YY 0042-2018High-frequency jet ventilators (English PDF)
高频喷射呼吸机
This is a limited preview
Buy now to download the full PDF (22 pages)
Need quotation for your management?
Get a formal quotation for YY 0042-2018 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
December 20, 2018
Implementation date
June 1, 2020
Scope
YY 0042-2018 is the English-translated version of 高频喷射呼吸机.
YY 0042-2018 is the Chinese pharmaceutical industry standard for the high-frequency jet ventilator, the ventilator that delivers small volumes at a high rate through an open inspiratory and expiratory pathway.
It specifies the basic safety, the performance, the marking and the information the manufacturer provides, covering the driving pressure and the frequency and their accuracy, the humidification and the gas supply, the alarms and the monitoring, the protection of the patient against excessive pressure and the electrical safety.
It applies to ventilators used under the supervision of medical staff for respiratory support, respiratory therapy and emergency resuscitation. Issued on 20 December 2018, it applies from 1 June 2020.
Found what you need?
Document preview — YY 0042-2018
National Standard of the People's Republic of China
- Replacing
- YY 0042-2007
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative References5
- 3 Terms and Definitions6
- 4 Gas-Source Supply6
- 5 Power Supply7
- 6 Mechanical Strength8
- 7 Waterproof8
- 8 Fireproof8
- 9 Cleaning, Disinfection or Sterilization8
- 10 Material Compatibility9
- 11 Accuracy of Work Data9
- 12 Prevention of Dangerous Outputs10
- 13 Connection11
- 14 Marking11
- 15 Information Provided by the Manufacturer12
1 Scope
This Standard specifies the basic safety, performance, marking and information provided by the manufacturer for high-frequency jet ventilator (hereinafter referred to as ventilator).
This Standard applies to high-frequency jet ventilator with both exhalation and inhalation open.
The device is suitable for use by patients who need respiratory support, respiratory therapy and emergency resuscitation under the supervision of medical staff.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) is applicable to this document.
GB/T 4999 Anesthetic Equipment and Breathing Machine - Vocabulary
GB 9706.1 Medical Electrical Equipment - Part 1: General Requirements for Safety
GB 18278.1 Sterilization of Health Care Products - Moist Heat - Part 1: Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices
GB 18279.1 Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices
GB 18280.1 Sterilization of Health Care Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices
GB 18281.1 (all parts) Sterilization of Health Care Products - Biological Indicators
YY/T 0466.1-2016 Medical Devices - Symbols to be Used with Medical Device Labels, Labelling and Information to be Supplied - Part 1: General Requirements
YY 0600.3-2007 Lung Ventilators for Medical Use - Particular Requirements for Basic Safety and Essential Performance - Part 3: Emergency and Transport Ventilators
YY 0601 Medical Electrical Equipment - Particular Requirements for Basic Safety and normally within the input pressure range of 280kPa~600kPa
4 Gas-Source Supply
It shall not cause any hazard when the pressure of the air inlet is up to 1000kPa. When the gas source pressure is 280kPa, the average gas flow measured at the high-pressure gas input port of the ventilator shall not exceed 60L/min (over 10s time-weighted average).
4.2 Alarm of abnormal gas-source supply
The ventilator shall have an alarm function of abnormal gas-source supply. When the gas-source supply pressure is too low to maintain the normal operation of the ventilator, the ventilator shall be able to generate a high-priority alarm or a continuous audible alarm in accordance with YY 0709; and the audible alarm signal shall at least last 120s.
4.3 One gas missing in the air-oxygen mixing system
In addition to the air-oxygen mixing system mixed with the atmosphere, if a gas is missing in the air-oxygen mixing system, the ventilator shall be able to automatically switch to make the remaining gas continue working; and at the same time, it shall generate an alarm signal of at least low priority in accordance with YY 0709.
5 Power Supply
5.1 Alarm on the abnormality of grid power supply
The ventilator that is powered only by the grid power supply shall have the function of alarming the abnormality of grid power supply. When the voltage of the grid power is too low to maintain the normal operation of the ventilator, a high-priority alarm or a continuous audible alarm in accordance with YY 0709 shall be activated; and the audible alarm signal shall last at least 120s.
5.2 DC power supply
In addition to the grid power supply, if the ventilator is additionally equipped with a DC power supply (internal power supply or external DC power supply), and can automatically switch to DC power supply when the grid power supply is abnormal to maintain its normal operation, then when the grid power supply is abnormal, the alarm on the abnormality of grid power supply shall not be activated. For the conversion of such power supply, it shall comply with the provisions of the information prompt signal or low-priority alarm signal specified in YY 0709.
The ventilator that is powered by DC power supply should activate a medium-priority alarm in accordance with YY 0709 when the battery is low; and should activate a high-priority alarm in accordance with YY 0709 before all power is lost.
If the ventilator is equipped with a DC power supply, a method for the operator to determine the status of the DC power supply shall be provided.
5.3 Prevent mis-operation of the power switch
A method to prevent mis-operation of the power switch shall be provided.
5.4 Power cord
If the ventilator does not have a backup power supply, its equipped grid power cord shall be a non-detachable power cord; or measures shall be provided to prevent the grid power cord from accidentally falling off the ventilator.
