- ICS
- 75.080
- Classification
- E 30
- Replacing
- GB/T 17040-2008
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- Introduction5
- 1 Scope6
- 2 Normative References6
- 3 Method Summary7
- 4 Conditions for Use of the Method7
- 5 Instruments and Equipment8
- 6 Reagents and Materials8
- 7 Preparation of Sample Cups10
- 8 Calibration10
- 9 Instrument Preparation13
- 10 Sampling13
- 11 Test Procedures14
- 12 Calculation14
- 13 Test Report14
- 14 Quality Control15
- 15 Precision and Bias15
- Appendix A (Informative) Interferences19
- Appendix B (Informative) Handling of Oxygenated Fuels21
- Appendix C (Informative) Precision for Diesel Fuel22
- Appendix D (Informative) Precision for Gasoline24
1 Scope
This Standard specifies a test method for the determination of sulfur content in petroleum and petroleum products using energy dispersive X-ray fluorescence spectrometry.
This Standard is applicable to the determination of petroleum and petroleum products that are single-phase and liquid at ambient or moderately elevated temperatures, or are soluble in hydrocarbon solvents; these include automotive gasoline, ethanol-blended gasoline, diesel fuel, biodiesel and its blends, jet fuel, kerosene, other distillate fuels, naphtha, residual fuel, crude oil, lubricating base oils, hydraulic fluids, and similar petroleum products.
This Standard is applicable to the determination of sulfur mass fractions in the range of 0.0017% (17 mg/kg) to 4.6% (46 000 mg/kg).
2 Normative References
The following documents are indispensable for the application of this document. In terms of references with a specified date, only the version with that specific date is applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB/T 4756 Method for manual sampling of petroleum liquids
GB/T 8170 Rules of rounding off for numerical values & expression and judgement of limiting values
GB/T 27867 Petroleum liquid - Automatic pipeline sampling
NB/SH/T 0843 Standard practice for evaluating analytical measurement system performance of petrochemical industry (Statistical techniques)
ASTM D6259 Practice for determination of a pooled limit of quantitation for a test method
ASTM D7343 Practice for optimization, sample handling, calibration, and validating of X-ray fluorescence spectrometry methods for elemental analysis of petroleum products and lubricants
3 Method Summary
The specimen is placed in the X-ray beam emitted from an X-ray source, and the intensity of the excited sulfur Kalpha characteristic X-rays (energy: 2.3 keV) is measured. The accumulated counts are compared with those of pre-prepared standard samples to determine the sulfur content expressed as a mass fraction (see ASTM D7343). At least three sets of standard samples are required to cover the following mass fraction ranges: 0.0% to 0.1%, 0.1% to 1.0%, and 1.0% to 5.0%.
4 Conditions for Use of the Method
This Standard assumes that the matrix of the standard samples matches that of the samples to be tested, or that differences in the matrix have been taken into account (see A.2). Differences in the carbon-to-hydrogen mass ratio (mC/mH) between the samples to be tested and the standard samples (see Appendix A), or the presence of other heteroatoms, may result in a matrix mismatch. Table 1 specifies the permissible contents of heteroatoms, as well as ethanol and methanol, in the samples to be tested.
This Standard may also be used for the analysis of oxygenated fuels containing ethanol and methanol at levels exceeding the ranges specified in Table 1; however, the precision and bias data do not apply (see Appendix B).
For samples with relatively high oxygen content (mass fraction greater than 3%), the samples shall be diluted or matrix-matched standard samples shall be used to ensure the accuracy of the determination results.
5 Instruments and Equipment
5.1 Energy Dispersive X-ray Fluorescence Spectrometer
5.1.1 Any energy dispersive X-ray fluorescence spectrometer suitable for this test method shall include, at a minimum, the components specified in 5.1.2 to 5.1.8 and yield equivalent test results for the samples.
5.1.2 X-ray excitation source: An X-ray tube with an effective energy greater than 2.5 keV.
When using a spectrometer with an X-ray tube as the excitation source, operation shall be in accordance with the instrument manufacturer's safety instructions.
