YY/T 0313-2014Medical polymer products—Requirement for package and information supplied by manufacturer (English PDF)

医用高分子产品 包装和制造商提供信息的要求

Preview PDF

This is a limited preview

Buy now to download the full PDF (23 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0313-2014 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

June 17, 2014

Implementation date

July 1, 2015

Scope

YY/T 0313-2014 is the English-translated version of 医用高分子产品 包装和制造商提供信息的要求.

YY/T 0313-2014 is the Chinese pharmaceutical industry standard that sets the requirements for the package of a medical polymer product and for the information the manufacturer supplies with it.

It fixes what the primary and the outer package must carry, among it the identification of the product and of the manufacturer, the batch number, the date of manufacture or of expiry, the state of sterility and the method of sterilization where it applies, and it sets out the instructions and the warnings that must accompany the product. Issued on 17 June 2014, it has applied since 1 July 2015.

Found what you need?

Document preview — YY/T 0313-2014

National Standard of the People's Republic of China

Replacing
YY/T 0313-1998

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions6
  • 4 Product classification8
  • 5 Package requirements8
  • 6 Requirements for the information supplied by the manufacturer12
  • Annex A (Informative) Guidelines for the information of medical devices required to be supplied for meeting the requirements of EU Council Directive 93/42/EEC16
  • References23

Foreword

This Standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This Standard replaces the YY/T 0313-1998 Package, label, transport and storage for medical polymer products. Compared with YY/T 0313-1998, the major changes in this Standard are as follows.

- MODIFY the standard name;

- MODIFY the "Scope" applicable to this Standard (SEE Chapter 1 in this edition);

- MODIFY the "Normative references" (SEE Chapter 2 in this edition);

- MODIFY partial contents of the "Terms and definitions" (SEE Sections 3.7, 3.8 and 3.17 in this edition and Sections 3.8, 3.17 and 3.18 in 1998 edition);

- RECLASSIFY the products; DELETE the "Disinfection products" and relevant contents (Sections 4.1 and 5.2 in 1998 edition);

- MODIFY the requirements for sterile package as "in line with GB/T 19633.1 and relevant standards" (SEE Section 5.2.3 in this edition and Section 5.3.3 in 1998 edition);

- Since the categories of corrugated boxes vary from three to two, due to GB 6543 updated in normative references, this Standard has also been modified accordingly (SEE Section 5.3.3 in this edition and Section 5.4.3 in 1998 edition);

- ADD the requirements for the information supplied by manufacturer (SEE Chapter 6 in this edition);

- The symbols for product package have been partially modified (SEE Section 6.3 in this edition);

- DELETE the Chapter 7 "Transport and storage" (Chapter 7 in 1998 edition);

- MODIFY the Annex A as "Guidelines for the information of medical devices required to be supplied for meeting the requirements of EU Council Directive 93/42/EEC" (SEE Annex A in this edition and Annex A in 1998 edition); and - DELETE Annex B (Annex B in 1998 edition).

Some of the elements of this document may be the subject of patent rights.

The issuing authorities of this document shall not be held responsible for identifying any or all such patent rights.

1 Scope

This Standard specifies the requirement for package and information supplied by manufacturer of medical polymer products.

Note: The provisions of national regulations and product standards take precedence over this Standard.

systems

2 Normative references

The following documents are essential to the application of this document.

For dated references, only the editions with the dates indicated are applicable to this document. For undated references, only the latest editions (including all the amendments) are applicable to this document.

GB/T 3102 (all parts) Quantities and units

GB/T 4892 Dimensions of rigid rectangular packages - Transport packages

GB/T 6543 Single and double corrugated boxes for transport packages

GB/T 7408 Data elements and interchange formats - Information interchange - Representation of dates and times

GB/T 19633.1 1 Packaging for terminally sterilized medical devices - Part

3 Terms and definitions

The terms and definitions defined in GB/T 19633.1 and the following ones are applicable to this document.

3.1 Contents

The contents refer to a generic term for product(s), divider(s), instructions and / or inner package and other articles in a package.

3.6 Transport package (outer package)

It refers to the packaging unit applicable to product transport.

3.7 Shelf package

It refers to the protective packaging unit that is placed on a shelf to ensure that the product meets the intended performance within the service life, in compliance with the manufacturer's storage conditions.

3.8 Divider

It refers to a generic term for the partitions, trays, brackets, cannulas and other protective packaging components that are not mutually extruded among the components with sufficient intensity designed to be a set of products or among the products.

3.12 Sterile package

It refers to the package that remains sterile inside. Once it ruptures, no internal sterility is guaranteed.

