Scope
GB/T 261-2021 (Determination of flash point - Pensky-Martens closed cup method) is available as an English-translated PDF.
GB/T 261-2021 — This Document specifies the method for determining the flash point of flammable liquids, liquids with suspended particles, liquids that tend to form films on the surface under test conditions, and other liquids using a Pensky-Martens closed cup flash point tester. NOTE 1: Although this Document can be used for kerosene with a flash point above 40°C, the flash point of kerosene is generally determined in accordance with GB/T 21789. The flash point of unused lubricating oils is usually determined in accordance with GB/T 3536. NOTE 2: This Document can be used for jet fuels with a flash point below 40°C, but the precision has not been verified. Warning: Since some mixtures do not obtain the flash point defined in this Document, but obtain a significantly increased flash (not the effect of halo) and a change in the flash from blue to yellow-brown, continued heating can cause the vapor around the test cup to burn. This Document is applicable to samples with a flash point range of 40°C ~ 370°C. This Document includes three test steps: Step-A, Step-B and Step-C. Step-A is applicable to distillate fuels (including diesel, biodiesel blends, heating oil and turbine fuel), unused lubricating oils, paints and varnishes and other homogeneous liquids not included in Step-B and Step-C. Step- B is applicable to residual fuel oil, diluted asphalt, used lubricating oil, liquids with suspended particles, liquids that tend to form films on the surface under the test conditions or liquids whose viscosity is not suitable for heating under the stirring speed and heating conditions specified in Step-A. Step-C is applicable to the determination of BD100 biodiesel samples. This Document does not apply to water-based paints and water-based varnishes. NOTE 3: When monitoring the lubricating oil system, in order to compare the flash points of unused lubricating oil and used lubricating oil, Step-A can also be used to determine the flash point of used lubricating oil, but this precision is only applicable to Step-B. C.3.3 Ignitor The diameter of the tip opening of the ignition tube flame jet device is about 0.7mm~0.8mm (see Figure C.4). The tip is stainless steel, or other suitable metal material. The ignitor is equipped with a mechanical operator. When the slide plate is in the "open" position, the tip is lowered so that the center of the flame nozzle opening hole is located between the upper and lower surface planes of the test cup cover and passes through a point on the center radius of the maximum opening A (see Figure C.3). The bead at the tip is made of suitable material, about 4mm, which is equivalent to the flame size, and is assembled in a prominent position on the test cup cover. C.3.4 Automatic ignition device For automatic ignition of the flame, the diameter of the ignitor tip opening is about 0.7mm~0.8mm. C.3.5 Stirring device It is mounted in the center of the test cup cover (as shown in Figure C.4), with two double- bladed metal paddles. The lower paddle surface is about 38mm away from the test cup cover; the two blades are 8mm wide and inclined at 45°. The upper paddle surface is about 19 mm from the test cup cover; the two blades are 8 mm wide and inclined at 45°. The two sets of paddles are fixed on the rotating shaft of the stirrer. When viewed from the bottom of the stirrer, the two blades of one set of paddles are at 0° and 180°; and the two blades of the other set of paddles are at 90° and 270°. The rotating shaft of the stirrer can be connected to the motor by a transmission flexible shaft or a suitable pulley block structure. C.4 Heating chamber and bath jacket The test cup shall be provided with heat by designing a suitable heating chamber; the effect of the heating chamber is equivalent to an air bath. The heating chamber shall consist of an air bath and a bath jacket that can hold the test cup. The air bath has a cylindrical inner side and meets the requirements shown in Figure C.1. The air bath can be flame-heated, electrically heated metal casting or resistance element heated. The air bath shall ensure that it does not deform at the specified temperature of the test. If the air bath is flame-heated or electrically heated metal casting, in actual use, the temperature of its bottom and side walls shall be consistent; and the thickness of the air bath shall be no less than 6 mm. If the air bath is flame heated, the casting shall be designed so that the combustion products of the flame cannot move up the walls of the test cup or into the test cup. If the heater is resistive, it shall be constructed so that all parts of the surface are heated evenly. ......
