YY/T 1718-2020Medical devices for human assisted reproductive technology in vitro—Embryo transfer catheters (English PDF)
人类体外辅助生殖技术用医疗器械 胚胎移植导管
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1718-2020 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 31, 2020
Implementation date
April 1, 2021
Scope
YY/T 1718-2020 is the English-translated version of 人类体外辅助生殖技术用医疗器械 胚胎移植导管.
YY/T 1718-2020 is the Chinese pharmaceutical industry standard for the embryo transfer catheter used in assisted reproduction, the soft catheter through which gametes, zygotes, cleavage stage embryos or blastocysts are placed in the uterine cavity or the fallopian tube.
It sets the composition, the requirements, the test methods, the markings, the packaging and the transport, covering the dimensions and flexibility of the catheter and its guide, the tip and its atraumatic form, the connection to the syringe, the freedom from leakage, the effect of the materials on embryos and sterility. It was issued on 31 March 2020 and has applied since 1 April 2021.
Found what you need?
Document preview — YY/T 1718-2020
National Standard of the People's Republic of China
- ICS
- 11.040.30
- Classification
- C 30
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Product structure, composition and materials
- 5 Design attributes
- 6 Requirements and test methods
- 7 Sterile products
- 8 Packaging
- 9 Signs and instructions
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
Please note that certain contents of this document may involve patents. The issuing agency of this document is not responsible for identifying these patents.
1 Scope
This standard specifies the composition, requirements, test methods, markings, packages of medical device embryo transfer catheters for human in vitro assisted reproductive technology. Loading and transportation.
This standard applies to embryo transfer catheters used to transfer gametes, zygotes, cleavage stage embryos or blastocysts into the uterine cavity or fallopian tube.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated references, the latest version (including all amendments) applies to this document.
GB/T 191 Packaging, Storage and Transportation Graphic Mark
GB/T 1962.1 Syringes, injection needles and other medical devices 6% (Luer) tapered joints Part 1: General requirements
GB/T 1962.2 Syringes, injection needles and other medical devices 6% (Luer) tapered joints Part 2: Locking joints
GB/T 14233.1-2008 Medical infusion, blood transfusion, and injection equipment inspection methods Part 1: Chemical analysis methods
GB/T 14233.2 Test methods for medical infusion, blood transfusion and injection equipment-Part 2: Biological test methods
GB/T 15812.1-2005 Non-intravascular catheters Part 1: General performance test methods
GB/T 16886.1 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing in the Process of Risk Management
GB/T 16886.4 Biological Evaluation of Medical Devices Part 4: Test Selection for Interaction with Blood
GB/T 16886.5 Biological Evaluation of Medical Devices Part 5: In Vitro Cytotoxicity Test
GB/T 16886.10 Biological Evaluation of Medical Devices Part 10: Irritation and Skin Sensitization Test
GB/T 16886.11 Biological Evaluation of Medical Devices Part 11: Systemic Toxicity Test
GB 18278.1 Sterilization Damp Heat of Medical and Healthcare Products Part 1: Development, Validation and Routine Control of the Sterilization Process of Medical Devices Claim
GB 18279.1 Sterilization of medical and health care products ethylene oxide Part 1: Development, validation and routine control of the sterilization process of medical devices System requirements
GB/T 18279.2 Sterilization of Medical and Health Care Products Ethylene Oxide Part 2: GB 18279.1 Application Guide
GB 18280.1 Sterilization radiation of medical and health care products Part 1: Development, validation and routine control of the sterilization process of medical devices Claim
GB 18280.2 Medical and health products sterilization radiation Part 2: Establishing sterilization dose
GB/T 18280.3 Sterilization Radiation of Medical and Health Care Products Part 3: Dose Measurement Guidelines GB/T 19974 The characteristics of sterilization factors for medical and health care products and the setting, confirmation and routine control of sterilization processes for medical devices General requirements
YY/T 0149-2006 Stainless steel medical equipment corrosion resistance test method
YY/T 0313 Medical polymer product packaging and information requirements provided by manufacturers
YY/T 0466.1 Medical devices are used for medical device labeling, marking and providing information symbols. Part 1: General requirements
YY/T 0995 Terms and definitions of medical devices used in human assisted reproductive technology
YY/T 1434 In vitro mouse embryo test of medical devices for human in vitro assisted reproductive technology
3 Terms and definitions
The following terms and definitions defined in YY/T 0995 apply to this document.
3.1 Embryo transfer catheter
Transfer the gametes, zygotes, cleavage stage embryos or blastocysts produced by in vitro fertilization (IVF) and transfer of gametes into the uterine cavity or fallopian tubes Special equipment for intra-tube transplantation.
4 Product structure, composition and materials
4.1 Product composition
The manufacturer shall describe in detail the specifications and models of each product in the same registration unit. The basic structure of the embryo transfer catheter is shown in Figure 1. Products may vary.
4.2 Materials
4.2.1 Raw materials
The manufacturer shall evaluate the raw materials of the main components of the product.
4.2.2 Containers and packaging
The safety of packaging materials should be guaranteed.
5 Design attributes
In order to achieve the expected performance, the design attributes should take into account the possible damage during the embryo transfer process when the product is used.
6 Requirements and test methods
6.1 Physical properties
6.1.1 Dimensions
The manufacturer should consider what is applicable in the following dimensions.
a) Length of inner and outer pipes;
b) The diameter of the inner and outer pipes;
c) Length of stainless steel pipe;
d) Stainless steel pipe diameter; e) Ferrule diameter. The manufacturer shall give the nominal size and tolerance.
6.1.2 Distance indicator
The inner tube, outer tube or stainless steel tube of the embryo transfer catheter should have a distance indicating mark, the width of the mark, the length of the mark and the nominal size of the interval, The tolerance does not exceed ±10%.
6.1.3 Appearance
The surface of all fittings should be smooth, no black spots, no impurities, the inner tube and outer tube should be clearly marked on the scale, and the inner tube, outer tube, and guide core should have a pointed circle at the front end Passivation.
6.1.4 Corrosion resistance
According to the boiling water test method in Chapter 5 of YY/T 0149-2006, there should be no traces of corrosion on the surface of the stainless steel pipe.
6.1.5 Ultrasonic detectable
The outer tube of the embryo transfer catheter can be clearly displayed under B-mode ultrasound.
6.1.6 Cone joint
The tapered joint is tested according to the GB/T 1962.1 or GB/T 1962.2 method, and should meet the requirements of GB/T 1962.1 or GB/T 1962.2 Regulations.
6.1.7 Flow
According to GB/T 15812.1-2005 Appendix E method for testing, the amount of liquid passed in 60s is not less than 2mL.
6.1.8 Tightness
There is no liquid leakage in the inner tube with 100kPa pressure.
6.1.9 Connection strength
The inner tube and the joint should be able to withstand 3N axial static tension, and it will not separate for 5s, and the outer tube and the joint should be able to withstand a 3N shaft Pull to the static state for 5 seconds without separation.
6.2 Chemical properties
6.2.1 Preparation of test solution
Prepare the test solution according to the method specified in GB/T 14233.1-2008.
6.2.2 Reducing substances
Test according to the method specified in GB/T 14233.1-2008, the test solution and blank solution consume potassium permanganate solution [c(KMnO4)= The volume difference of 0.002mol/L] should not exceed 1.5mL.
6.2.3 Metal ions
Test according to the method specified in GB/T 14233.1-2008, the total extractable metal content in the test solution shall not exceed 1 µg/mL, and the content of cadmium The amount should not exceed 0.1µg/mL.
6.2.4 pH
According to the method specified in GB/T 14233.1-2008, the pH difference between the test solution and the control solution of the same batch shall not be greater than 1.0.
6.2.5 Evaporation residue
According to the method specified in 5.5 of GB/T 14233.1-2008, the total amount of evaporation residue should not exceed 2mg.
6.2.6 UV absorbance
Test according to the method specified in 5.7 of GB/T 14233.1-2008, within the wavelength range of 250nm~320nm, the absorbance of the test solution should not Greater than 0.1.
6.2.7 Residual ethylene oxide
Test according to the method specified in 9.4 of GB/T 14233.1-2008, the residual amount of ethylene oxide in each embryo transfer catheter should not exceed 0.5mg.
6.3 Biological performance
6.3.1 Sterility
According to GB/T 14233.2, the embryo transfer catheter should be sterile.
6.3.2 Bacterial endotoxin
Test according to the provisions of GB/T 14233.2, bacterial endotoxin should not be greater than 5EU/set.
6.4 Biological evaluation
6.4.1 General
The embryo transfer catheter should be biologically evaluated in accordance with the provisions of GB/T 16886.1, and should be For material manufacturing, the following items should be evaluated.
6.4.2 Pyrogen
According to GB/T 14233.2, the product should be pyrogen-free.
6.4.3 Cytotoxicity
According to the method specified in GB/T 16886.5, the cytotoxicity reaction is not greater than Grade I.
6.4.4 Intradermal stimulation
According to the method specified in GB/T 16886.10, there should be no intradermal irritation.
6.4.5 Skin sensitization
According to the maximum dose method in GB/T 16886.10, there should be no sensitization.
6.4.6 Hemolysis
According to the method specified in GB/T 16886.4, the hemolysis rate should be less than 5.0%.
6.4.7 Acute systemic toxicity
According to the method specified in GB/T 16886.11, there should be no acute systemic toxicity.
6.4.8 Mouse embryo test
The extract was prepared and tested according to the regulations in YY/T 1434, and the blastocyst formation rate was >=80%.
7 Sterile products
The product sterilization process should be confirmed and routinely controlled, and the sterility assurance level (SAL) of 10^-6 should be reached.
a) If the product is sterilized by ethylene oxide, it should meet the requirements of GB 18279.1 and GB/T 18279.2;
b) If the product is sterilized by moist heat, it should meet the requirements of GB 18278.1;
c) If the product is sterilized by radiation, it should meet the requirements of GB 18280.1, GB 18280.2 and GB/T 18280.3;
d) If the product uses other sterilization processes, it should meet the requirements of GB/T 19974.
8 Packaging
8.1 General
The packaging of the product should comply with the relevant regulations of YY/T 0466.1, GB/T 191, and YY/T 0313.
8.2 Single package
Each embryo transfer catheter should be enclosed in a single package.
When using ethylene oxide sterilization, the embryo transfer catheter should be made of dialysis paper on one side and plastic material or full dialysis paper packaging material.
Encourage the use of packaging materials with better dialysis effects, and packaging materials must not have harmful effects on the contents. The material and design of this packaging should make sure.
a) Under dry, clean and fully ventilated storage conditions, the contents can be guaranteed to be sterile;
b) When taken out of the package, the risk of contamination of the contents is minimal;
c) During normal handling, transportation and storage, the single package should fully protect the contents;
d) Once opened, the package cannot be easily resealed, and there should be obvious signs of tearing.
8.3 Medium packaging
One or more single packages should be packed in one middle package.
During normal handling, transportation and storage, the intermediate packaging should adequately protect the contents.
The inspection certificate and product instruction manual should be placed in the middle package.
8.4 Large packaging
One or more medium packages should be packed in one large package.
During normal handling, transportation and storage, the bulk packaging should be able to adequately protect the contents.
9 Signs and instructions
9.1 General
Symbols should comply with the requirements of YY/T 0466.1.
9.2 Single package
The single package should have at least the following signs.
a) Product name and marking according to Chapter 4 product structure, composition and material;
b) The words or equivalent signs of "sterile" and the method of sterilization;
c) Production batch number or production date;
d) Expiration year and month or expiration date;
e) "One-time use" or equivalent mark;
f) The name and address of the manufacturer and/or supplier.
9.3 Medium packaging
At least the following signs should be on the middle package.
a) Product name and marking according to Chapter 4 product structure, composition and material;
b) The words or equivalent signs of "sterile" and the method of sterilization;
c) Production batch number or production date;
d) Expiration year and month or expiration date;
e) "One-time use" or equivalent mark;
f) The name and address of the manufacturer and/or supplier.
9.4 Large packaging
If the medium package is packed in a large package, the large package should have at least the following signs.
a) Product name and marking according to Chapter 4 product structure, composition and material;
b) The words or equivalent signs of "sterile" and the method of sterilization;
c) Production batch number or production date;
d) Expiration year and month or expiration date;
e) "One-time use" or equivalent mark;
f) The name and address of the manufacturer and/or supplier;
g) Requirements for handling, storage and transportation.
9.5 Instructions
The instructions should specify the precautions, warning information and contraindications of the operation, and prompt the related risks and safety precautions to be taken.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB/T 1962.1Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirement
- GB/T 1962.2Conical fittings with a 6 %(Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- GB/T 14233.1-2008Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 14233.2Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods
- GB/T 15812.1-2005Catheters other than intravascular catheters -- Part 1: Test methods for common for properties
Similar standards
How to Download YY/T 1718-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 18278.1-2015 — Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
GB 18279.1-2015 — Sterilization of health care products—Ethylene oxide—Part 1: Requirements for development,validation and routine control of a sterilization process for medical devices
GB 18280.1-2015 — Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices