GB 18278.1-2015Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices (English PDF)

医疗保健产品灭菌 湿热 第1部分:医疗器械灭菌过程的开发、确认和常规控制要求

Preview PDF

This is a limited preview

Buy now to download the full PDF

Need quotation for your management?

Get a formal quotation for GB 18278.1-2015 in 30 minutes.

Issued by

General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China; Standardization Administration of China

Level / Type

National · Mandatory

Issue date

December 10, 2015

Implementation date

January 1, 2017

Scope

GB 18278.1-2015 is the English-translated version of 医疗保健产品灭菌 湿热 第1部分:医疗器械灭菌过程的开发、确认和常规控制要求.

GB 18278.1-2015 is the Chinese standard "Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices".

Its scope clause reads: GB 18278 of the provisions of this part of the development of medical devices moist heat sterilization, validation and routine control. Note.

Although the scope of this section is limited to medical devices, but require it to provide provisions and guidelines applicable to other health-related products.

This section contains the following heat sterilization process, but not limited to. a) saturated steam - Gravity exhaust system; b) saturated steam - powered exhaust system; c) air-vapor gas mixture; d) water spray; e) water immersion.

Note. See Appendix E. This section does not provide for the development of spongiform encephalopathy (disease such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease) pathogen inactivation process, recognition and often Regulatory control.

For dealing with potential contamination by these pathogens in the material, some countries have developed special specifications. Note. See YY/T 0771.1, YY/T 0771.2, ISO 22442-3.

This section does not apply to moist heat sterilization and other agents (such as. formaldehyde) sterilization sterilization process as a combination of factors.

This section does not indicate specific requirements detailed in "sterile" medical instruments. Note. It should be noted that different countries and regions for the labeling of medical devices "sterile" have different requirements, for example, YY/T 0615.1 or ANSI/AAMIST67.

This section does not provide for control of all stages of production of medical devices quality management system. Note.

This section is not required for the manufacture of a complete quality management system requirements, it is only the minimum requirements for the control of the sterilization process required for quality management The basic elements of the system, these elements normative references in the text to the appropriate location (see especially Chapter 4) as.

We should focus on controlling medical equipment production All stages (including the sterilization process including) quality management systems (see ISO 13485).

Some countries may require medical devices By the implementation of a complete quality management system, and evaluated by third parties. This section does not specify the moist heat sterilization facility design and operation of related occupational safety requirements.

Note. The requirements for the safe operation of the provisions of GB 4793.4.Moreover, some countries also have safety regulations. Its clauses include terms and definitions.

Found what you need?

Document preview — GB 18278.1-2015

National Standard of the People's Republic of China

ICS
11.080.01
Classification
C47

Issued by: General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China; Standardization Administration of China

Contents

  • Preface
  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Quality Management System Elements
  • 5 sterilization factor characterization
  • 6 features described processes and equipment
  • 7 Products OK
  • 8 process determines
  • 9 Confirm
  • 10 Routine monitoring and control
  • 11 sterilized product release
  • 12 to maintain the effectiveness of the process
  • Appendix A (informative) Guide
  • Annex D (normative) based on the reference microorganism inactivation traditional process determines (overkill approach)
  • Annex E (informative) run cycle
  • References

1 Scope

GB 18278 of the provisions of this part of the development of medical devices moist heat sterilization, validation and routine control.

Note. Although the scope of this section is limited to medical devices, but require it to provide provisions and guidelines applicable to other health-related products.

This section contains the following heat sterilization process, but not limited to.

a) saturated steam - Gravity exhaust system;

b) saturated steam - powered exhaust system;

c) air-vapor gas mixture;

d) water spray;

e) water immersion.

Note. See Appendix E.

This section does not provide for the development of spongiform encephalopathy (disease such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease) pathogen inactivation process, recognition and often

Regulatory control. For dealing with potential contamination by these pathogens in the material, some countries have developed special specifications.

Note. See YY/T 0771.1, YY/T 0771.2, ISO 22442-3.

