GB/T 4336-2016Carbon and low-alloy steel—Determination of multi-element contents—Spark discharge atomic emission spectrometric method (routine method) (English PDF)
碳素钢和中低合金钢 多元素含量的测定 火花放电原子发射光谱法(常规法)
This is a limited preview
Buy now to download the full PDF (43 pages)
Need quotation for your management?
Get a formal quotation for GB/T 4336-2016 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
February 24, 2016
Implementation date
November 1, 2016
Scope
GB/T 4336-2016 is the English-translated version of 碳素钢和中低合金钢 多元素含量的测定 火花放电原子发射光谱法(常规法).
GB/T 4336-2016 specifies the determination of the carbon, silicon, manganese, phosphorus, sulfur, chromium, nickel, tungsten, molybdenum, vanadium, aluminum, titanium, copper, niobium, cobalt, boron, zirconium, arsenic and tin contents of carbon and low alloy steels by the spark discharge atomic emission spectrometric method (routine method).
Clause 1 states that it applies to the sample analysis of casting or forging carbon and low-alloy steels made by electric furnace, induction furnace, electroslag furnace, converter furnace and the like, and refers the measurement range of each element to Table 1, whose values the page does not print.
Its fourteen clauses are Scope; Normative references; Principle; Instruments; Sampling and sample preparation; Standard samples, standardization samples and control samples; Preparation of instruments; Calibration; Analysis conditions and analysis procedures; Calculation of analysis results; Precision; Acceptability of measurement results and determination of final report result; Determination of the accuracy of laboratory measurement results; and Test report, followed by informative Annex A with additional information on the precision test, informative Annex B with precision-related data, and a No.1 Amendment List.
Clause 3 gives the principle: a discharge is generated between the prepared bulk sample and the counter electrode under the action of the spark light source, plasma forms at high temperature in the inert atmosphere, the activated atoms of the measured element emit a characteristic spectrum as electrons fall from high to low energy levels, and the spectral intensity of the selected analytical line and of the internal standard line is measured.
Clause 4 lists the units of the spectrometer, among them a stable spark excitation light source, a spark room designed for argon and mounted directly on the spectrometer, with an argon flush spark bracket for the flat sample and the rod-shaped counter electrode, a spectrometer whose reciprocal of the first-grade spectral line dispersion is less than 0.6 nm/mm, with a focal length of 0.35 m to 1.0 m and a wavelength range of 165.0 nm to 410.0 nm, working below 3 Pa or filled with high-purity inert gas of purity not less than 99.999 %, and a photometric system with a photoelectric conversion detector, integrating capacitors, a measuring unit and the necessary switching circuits.
Clause 5 refers sampling and sample preparation to GB/T 20066, keeps the deoxidizer content below 0.35 % when aluminum deoxidation is used, and has samples taken from the mould usually cut at one third of the lower end or, when uncut, skimmed by 1 mm of surface; clause 6 distinguishes standard samples used to plot the calibration curve, standardization samples used to correct instrument drift and control samples used to correct the results of the analytical sample or for type standardization correction.
Clause 7 asks for a shock-proof and clean laboratory at 15 degrees C to 30 degrees C with a relative humidity below 80 % and an indoor temperature change within 5 degrees C in the same standardization cycle, a supply voltage change of less than 10 % and a frequency change of less than 2 % with a dedicated ground line, and adequate power time for the excitation light source and the photometric system; clause 8 sets out the calibration curve method, in principle with standard samples of more than 5 levels each activated at least 3 times, the original calibration curve method and the control sample method.
Clause 9 refers the analysis conditions to Table 2 and the analytical and internal marking lines to Table 3 and activates each analytical sample at least 2 times, clause 10 reads the element content off the calibration curve and keeps the result within the content range of the standard samples used, clause 11 records a precision test run by 15 laboratories on 11 to 22 levels of 14 elements in low-alloy steels in 2013 and by 12 laboratories on 18 to 36 levels of 5 elements in low-alloy steels in 2014, and clause 12 accepts two independent measurement results whose absolute difference is not greater than r and reports their arithmetic mean; clause 2 cites GB/T 6379.1, GB/T 6379.2 and GB/T 20066.
It replaces GB/T 4336-2002, was issued on February 24, 2016 and has applied since November 1, 2016.
Found what you need?
