YY/T 0321.2-2021Single-use needle for anaesthesia (English PDF)

一次性使用麻醉用针

Preview PDF

This is a limited preview

Buy now to download the full PDF (21 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0321.2-2021 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

March 9, 2021

Implementation date

April 1, 2022

Scope

YY/T 0321.2-2021 is the English-translated version of 一次性使用麻醉用针.

YY/T 0321.2-2021 is the Chinese pharmaceutical industry standard for single-use anaesthesia needles with a nominal outside diameter from 0.4 mm to 2.1 mm, the needles of the puncture kits used for epidural and spinal anaesthesia.

It is the second part of the YY/T 0321 series and sets the classification and labelling of the needles, the materials, and their physical, chemical and biological properties, together with marking, packaging, transport and storage.

The physical requirements cover the point and its type, the patency and stiffness of the tube, the strength of the joint to the hub, the conical fitting, and the freedom from leakage.

It replaces YY 0321.2-2009, was issued on 9 March 2021 and has applied since 1 April 2022.

Found what you need?

Document preview — YY/T 0321.2-2021

National Standard of the People's Republic of China

Replacing
YY 0321.2-2009

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Classification and labeling6
  • 4 Materials9
  • 5 Physical properties9
  • 6 Chemical properties15
  • 7 Biological properties15
  • 8 Marking16
  • 9 Packaging17
  • 10 Transportation and storage17
  • Appendix A (Normative) Flow test method18
  • Appendix B (Normative) Particle content test method19
  • Appendix C (Normative) Preparation method of test solution for chemical properties .20
  • References21

Foreword

YY 0321 consists of the following parts:

-- Single-use puncture set local anaesthesia;

-- Single-use needle for anaesthesia;

-- Single-use filter for anaesthesia. This Part is Part 2 of YY 0321.

This Part was drafted in accordance with the rules given in GB/T 1.1-2009. This Part replaces YY 0321.2-2009 Single-use needle for anaesthesia.

Compared with YY 0321.2-2009, the main technical changes, except editorial changes, are as follows:

-- Add a schematic diagram of matching type of cutting edge and lumbar puncture needle (see Figure 1; Figure 1 of the 2009 edition);

-- Modify the needle tube length of "basic dimensions of needle for anaesthesia" (see 5.1; 3.4 of the 2009 edition);

-- Modify the requirements for the "scale line" of the needle tube (see 5.3.5; 5.2.5 of the 2009 edition);

-- Modify the requirements for "the inner arc of the bevel of the needle tip of the epidural puncture needle" (see 5.5.2; 5.4.2 of the 2009 edition);

-- Modify the requirements and test methods for "connection firmness" (see 5.7; 5.6 of the 2009 edition);

-- Add the requirements for "the total content of barium, chromium, copper, lead and tin in the test solution" in metal ions (see 6.2);

-- Delete "Note 2" on the applicability of the sterility test method in sterility (see 7.1 of the 2009 edition);

-- Modify the requirements for "marking" and delete the requirements for "instructions for use" (see Chapter 8; Chapter 8 of the 2009 edition);

-- Modify the requirements for "packaging" (see Chapter 9; Chapter 9 of the 2009 edition);

-- Delete the inspection rules of type inspection (see Appendix D of the 2009 edition).

1 Scope

This Part of YY 0321 specifies the classification and labeling, materials, physical properties, chemical properties, biological properties, marking, packaging, transportation and storage of single-use needle for anaesthesia with a nominal outer diameter of 0.4 mm ~ 2.1 mm.

This Part applies to single-use needle for anaesthesia for puncture and drug injection during human epidural nerve block (referred to as epidural anesthesia), subarachnoid space block (referred to as lumbar anesthesia), combined epidural and lumbar anesthesia, and local anesthesia of nerve block.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies to this document. For undated references, the latest edition (including any amendment) applies to this document.