5.1.3 Removable sample cup: These shall be designed to hold the sample and fit the geometry of the X-ray fluorescence spectrometer, featuring a replaceable X-ray-transparent polyester film window; the minimum sample depth shall be 4 mm, and the minimum diameter shall be 10 mm.
Disposable sample cups are preferred for measuring samples with low sulfur content (mass fraction less than 50 mg/kg).
5.1.4 X-ray detector: It exhibits relatively high sensitivity to the sulfur Kalpha characteristic X-rays (energy: 2.3 keV) and has a resolution of no more than 800 eV.
5.1.5 Filter: A device capable of distinguishing the sulfur Kalpha spectral line from other X-rays.
5.1.6 Signal conditioning and data processing electronic system: It is capable of X-ray intensity counting, background X-ray subtraction, spectral line smoothing and correction, and conversion of sulfur X-ray intensity into mass fraction.
5.1.7 Under optimized conditions, the sensitivity of the spectrometer shall be sufficient to measure samples with a sulfur content of approximately 0.05% (mass fraction); the error shall be attributable to counting statistics, and the standard deviation shall not exceed 0.5%.
5.1.8 Display or printer: It is capable of displaying or printing the sulfur mass fraction.
5.2 Analytical Balance
The readability shall be 0.1 mg, and the maximum weighing capacity shall be 100 g.
6 Reagents and Materials
6.1 Reagent purity: All chemical reagents used in this Standard shall be of analytical reagent grade or higher purity.
6.2 Di-n-butyl sulfide (DBS): Not lower than guaranteed reagent grade; molecular formula: (C4H9)2S. In this Standard, it is used as a reference material with a certified sulfur content [typically 21.91% by mass]; this value is used to calculate the precise sulfur content of the prepared standard samples (see 8.1). Warning - n-Butyl sulfide is a flammable and toxic compound.
NOTE: It is necessary to know the sulfur content of the di-n-butyl sulfide, rather than just its purity, because the impurities may also be sulfur-containing compounds.
6.3 Drift monitoring samples (optional): Any sulfur-containing substance that remains stable under repeated X-ray irradiation - such as polysulfide oil, glass, or metal samples - may be used as a drift monitoring sample. Liquids, pressed powders, and solid materials that decompose under repeated X-ray irradiation shall not be used. Available sulfur-containing substances include liquid petroleum products suitable for long-term use, metal alloys, or fused glass discs. With an appropriate counting time, the count rate exhibited by the monitoring sample shall show a relative standard deviation of less than 1%. The count rate of the monitoring sample shall be determined during calibration (see 8.2.1) and re-determined during analysis (see 11.2). The count rate will be used to calculate the drift correction factor (see 14.6).
Although the calculation can be performed manually, drift correction is usually carried out automatically by software. For X-ray spectrometers with relatively good stability, the magnitude of the drift correction factor does not change significantly overall.
6.4 Polysulfide oil: Nonyl polysulfide with a known sulfur content, typically diluted with a hydrocarbon-based solvent. Warning - It may cause skin sensitization.
NOTE: Polysulfide oil is a high-molecular-weight oil with a sulfur content of up to 50% by mass. It retains excellent physical properties - such as low viscosity, low volatility, and a relatively long shelf life - even after thorough mixing with white oil. The sulfur content of the polysulfide oil is determined by diluting it with white oil and then directly comparing it against a corresponding certified reference material.
6.5 White oil: Sulfur content less than 2 mg/kg. In this Standard, it serves as the diluent for preparing standard samples (the sulfur content of any other diluents used in this Standard is also less than 2 mg/kg). When performing low-sulfur testing (mass fraction less than 200 mg/kg), the calculation of the sulfur content in the standard sample shall include the sulfur content of the matrix material (see 8.1). If the sulfur content of a solvent or reagent is unknown, it must be determined, or a reagent of higher purity must be used to prepare the standard sample.
6.6 X-ray transparent films: Any film that is resistant to the sample's chemical action, sulfur-free, and allows sufficient X-ray transmission may be used. Suitable options include polyester, polypropylene, polycarbonate, and polyimide films. However, samples with high aromatic hydrocarbon content may dissolve polypropylene, polycarbonate, and polyester films.
