YY/T 0706-2017Particular specifications for mammographic X-ray equipment (English PDF)
乳腺X射线机专用技术条件
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0706-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 28, 2017
Implementation date
April 1, 2018
Scope
YY/T 0706-2017 is the English-translated version of 乳腺X射线机专用技术条件.
YY/T 0706-2017 is the Chinese pharmaceutical industry standard for the mammography machine.
It gives the terms and definitions, the classification and composition, the requirements and the test methods, covering the X-ray tube and its focal spots, the accuracy and reproducibility of the loading factors and of the automatic exposure control, the compression device and its force and display, the imaging performance in terms of resolution and contrast, the average glandular dose and the radiation protection of the operator.
It does not apply to the biopsy, stereotactic and tomosynthesis attachments of such a machine. Issued on 28 March 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 0706-2017
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C 43
- Replacing
- YY/T 0706-2008
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Classification and composition
- 5 Requirements
- 6 Test method
- Appendix A (informative) Low contrast test phantom
- Appendix B (informative) Breast imaging phantom
- Appendix C (Normative Appendix) Mammary mean dose gland calculation
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces YY/T 0706-2008 "Technical Conditions for Mammography X-ray Machines".
Compared with YY/T 0706-2008, the main technical changes of this standard are as follows.
--- Modified the "scope" (see Chapter 1);
---Modified "normative references" (see Chapter 2);
---Modified "terms and definitions" (see Chapter 3);
--- Modified "Classification and Composition" to increase the content of digital mammography machines (see Chapter 4);
--- Revised "specified by product standards" to "specified by the manufacturer" (see Chapter 5, Chapter 6);
--- Modified "Load factor control and display" (see 5.3);
--- Modified the "Automatic Exposure Control (AEC) System" (see 5.4.1);
--- Modified "artifacts" (see 5.4.2);
---Modified the "Imaging Performance" section (see 5.4.3, 5.4.4, 5.4.5, 5.4.6, 5.4.7, 5.4.8);
---Modified "mechanical movement" (see 5.5.1);
---Modified "compression device" (see 5.5.2);
--- Modified the "pressure plate" (see 5.5.3);
---Modified "braking force" (see 5.5.4);
---Modified the "mobile performance of mobile mammography machine" (see 5.5.7);
--- Modified "bearing weight" (see 5.5.8);
---Modified "noise" (see 5.5.9);
--- Added "angle indication value" (see 5.5.6);
---Modified "high-voltage cable plugs and sockets" (see 5.7);
--- Removed "Continuous loading 3min";
---Modified part of "Load factor control and display" (see 6.3);
--- Modified the "Automatic Exposure Control (AEC) System" (see 6.4.1);
--- Modified "artifacts" (see 6.4.2);
--- Removed "total filtration", "material decay rate between the upper surface of the breast tray and the image receptor plane", "active anti-scatter filter "Dilution of wire shadow", "high contrast resolution";
--- Added "afterimage", "anti-scatter filter grid", "spatial resolution low contrast resolution", "mammary gland dose", "digital Image uniformity" (see 6.4.3, 6.4.4, 6.4.5, 6.4.6, 6.4.7, 6.4.8);
---Modified "mechanical device performance" (see 6.5.1, 6.5.2, 6.5.3, 6.5.4, 6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9);
---Modified "high-voltage cable plugs, sockets", "environmental requirements", "safety and electromagnetic compatibility requirements" (see 6.7, 6.9, 6.10);
--- Added Appendix A, Appendix B and Appendix C.
1 Scope
This standard specifies the terms and definitions, classification and composition, requirements and test methods of mammography machines (hereinafter referred to as mammography machines).
This standard applies to mammography machines.