Document preview — GB/T 4336-2016
National Standard of the People's Republic of China
- ICS
- 77.080.20
- Classification
- H 11
- Replacing
- GB/T 4336-2002
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope5
- 2 Normative references6
- 3 Principle6
- 4 Instruments6
- 5 Sampling and sample preparation7
- 6 Standard samples, standardization samples and control samples8
- 7 Preparation of instruments9
- 8 Calibration10
- 9 Analysis conditions and analysis procedures11
- 10 Calculation of analysis results15
- 11 Precision15
- 12 Acceptability of measurement results and determination of final report result16
- 13 Determination of the accuracy of laboratory measurement results19
- 14 Test report19
- Annex A (informative) Additional information of precision test20
- Annex B (informative) Precision-related data No.1 Amendment List33
1 Scope
This Standard specifies the determination of carbon, silicon, manganese, phosphorus, sulfur, chromium, nickel, tungsten, molybdenum, vanadium, aluminum, titanium, copper, niobium, cobalt, boron, zirconium, arsenic and tin contents in carbon and low alloy steels by spark discharge atomic emission spectrometric method (routine method).
This Standard applies to the sample analysis of casting or forging carbon and low-alloy steels by electric furnace, induction furnace, electroslag furnace, converter furnace, etc.; see Table 1 for the measurement range of each element.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
GB/T 6379.1 Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions
GB/T 6379.2 Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
GB/T 20066 Steel and iron - Sampling and preparation of samples for the determination of chemical composition
3 Principle
Discharge is generated between the prepared bulk sample and the counter electrode under the action of the spark light source, and the plasma is generated in the high temperature and the inert atmosphere. When the atoms of the measured element are activated, the electrons are transitioned between different energy levels in the atom, and the characteristic spectrum is generated when the transition from the high-energy level to the low-energy level. Measure the spectral intensity of the characteristic spectrum of the selected analytical element and the internal standard element.
4 Instruments
The spark discharge atomic emission spectrometer consists mainly of the following units.
4.1 Excitation light source
The excitation light source shall be a stable spark excitation light source.
4.2 Spark room
The spark room is specially designed for use with argon, and the spark room is mounted directly on the spectrometer, with an argon flush spark bracket to place the flat sample and the rod-shaped counter electrode. The argon gas path in the spark chamber shall be able to replace the air in the optical path between the analysis gap and the condenser and provide an argon atmosphere for the analysis gap.
4.5 Spectrometer
Generally, for the spectrometer, the reciprocal of the first-grade spectral line dispersion of shall be less than 0.6 nm/mm, the focal length is 0.35 m ~ 1.0 m, the wavelength range is 165.0 nm ~ 410.0 nm, the vacuum of the spectrometer shall work below 3 Pa, or being filled with high-purity inert gas (the gas does not the absorb spectral line with a wavelength less than 200 nm, and the purity is not less than 99.999 %).
4.6 Photometric system
The photometric system shall include a photoelectric conversion detector for receiving a signal, an integrating capacitor capable of storing each of the output electrical signals, a measuring unit for directly or indirectly recording the voltage or frequency on the integrator, and a necessary switching circuit device provided for the desired timing.
5 Sampling and sample preparation
5.1 Sampling
The sampling and sample preparation is carried out according to the specifications of GB/T 20066. When taking samples, it shall ensure that the analytical sample is taken evenly, without shrinkages and cracks. When taking casting samples, the molten steel shall be injected into the specified mold; when adopting aluminum deoxidation, the deoxidizer content shall not exceed 0.35 %; when taking samples form steel products, it shall select representative parts.
5.2 Sample preparation
Samples taken from the mold are usually cut in the height of 1/3 of the lower end. For uncut samples, the surface shall be removed by 1 mm in thickness. The cutting equipment used shall be the cutting machine with resin cutting blades, metal cutting machines, etc.
6 Standard samples, standardization samples and control samples
6.1 Standard samples
The standard sample is used to plot the calibration curve, its chemical properties and organizational structure shall be similar to those of the analytical samples, and shall cover the content range of the analytical elements and maintain the appropriate gradient; the content of the analytical elements shall be determined with accurate and reliable methods.
6.2 Standardization samples
The change in the state of the instrument may result in the deviation of the measurement results, to directly use the original calibration curve to obtain accurate results, 1 ~ 2 samples are used to standardize the instrument, these samples are called standardization samples. The sample shall be very uniform and have an appropriate content; the sample can be selected from standard samples or specially smelted. When using two-point standardization, the content is the content on each element calibration curve, respectively, near the upper and lower limits.
6.3 Control samples
The control sample is the uniform sample with similar metallurgical process, organizational structure and chemical composition to those of the analytical sample, being used for correcting the measurement results of the analytical sample or for type standardization correction.
7 Preparation of instruments
7.1 Storage of instruments
The spectrometer, as recommended by the equipment manufacturer, shall be placed in shock-proof and clean laboratories, usually the indoor temperature maintains at 15 °C ~ 30 °C and the relative humidity shall be less than 80 %. In the same standardization cycle, the indoor temperature change does not exceed 5 °C.
7.2 Power supply
To ensure the stability of the instrument, the power supply voltage change shall be less than ± 10 %, the frequency change shall be less than ± 2 % and the AC power supply shall be the sine wave. According to the use requirements of the instrument, a dedicated ground line shall be equipped.
