YY/T 1698-2020Medical devices for human assisted reproductive technology in vitro—Assisted reproduction aspiration needles (English PDF)

人类体外辅助生殖技术用医疗器械 辅助生殖穿刺取卵针

Preview PDF

This is a limited preview

Buy now to download the full PDF (21 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1698-2020 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

June 30, 2020

Implementation date

June 1, 2021

Scope

YY/T 1698-2020 is the English-translated version of 人类体外辅助生殖技术用医疗器械 辅助生殖穿刺取卵针.

YY/T 1698-2020 is the Chinese pharmaceutical industry standard for the oocyte retrieval needle used in assisted reproduction, the needle passed through the vaginal wall under ultrasound to aspirate follicles.

It sets the specifications and models, the composition, the materials and the design attributes, the requirements and the test methods, the sterilization requirements for a sterile product, and the packaging, marking and instructions, covering the point and the patency of the needle, its echogenicity, the strength of the joints, the connection to the aspiration line and the effect of the materials on gametes. Issued on 30 June 2020, it has applied since 1 June 2021.

Found what you need?

Document preview — YY/T 1698-2020

National Standard of the People's Republic of China

Classification
C 30

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Specifications, models, composition and materials
  • 5 Design attributes
  • 6 Requirements and test methods

Foreword

This standard was drafted in accordance with the rules given in GB/T 1.1-2009.

1 Scope

This standard specifies the specifications, models, composition, materials, and design of the needles for assisted reproductive puncture and egg retrieval for medical devices used in human in vitro assisted reproductive technology.

Design attributes, requirements and test methods, sterilization requirements for sterile products, packaging, markings and instructions.

This standard is applicable to the medical device assisted reproductive puncture and egg retrieval needle for human in vitro assisted reproductive technology (hereinafter referred to as the puncture egg retrieval needle).

Under the guidance of ultrasound, the ovaries are punctured and the follicular fluid is sucked to obtain oocytes.

2 Normative references

The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated reference documents, the latest version (including all amendments) is applicable to this document.

GB/T 191 Graphical signs for packaging, storage and transportation

GB/T 1962.1 Syringes, injection needles and other medical devices 6% (Ruhr) tapered joints Part 1: General requirements

GB/T 1962.2 Syringes, injection needles and other medical devices 6% (Luer) tapered joints Part 2: Locking joints

GB/T 14233.1-2008 Medical transfusion, blood transfusion, injection equipment inspection methods Part 1: Chemical analysis methods

GB/T 14233.2 Test methods for medical infusion, blood transfusion and injection equipment Part 2: Biological test methods

GB/T 15812.1-2005 Non-intravascular catheters Part 1: General performance test methods

GB/T 16886.1 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing in the Process of Risk Management

GB/T 16886.4 Biological Evaluation of Medical Devices Part 4: Test Selection for Interaction with Blood

GB/T 16886.5 Biological Evaluation of Medical Devices Part 5: In Vitro Cytotoxicity Test

GB/T 16886.10 Biological Evaluation of Medical Devices Part 10: Irritation and Skin Sensitization Test

GB/T 16886.11 Biological Evaluation of Medical Devices Part 11: Systemic Toxicity Test

GB 18278.1 Sterilization of Medical and Health Products Moisture Heat Part 1: Development, Validation and Routine Control of the Sterilization Process of Medical Devices Claim

GB 18279.1 Sterilization of Medical and Healthcare Products Ethylene Oxide Part 1: Development, Validation and Routine Control of the Sterilization Process of Medical Devices System requirements

GB/T 18279.2 Sterilization of Medical and Health Care Products Ethylene Oxide Part 2: GB 18279.1 Application Guide

GB 18280.1 Sterilization radiation of medical and health care products Part 1: Development, validation and routine control of the sterilization process of medical devices Claim

GB 18280.2 Sterilization Radiation of Medical and Health Care Products Part 2: Establishing the Sterilization Dosage

GB/T 18280.3 Medical and health care products sterilization radiation Part 3: Dose measurement guide

GB/T 18457-2015 Stainless steel needle tube for manufacturing medical equipment

GB/T 1974 The characteristics of sterilization factors for medical and health care products and the development, validation and routine control of the sterilization process of medical devices General requirements

YY/T 0313 Medical polymer product packaging and information provided by the manufacturer

YY/T 04666.1 Medical devices used for medical device labeling, marking and information symbols. Part 1: General requirements

YY/T 0995 Terms and Definitions of Medical Devices for Human Assisted Reproductive Technology

YY/T 1434 In vitro mouse embryo test of medical devices for human in vitro assisted reproductive technology

3 Terms and definitions

The following terms and definitions defined in YY/T 0995 apply to this document.