GB/T 1962.1, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirement

GB/T 1962.2, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings

GB/T 6682, Water for analytical laboratory use - Specification and test methods

GB/T 14233.1-2008, Test methods for infusion, transfusion, injection equipment for medical use - Part 1: Chemical analysis methods

GB/T 14233.2, Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods

GB/T 16886.1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

GB/T 18457, Stainless steel needle tubing for the manufacture of medical devices

YY/T 0313, Medical polymer products - Requirement for package and information supplied by manufacturer

YY/T 0466.1, Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements

3 Classification and labeling

3.1 Classification

Single-use needle for anaesthesia (hereinafter referred to as needle for anaesthesia) can be divided into the following categories:

a) epidural puncture needle;

b) lumbar puncture needle;

c) combined anesthesia trocar;

d) nerve block puncture needle.

3.2 Labeling

3.2.1 Product model

The product model labeling is composed of product designation and classification code. Product designation: Represent needle for anaesthesia by AN.

Classification code: Represent epidural puncture needle by E; represent lumbar puncture needle by S; represent combined anesthesia trocar needle by E/S II; represent nerve block puncture needle by N.

Note: For the lumbar puncture needles with Type II cutting edge, the cutting-edge type code should be added after the classification code. Example 1: Single-use needle for anaesthesia - epidural puncture needle:

AN-E Example 2: Single-use needle for anaesthesia - combined anesthesia trocar of epidural puncture needle and type II cutting edge lumbar puncture needle: AN-E/S II

3.2.2 Product specifications

The product specification labeling is represented by the nominal outer diameter of the needle tube × the length of the needle tube. Dimensions in millimeters.

5 Physical properties

5.3.3 Resistance to corrosion

When testing according to GB/T 18457, there shall be no traces of corrosion on the soaked part of the needle tube.

5.3.4 Outer surface

Observed with normal vision or corrected vision, the outer surface of the needle tube shall be smooth and free of impurities during metal processing.

5.3.5 Graduation line

If the needle tube of the needle for anesthesia has a scale mark, the graduation line shall be clear without broken lines, and the size of the graduation line shall meet the requirements in Figure 1. Immerse the needle tube in water at 18 °C ~ 28 °C for 24 hours; then take it out; wipe the graduation line with white medical gauze. There shall be no color falling off.

5.4 Lining core

5.4.1 The lining core shall have a smooth surface, and shall pass through the inner cavity of the needle for anesthesia; there shall be no blockage in the hand feeling test.

5.4.2 The corrosion resistance of the lining core made of stainless steel shall meet the requirements of 5.3.3.

5.5 Needle tip

5.5.1 When inspecting with a 3x magnifying glass, the needle tip shall be free of burrs, flat ends and hooks.

5.5.2 There shall be no sharp edges on the inner arc of the bevel of the needle tip of the epidural puncture needle. Use an anesthesia catheter of the corresponding specification to insert it from the tapered hole of the needle file until it passes through the bevel of the needle tip by 10 mm, so that the needle tip touches a vertical plane, and the inner arc of the needle forms an included angle of 90° ~ 95° with the catheter. Slowly withdraw the catheter without breaking it. Nominal outside diameter of the needle tube (Specifications)

Distance between rigid support and point of application of bending force The clamp for fixing the needle tube must not damage the needle tube.

6 Chemical properties

6.1 Preparation method of test solution

See Appendix C for the preparation method of test solution.

6.2 Metal ions

When testing according to 5.6.1 in GB/T 14233.1-2008, the color of the test solution shall not exceed the standard control solution with a mass concentration rho(Pb2+) = 5 µg/mL.

When using the atomic absorption spectrophotometry (AAS) in GB/T 14233.1 or an equivalent method for testing, the total content of barium, chromium, copper, lead and tin in the test solution shall not exceed 1 µg/mL, and the content of cadmium shall not exceed 0.1 µg/mL.

6.3 Limits for acidity and alkalinity

When testing according to 5.4.1 in GB/T 14233.1-2008, the difference of pH between the test solution and the same batch of blank solution shall not exceed 1.0.

6.4 Residue of ethylene oxide

If it is sterilized by ethylene oxide, when testing according to Chapter 9 of GB/T 14233.1-2008, the residual amount of ethylene oxide in each anesthesia needle shall not exceed 0.1 mg.

