YY 0832.1-2011Stereotactic and planning system for radiotherapy with X-Radiation - Part 1: Stereotactic and planning system for radiotherapy with X-Radiation for head lesion (English PDF)
X射线放射治疗立体定向及计划系统 第1部分:头部X射线放射治疗立体定向及计划系统
This is a limited preview
Buy now to download the full PDF (18 pages)
Need quotation for your management?
Get a formal quotation for YY 0832.1-2011 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
December 31, 2011
Implementation date
June 1, 2013
Scope
YY 0832.1-2011 is the English-translated version of X射线放射治疗立体定向及计划系统 第1部分:头部X射线放射治疗立体定向及计划系统.
YY 0832.1-2011 is the Chinese pharmaceutical industry standard for the stereotactic and planning system used in X-radiation radiotherapy of lesions of the head.
It is the first part of the YY 0832 series and specifies its field of application, the terms and the performance requirements, covering the stereotactic frame and the accuracy of the coordinates it defines, the imaging data and their reconstruction, the accuracy of the dose calculation and of the target definition, and the transfer of the plan to the treatment machine and its verification. Issued on 31 December 2011, it has applied since 1 June 2013.
Found what you need?
Document preview — YY 0832.1-2011
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Terms and Definitions5
- 4 Requirements5
- 5 Test Methods7
1 Scope
This Part of YY 0832 specifies the scope, terms, performance requirements of the stereotactic and planning system for radiotherapy with X-radiation for head lesion.
This Part is applicable to the stereotactic and planning system for radiotherapy with X- radiation for head lesion (hereinafter referred to as the system). Such system shall be used in conjunction with a medical electron accelerator to perform stereotactic radiotherapy for small lesions in the head and neck.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.
GB 9706.5 Medical Electrical Equipment - Part 2. Particular Requirements for the Safety of Electron Accelerators in the Range 1 MeV to 50 MeV
GB 15213 Medical Electron Accelerators - Functional Performance Characteristics and Test Methods
GB/T 17857 Medical Radiology - Terminology (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry)
GB/T 18987 Radiotherapy Equipment - Coordinates, Movements and Scales
YY 0637 Medical Electrical Equipment - Requirements for the Safety of Radiotherapy Treatment Planning Systems
3 Terms and Definitions
For the purposes of this document, the terms and definitions given in GB 9706.5, GB 15213, GB/T 17857 apply.
4 Requirements
4.1 Coordinates
The system shall adopt the requirements for coordinates, movements and scales in GB/T 18987. If other coordinates are adopted, they shall be indicated in the random file as required in 4.2a).
4.2 Random file
The random file shall include at least the following contents.
a) The coordinates adopted by the system; as well as the relationship between such coordinates and the coordinates defined in GB/T 18987;
b) The medical electron accelerator (hereinafter referred to as the accelerator) used in conjunction with shall meet the requirements of GB 15213 and GB 9706.5;
c) The energy range of the accelerator used in conjunction with;
d) The parameters of collimator, including the quantity, No., height, nominal dimensions of the two end faces;
e) The size of the radiation filed for the accelerator used in conjunction with, after assembly of the collimator;
f) The size of radiation field of each collimator along the direction of the two main axes of the isocenter of accelerator used in conjunction with;
g) The value of penumbra of each collimator along the direction of the two main axes of the isocenter of accelerator used in conjunction with;
h) The parameters, dimensions, and relevant information of the stereotactic device used in conjunction with;
i) In order to meet the isocenter accuracy and other requirements of the accelerator used in conjunction with as required in 4.6;
j) In order to meet the imaging parameters required in 4.8, as well as the test methods whether the image satisfies the therapy plan.
The target reconstruction position error calculated by the therapy planning software shall be no greater than 1.5mm.
4.8.3 Area coincidence rate
The coincidence rate BETWEEN the area enclosed by 80% isodose line calculated by the therapy planning software AND the area enclosed by the film calculated under the same conditions shall be greater than 90%.
4.9 Security requirements of the therapy planning software
The therapy planning software in the system shall meet the requirements of YY 0637.
5 Test Methods
5.1 Coordinates
The inspection of the random file shall meet the requirements of 4.1.
5.2 Requirements for the random file
The inspection of random file shall meet the requirements of 4.2.
5.3 Size deviation of radiation field
It shall be performed by the following methods.
a) Select a collimator for therapy; place the film in a plane that is through the isocenter and perpendicular to the radiation beam axis; the equivalent water mold thickness on the film is at least 5cm; while the equivalent water mold thickness below the film is 5cm. Set the angle between accelerator frame and the speed-limiting system into 0°; select the appropriate energy and dose for the accelerator used in conjunction with; sequentially expose the film from the lower to the higher dose; so that calibrate the film gray scale value corresponding to different absorbed dose.
b) Use the appropriate mathematical model according to the absorbed dose value of the irradiation and the gray scale value of the corresponding film; draw a gray scale-dose response curve of the used radiation film;
c) According to the conditions of a), select appropriate dose to sequentially expose the collimator of each size; the irradiation dose shall be within the appropriate response area of the gray scale-dose response curve of the used film; process the film after exposure;
Yi - coordinate along the Y-axis measure after the ith repetitive positioning, in mm;
Z0 - coordinate along the Z-axis measured at the first time, in mm;
Yi - coordinate along the Z-axis measure after the ith repetitive positioning, in mm
5.8 Performance of therapy planning software
5.8.1 Dose calculation error of therapy planning software
It shall be performed as per the following methods.
a) The ionization chamber used for measurement shall meet the requirements of the measuring conditions. Use head model, separately take the three references points, i.e. model center and two isocenters as the targets to do three therapy plans; set the radiation dose of the target within the high-dose and low-gradient area; and finish the design of the therapy plan;
b) Insert the ionization chamber into the model; so that its effective measurement points overlap with the target;
c) Separately place the three targets at the accelerator isocenters; use the collimator adopted in the therapy planning and the radiation conditions to irradiate; measure the absorbed dose of the three targets;
d) Calculate the error between the absorbed dose value calculated by the therapy planning software as per the Formula (4) and the actually-measured absorbed dose value; the maximum value shall meet the requirements of 4.8.1. Where.
deltaP - the relative percentage error between he actually-measured absorbed dose value and the absorbed dose value calculated by the therapy planning, in %;
P0 - absorbed dose value of target calculated by the therapy planning software, in Gy;
Pi - actually-measured absorbed dose value (i=1~3) of target, in Gy.
5.8.2 Target position calculation error of the therapy planning software
It shall be performed as per the following methods. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.5Medical electrical equipment -- Part 2-1: Particular requirements for the safety of electron accelerators in the range 1 MeV to 50 MeV
- GB 15213Medical electron accelerators-Functional performance characteristics and test methods
- GB/T 17857Medical radiology. Terminology (Equipment for radiotherapy, nuclear medicine and radiation dosimetry)
- GB/T 18987Radiotherapy equipment -- Coordinates, movements and scales
- YY 0637Medical electrical equipment—Requirements for the safety of radiotherapy treatment planning systems
Similar standards
How to Download YY 0832.1-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 15213-2016 — Medical electron accelerators-Functional performance characteristics and test methods
GB 9706.5-2008 — Medical electrical equipment -- Part 2-1: Particular requirements for the safety of electron accelerators in the range 1 MeV to 50 MeV
GB/T 17857-1999 — Medical radiology. Terminology (Equipment for radiotherapy, nuclear medicine and radiation dosimetry)