Valid

YY 0637-2013Medical electrical equipment—Requirements for the safety of radiotherapy treatment planning systems (English PDF)

医用电气设备 放射治疗计划系统的安全要求

Preview PDF

This is a limited preview

Buy now to download the full PDF (38 pages)

Need quotation for your management?

Get a formal quotation for YY 0637-2013 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

October 21, 2013

Implementation date

October 1, 2014

Scope

YY 0637-2013 is the English-translated version of 医用电气设备 放射治疗计划系统的安全要求.

YY 0637-2013 is the Chinese pharmaceutical industry standard for the safety of radiotherapy treatment planning systems, identical to IEC 62083:2009.

It covers the accompanying documents and general operational safety, including units for distances and radiation quantities, date and time format, protection against unauthorized use and modification, data limits, correctness of data transfer, coordinate systems and data archiving.

Further clauses deal with equipment and brachytherapy source modelling, patient anatomy modelling and contouring, treatment plan preparation, identification and electronic signatures, and the accuracy of absorbed dose calculation algorithms.

The treatment plan report, software versions, use errors and hardware safety are also addressed. It replaces YY 0637-2008. Issued on 21 October 2013, it has applied since 1 October 2014.

Found what you need?

Document preview — YY 0637-2013

National Standard of the People's Republic of China

Replacing
YY 0637-2008

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms, definitions and abbreviations
  • 3.1 Terms and definitions
  • 3.2 Abbreviations
  • 4 General
  • 4.1 Development
  • 4.2 Testing during installation
  • 5 ACCOMPANYING DOCUMENTS
  • 6 General requirements for operational safety
  • 6.1 Distances and linear, and angular dimensions
  • 6.2 RADIATION quantities
  • 6.3 Date and time format
  • 6.4 Protection against unauthorized use
  • 6.5 Data limits
  • 6.6 Protection against unauthorized modification
  • 6.7 Correctness of data transfer
  • 6.8 Coordinate systems and scales
  • 6.9 Saving and archiving data
  • 7 RADIOTHERAPY TREATMENT EQUIPMENT MODELLING and BRACHYTHERAPY SOURCE MODELLING
  • 7.1 EQUIPMENT MODEL
  • 7.2 BRACHYTHERAPY SOURCE MODEL
  • 7.3 Dosimetric information
  • 7.4 EQUIPMENT MODEL, BRACHYTHERAPY SOURCE MODEL acceptance
  • 7.5 EQUIPMENT MODEL, BRACHYTHERAPY SOURCE MODEL deletion
  • 8 ANATOMY MODELLING
  • 8.1 Data acquisition
  • 8.2 Coordinate systems and scales
  • 8.3 Contouring of regions of interest
  • 8.4 PATIENT ANATOMY MODEL acceptance
  • 8.5 PATIENT ANATOMY MODEL deletion
  • 9 TREATMENT PLANNING
  • 9.1 General requirements
  • 9.2 TREATMENT PLAN preparation
  • 9.3 TREATMENT PLAN identification
  • 9.4 TREATMENT PLAN deletion
  • 9.5 Electronic signatures
  • 10 ABSORBED DOSE distribution calculation
  • 10.1 Algorithms used
  • 10.2 Accuracy of algorithms
  • 11 TREATMENT PLAN report
  • 11.1 Incomplete TREATMENT PLAN report
  • 11.2 Information on the TREATMENT PLAN report
  • 11.3 Transmitted TREATMENT PLAN information
  • 12 General hardware diagnostics
  • 13 Data and code
  • 14 Human errors in software design
  • 15 Change in software versions
  • 16 USE ERRORS
  • Annex A (Normative) Hardware safety
  • Annex B (Informative) Imported and exported data
  • Bibliography

Foreword

This Standard is drafted according to the rules specified in GB/T 1.1-2009.

The Standard replaces YY0637-2008 Medical Electrical Equipment - Requirements for the Safety of Radiotherapy Treatment Planning Systems. Compared with YY0637-2008, the main technical changes of this Standard are as follows.

- Add 3.2 Abbreviations;

- Delete Chapter 3 of the previous standard;

- Split the 8.1 of the previous standard INTO 7.1 and 7.2 of this Standard;

- Delete Chapter 14 of the previous standard.

The Standard is identical to IEC62083:2009 Medical Electrical Equipment - Requirements for the Safety of Radiotherapy Treatment Planning Systems.

Our national documents that are consistent with the international documents of the Normative References of this Standard are as follows. [Translator. NIL here] For ease of use, this Standard is modified as follows.

-- Modify the "international standard" to "this Standard";

-- Replace the comma "," with a decimal point ".";

-- Delete the foreword of international standard. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 38 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY 0637

EditionTitleRevisionStatus
YY 0637-2013Medical electrical equipment—Requirements for the safety of radiotherapy treatment planning systemscurrent editionCurrent
YY 0637-2008Medical electrical equipment—Requirements for the safety of radiotherapy treatment planning systemsprevious editionSuperseded

This page sells the current edition, YY 0637-2013. Earlier editions are listed for reference only.

How to Download YY 0637-2013

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
38 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0637-2013

$125.00