3.13 Double-wall package

A layer of sealed package is supplemented outside the sealed package, which can be structurally unsealed in layers.

3.14 Single-service package

It refers to the package in which the quantity of the contents (such as liquid, particles, etc.) is for single use only.

3.15 Multi-service package

It refers to the package in which the quantity of the contents (such as liquid, particles, etc.) can be consumed for multiple times.

4 Product classification

4.1 When designing the product package, TAKE account of the product's sanitary requirements first, then the requirements for the package combined with the product's specific circumstances and physical properties. Therefore, it is necessary to classify the products.

4.2 The products are classified as general products and non-sterile products according to the sanitary requirements.

4.3 According to the physical properties, the products are classified as.

5 Package requirements

5.2.4 For sterile supply of powder, liquid or ointment material, a single-service package shall be used. If the single-service package cannot be used as a sterile package unit, a layer of unit package shall be supplemented as a sterile package.

5.7.1 For the spare parts and raw materials produced or cleaned in a clean environment and provided for other manufacturers, the double-wall packages shall be used as their minimum packages.

5.7.2 For the products whose quality will be affected by moisture, their

packages shall be kept away from moisture. Moisture absorption agent shall be added if necessary.

5.1 General requirements

5.1.1 The product packages shall be applicable to the storage and transport processes. Ensure that the product quality is not affected during normal storage and transport.

5.1.3 The product package shall be convenient for the use of the product.

5.1.4 The size of the rectangular package of the product shall be prioritized with the size specified in GB/T 4892.

5.3 Products in Class

A 5.3.1 Each product in Class A shall have a minimum package and a shelf package.

5.3.2 If the contents are required to be kept clean before use, a sealed package shall be used as their minimum package.

5.4 Products in Class B

5.4.1 Each product in Class B shall have a minimum package or a unit package and a shelf package.

5.4.2 If the contents are required to be kept clean before use, a sealed package shall be used as their minimum package or unit package.

5.4.3 When the unit package is relatively small, a rigid package shall be used as a multi-unit package.

5.5 Products in Class C

5.5.1 Each product in Class C shall have a minimum package or a unit package, a multi-unit package and / or a shelf package. When the product's unit package is relatively large, a multi-unit package may not be used.

5.5.2 For a complete set of products consisting of a plurality of separate

components, its unit package shall be designed with a divider to prevent mutual extrusion.

5.6 Products in Class D

5.6.1 Each product in Class D shall have at least a primary package (a single-service package or a multi-service package), a multi-unit package and / or a shelf package. When the primary package is relatively large, a multi-unit package may not be used.

5.6.2 The primary package shall be prioritized with the design of the convenient-for-use package.

6 Requirements for the information supplied by the manufacturer

6.1 Product identification information

6.1.1 Product category

When the category of medical devices belonging to the product needs to be explained in the information supplied, priority shall be given to the denomination in YY/T 0468.

6.2 Product service information

6.2.1 General requirements

Any way of supplying information with the product shall take account of the prospective users, service conditions, the safety and effectiveness of the use of single device, etc.

6.2.2 Special requirements
6.2.2.1 Applicability

The service information shall be applicable to the products in packages.

When some products are supplied in multiple forms (such as sterile and non-sterile) or of multiple specifications, information shall be separately supplied for various forms or specifications, respectively.

6.2.2.3 Readability

The information expected to be visually recognized shall be readable, taking account of special dimensions and service conditions of particular products for corresponding adjustment, if necessary.

6.2.2.5 Safety

As long as it is actually needed, the media for supplying information shall be guaranteed not to be destroyed or deliberately altered by the personnel other than the manufacturer, whether for malicious purposes or not.

6.2.2.6 Normalization

The dates shall be expressed in the format of YYYY-MM-DD, YYYY-MM or YYYY given in GB/T 7408.

6.2.2.7 Changes in information supplied

For any changes in the information that has been supplied for the user, PASS on the changes clearly to the user if it is important for patient safety.

6.3 Symbols on the product packages

6.3.1 Unit package or primary package
6.3.2 Symbols on the shelf packages

The following symbols shall be generally marked on the shelf packages. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.

Referenced standards

Cited by

Similar standards

Editions of YY/T 0313

EditionTitleRevisionStatus
YY/T 0313-2014Medical polymer products—Requirement for package and information supplied by manufacturercurrent editionCurrent
YY/T 0313-1998Medical polymer products—Requirement for package and information supplied by manufacturerprevious editionSuperseded

This page sells the current edition, YY/T 0313-2014. Earlier editions are listed for reference only.

How to Download YY/T 0313-2014

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
23 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0313-2014

$95.00