6 Mechanical Strength
The ventilator that can be used for out-of-hospital transport shall be able to withstand the stress caused by rough operation during working; and the ventilator shall continue to work during and after the vibration impact test according to 21.6a), 21.6b) and 21.6aa) of YY 0600.3-2007.
7 Waterproof
The waterproof rating of ventilators that can be used for out-of-hospital transport shall be IPX4.
8 Fireproof
In normal or single fault conditions, combustible materials shall not be in both of the following conditions at the same time:
--- the material temperature reaches the minimum ignition temperature;
--- the presence of oxidizing agents.
9 Cleaning, Disinfection or Sterilization
Accessories and components in the ventilator breathing system that may be contaminated with rebreathing gases shall be removable for cleaning, disinfection, or sterilization.
If the label indicates that the product is sterile, the product should be sterilized by appropriate and effective methods in accordance with the relevant descriptions in GB 18278.1, GB 18279.1,
GB 18280.1 and GB 18281.
Packaging systems without passing through the sterilized equipment shall maintain product cleanliness prior to sterilization, use; and minimize the risk of bacterial contamination. (0.1kPa + 10% of the setting value).
11 Accuracy of Work Data
11.6 Sigh
If the ventilator has a sigh function, the tidal volume during sighing shall be no less than 1.5 times the tidal volume during non-sighing.
11.7 Measurement of airway pressure
A device to measure the airway pressure at the patient's connection port shall be provided; and the accuracy of the pressure measuring device shall be ± (2% full scale + 8% actual reading).
NOTE: The actual measurement can be carried out at any part of the ventilator breathing system; but the displayed pressure value shall be the pressure corresponding to the patient connection port.
11.8 Oxygen concentration monitoring
The ventilator should be equipped with an oxygen concentration monitor that complies with the provisions of YY 0601 for measuring the oxygen concentration at the inspiratory branch or the patient connection port.
12 Prevention of Dangerous Outputs
12.1 Protection against maladjustment
Protection shall be provided to prevent maladjustment that could produce dangerous outputs.
12.2 Maximum ultimate pressure
Under normal use and single fault conditions, the pressure at the patient connection port shall not exceed 12.5kPa.
12.3 Alarm on high airway pressure
The ventilator shall have a function of alarming the high airway pressure. When the airway pressure reaches a preset limit value, the ventilator shall be able to generate a high-priority alarm signal in accordance with YY 0709.
12.4 Alarm on respiratory system integrity
The ventilator shall have a function of alarming the respiratory system integrity; and the alarm signal shall be a high-priority alarm in accordance with YY 0709.
NOTE: Alarms with low airway pressure, low expiratory volume, and pipeline detachment can be considered to meet the requirements of this clause. symbols of reference number 5.4.2 in YY/T 0466.1-2016), if applicable;
--- the word "sterile" or an equivalent symbol (such as the graphic symbols of reference numbers 5.2.1 ~ 5.2.5 in YY/T 0466.1-2016), if applicable;
--- recommended cleaning, disinfection or sterilization methods, if applicable;
--- The packaging containing rubber shall be clearly marked with the words "natural rubber latex" or an equivalent symbol (such as the graphic symbol of reference number 5.4.5 in YY/T 0466.1-2016).
14 Marking
14.4 All components that are sensitive to the direction of airflow, if the operator can disassemble and install them without tools; measures shall be taken to prevent misconnection. If not, the components shall be marked with clear, easy-to-recognize and permanently attached arrow to indicate the airflow direction.
14.5 If the needle advancing method is used to adjust the oxygen concentration, the advancing direction of the needle when the oxygen concentration increases shall be marked.
15 Information Provided by the Manufacturer
The manufacturer shall provide the following information:
a) the manufacturer's address or the name and address of an authorized representative;
b) cleaning, disinfection or sterilization for reuse, and details of any other means of restriction, if applicable;
c) if the ventilator has internal and/or external backup power supply, the manufacturer shall indicate the corresponding parameters in the relevant documents, including at least:
- voltage;
- current;
- amp-hour rating or operating time from a fully charged power supply;
- a method for determining the status of the backup power supply.
d) if the ventilator is supplied by high-pressure gas, the supply pressure and flow range shall be stated;
e) a warning that the ventilator shall not be covered or placed in a location that affects the operation and performance of the ventilator (e.g., the ventilator shall not be placed next to a barrier, which shall prevent the flow of cool air and overheat the equipment), if applicable; […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 22 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 4999Anaesthetic and respiratory equipment Vocabulary
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB 18278.1Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
- GB 18279.1Sterilization of health care products—Ethylene oxide—Part 1: Requirements for development,validation and routine control of a sterilization process for medical devices
- GB 18280.1Sterilization of health care products - Radiation - Part 1: Requirements for the development
- YY/T 0466.1-2016Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
GB 18281.1
Similar standards
Editions of YY 0042
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0042-2018 | High-frequency jet ventilators | current edition | Current |
| YY 0042-2007 | High-frequency jet ventilators | previous edition | Superseded |
This page sells the current edition, YY 0042-2018. Earlier editions are listed for reference only.
How to Download YY 0042-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 18278.1-2015 — Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
GB 18279.1-2015 — Sterilization of health care products—Ethylene oxide—Part 1: Requirements for development,validation and routine control of a sterilization process for medical devices
GB 18280.1-2015 — Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
Secure payment via Stripe
Payments accepted
YY 0042-2018
$105.00