This standard does not apply to biopsy devices, stereotactic devices and tomographic devices of mammography machines.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety
GB 9706.11 Medical electrical equipment - Part 2. Safety requirements for medical diagnostic X-ray source components and X-ray tube assemblies
GB 9706.12 Medical electrical equipment Part 1. Safety general requirements III. Parallel standard diagnostic X-ray equipment radiation protection common enquiries
GB 9706.14 Medical electrical equipment - Part 2. Particular requirements for safety of X-ray equipment
GB 9706.15 Medical electrical equipment - Part 1-1 - Safety common requirements - Parallel standard. Medical electrical system safety requirements
GB 9706.24-2005 Medical electrical equipment - Part 2-45. Mammography equipment and mammography stereo positioning device Safety specific requirements
GB/T 10149 medical X-ray equipment terminology and symbols
GB/T 17006.9 Evaluation and routine testing of medical imaging systems - Part 2-10. Stability test mammography device
GB/T 19042.2-2005 Evaluation and routine testing of the medical imaging sector - Part 3-2. Mammography X-ray equipment Performance acceptance test
YY 0076-1992 Technical conditions for coating classification of metal parts
YY/T 0106 General technical conditions for medical diagnostic X-ray machines
YY/T 0291 Environmental requirements and test methods for medical X-ray equipment
YY/T 0480 Diagnostic characteristics of X-ray imaging equipment and mammography anti-scatter grid
YY 0505 Medical Electrical Equipment - Part 1-2. Safety General Requirements Parallel Standards. Electromagnetic Compatibility Requirements and Testing
YY/T 1541 Mammography X-ray machine high voltage cable assembly and socket technical conditions
3 Terms and definitions
The following terms and definitions as defined in GB/T 10149 apply to this document. A device that supports components such as X-ray tube heads.
3.2 Compression device
A special device for compressing the breast on a mammary machine, including a compression plate, a patient support, and a device for controlling compression movement. A transparent plate on the mammary machine that is used to compress the patient's breast.
3.4 Average gland dose averageglandulardose; AGD
Using the prescribed calculation method, the average absorption of the breast gland tissue (excluding the skin) whose known composition is uniformly compressed is calculated. dose. A physical quantity that describes the ability of a system to distinguish between different contrast objects and the inherent noise of an image in a digital image.
3.6 Artifact
A structure that is clearly visible on an image. It does not reflect the internal structure of the object, nor can it be explained by noise or system modulation transfer functions. [GB/T 19042.2-2005, definition 3.3.1]
3.7 Afterimage erasurethoroughness
Part of the previous image that was generated in the subsequent image due to incomplete signal removal after the previous image signal was read All.
3.8 Unprocessed image
After pixel defect correction and field correction, the image is not post-image processed.
4 Classification and composition
4.1 Classification
4.1.1 According to the product structure, it can be divided into fixed and mobile mammography machines.
4.1.2 According to the image data output format, it can be divided into screen (analog) and digital mammography.
4.2 Composition
The mammography machine should have at least the following main components.
a) X-ray generating device;
b) the projection frame;
c) compression device;
d) image receiving device;
e) Breast image acquisition workstation (for digital mammography).
5 Requirements
5.1 Working conditions
5.1.1 Environmental conditions
Unless otherwise specified, the working environment of the mammary machine should meet.
a) Ambient temperature. 10 ° C ~ 40 ° C;
b) Relative humidity. 30%~75%;
c) Atmospheric pressure. 700hPa~1060hPa.
5.1.2 Power conditions
The manufacturer shall state in the random document the power conditions used for the product, and the operating power conditions shall satisfy.
a) power supply voltage and number of phases. specified by the manufacturer, the network voltage fluctuations should not exceed ± 10% of the nominal value;
b) Power frequency. 50Hz ± 1Hz;
c) Power supply resistance. It should meet the requirements of 10.2.2 of GB 9706.24-2005;
d) Power supply capacity. specified by the manufacturer.
5.2 Electric power
5.2.1 Maximum output electric power
The mammography machine should specify the corresponding combination of X-ray tube voltage and X-ray tube current that results in the maximum output electrical power.
5.2.2 Nominal electric power
It should be specified in kilowatts (kW) that the X-ray generator can provide when the loading time is 0.1 s and the X-ray tube voltage is 30 kV.
The unit's maximum constant electrical power output is given as the nominal electrical power. If this value cannot be pre-selected, the X shot closest to 30kV can be used.
The line voltage value and the closest load time value, but not less than 0.1s.
The nominal electrical power should be given in combination with the X-ray tube voltage and X-ray tube current and load time.
5.3 Loading factor control and display
5.3.1 X-ray tube voltage
5.3.1.1 The mammography machine shall specify the voltage adjustment range and adjustment mode of the X-ray tube. The maximum X-ray tube voltage value shall not exceed 50kV.
5.3.1.2 The manufacturer shall specify the deviation of the voltage value of the X-ray tube, but at least it shall comply with 50.103.1a of GB 9706.24-2005. Claim.
5.3.2 X-ray tube current
5.3.2.1 If the X-ray tube current display is provided, the mammography machine shall specify the X-ray tube current adjustment range and adjustment method.
5.3.2.2 The manufacturer shall specify the deviation of the X-ray tube current value, but at least it shall comply with the requirements of 50.103.2 of GB 9706.24-2005.