This section does not apply to moist heat sterilization and other agents (such as. formaldehyde) sterilization sterilization process as a combination of factors.

This section does not indicate specific requirements detailed in "sterile" medical instruments.

Note. It should be noted that different countries and regions for the labeling of medical devices "sterile" have different requirements, for example, YY/T 0615.1 or ANSI/AAMIST67.

This section does not provide for control of all stages of production of medical devices quality management system.

Note. This section is not required for the manufacture of a complete quality management system requirements, it is only the minimum requirements for the control of the sterilization process required for quality management

The basic elements of the system, these elements normative references in the text to the appropriate location (see especially Chapter 4) as. We should focus on controlling medical equipment production

All stages (including the sterilization process including) quality management systems (see ISO 13485). Some countries may require medical devices

By the implementation of a complete quality management system, and evaluated by third parties.

This section does not specify the moist heat sterilization facility design and operation of related occupational safety requirements.

Note. The requirements for the safe operation of the provisions of GB 4793.4.Moreover, some countries also have safety regulations.

2 Normative references

The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein

Member. For undated references, the latest edition (including any amendments) applies to this document.

ISO 11138-1 Sterilization of health care products - Biological indicators - Part 1.General (Sterilizationofhealthcareprod- ucts-Biologicalindicators-GeneralRequirements)

ISO 11138-3 Sterilization of health care products - Biological indicators - Part 3.Biological indicators for moist heat sterilization (Sterilization ofhealthcareproducts-Biologicalindicators-Biologicalindicatorsformoistheatsterilizationproces- ses)

ISO 11140-1 Sterilization of health care products - Chemical indicators - Part 1.General (Sterilizationofhealthcare products-Chemicalindicators-Part 1.Generalrequirements)

Class 2 indicators for steam penetration test BD. ISO 11140-3 Sterilization of health care products - Chemical indicators - Part 3

Object system (Sterilizationofhealthcareproducts-Chemicalindicators-Part 3.Class2indicatorsystems foruseintheBowieandDick-typesteampenetrationtest)

ISO 11140-4 Sterilization of health care products - Chemical indicators - Part 4.BD II used to replace steam penetration test finger

Was shown (Sterilizationofhealthcareproducts-Chemicalindicators-Part 4.Class2indicatorsasanalter- nativetotheBowieandDick-typetestfordetectionofsteampenetration)

ISO 11140-5 Sterilization of health care products chemical indicators - Part 5.Class II for BD air exclusion means test

It was shown (Sterilizationofhealthcareproducts-Chemicalindicators-Part 5.Class2indicatorsforBowie andDick-typeairremovaltests)

ISO 10012 Measurement management systems for measurement processes and measuring equipment (Measurementmanagementsystems-

Requirementsformeasurementprocessesandmeasuringequipment)

ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1.materials, sterile barrier systems and packaging systems requirements (Packa- gingforterminalysterilizedmedicaldevices-Part 1.Requirementsformaterials, sterilebarriersys- temsandpackagingsystems)

ISO 11607-2 Packaging for terminally sterilized medical devices - Part 2.forming, sealing and assembly processes requires confirmation of (Packagingforterminalysterilizedmedicaldevices-Part 2.Validationrequirementsforforming, sealingandassemblyprocesses)

Sterilization Microbiological methods - Part 1 ISO 11737-1 medical devices. Determination of the total number of microorganisms on the product (Sterilizationofmedicaldevices-Microbiologicalmethods-Part 1.Determinationofapopulationof microorganismsonproducts)

Learn sterilization methods - Part 2 microorganisms ISO 11737-2 medical devices. sterility test confirmed the sterilization process (Sterilizationofmedicaldevices-Microbiologicalmethods-Part 2.Testsofsterilityperformedinthe validationofasterilizationprocess)

ISO 13485.2003 Medical Device Quality Management System Requirements for regulatory purposes (Medicaldevices-Qualityman- agementsystems-Requirementsforregulatorypurposes)

Sterilization process ISO 17664 medical device manufacturer providing reusable information (Sterilizationof sterilized medical devices medicaldevices-Informationtobeprovidedbythemanufacturerfortheprocessingofresterilizable medicaldevices)

3 Terms and Definitions

The following terms and definitions apply to this document.