3.1 Puncture and egg retrieval needle

In the process of in vitro fertilization-embryo transfer and its derivative technology, it is used to puncture follicles in the ovary to obtain oocytes. instrument.

4 Specifications, models, composition and materials

4.1 Specifications

The manufacturer shall describe in detail the specifications and models of each product in the same registration unit, generally divided into single-chamber, double-chamber and two-section tube types. Single cavity take The egg needle is composed of a puncture needle, a puncture needle protective cover, a handle, a suction catheter, a catheter sleeve, a silicone plug, a vacuum catheter and an inner conical connector. Of which double cavity The egg retrieval needle also includes a flushing catheter. The two-section tube type egg retrieval needle is different in the main puncture needle, which is divided into the front section and the back section.

4.2 Product composition

The specific names of the various components of the puncture needle product are shown in Figure 1 to Figure 3, and the names and functions of the accessories are shown in Table 1. Description.

4.3 Materials

4.3.1 Raw materials

The manufacturer shall evaluate the raw materials that make up the parts of the product.

4.3.2 Containers and packaging

The safety of packaging materials should be guaranteed.

5 Design attributes

In order to achieve the expected performance requirements, the design attributes should take into account the damage to the patient during the puncture process when the product is used, and the negative pressure suction Potential damage to oocytes may be.

6 Requirements and test methods

6.1 Physical properties

6.1.1 Overall
6.1.1.1 Dimensions

The manufacturer should consider what is applicable in the following dimensions.

a) Effective length of puncture needle;

b) The inner and outer diameter of the puncture needle;

c) The length of the suction catheter;

d) The length of the vacuum duct;

e) Length of irrigation catheter;

f) The thickness of the catheter wall. The manufacturer shall give the nominal size and tolerance.

6.1.1.2 Vacuum leakage

Remove the protective cover of the puncture and egg retrieval needle, connect it to the hydraulic system, and fill the system with de-bubble distilled water or deionized water at (23±2)°C.

Exhaust air, occlude the test sample, apply the negative pressure specified by the company, and keep the system stable for 120s. Keep the system under negative pressure 120s. During this period, there should be no leakage of the entire assembly (that is, the formation of one or more bubbles) and no droplets should occur.

6.1.1.3 Flow

According to GB/T 15812.1-2005 Appendix E method for testing, the amount of liquid passed in 60s is not less than 2mL.

6.1.1.4 Connection strength

The connection between the pipe and the handle, and the connection between the pipe and the conical joint should be able to withstand a 3N axial static tension without separation for 5 seconds.

6.1.2 Puncture needle
6.1.2.1 Appearance

Observed with normal or corrected vision, the outer surface of the needle tube should be smooth and free of defects.

6.1.2.2 Cleanliness

Observe with normal or corrected vision, the outer surface of the needle tube should be free of impurities produced during metal processing.

6.1.2.3 Rigidity

When testing according to GB/T 18457-2015, the deflection value of the needle tube should meet the requirements.

6.1.2.4 Toughness

When testing according to GB/T 18457-2015, the needle tube shall not be broken.

6.1.2.5 Corrosion resistance

When testing in accordance with GB/T 18457-2015 Appendix F, there should be no traces of corrosion on the part where the needle tube is immersed.

6.1.2.6 Ultrasonic detectable

The puncture needle is tested by the applicable model, and the marked position under the B-ultrasound is clearly marked.

6.1.2.7 Protective sleeve for puncture needle

Observe with normal or corrected vision. The surface should be smooth, free of black spots and impurities, and the protective cover should not fall off naturally and be easy to remove.

6.1.3 Handle
6.1.3.1 Appearance

Observe with normal or corrected vision, the surface should be smooth, free of impurities, and there should be no obvious deformation.

6.1.3.2 Positioning

The handle must have an obvious mark corresponding to the direction of the needle tip blade surface, and the mark should be clear and intuitive.

6.1.4 Piping system

Observe with normal or corrected vision. The surface should be smooth, free of black spots and impurities, and the entire pipeline should be free of creases.

6.1.5 Silicone plug
6.1.5.1 Appearance

Observed with normal or corrected vision, the surface should be smooth, free of black spots, and free of impurities, and there should be no obvious deformation.

6.1.5.2 Egg protection measures

When the oocytes enter the constant temperature test tube through the suction catheter, in order to prevent oocyte damage at the end of the suction catheter, the suction catheter should be The end or the silicone plug is equipped with egg protection measures, such as the flaring of the end of the suction catheter or the silicone plug into the test tube.

6.1.6 Within

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.

Referenced standards

Normative references

YY/T 04666.1

Similar standards

How to Download YY/T 1698-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
21 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1698-2020

$175.00