YY/T 1814-2022Implants for surgery-Synthetic non-absorbable patch-Hernia repair patch (English PDF)
外科植入物 合成不可吸收补片 疝修补补片
This is a limited preview
Buy now to download the full PDF (20 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1814-2022 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
May 18, 2022
Implementation date
June 1, 2023
Scope
YY/T 1814-2022 is the English-translated version of 外科植入物 合成不可吸收补片 疝修补补片.
YY/T 1814-2022 is the Chinese pharmaceutical industry standard for the synthetic non-absorbable patch used to repair a hernia, the mesh implanted to reinforce the abdominal wall.
It gives the terms and definitions, the requirements and the test methods, and the rules for packaging, marks, transport and storage.
The requirements cover the polymer the mesh is made of and its identification, the structure and the areal density, the thickness and the pore size, the tensile strength and the burst strength, the suture retention strength, the sterility and the biological evaluation, together with what the instructions must tell the surgeon. Issued on 18 May 2022, it has applied since 1 June 2023.
Found what you need?
Document preview — YY/T 1814-2022
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Technical requirements5
- 5 Inspection methods7
- 6 Marking requirements11
- 7 Packaging, transport, and storage12
- Bibliography13
1 Scope
This document specifies the terms and definitions, requirements, test methods, packaging, marks, transportation, and storage requirements for hernia repair patch products.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 3917.2, Textiles -- Tear properties of fabrics -- Part 2: Determination of tear force of trouser-shaped test specimens (Single tear method)
GB/T 3923.1, Textiles -- Tensile properties of fabrics -- Part 1: Determination of maximum force and elongation at maximum force using the strip method
GB/T 14233.1-2008, Test methods for infusion, transfusion, injection equipment for medical use -- Part 1: Chemical analysis methods
GB/T 14233.2-2005, Infusion, transfusion, injection equipment for medical use -- Part 2: Biological test methods
GB/T 16886 (all parts), Biological evaluation of medical devices
YY/T 0313, Medical polymer products -- Requirement for package and information supplied by manufacturer
YY/T 0500-2021, Cardiovascular implants and extracorporeal systems-vascular prostheses-tubular vascular grafts and vascular patches Pharmacopoeia of the People's Republic of China (Edition 2020)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1 Hernia
Any organ or tissue that has departed from its original location and entered another site by passing through normal or abnormal - whether congenital or acquired - weak points, defects, or orifices within the human body.
3.2 Hernia repair
This method utilizes synthetic materials as patches to reinforce weak points, defects, or gaps where a hernia has formed. It overcomes the disruption to normal anatomical structures inherent in traditional surgery (i.e., suture-only repair without the use of a patch) and features distinct anatomical layering.
3.3 Non-absorbable material
Materials that are not absorbed by the human body after implantation (currently, those commonly used in clinical practice include polypropylene, polyester, polyvinylidene fluoride, etc.).
4 Technical requirements
4.1 Physical properties
4.1.1 Appearance
It shall be free of stains or foreign matter, as well as tears, abrasions, and particulate matter.
4.1.2 Thickness
It shall comply with the manufacturer's specifications.
4.1.3 Tensile strength
The longitudinal and transverse tensile strengths shall be >= 16 N/cm.
4.1.4 Tensile elongation
The tensile elongation at a tensile strength of 16 N/cm - both in the longitudinal and transverse directions - shall comply with the manufacturer's specifications.
4.1.5 Seam strength
The seam strength shall be >= 16 N.
4.1.6 Bursting strength
The bursting strength shall be >= 20 kPa.
4.1.7 Tear strength
The tear strength shall be >= 16 N.
4.1.8 Pore size
The manufacturer shall describe the pore size of the hernia repair patch. The pore size of the hernia repair patch shall fall within the range specified by the manufacturer.
4.1.9 Connection force
For multi-layer hernia repair patches, the manufacturer shall clearly specify the method of connection, and the connection strength shall be >= 16 N.
4.2 Chemical properties
4.2.1 General provisions
The chemical properties of the hernia repair patch shall be subjected to the following tests in the supplied state. If substances that may affect the pass/fail determination of the test results specified in 4.2.3, 4.2.4, 4.2.5, and 4.2.6 have been introduced during the product manufacturing process, such test items shall be specifically identified, subject to additional requirements, and tested in accordance with the provisions of 4.2.8.
4.2.2 pH
The difference between the pH value of the hernia repair patch test solution and that of the blank solution shall not exceed 1.5.
4.2.3 Trace elements and total heavy metals
The content of nickel, cobalt, copper, arsenic, mercury, lead, chromium, cadmium, and other elements introduced during the manufacturing process - as well as the total heavy metal content - in hernia repair patches shall be less than the specified limit values (µg/g).
NOTE. The maximum limits for trace elements may be established by [2-6] in Bibliography.
4.2.4 Reducing substances
The difference between the volumes of 0.002 mol/L potassium permanganate solution consumed by the hernia repair patch test solution and the blank solution shall not exceed 2.0 mL.
4.2.5 UV absorbance
The absorbance of the hernia repair patch test solution shall not exceed 0.1 within the wavelength range of 220 nm~340 nm.
4.2.6 Evaporation residue
The residue in 50 mL of the test solution shall not exceed 5 mg.
4.2.7 Infrared spectroscopy
The manufacturer shall provide a standard infrared spectrum corresponding to the hernia repair patch and clearly identify the positions of the characteristic peaks. The deviation of the measured absorption band wavenumbers from the specified wavenumbers shall be less than 0.5%.
4.2.8 Supplementary test items
Manufacturers shall provide a detailed description of any unique chemical substances utilized during the product manufacturing process (e.g., polymerization aids, coatings, pharmaceutical ingredients, etc., that would significantly impact the test results of the aforementioned inspection items), and shall additionally specify the corresponding inspection parameters (including sample preparation methods, test methods, result calculation procedures, etc.).
4.2.9 Ethylene oxide residue
For hernia repair patches sterilized with ethylene oxide, the residual ethylene oxide content shall not exceed 10 µg/g.
4.3 Biological properties
4.3.1 General principles
It shall comply with the relevant requirements of GB/T 16886.1.
4.3.2 Sterility test
It shall comply with sterility requirements.
4.3.3 Bacterial endotoxin test
It shall be less than 0.5 EU/mL.
4.3.4 Biological evaluation
Biological evaluation is conducted in accordance with the principles of GB/T 16886.1.
5 Inspection methods
5.1 Physical test
5.1.1 Appearance
Visual inspection. The results shall comply with the requirements of 4.1.1.
5.1.2 Thickness
Measure using general-purpose measuring instruments. The results shall comply with the requirements of 4.1.2.
5.1.3 Tensile strength
The procedure shall be carried out in accordance with the method specified in GB/T 3923.1.The results shall comply with the requirements of 4.1.3.
5.1.4 Tensile elongation
The procedure shall be carried out in accordance with the method specified in GB/T 3923.1.The results shall comply with the requirements of 4.1.4.
5.1.5 Seam strength
A suture of sufficient tensile strength is passed through the hernia repair patch at a point 5 mm~10 mm from one edge, forming a semi-loop. The suture is then withdrawn at a speed of 50 mm/min~200 mm/min. The type of suture used shall be recorded. The magnitude of the tensile force required to either pull the suture out of the patch or cause damage to the patch constitutes the patch's suture strength. The results shall comply with the requirements specified in 4.1.5.
5.1.6 Bursting strength
The procedure shall be conducted in accordance with the method specified in A.5.3.5 of YY/T 0500-2021.The results shall comply with the requirements of 4.1.6.
5.1.7 Tear strength
The procedure shall be carried out in accordance with the method specified in GB/T 3917.2.The results shall comply with the requirements of 4.1.7.
5.1.8 Pore size
The procedure shall be conducted in accordance with the method specified in A.5.2.1 of YY/T 0500-2021.The results shall comply with the requirements of 4.1.8.
5.1.9 Connection force
Secure the connecting sections of the multilayer hernia repair system separately within the grips of a materials testing machine. Apply tension at a speed of 100 mm/min. The tensile force required to cause the connection point to rupture constitutes the connection strength of the multilayer hernia repair system. The results shall comply with the requirements specified in 4.1.9.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 20 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 3917.2Textile -- Tear properties of fabrics -- Part 2: Determination of tear force of trouser-shaped test specimens (Single tear method)
- GB/T 3923.1Textiles -- Tensile properties of fabrics -- Part 1: Determination of maximum force and elongation at maximum force using the strip method
- GB/T 14233.1-2008Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 14233.2-2005Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods
- YY/T 0313Medical polymer products—Requirement for package and information supplied by manufacturer
- YY/T 0500-2021Cardiovascular implants and extracorporeal systems. vascular prostheses. tubular vascular grafts and vascular patches
GB/T 16886
Similar standards
How to Download YY/T 1814-2022
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14233.1-2008 — Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use — Part 1: Chemical Analysis Methods
GB/T 14233.2-2005 — Test methods for infusion, transfusion, injection equipment for medical use-Part 2: Biological test methods
GB/T 16886.1-2022 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process