5.3.3 Loading time
5.3.3.1 If the loading time display is provided, the mammography machine shall specify the loading time adjustment range and adjustment method.
5.3.3.2 The manufacturer shall specify the deviation of the load time value, but at least it shall comply with the requirements of 50.103.3 of GB 9706.24-2005.
5.3.4 Current time product
5.3.4.1 If the current time product is displayed, the mammography machine shall specify the current time product adjustment range and adjustment mode.
5.3.4.2 The manufacturer shall specify the deviation of the current time product value, but at least it shall comply with the requirements of 50.103.4 of GB 9706.24-2005.
5.3.5 Anti-overload
The mammography machine should have anti-overload measures to ensure that the loading factor selection does not exceed the rated capacity of the X-ray tube. Should comply with the instruction manual The combination of the maximum load factors given in.
5.3.6 Focus
5.3.6.1 The manufacturer shall specify the nominal value of the X-ray tube assembly.
5.3.6.2 When using the normal exposure mode, the nominal value of the X-ray tube assembly should not be greater than 0.4; when using the magnified photographic exposure mode, X-ray The nominal value of the conduit assembly focus should not be greater than 0.2.
5.3.6.3 When multiple focus is provided, the preselected focus shall be displayed prior to exposure, except that the automatic exposure control (AEC) is used to adjust the focus.
5.3.6.4 If the focus is automatically selected, the focus of the actual exposure should be displayed after exposure.
5.3.7 Target surface material
5.3.7.1 The mammography machine shall specify the target surface material used for the X-ray tube.
5.3.7.2 When multiple targets are provided, in addition to using AEC to modulate the target, the preselected target should be displayed prior to exposure.
5.3.7.3 If the target is automatically selected, the actual exposed target should be displayed after exposure.
5.3.8 Additional filtration
5.3.8.1 The mammography machine shall specify the additional filtration used.
5.3.8.2 When multiple filtrations are provided, in addition to using AEC to adjust for additional filtration, a pre-selected additional filtration shall be displayed prior to exposure.
5.3.8.3 If additional filtering is automatically selected, additional filtering of the actual exposure should be displayed after exposure.
5.4 Imaging performance
5.4.1 Automatic Exposure Control (AEC) System
5.4.1.1 The mammography machine shall provide an automatic exposure control (AEC) system.
5.4.1.2 The implementation of the automatic exposure control function should be described in the random file.
5.4.1.3 The repeatability of the digital mammography automatic exposure control is measured by the deviation of the dose or milliampere second, and the deviation value should not be greater than ±5%.
5.4.1.4 The manufacturer shall specify the automatic exposure of the digital mammography machine to different thicknesses of polymethyl methacrylate (PMMA) phantoms. The contrast-to-noise ratio (CNR) minimum of the response.
5.4.2 Artifacts
There should be no visible artifacts.
5.4.3 Afterimage
For digital mammography, the residual coefficient value should be less than 0.3.
5.4.4 Anti-scatter grid
5.4.4.1 The configuration of the anti-scatter grid should be given in the random file (if there is a grid or no gate, moving grid or static grid).
5.4.4.2 There shall be no visible visible shadows within the specified loading time and/or phantom thickness range.
5.4.4.3 When the maximum compressive force is applied to the dynamic grid, the movement of the grid shall not be affected.
5.4.4.4 In the magnification photography mode, the grid should be removed without being left in the radiation beam.
5.4.5 Spatial resolution
5.4.5.1 The spatial resolution of the screen type mammography machine should not be less than 12 lp/mm.
5.4.5.2 The spatial resolution of the digital mammography machine should not be less than 5 lp/mm.
5.4.6 Low contrast resolution
For digital mammography machines, manufacturers should specify low contrast resolution.
5.4.7 Average glandular gland dose
5.4.7.1 The mammography should show the average glandular dose of the mammary gland per irradiation.
5.4.7.2 The digital mammography machine shall display the average gland dose for each captured image.
5.4.7.3 The average gland dose in the normal photography mode of the mammary gland should meet the limits in Table 1.
5.4.8 Digital image uniformity
The maximum deviation of the gray value between the central area and the surrounding area specified by the image receptor shall not exceed ±10%.
5.5 Mechanical performance
5.5.1 Mechanical movement
5.5.1.1 The range of travel of the projection frame in the vertical direction shall be specified, and the deviation of the stroke range shall be less than ± 5%.
5.5.1.2 The angle of rotation of the projection frame and the deviation value shall be specified. The angular deviation value shall be less than ±2°.