3.1 Air Detector

Means for detecting whether or not the presence of non-condensable gases in the steam and condensate, or sterilization chamber.

3.2 Automatic controller

According to a predetermined program operating parameters, control sterilizer in order to run the program at all stages of the device required.

3.3 Bioburden

The total number of products and/or sterile barrier system surfaces and/or internal viable microorganisms.

[GB/T 19971-2015, Definition 2.2]

3.4 Biological indicators

Provisions sterilization process has a specific resistance, the test system containing viable microorganisms.

[GB/T 19971-2015, the definition 2.3]

3.5 Calibration

Under specified conditions, to determine the magnitude of measuring instruments or measuring systems indicated, or the amount of material measure or reference material represents

Value, and the corresponding reproduced by the standard set of operations between values. Its purpose is determined by comparison with a standard measuring device

Indication.

[VIM.1993, definitions 6.11]

3.6 Chemical indicators (non-biological indicators) chemicalindicator (non-biologicalindicator)

Depending on the chemical or physical changes after exposure to a particular process, a reflection of changes in process variables or more predetermined test systems.

[GB/T 19971-2015, the definition 2.6]

3.7 Sealed packaging products

At any stage of the sterilization process, without direct contact with the sterilizer product environment.

Note. The internal environment of the sterilizer only for heating and cooling purposes, and not to obtain sterilization effect, for example, sealed bottle of solution.

3.8 Correct correction

Measures taken to eliminate substandard have been found.

Note 1.The correction can be made in conjunction with the implementation of corrective action together.

Note 2.The example rework or downgrading as corrected.

[GB/T 19000-2008, the definition 3.6.6]

3.9 Corrective measures

Measures to eliminate substandard or other undesirable situation discovered the reason taken.

Note 1.a failure can have several causes.

NOTE 2.Corrective action is taken to prevent recurrence whereas preventive action is taken to prevent occurrence.

Note 3.The correction and corrective action are different.

[GB/T 19000-2008, the definition 3.6.5]

3.10 D value

D10 value D10value

Under specified conditions, the inactivation of 90% of the test organisms or the time required dosage.

NOTE. For this section is, D value refers to the sterilization process test microorganisms by 90%.

[GB/T 19971-2015, the definition 2.11]

3.11 Development

Elaboration of a technical norms.

[GB/T 19971-2015, the definition 2.13]

3.12 Environmental Control

Application engineering and/or system procedures, specific environment maintained within the specified limits.

Note. including air filters, liquid filters, surface disinfection, protective clothing, and management procedures.

[GB/T 19971-2015, the definition 2.16]

3.13 Equilibration time

Measured from the reference point to achieve sterilization temperature, all point to the time of sterilization loads have reached the sterilizing temperature desired.

3.14 Setting establish

Through theoretical determination, and after trials.

[GB/T 19971-2015, the definition 2.17]

3.15 Exposure time

Time process parameters is maintained within a predetermined tolerance range.

[GB/T 19971-2015, the definition 2.18]

3.16 Failure fault

One or more process parameters exceed a predetermined tolerance range.

[GB/T 19971-2015, the definition 2.19]

3.17 F0 value

Z value corresponds to the reference microorganisms at 10 °C, 121.1 °C under the sterilization process for sterilization of time, in minutes (min).

3.18 Health care products

Medical devices (including in vitro diagnostic medical devices) or pharmaceutical products (including biopharmaceuticals).

[GB/T 19971-2015, the definition 2.20]

3.19 Duration holdingtime

Sterilization chamber measuring point and reference point for all load sterilization temperature holding time in the continuous sterilization temperature range.

3.20 Installation identification instalationqualification; IQ

Proof equipment according to specifications provided and installed, and documented process.

[GB/T 19971-2015, the definition 2.22]

3.21 Load configuration

Provisions in article combination sterilizing chamber, including the sterilization chamber fixing member, and the number of items to be sterilized, type, distribution and direction.

3.22 Maintenance

For an article is maintained at or restored to its state with the desired function of a combination of all technical and related management behavior.

3.23 Medical Devices

Manufacturer's intended use for one or more specific purposes for human instrument, whether used alone or in combination, provided

Equipment, appliances, machinery, appliances, implant, in vitro reagent or calibrator, software, material or other similar or related items. These objects are.

--- Disease diagnosis, prevention, monitoring, treatment or mitigation;

--- Injury diagnosis, monitoring, treatment, mitigation or compensation;

--- Study the anatomy or a physiological process, replace or adjust;

--- Support or sustain life;

--- Control of conception;

--- Sterilization of medical devices;

--- Based on a sample taken from the body of in vitro examination of ways to provide medical information.

Its role in the design of the main body surface or in vivo effect is not pharmacological, immunological or metabolic means available, but these may have

It means to participate and play a supporting role.

[YY/T 0287-2003, the definition 3.7]

Note. This is defined by YY/T 0287-2003, developed by the GHTF (GHTF2002).

3.24 Measuring chain

Series feature measurement apparatus or measurement system, which constitutes the measurement signal from the input (measured by the magnitude) to the output (measurement result) of path.

3.25 Microbe microorganism

Under a microscope to see the small individuals, including bacteria, fungi, protozoa and viruses.

Note. Specific criteria may not need to inactivate all the microorganisms to prove the effect of the sterilization process, according to the above definition to confirm the sterilization process and/or normally

Regulatory control.

[GB/T 19971-2015, the definition 2.26]

3.26 Heat moistheat

Steam or liquid water in the presence of water in the manner of thermal energy to achieve the purpose of inactivation of microorganisms.

3.27 Non-condensable gases

Air and/or other gas saturated steam process conditions can not be liquefied.

3.28 Run identification operationalqualification; OQ

Proving devices installed by running the program is run within a predetermined limit, and the formation process of the file.

[GB/T 19971-2015, the definition 2.27]

3.29 Operating cycle

All stages of the program in accordance with the provisions of the order execution.

[IEC 61010-2-040.2005]

3.30 Packaging systems

Sterile barrier system and protective packaging portfolio.

[GB/T 19971-2015, the definition 2.28]

3.31 Performance appraisal performancequalification; PQ

Proof device has been installed according to procedures and operations, according to predetermined criteria throughout the production to meet the technical specifications of the product and form

Process file.

[GB/T 19971-2015, the definition 2.30]

3.32 Precautions preventiveaction

Measures to eliminate potential nonconformity or other potentially undesirable situation reason taken.

Note 1.A potential failure can have several causes.

Note 2.Preventive action is taken to prevent occurrence whereas corrective action is taken to prevent recurrence.

[GB/T 19000-2008, the definition 3.6.4]

3.33 Sterilizing time plateauperiod

Equilibration time plus hold time.

3.34 Process challenge device processchalengedevice; PCD

Designed to simulate product to be sterilized, and sterilization process to determine the formation of the resistance for the effective performance of the estimating process.

[GB/T 19971-2015, the definition 2.33]

3.35 Process parameters

Specified by the process variable value obtained.

Note. The sterilization process specifications include process parameters and their tolerances.

[GB/T 19971-2015, the definition 2.34]

3.36 Process variables

Conditions within the scope of the sterilization process, the change can affect the germicidal efficacy.

Example. time, temperature, pressure, concentration, humidit...

GB 18279-2023 | GB/T 18279.2-2015 | GB/T 18282.6-2026 | […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.

Referenced standards

Normative references

ISO 11138 · ISO 11140 · ISO 10012 · ISO 11607 · ISO 11737 · ISO 13485.2003 · ISO 17664

Cited by

How to Download GB 18278.1-2015

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB 18278.1-2015

$125.00