YY/T 0334-2022General specification for surgical implants made of silicone elastomer (English PDF)
硅橡胶外科植入物通用要求
This is a limited preview
Buy now to download the full PDF (19 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0334-2022 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
August 17, 2022
Implementation date
September 1, 2023
Scope
YY/T 0334-2022 is the English-translated version of 硅橡胶外科植入物通用要求.
YY/T 0334-2022 is the Chinese pharmaceutical industry standard that sets the general requirements for surgical implants made of silicone elastomer.
It covers the appearance of the implant, its chemical and biological requirements, sterility, and packaging and marking, with normative appendices giving the test method for trace elements, the preparation of the test solution and the determination of the evaporation residue.
It is used to test and evaluate implants made of silicone elastomer; it does not fix the physical and mechanical properties of any particular implant, which belong to the standard for that product.
It replaces YY 0334-2002, was issued on 17 August 2022 and has applied since 1 September 2023.
Found what you need?
Document preview — YY/T 0334-2022
National Standard of the People's Republic of China
- Replacing
- YY 0334-2002
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction5
- 1 Scope6
- 2 Normative References6
- 3 Terms and Definitions6
- 4 Appearance7
- 5 Requirements7
- 6 Sterility9
- 7 Packaging and Marking9
- Appendix A (normative) Test Method for Trace Elements10
- Appendix B (normative) Preparation Method for Test Solution11
- Appendix C (normative) Test Method for Evaporation Residue12
- Appendix D (normative) Test Method for pH13
- Appendix E (normative) Test Method for Peroxide14
- Appendix F (normative) Test Method for Reducing Substances (easily oxidized)16
- Appendix G (normative) Test Method for Ultraviolet Absorption18
- Bibliography19
1 Scope
This document specifies general requirements for the chemical and biological properties, sterility, packaging and marking of surgical implants made of silicone elastomer.
This document is applicable to the test and evaluation of surgical implants made of silicone elastomer.
NOTE 1.this document does not specify specific indicators of the physical and mechanical properties of the implants.
NOTE 2.for surgical implants made of silicone elastomer that have added color masterbatch and developer, etc. in order to increase the functionality of the products, the manufacturer may refer to this document based on risk assessment.
2 Normative References
The contents of the following documents constitute indispensable clauses of this document through normative references in the text. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB/T 601 Chemical Reagent - Preparations of Standard Volumetric Solutions
GB/T 14233.1 Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use - Part 1: Chemical Analysis Methods
GB/T 16886.1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process
YY/T 0313 Medical Polymer Products - Requirement for Package and Information Supplied by Manufacturer Pharmacopoeia of the People's Republic of China (2020) Volume IV
3 Terms and Definitions
The following terms and definitions are applicable to this document.
3.1 Heat-vulcanization; heat-curing
Through heating, the silicone compound (3.3) is cross-linked into an elastic material (silicone elastomer).
3.2 Silicone elastomer
A synthetic elastomer obtained by cross-linking silicon-reinforced silicone polymer chains consisting essentially of repeating diorganosiloxane units.
3.3 Silicone compound
A material composed of polysiloxane mixed with suitable fillers and cross-linking agents, without cross-linking.
3.4 Silicone, polysiloxane
A polymer whose main chain is alternately linked by silicon and oxygen atoms, and whose side chains contain organic groups.
3.5 Post-curing; secondary curing
After vulcanization, a process performed at a certain temperature to enhance the physical properties of the material and / or remove excess decomposition products.
4 Appearance
For transparent or semitransparent materials, under 10 times magnification, visually inspect the outer surface of the implants made of silicone elastomer, and there shall be no impurities or other foreign particles.
5 Requirements
5.1 General
The implants made of silicone elastomer shall be tested as supplied. If the implants made of silicone elastomer have two or more separable different types of silicone elastomer, each type of silicone elastomer shall be respectively sampled and tested.
5.2 Biological Evaluation
Carry out biological evaluation of the implants made of silicone elastomer in accordance with
GB/T 16886.1.The evaluation results shall have no unacceptable biological hazards.
5.3 Material Requirements
5.3.3 Identification
In accordance with the 0402 infrared spectrophotometry of the Pharmacopoeia of the People's Republic of China (2020) Volume IV, carry out the test. The infrared spectrum of the implants made of silicone elastomer shall be consistent with the standard control spectrum provided by the manufacturer.
5.4 Requirements for Soluble Substances
5.4.1 Solution appearance
In accordance with B.3, take the filtrate. In accordance with the 0901 solution color inspection method and the 0902 clarity inspection method in Pharmacopoeia of the People's Republic of China (2020) Volume IV, carry out the test. The test solution shall be colorless and clear.
5.4.3 pH
When tested in accordance with Appendix D, the difference in pH value between the test solution and the blank solution shall not be greater than 1.5.
5.4.4 Catalyst residue
For the peroxide type of implants made of silicone elastomer, when tested in accordance with Appendix E, the difference in the volume of the sodium thiosulfate standard titration solution [c (Na2S2O3) = 0.01 mol/L] consumed by the tested solution and the blank solution shall not exceed 0.2 mL.
5.4.5 Reducing substances (easily oxidized)
When tested in accordance with Appendix F, the difference in the volume of the potassium permanganate standard titration solution [c(1/5KMnO4) = 0.01 mol/L] consumed by the test solution and the blank solution shall not exceed 3.0 mL.
5.4.6 Ultraviolet absorption
When tested in accordance with Appendix G, the absorption value in the wavelength range of 220 nm ~ 340 nm shall not exceed 0.4.
5.4.7 Heavy metals
In accordance with B.3, take the filtrate. When tested in accordance with GB/T 14233.1, the heavy metal contents shall not exceed 1 mg/L.
6 Sterility
The surgical implants made of silicone elastomer shall undergo a validated sterilization process.
If ethylene oxide is used for sterilization, the residual amount of ethylene oxide for the sterilization of the implants made of silicone elastomer shall be controlled.
7 Packaging and Marking
The packaging and marking of the surgical implants made of silicone elastomer shall comply with the relevant stipulations of YY/T 0313.Permanent surgical implants shall be uniquely and traceably identified.
Appendix A (normative) Test Method for Trace Elements
A.1 Principle
The sample is digested in a sealed mode through nitric acid and hydrofluoric acid to prepare a specimen solution. Use atomic absorption spectrophotometry or inductively coupled plasma atomic emission spectrometry or inductively coupled plasma mass spectrometry to directly determine the contents of arsenic, cadmium, lead, chromium, iron and platinum.
A.2 Preparation of Reagents and Solutions
A.2.1 Standard solutions of arsenic, cadmium lead, chromium, iron and platinum. in accordance with the above-mentioned conventional methods, respectively prepare stock solutions containing 1 mg/mL of arsenic, cadmium, lead, chromium, iron and platinum. Before use, use nitric acid solution with a volume fraction of 1% to dilute to the required concentration. A.2.2 Nitric acid. superior-grade purity.
A.3 Preparation of Test Solution
Accurately weigh-take 1.0 g of the sample, place it in a polytetrafluoroethylene cup, add 5 mL of nitric acid and 5 mL of hydrofluoric acid, cover and seal it; tighten the stainless-steel sheath and heat it in the oven at 160 C for 2 hours; take it out and let it cool to room temperature.
Take out the polytetrafluoroethylene cup, and heat it uncovered on an electric hot plate at low temperature, until it approaches almost dryness. Along the wall of the cup, add 0.5 mL of nitric acid, evaporate again to nearly dryness. Add 5 mL of nitric acid with a volume fraction of 1%, heat it to nearly boiling, and let it cool. Use a polyethylene tube to absorb the solution, transfer the solution into a 25 mL volumetric flask. Use nitric acid with a volume fraction of 1% to wash the polytetrafluorethylene cup several times, put the washing solution into a volumetric flask, and use nitric acid with a volume fraction of 1% to dilute it to the scale.
The microwave digestion method can also be used to prepare the test solution.
A.4 Test Method
In accordance with the 0406 atomic absorption spectrophotometry or the 0411 inductively coupled plasma atomic emission spectrometry or the 0412 inductively coupled plasma mass spectrometry in Pharmacopoeia of the People's Republic of China (2020) Volume IV, determine the digested solution.
Appendix B (normative) Preparation Method for Test Solution
B.1 Reagents
B.1.1 Distilled or deionized water. B.1.2 n-hexane (analytically pure).
B.2 Instruments
B.2.1 Reflux device made of borosilicate glass, with a capacity of 250 mL ~ 500 mL.
B.2.2 Vacuum filtration device (including borosilicate sand core filtration device, 0.45 m aqueous filter membrane and vacuum filtration pump). B.2.3 Analytical balance (accurate to 0.01 g).
B.3 Distilled Water Extraction
Cut the sample into pieces smaller than 10 mm, use a balance to weigh 12.50 g 0.25 g of sample, accurate to 0.01 g, and record the sample mass. Place the sample in a reflux device, add 250 mL of distilled water or deionized water, and heat to reflux for 5 hours.
After stopping heating, cool the reflux device and immediately use a vacuum filtration device to filter it. If the filtrate needs to be placed, place it in a covered borosilicate container.
B.4 N-hexane Extraction
Cut the sample into pieces smaller than 5 mm, use a balance to weigh 5.00 g 0.50 g of sample, accurate to 0.01 g, and record the sample mass. Place the sample in a reflux device, add 250 mL of n-hexane, and heat to reflux for 4 hours.
After stopping heating, cool the reflux device and immediately use a vacuum filtration device to filter it. If the filtrate needs to be placed, place it in a covered borosilicate container.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 601Chemical Reagent - Preparations of Standard Volumetric Solutions
- GB/T 14233.1Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- YY/T 0313Medical polymer products—Requirement for package and information supplied by manufacturer
Similar standards
Editions of YY/T 0334
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0334-2022 | General specification for surgical implants made of silicone elastomer | current edition | Current |
| YY 0334-2002 | General specification for surgical implants made of silicone elastomer | previous edition | Superseded |
This page sells the current edition, YY/T 0334-2022. Earlier editions are listed for reference only.
How to Download YY/T 0334-2022
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14233.1-2008 — Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use — Part 1: Chemical Analysis Methods
GB/T 14233.1-2022 — Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods