Scope

YY/T 0310-2015 is the English-translated version of X射线计算机体层摄影设备通用技术条件.

YY/T 0310-2015 is the Chinese standard "General specifications for X-ray equipment for computer tomography". Its scope clause reads: This Standard specifies the terms and definitions, classification, composition, requirements and test methods of X-ray equipment for computed tomography (hereinafter referred to as CT scanner).

This Standard is applicable to CT scanner, including CT scanner which provides image data for radiation treatment planning. Its clauses include terms and definitions; classification and composition; requirements; test methods.

It was issued by the China Food and Drug Administration on MARCH 2, 2015 and took effect on JANUARY 1, 2016.

Found what you need?

Document preview — YY/T 0310-2015

National Standard of the People's Republic of China

Replacing
YY 0310-2005

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions5
  • 4 Classification and composition5
  • 5 Requirements6
  • 6 Test methods13

Foreword

This Standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This Standard replaces YY 0310-2005 General specifications for X-ray equipment for computed tomography.

Compared with YY 0310-2005, the main technical changes are as follows.

- Modified the description of application scope, including CT scanner which provides image data for radiation treatment planning;

- Added YY 0505 and YY/T 0708 as normative references;

- Modified environmental conditions in requirements to conditions required by GB 9706.1;

- Added requirements for scanning frame rotation speed;

- Modified the original scanning time to exposure time;

- Deleted indicating instrument (5.5.5 of 2005 edition);

- Added requirements and test methods of radiation treatment planning (RTP).

1 Scope

This Standard specifies the terms and definitions, classification, composition, requirements and test methods of X-ray equipment for computed tomography (hereinafter referred to as CT scanner).

This Standard is applicable to CT scanner, including CT scanner which provides image data for radiation treatment planning.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB 9706.1 Medical electrical equipment - Part 1: General requirements for safety

GB 9706.11 Medical electrical equipment - Part 2: Particular requirements for the safety of

GB 9706.12 Medical electrical equipment - Part 1: General requirements for safety 3.collateral STANDARD. General requirements for radiation protection in diagnostic X-ray equipment

GB 9706.14 Medical electrical equipment - Part 2: Particular requirements for the safety of associated equipment of X-ray equipment

GB 9706.15 Medical electrical equipment - Part 1: General requirements for safety - 1.Collateral standard. Safety requirements for medical electrical systems

GB 9706.18 Medical electrical equipment - Part 2: Particular requirements for the safety of X-ray equipment for computed tomography

GB/T 10149 Terminology and symbol for medical X-ray equipment

GB/T 10151 Medical diagnostic X-ray equipment - Specifications for high voltage cable plugs and sockets

GB/T 19042.5-2006 Evaluation and routine testing in medical imaging departments - Part 3-5.Acceptance tests-Imaging performances of computed tomography X-ray equipment

YY 0076-1992 Coating classifications for metal product - Technical conditions

YY/T 0291 Environmental requirements and test methods for medical x-ray equipment

YY 0505 Medical electrical equipment Part 1-2.General requirements for safety Collateral standards. Electromagnetic compatibility Requirements and tests

YY 0637 Medical electrical equipment - Requirements for the safety of radiotherapy treatment planning systems

YY/T 0708 Medical electrical equipment - Part 1-4.General requirements for safety - collateral standard.

3 Terms and definitions

Terms and definitions defined in GB/T 10149, GB 9706.18, GB/T 19042.5-2006 and YY 0637 apply to this document.

4 Classification and composition

4.1 Classification

Security classification shall comply with the requirements on classification in GB 9706.1.

4.2 Composition

CT scanner shall be composed of the following parts.

a) Scanning frame;

b) X-ray generator;

c) Patient support;

d) Console;

M1, M2 - Marks on patient support surface; d1, d2 - Distance between marking point and Z-axis.

5 Requirements

5.7.2 Patient support surface

Patient support surface shall be plane, or the attachment which can make it as plane shall be indicated in accompanying file and make it obtained.

Positioning aids of therapy system are allowed to use by patient support.

5.7.3 Diagnostic bed sag (patient support rigidity)

It shall stipulate the diagnostic bed sag within 40 cm (typical scan length plus the amount of movement of scanning plane).

5.7.4 Integrated positioning light used for patient positioning

If positioning light is integrated in CT scanner, accompanying file shall verify if the intended use of positioning light can be used for patient positioning in RTP.

If the intended use of positioning light can be used for patient positioning in RTP, it shall have the following accuracy.

The accuracy of axial positioning light at isocenter shall be ± 1 mm. The accuracy at ± 250 mm in X direction shall be ± 2 mm.

Sagittal and coronal positioning lights shall be extended to within scanning plane. The accuracy related to rotation axis shall be ± 1 mm.

Marking line width of positioning light at isocenter (FWHM) shall not exceed 1 mm.

5.7.5 Typical scanning mode that generates RTP

The accompanying file shall stipulate the typical CT operating conditions which generate images for RTP. The agreement not applicable to generate RTP image shall be indicated in the accompanying file.

For RTP's typical CT operating conditions, the accompanying file shall provide measurement results of noise, average CT value and uniformity measured by the methods in GB/T 19042.5-2006.

5.7.6 HU value conversion

HU value of each model measured under CT operating conditions indicated in 5.7.5, and the conversion of electronic and mass density values related to

a) CT scanner appearance shall look neat, beautiful; the surface shall be smooth and clean; the color shall be even; there must no damage spots, cracks and other defects;

b) The main electroplating pieces of CT scanner shall comply with requirements on grade 2 appearance in YY 0076-1992.

5.9 Environment test

It shall comply with requirements of YY/T 0291. The final test items shall at least contain 5.2.1, 5.2.2, 5.2.3, 5.2.5, 5.2.8 and 5.5.3.

5.10 Safety

It shall comply with requirements of GB 9706.1, GB 9706.11, GB 9706.12, GB 9706.14, GB 9706.15, GB 9706.18, YY 0505 and YY/T 0708.

6 Test methods

6.1 Test conditions

CT scanner shall be tested under the following conditions.

a) Unless otherwise specified, all performance tests shall be conducted under conditions stipulated in 5.1;

b) CT body model which complies with American Association of Physicists in Medicine (AAPM) standard is recommended;

c) Unless otherwise specified, it shall use CT conditions of typical head and body to scan.

6.2 Performance

6.2.1 Image noise

Under typical head scanning conditions (recommended to use 10 mm slice thickness), use head CTDI body model to measure center dose.

If the equipment does not have 10 mm slice thickness, the product standard shall stipulate the conversion factor of image noise.

Place the homogeneous medium body model (20 cm water model) within scan field of view. Make body model axis coincide with the rotation axis of scanning frame. After scanning, select a region of interest of which the diameter is about 40% of image diameter. Measure the standard deviation SD of CT value in this region. Use formula (1) to calculate noise value N. reconstruct. Adjust the window width and window level. Observe the image through monitor. Take the diameter of the distinguishable minimum hole group as reference.

6.2.6 Operating noise

In the fastest scan state, use sound level meter to respectively measure the operating noise at 1.5 m height from ground, 1 m from the front, back, left and right of scanning frame surface. Take the maximum.

6.2.7 Artifact

Place water model within scan field of view. Make body model axis coincide with the rotation axis of scanning frame. Select a group of CT operating conditions to scan. Observe water model's CT image when window level is 0, window width are respectively 50 HU, 100 HU.

6.2.8 Slice thickness

This Standard recommends the following method. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.

Referenced standards

Normative references

GB 9706.11 · GB 9706.14 · GB 9706.18

Similar standards

Editions of YY/T 0310

EditionTitleRevisionStatus
YY/T 0310-2015General specifications for X-ray equipment for computed tomographycurrent editionCurrent
YY 0310-2005General specifications for X-ray equipment for computed tomographyprevious editionSuperseded

This page sells the current edition, YY/T 0310-2015. Earlier editions are listed for reference only.

How to Download YY/T 0310-2015

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
19 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0310-2015

$125.00