Valid

YY/T 0973-2016Radiotherapy treatment planning system for automatically- controlled brachytherapy afterloading equipment-Characteristics and test (English PDF)

自动控制式近距离后装设备放射治疗计划系统性能和试验方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0973-2016 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

March 23, 2016

Implementation date

January 1, 2017

Scope

YY/T 0973-2016 is the English-translated version of 自动控制式近距离后装设备放射治疗计划系统性能和试验方法.

YY/T 0973-2016 is the Chinese pharmaceutical industry standard for the treatment planning system of an automatic remote afterloading brachytherapy unit.

It sets the performance requirements and the test methods, covering the accuracy of the dose calculation around the source, the reconstruction of the applicator and catheter geometry, the dwell positions and times, the handling of the source data and the plan transfer and verification.

It applies to planning systems that use an iridium-192 afterloading source. Issued on 23 March 2016, it has applied since 1 January 2017.

Found what you need?

Document preview — YY/T 0973-2016

National Standard of the People's Republic of China

ICS
11.040.60
Classification
C 43

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Preface
  • 1 Scope
  • 2 normative reference document
  • 3 Terms and definitions
  • 4 Requirements
  • 4.1 length reconstruction deviation
  • 4.2 Volume Reconstruction Deviation
  • 4.3 Source Length Reconstruction Geometric accuracy
  • 4.4 Dose calculation deviation
  • 4.5 dose distribution accuracy
  • 4.6 random file
  • 5 Test Method
  • 5.1 Length Reconstruction Deviation
  • 5.2 Volume Reconstruction Deviation
  • 5.3 Source length reconstruction Geometric accuracy
  • 5.4 Dose calculation deviation
  • 5.5 Dose Distribution Accuracy
  • 5.6 random file
  • Appendix A (informative) Description of the dose rate formula
  • Appendix B (informative) 192Ir Dose data

Foreword

This standard is drafted in accordance with the rules given in GB/T 1.1-2009.

1 Scope

This standard specifies the performance of the automated controlled follow-up treatment equipment after radiotherapy planning system (hereinafter referred to as the treatment planning system) And test methods.

This standard applies to the treatment planning system using a 192Ir radioactive back-up treatment source.

For the use of other back-up treatment sources or other packaging sources, refer to the use of this standard.

2 Normative reference documents

The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.

GB/T 17857 Medical Radiology Terminology (Radiation Therapy, Nuclear Medicine and Radiation Dosage Equipment)

3 Terms and definitions

GB/T 17857 Definitions of terms and definitions apply to this document.

4 Requirements

4.1 Length reconstruction bias

The length of the treatment planning system should not be greater than 1 mm.

4.2 Volume Reconstruction Deviation

The volume of the treatment planning system should not be greater than 5%.

4.3 Source Length Reconstruction Geometric accuracy

Choose a different source device, the source source of the source run track in line with the source shape, reconstruction geometry should not be greater than ± 1mm.

4.4 Dose calculation deviation

The deviation between the calculated and the nominal values of the treatment planning system should not be greater than 5%.

4.5 Dose distribution accuracy

The area covered by the specific isodose curve calculated by the treatment planning system is not equal to the area covered by the corresponding isodose curve Should be less than 90%.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 0973-2016

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0973-2016

$235.00