Document preview — GB/T 261-2021
National Standard of the People's Republic of China
Issued by: SAMR; SAC
Contents
- Foreword...3
- Introduction...6
- 1 Scope...7
- 2 Normative References...8
- 3 Terms and Definitions...8
- 4 Overview of Method...8
- 5 Reagents and Materials...9
- 6 Instrument...9
- 7 Preparation of Instrument...10
- 8 Sampling...11
- 9 Sample Processing...11
- 10 Test Steps...12
- 11 Calculations...15
- 12 Expression of Results...16
- 13 Precision...16
- 14 Test Report...17
Foreword
This Document was drafted as per the rules specified in GB/T 1.1-2020 Directives for
Standardization - Part 1: Rules for the Structure and Drafting of Standardizing Documents.
This Document replaced GB/T 261-2008 Determination of Flash Point - Pensky-Martens
Closed Cup Method. Compared with GB/T 261-2008, the major technical changes of this
Document are as follows besides the structural adjustments and editorial modifications:
a) Change the scope from "applicable to samples with a flash point higher than 40°C" to
"applicable to samples with a flash point within the range of 40°C to 370°C" (see Clause
1 of this Edition; Clause 1 of the 2008 Edition);
b) Add Step A to the scope to indicate that it is applicable to biodiesel blended fuel samples
(see Clause 1 of this Edition);
c) Add Step C to apply to BD100 biodiesel samples, including the scope, sampling, test
conditions, test steps and precision, etc. (see Clauses 1, 8, 10 and 13 of this Edition);
d) Change and add the relevant contents of some instruments; including the requirements
for the built-in pressure gauge, barometer, temperature measuring device, explosion-
proof oven, and instrument installation location of automatic instruments (see Clause 6
of this Edition; Clause 6 of the 2008 Edition);
e) Refine and supplement the sampling and sample handling contents (see Clause 8 of this
Edition; Clause 8 of the 2008 Edition);
f) Expand the atmospheric pressure range for accurate correction of flash point in the flash
point correction formula (see Clause 11 of this Edition; Clause 11 of the 2008 Edition);
g) Add The tolerance values of calibration standard substances to Table B.1 of Appendix B
(see Appendix B of this Edition);
h) In order to cooperate with the use of automatic temperature measuring devices, delete
Appendix D Thermometer Adapter in GB/T 261-2008; and add relevant content of
electronic temperature measuring equipment (see D.1 in Appendix D of this Edition).
This Document used redrafting method to modify and adopt ISO 2719:2016 Determination of
Flash Point - Pensky-Martens Closed Cup Method.
Compared with ISO 2719:2016, this Document made partial adjustment in structure; Appendix
A list the comparison table of clauses between this Document and ISO 2719:2016.
The major technical differences and causes between this Document and ISO 2719:2016 are as
follows:
Determination of Flash Point - Pensky-Martens Closed Cup
Method
Warning: People who use this Document shall have practical working experience in
regular laboratory. The use of this Document may involve certain dangerous materials,
equipment, and operations. This Document does not point out all possible safety issues.
Users are responsible for taking appropriate safety and health measures and ensuring
that the conditions stipulated by relevant national laws and regulations are met.
1 Scope
This Document specifies the method for determining the flash point of flammable liquids,
liquids with suspended particles, liquids that tend to form films on the surface under test
conditions, and other liquids using a Pensky-Martens closed cup flash point tester.
NOTE 1: Although this Document can be used for kerosene with a flash point above 40°C, the flash point
of kerosene is generally determined in accordance with GB/T 21789. The flash point of unused
lubricating oils is usually determined in accordance with GB/T 3536.
NOTE 2: This Document can be used for jet fuels with a flash point below 40°C, but the precision has
not been verified.
Warning: Since some mixtures do not obtain the flash point defined in this Document, but
obtain a significantly increased flash (not the effect of halo) and a change in the flash from
blue to yellow-brown, continued heating can cause the vapor around the test cup to burn.