7 Biological properties

7.1 General

Where there is a production of new products and change in raw materials, the biological evaluation of cytotoxicity, intradermal irritation, skin sensitization, systemic acute toxicity and blood compatibility shall be carried out according to the provisions of GB/T 16886.1.

7.2 Sterility

Each single packaged needle for anesthesia shall pass a sterility validation and routine control to render the product sterile.

Note: See GB 18278.1, GB 18279.1 and GB 18280.1 for appropriate sterilization confirmation and routine control.

7.3 Bacterial endotoxins

When testing according to the gel method in GB/T 14233.2, the bacterial endotoxin limit of single-use needle for anaesthesia shall not exceed 2.15 EU per piece for lumbar puncture needle, and shall not exceed 20 EU per piece for epidural puncture needle and nerve block puncture needle.

8 Marking

8.1 Single package

The single package shall at least include the following information:

a) manufacturer and/or distributor name and address;

b) product name, model, specifications;

c) batch number or production date, or graphic symbols conforming to YY/T 0466.1;

d) expiration date (year and month), with appropriate text attached, or graphic symbols conforming to YY/T 0466.1;

e) words "single-use", or equivalent instructions, or graphic symbols conforming to

YY/T 0466.1;

f) words "sterile" and "sterilization method", or graphic symbols conforming to

YY/T 0466.1;

g) words such as "damaged package, prohibited use" and "destroy after use", or graphic symbols conforming to YY/T 0466.1.

8.2 Shelf or multi-unit packaging

Shelf packs or multi-unit packs (if used) shall be marked with at least the following information:

a) manufacturer and/or distributor name and address;

b) product name, model, specifications;

c) batch number or production date, or graphic symbols given in YY/T 0466.1;

d) expiration date (year and month), with appropriate text attached, or graphic symbols given in YY/T 0466.1;

e) words "single-use", or equivalent instructions, or graphic symbols in accordance with YY/T 0466.1;

Appendix A (Normative) Flow test method

A.1 Principle

Pass water under a certain pressure through the inner cavity of the tested piece, and calculate the amount of water flowing out of the tested piece per unit time.

A.2 Test equipment

A.2.1 Flow tester: The water pressure can be adjusted within 0 kPa ~ 300 kPa, the pressure tolerance is ±5%, and the water outlet has an outer conical joint.

A.2.2 A 25 mL graduated cylinder with a minimum scale of 0.5 mL.

A.3 Test steps

A.3.1 Turn on the power of the instrument.

A.3.2 Install the tested piece on the outer conical joint of the water outlet of the instrument. A.3.3 Adjust the water pressure to the standard value. A.3.4 Select the test time.

A.3.5 Start the test switch, let water pour into the measuring cylinder from the outlet of the tested piece, and it will automatically shut down when the stipulated time is reached. A.3.6 Read the water volume in the graduated cylinder.

Appendix B (Normative) Particle content test method

B.1 Principle

Evaluate particle contamination by flushing the test piece, collecting particles in the eluent of the test piece, and counting them.

B.2 Test equipment

The particle counter (resistance method or light resistance method) has a stirring system, and the sampling volume at a time is not less than 1 mL.

B.3 Flushing solution

Water for injection, purified water or sodium chloride solution with a mass concentration of 9 g/L. Filter through a microporous membrane with a pore size of 0.2 µm.

B.4 Test steps

Use a clean syringe to take 60 mL of flushing solution, and flow it into the sample pool of a clean counter through the measured piece to obtain the eluent. Count the particles larger than or equal to 5.0 µm in the eluate in the sample pool, and the total sampling volume shall not be less than 15 mL.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 0321.2

EditionTitleRevisionStatus
YY/T 0321.2-2021Single-use needle for anaesthesiacurrent editionCurrent
YY 0321.2-2009Single-use needle for anaesthesiaprevious editionSuperseded

This page sells the current edition, YY/T 0321.2-2021. Earlier editions are listed for reference only.

How to Download YY/T 0321.2-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
21 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0321.2-2021

$245.00