YY/T 0291-2016Environmental requirements and test methods for medical X-ray equipment (English PDF)

医用X射线设备环境要求及试验方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (18 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0291-2016 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

March 23, 2016

Implementation date

January 1, 2017

Scope

YY/T 0291-2016 is the English-translated version of 医用X射线设备环境要求及试验方法.

YY/T 0291-2016 is the Chinese pharmaceutical industry standard that sets the environmental requirements and the test methods for medical X-ray equipment: the groups of climatic and mechanical conditions such equipment must withstand in transport, storage and use, the tests for temperature, humidity, low pressure, vibration and shock, and the checks of function and appearance that follow each test. Issued on 23 March 2016, it has applied since 1 January 2017.

Found what you need?

Document preview — YY/T 0291-2016

National Standard of the People's Republic of China

Replacing
YY/T 0291-2007

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative References5
  • 3 Terms and definitions5
  • 4 Environment grouping6
  • 5 Requirements6
  • 6 Test methods8
  • Annex A (Normative) Test requirements and test items12
  • Annex B (Informative) Environment test sequence chart14

Foreword

This Standard was drafted in accordance with the rules given in GB/T 1.1-2009.

Compared with YY/T 0291-2007, the major technical changes of this Standard are as follows.

-- it adds in the normative references JTG B01-2014, Technical Standard of Highway Engineering;

-- it adds the clause of terms and definitions;

-- it modifies the expression of environment grouping (see 4; and see 3 of 2007 edition);

-- it modifies the expression of requirements (see 5; and see 4.1, 4.2 and 4.3 of 2007 edition);

-- it modifies the expression of special circumstances (see 6.3; and see 5.1 of edition

2007);

-- it modifies the expression of test sequence (see 6.1.2; and see 5.2.2 of 2007 edition);

-- it modifies the expression of test content (see 6.2; and see 5.2.3 of edition 2007);

-- it modifies the expression of environment test methods (see 6.2; and see 5.3 of 2007 edition);

-- it modifies the expression of requirements for recovery time and requirements for power supply adaptability (see Table A.1; and see Table A.1 of 2007 edition);

-- it adds the environment test sequence chart Annex B as an informative annex.

1 Scope

This Standard specifies the environment grouping, requirements and test methods for medical X-ray equipment (hereinafter referred to as equipment). This Standard applies to medical X-ray equipment.

2 Normative References

The following referenced documents are indispensable for the application of this document. For dated references, only the edition dated applies to this document. For undated references, the latest edition of the referenced documents (including all amendments) applies to This Standard.

GB 9706.1-2007, Medical electrical equipment - Part 1. General requirements for safety

GB 9706.3-2000, Medical electrical equipment - Part 2. Particular requirements for the safety of high-voltage generators of diagnostic X-ray generators

GB 9706.18-2006, Medical electrical equipment - Part 2. Particular requirements for the safety of X-ray equipment for computed tomography

GB 9706.24-2005, Medical electrical equipment - Part 2-45. Particular requirements for the safety of mammographic X-ray equipment and mammographic sterotactic devices

GB/T 10149-1988, Terminology and symbol for medical X-ray equipment

GB/T 14710-2009, Environmental requirement and test methods for medical electrical equipment JTG B01-2014, Technical Standard of Highway Engineering

3 Terms and definitions

For the purposes of this document, the terms and definitions defined in GB/T 10149- 1988 apply.

After test, the fasteners of equipment shall not become loose, which shall be tested in accordance with the test items specified in product standard.

5 Requirements

5.2.4 Power supply adaptability

5.2.4.1 The power supply of equipment is set at 110% and 90% of rated working voltage, supply frequency 50 Hz ± 1 Hz (if necessary). Maintain the voltages for at least 15 min and then test the relevant test items specified in product standard.

5.2.4.2 For equipment having special requirements for power supply frequency and voltage, its working range of frequency and voltage can be specified otherwise in product standard.

6 Test methods

6.1 Test procedure and test sequence

6.1.1 Test procedure

Each test normally includes the following procedures.

a) pretreatment (if necessary);

b) initial detection (if necessary);

c) test;

d) intermediate detection (if necessary);

e) recovery (if necessary); f) final detection.

6.1.2 Test sequence

When multiple tests are conducted for the same equipment, tests shall be normally conducted in the following sequence.

a) rated working low temperature test;

b) low temperature storage test;

c) rated working high temperature test;

d) high temperature storage test;

e) rated working damp heat test;

f) damp heat storage test;

Under the condition of normal package of equipment, equipment shall be fixed to the rear part of truck in accordance with the sign "UP" location. The load of truck during test is 1/3 of the rated load. Test can be conducted using truck or transportation test apparatus.

After test, test whether equipment fasteners become loose and conduct test items in accordance with product standard.

6.2.4 Power supply adaptability test

This test is normally conducted during rated working low temperature test and rated working high temperature test. During test, connect the power cord of equipment to an adjustable power supply; maintain the output frequency of the adjustable power supply at the rated product frequency; set voltages at 110% and 90% of rated product voltage; maintain the voltages at least 15 min; and then conduct relevant test items specified in product standard.

6.3 Special circumstances

Consider the following circumstances.

a) climatic environment test only applies to those parts or components which may be subjected to the climatic conditions simulated by test;

b) equipment shall be placed in test chamber unpackaged, in the ready-for-use state, at a normal working position;

c) equipment accessories shall be tested along with equipment, unless they are specified otherwise in product standard;

d) the components which can be detached without using tools shall be detached and tested along with major parts;

e) if test can't be conducted without disassembling and reassembling, the time intervals between treatment and test and between test and detection shall be decided in accordance with the time of equipment handling, assembling and debugging;

f) if equipment (such as large equipment or equipment having special requirements for service environment) is inappropriate for some environment test items (such as rated working low temperature test, rated working high temperature test, rated working damp heat test and so on), the manufacturer shall provide specific test requirements and indicate in product standard;

g) if product standard has special requirements for environment test conditions, they shall be implemented as specified in product standard. When it is impossible to test the whole machine, product standard shall specify which key parts or components are to be tested;

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.

Referenced standards

Cited by

Similar standards

Editions of YY/T 0291

EditionTitleRevisionStatus
YY/T 0291-2016Environmental requirements and test methods for medical X-ray equipmentcurrent editionCurrent
YY/T 0291-2007Environmental requirements and test methods for medical X-ray equipmentprevious editionSuperseded

This page sells the current edition, YY/T 0291-2016. Earlier editions are listed for reference only.

How to Download YY/T 0291-2016

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
18 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0291-2016

$115.00