This Document is applicable to samples with a flash point range of 40°C ~ 370°C. This
Document includes three test steps: Step-A, Step-B and Step-C. Step-A is applicable to distillate
fuels (including diesel, biodiesel blends, heating oil and turbine fuel), unused lubricating oils,
paints and varnishes and other homogeneous liquids not included in Step-B and Step-C. Step-
B is applicable to residual fuel oil, diluted asphalt, used lubricating oil, liquids with suspended
particles, liquids that tend to form films on the surface under the test conditions or liquids whose
viscosity is not suitable for heating under the stirring speed and heating conditions specified in
Step-A. Step-C is applicable to the determination of BD100 biodiesel samples. This Document
does not apply to water-based paints and water-based varnishes.
NOTE 3: When monitoring the lubricating oil system, in order to compare the flash points of unused
lubricating oil and used lubricating oil, Step-A can also be used to determine the flash point of used
lubricating oil, but this precision is only applicable to Step-B.
C.3.3 Ignitor
The diameter of the tip opening of the ignition tube flame jet device is about 0.7mm~0.8mm
(see Figure C.4). The tip is stainless steel, or other suitable metal material. The ignitor is
equipped with a mechanical operator. When the slide plate is in the "open" position, the tip is
lowered so that the center of the flame nozzle opening hole is located between the upper and
lower surface planes of the test cup cover and passes through a point on the center radius of the
maximum opening A (see Figure C.3).
The bead at the tip is made of suitable material, about 4mm, which is equivalent to the flame
size, and is assembled in a prominent position on the test cup cover.
C.3.4 Automatic ignition device
For automatic ignition of the flame, the diameter of the ignitor tip opening is about
0.7mm~0.8mm.
C.3.5 Stirring device
It is mounted in the center of the test cup cover (as shown in Figure C.4), with two double-
bladed metal paddles. The lower paddle surface is about 38mm away from the test cup cover;
the two blades are 8mm wide and inclined at 45°. The upper paddle surface is about 19 mm
from the test cup cover; the two blades are 8 mm wide and inclined at 45°. The two sets of
paddles are fixed on the rotating shaft of the stirrer. When viewed from the bottom of the stirrer,
the two blades of one set of paddles are at 0° and 180°; and the two blades of the other set of
paddles are at 90° and 270°.
The rotating shaft of the stirrer can be connected to the motor by a transmission flexible shaft
or a suitable pulley block structure.
C.4 Heating chamber and bath jacket
The test cup shall be provided with heat by designing a suitable heating chamber; the effect of
the heating chamber is equivalent to an air bath. The heating chamber shall consist of an air
bath and a bath jacket that can hold the test cup.
The air bath has a cylindrical inner side and meets the requirements shown in Figure C.1. The
air bath can be flame-heated, electrically heated metal casting or resistance element heated. The
air bath shall ensure that it does not deform at the specified temperature of the test.
If the air bath is flame-heated or electrically heated metal casting, in actual use, the temperature
of its bottom and side walls shall be consistent; and the thickness of the air bath shall be no less
than 6 mm. If the air bath is flame heated, the casting shall be designed so that the combustion
products of the flame cannot move up the walls of the test cup or into the test cup.
If the heater is resistive, it shall be constructed so that all parts of the surface are heated evenly.
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.
Referenced standards
Cited by
- GB 11118-2026Hydraulic fluids (L-HL, L-HM, L-HV, L-HS and L-HG type)
- GB 23971-2025Organic heat transfer fluids
- GB 46037-2025Non-traditional machines - Safety technical requirements
- GB 6537-2025No.3 jet fuel
- GB 6944-2025Classification and code of dangerous goods
- GB/T 24747-2023Safety technology conditions for organic heat transfer fluid
How to Buy GB/T 261-2021
- 1Add to cart. Click the "Buy GB/T 261-2021" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 46830.2-2026 — Plastics - Liquid crystal polymer (LCP) moulding and extrusion materials - Part 2: Preparation of test specimens and determination of properties
GB/T 28593-2026 — Sand and dust weather warning levels
GB/T 19608.3-2026 — Classification of special environmental conditions - Part 3: Plateau
Secure payment via Stripe
Payments accepted
GB/T 261-2021
$220.00