5.5.1.3 The projection bracket shall be locked at any position on the full stroke.
5.5.1.4 Focus---The distance of the image receiving surface (FID) should not be less than 600mm.
5.5.1.5 When the power supply is interrupted, the mechanical movement is locked except for the compression plate.
5.5.2 Compression device
5.5.2.1 The pressure should be stable when under pressure, and the pressure value should not exceed 10% within at least 1 min.
5.5.2.2 Where applicable, the equipment is capable of displaying the applied force and its indication value deviation shall be less than ±20N.
5.5.3 Pressure plate
The compression plate shall comply with the relevant requirements of GB 9706.24-2005.
5.5.4 Braking force
The linear motion and the rotational motion part of the mechanical device should have a braking device (the suspension device requires the balancing without the need for braking). Outside), its braking force should not be less than 100N.
5.5.5 Length indication value
The deviation between the indicated value of the length (excluding the compression thickness) and the actual value shall not be greater than ± 5% of the indicated value.
5.5.6 Angle indication value
The deviation of the indicated value of the angle from the actual value is specified by the manufacturer.
5.5.7 Mobile performance of mobile mammography
The human-moving mammography machine moves on a flat concrete floor, and the starting push (pull) force applied to the handlebar should not exceed 250N.
5.5.8 Load bearing
Devices that support adult patients (eg, pedals, chairs, beds) can work properly with 100kg of mass.
5.5.9 Noise
When the mammography machine is running under no-load conditions (non-load-bearing state), the noise generated should not exceed 70dB (A-weighted network) (not including 3s) Non-persistent and aperiodic noise within).
5.6 Software Features
If applicable, the manufacturer shall specify the functionality of the operating software.
5.7 High voltage cable plug and socket
Unless otherwise specified, the requirements for type and basic dimensions, marking, and connection in YY/T 1541 shall be met.
5.8 Appearance
5.8.1 The shape of the mammary machine should be neat and beautiful, the surface should be smooth and smooth, the color should be uniform, and there should be no defects such as scars and cracks.
5.8.2 The main plating parts of the mammography machine shall comply with the requirements of Class 2 appearance in YY 0076-1992.
5.9 Environmental test
Unless otherwise specified, the product shall comply with the requirements of YY/T 0291. The initial, intermediate or final test item shall include at least 5.3.1 a), 5.3.2 a), 5.3.3 a), 5.3.4 a), 5.5.4 (when electrical braking) and 5.6 requirements.
5.10 Safety and Electromagnetic Compatibility Requirements
Should comply with GB 9706.1, GB 9706.11, GB 9706.12, GB 9706.14, GB 9706.15, GB 9706.24 and YY 0505 Claim.
6 Test methods
6.1 Test conditions
6.1.1 Environmental conditions
Should comply with the provisions of 5.1.1.
6.1.2 Power conditions
The test power supply conditions for the mammography machine are as follows.
a) The grid voltage and phase number shall comply with the manufacturer's specifications, and the grid voltage fluctuation shall not exceed ± 5% of the nominal value;
b) Power frequency. 50Hz ± 1Hz;
c) The power supply resistance shall comply with the provisions of 5.1.2c);
d) The power supply capacity shall comply with the provisions of 5.1.2d).
6.2 Electric power
6.2.1 Maximum output electric power
Loads are combined according to the loading factors that cause the maximum output power, and the presence or absence of abnormalities is observed.
6.2.2 Nominal electric power
The combination of X-ray tube voltage, X-ray tube current, and loading time leading to the nominal electric power was observed to observe whether there was any abnormality.
6.3 Loading factor control and display
6.3.1 X-ray tube voltage
6.3.1.1 X-ray tube voltage adjustment According to the X-ray tube voltage adjustment range and adjustment method specified in 5.3.1.1, observe the actual operation.
6.3.1.2 The deviation test of the X-ray tube voltage value shall be carried out in accordance with the provisions of 50.104.1 of GB 9706.24-2005.
6.3.2 X-ray tube current
6.3.2.1 X-ray tube current adjustment According to the X-ray tube current adjustment range and adjustment method specified in 5.3.2.1, observe the actual operation.
6.3.2.2 Deviation test of X-ray tube current value shall be carried out in accordance with the provisions of 50.104.2 of GB 9706.24-2005.
6.3.3 Loading time
6.3.3.1 Loading time adjustment According to the loading time adjustment range and adjustment method specified in 5.3.3.1, observe the actual operation.