7.3 Excitation light source
In order to make the electrical part of the excitation light source work stably, before starting work, it shall be given the appropriate power time.
Use a voltage regulator or a voltage regulator device to adjust the power supply voltage to the desired value of the instrument.
7.6 Photometric system
After being shut down and then restarted, generally adequate power time shall be endured to make the photometric system work stably.
Select the appropriate pre-combustion time of the analytical element by making a pre-combustion curve. The integration time is determined by the test basing on the analytical accuracy.
8 Calibration
8.1 Calibration curve method
Under the selected working conditions, activate a series of standard samples; in principle, use the standard samples of more than 5 levels, and activate at least 3 times per sample; plot a curve indicating the relation between the luminous intensity (or intensity ratio) and the content (or content ratio) of the analytical element as a calibration curve. Use this calibration curve to measure the element content in the sample.
8.2 Original calibration curve method
The original calibration curve method is to use the calibration curve method to plot the calibration curve. When temperature, humidity, vibration and other factors lead to displacement of the spectral line of the spectral instrument, or changes in luminous intensity leads to the calibration curve drifts, use the standardization sample to correct the drift of the calibration curve, to resume the strength of the corrected element to the strength when the calibration curve is initially established.
8.3 Control sample method
Due to the difference in the smelting process and the organizational structure between the analytical sample and the standard sample used to plot the calibration curve, the calibration curve is often changed.
9 Analysis conditions and analysis procedures
9.1 Analysis conditions
The recommended analysis conditions of this Standard are as shown in Table 2, and the analytical lines and internal marking lines are listed in Table 3.
9.2 Analysis procedures
9.2.1 Prepare the instrument in accordance with the requirements in 7.2 to 7.6.
9.2.2 Before the analysis work, activate a piece of sample 2 ~ 5 times, to confirm that the instrument is in the best working condition.
9.2.3 Standardization of the calibration curve. under the selected working condition, activate the standardization sample, and each sample is activated at least 3 times to correct the calibration curve. When there are significant changes in the instrument or the original calibration curve exceeds the calibration range due to the drift, the calibration curve needs to be replotted.
9.2.6 Activate the analytical sample according to the working conditions selected by 9.2.2, each sample is activated at least 2 times (the sample is activated once to obtain an independent measurement result;
10 Calculation of analysis results
According to the relative strength (or absolute intensity) of the analytical line, obtain the content of the analytical element from the calibration curve.
The analysis result of the elements to be measured shall be within the content range of a series of standard sample used for the calibration curve.
11 Precision
The precision test of this Standard has been carried out by 15 laboratories to 11 to 22 levels of 14 elements in low-alloy steels in 2013, and by 12 laboratories to 18 to 36 levels of 5 elements in low-alloy steels in 2014.Under the repeatability condition as specified in GB/T 6379.1, each laboratory shall measure the content of each level of element 2 times.
12 Acceptability of measurement results and determination of final report result
12.1 Under the repeatability conditions, if the absolute value of the difference between two independent measurement results is not greater than r, the two measurement results can be accepted. The final report result is the arithmetic mean of two independent measurement results.
12.2 Under the repeatability conditions, if the absolute value of the difference between two independent measurement results is greater than r, the laboratory shall measure one or two more results. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 43 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 6379.1Accuracy (trueness and precision) of measurement methods and results -- Part 1: General principles and definitions
- GB/T 6379.2Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
- GB/T 20066Steel and iron -- Sampling and preparation of samples for the determination of chemical composition
Cited by
- GB/T 13237-2026Cold rolled quality carbon structural steel sheet and strip
- GB/T 31304-2026Epoxy-coated high strength wire stay cable
- GB/T 31938-2026Steel plate, sheet and strip for ore and slurry pipelines
- GB/T 3273-2026Hot rolled steel sheet and strip for automobile frames
- GB/T 32963-2026Zinc-aluminium alloy coated steel wire cable
- GB/T 47975-2026Seamless steel pipes for trenchless drill pipe
Similar standards
Editions of GB/T 4336
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 4336-2016 | Carbon and low-alloy steel—Determination of multi-element contents—Spark discharge atomic emission spectrometric method (routine method) | current edition | Current |
| GB/T 4336-2002 | Carbon and low-alloy steel—Determination of multi-element contents—Spark discharge atomic emission spectrometric method (routine method) | previous edition | Superseded |
This page sells the current edition, GB/T 4336-2016. Earlier editions are listed for reference only.
How to Download GB/T 4336-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 20066-2006 — Steel and iron -- Sampling and preparation of samples for the determination of chemical composition
GB/T 6379.1-2004 — Accuracy (trueness and precision) of measurement methods and results -- Part 1: General principles and definitions
GB/T 6379.2-2004 — Measurement methods and results -- Accuracy (trueness and precision) -- Part 2: Determine the standard methods of measurement repeatability and reproducibility of the basic method