6.7 Helium (optional): Purge gas; select the appropriate specifications according to the instrument manufacturer's recommendations when required.
6.8 Counting gas: For instruments equipped with a flow-proportional counter, the purity of the counting gas shall meet the instrument manufacturer's specification requirements.
6.9 Calibration check sample: One or more liquid petroleum products or standard samples with known sulfur content (e.g., polysulfide oil, di-n-butyl sulfide, thiophene) that were not used to establish the calibration curve. Calibration check samples can be used to determine the accuracy and precision of the initial calibration curve (see Clause 8).
6.10 Quality control (QC) samples: Stable petroleum products or other solid samples analyzed periodically to ensure that the determination system is in a satisfactory state (see Clause 14). It is recommended to verify the system status using QC samples and quality control charts. The quality control process shall be carried out independently by each laboratory. Samples that are typical and remain stable over a long period may be used as QC samples. Solid materials are recommended for use as drift monitoring samples.
7 Preparation of Sample Cups
7.1 If reusable sample cups are used, ensure that they are clean and dry before use. Disposable sample cups shall not be reused. Polyester or polycarbonate films (see 6.6) with a thickness of less than 10 µm are commonly selected as window films due to their high transmittance for characteristic sulfur X-rays. A new window film shall be used for the determination of each sample. Avoid touching the interior of the sample cup, the window film exposed to X-rays, and the instrument's window. When measuring samples with relatively low sulfur content, oils from fingers can affect the determination results. Wrinkles in the window film can affect the transmittance of sulfur X-rays. Therefore, to ensure reliable determination results, the window film shall be kept flat and clean.
7.2 Recalibrate the instrument whenever the type or thickness of the window film changes, or before using a new batch of polyester film.
7.3 For samples with high aromatic hydrocarbon content, it is advisable to use a film free of impurity elements as the window film; a 6 µm thick polyimide film is the preferred choice.
8 Calibration
8.1 Standard Samples
8.1.1 Prepare standard samples by diluting di-n-butyl sulfide with sulfur-free white oil or another suitable matrix material (see 6.5). The sulfur content of the unknown sample shall fall within the range of the calibration curve used. Table 2 lists the recommended sulfur contents for standard samples. When the sulfur mass fraction of the standard sample is less than 0.02% (200 mg/kg), the sulfur content of the matrix material must be taken into account, as shown in Formula (1). Accurately weigh the recommended amounts of di-n-butyl sulfide and matrix diluent, calculate the sulfur content of the prepared standard sample, and input this value into the instrument for calibration. The sulfur content in the standard sample can be calculated using Formula (1). […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 4756Method for manual sampling of petroleum liquids
- GB/T 8170Rules of rounding off for numerical values & expression and judgement of limiting values
- GB/T 27867Petroleum liquid -- Automatic pipeline sampling
NB/SH/T 0843 · ASTM D6259 · ASTM D7343
Cited by
- GB 17411-2026Marine fuels
- GB 11121-2025Gasoline engine oils
- GB/T 384-2025Determination the heat of combustion for hydrocarbon fuels — Bomb calorimeter method
- GB/T 41961-2022Treatment and disposal methods for used mineral oil lubricant
- GB 36170-2018Crude oils
- GB/T 33956-2017Methods of determination and calculation of heat balance in continuous reheating furnace for steel rolling
Similar standards
Editions of GB/T 17040
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 17040-2019 | Determination of sulfur in petroleum and petroleum products—Energy dispersive X-ray fluorescence spectrometry | current edition | Current |
| GB/T 17040-2008 | Determination of sulfur in petroleum and petroleum products—Energy dispersive X-ray fluorescence spectrometry | previous edition | Superseded |
This page sells the current edition, GB/T 17040-2019. Earlier editions are listed for reference only.
How to Download GB/T 17040-2019
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 27867-2011 — Petroleum liquid -- Automatic pipeline sampling
GB/T 4756-2015 — Method for manual sampling of petroleum liquids
GB/T 47560-2026 — Wind energy generation systems - Technical specification for the yaw system of wind turbines
Secure payment via Stripe
Payments accepted
GB/T 17040-2019
$235.00