6.3.3.2 Deviation test of loading time value shall be carried out in accordance with the provisions of 50.104.3 of GB 9706.24-2005.
6.3.4 Current time product
6.3.4.1 Current time product adjustment According to the current time product adjustment range and adjustment method specified in 5.3.4.1, observe the actual operation.
6.3.4.2 The deviation test of current time product shall be carried out in accordance with the provisions of 50.104.4 of GB 9706.24-2005.
6.3.5 Anti-overload
Adjust the parameters of the mammography machine to the maximum load factor combination value specified in the instruction manual, and then adjust any load factor to increase adjacently.
When the gear is added, the loading factor condition is no longer increased or is in an overload state. For mammography machines with continuous adjustment of X-ray tube voltage, anti-overload When the operating voltage value of the device is higher than the X-ray tube voltage value specified in the instruction manual (but not exceeding 5kV), it should not be exposed and the alarm is displayed.
6.3.6 Focus
6.3.6.1 Check the random file check.
6.3.7 Target surface material
6.3.7.1 Check the random file check.
6.3.8 Additional filtration
6.3.8.1 Check the random file check.
6.4 Imaging performance
6.4.1 Automatic exposure control (AEC) system
6.4.1.3 Place a 40mm thick PMMA phantom on the patient support, with the phantom edge aligned with the chest wall side of the patient support; compression plate pressure On the phantom; set the commonly used tube voltage and target/filtering for AEC exposure; repeat exposure 5 times to calculate the current time of each measurement The maximum deviation of the product or dose value from the corresponding mean.
6.4.1.4 Place 20mm, 40mm, 60mm thick PMMA phantoms on the patient support, phantom edge and patient support chest The wall side is aligned, the compression plate is pressed on the phantom, AEC exposure is performed, and the exposure parameters such as tube voltage, current time product and target/filtering are recorded. Will be thick An aluminum sheet having a degree of 0.20±0.01 mm (purity of not less than 99.9% and a size of 30 mm×30 mm) is placed on the upper surface of the PMMA phantom, so that Manual exposure with the same exposure parameters as before (if not exactly the same, the closest exposure parameter is used). When getting each thickness The unprocessed image of the exposure, respectively measuring the area of interest of the area of 400 mm² in the area 2 shown in Fig. 1, the gray value of the aluminum image Mean and standard deviation (mAl, sigmaAl) and an area of interest of 400 mm² in region 1, the mean and standard deviation of the gray value of the background image (mBG, sigmaBG), calculate the CNR value according to equation (1). mBG-mAl sigma2BG sigma2Al (1)
6.4.1.2 Review random files
Note. Automatic exposure control should be provided in one possible combination, for example. with grid, without grid, magnified photography, non-magnified photography and various targets/filters The combination is used to achieve its function; in addition, the probe for measuring the dose should avoid the sensor that affects the AEC.
In the formula. mAl--- is the average value of the gray value of the region 2;
sigmaAl --- is the standard deviation of the gray value of the region 2; mBG--- is the mean value of the gray value of the region 1;
sigmaBG--- is the standard deviation of the grayscale value of the region 1.
Figure 1 Schematic diagram of CNR measurement
6.4.2 Artifacts
a) Screen type mammography machine is carried out according to the provisions of 5.10.2 of GB/T 19042.2-2005.
b) Digital mammography machine, placing a uniform filter plate i...
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB 9706.12Medical electrical equipment Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
- GB 9706.15Medical electrical equipment -- Part 1: General requirements for safety. 1 -- Collateral standard: Safety requirements for medical electrical systems
- GB/T 10149Terminology and symbol for medical X-ray equipment
- GB/T 17006.9Evaluation and routine testing in medical imaging departments -- Part 2-10: Constancy tests -- X-ray equipment for mammography
- GB/T 19042.2-2005Evaluation and routine testing in medical imaging departments - Part 3-2: Mammographic X-ray equipment imaging performance acceptance tests
GB 9706.11 · GB 9706.14 · GB 9706.24-2005
Similar standards
Editions of YY/T 0706
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0706-2017 | Particular specifications for mammographic X-ray equipment | current edition | Current |
| YY/T 0706-2008 | Particular specifications for mammographic X-ray equipment | previous edition | Superseded |
This page sells the current edition, YY/T 0706-2017. Earlier editions are listed for reference only.
How to Download YY/T 0706-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB 9706.12-1997 — Medical electrical equipment Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
GB 9706.15-2008 — Medical electrical equipment -- Part 1: General requirements for safety. 1 -- Collateral standard: Safety requirements for medical electrical systems