YY/T 1142-2013Methods of measuring the frequency of medical ultrasonic equipment and probe (English PDF)
医用超声设备与探头频率特性的测试方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1142-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 1142-2013 is the English-translated version of 医用超声设备与探头频率特性的测试方法.
YY/T 1142-2013 is the Chinese pharmaceutical industry standard that gives the methods for measuring the frequency of medical ultrasonic equipment and of its probes, over the frequency range it states.
It sets out the hydrophone and the measuring arrangement, the acquisition of the pressure waveform and its spectral analysis, and the way the acoustic working frequency and the bandwidth are derived and reported. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY/T 1142-2013
National Standard of the People's Republic of China
- Replacing
- YY/T 1142-2003
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Test conditions6
- 5 Test methods8
1 Scope
This standard specifies the testing methods for the frequency characteristics of medical ultrasonic equipment (hereinafter referred to as equipment) and probes with a frequency range of 0.5 MHz~15 MHz, and the calculation method of relevant parameters.
This standard applies to all types of medical ultrasonic equipment and probes operating in continuous wave, quasi-continuous wave, or pulse wave states.
2 Normative references
The following documents are essential for the application of this document. For the dated referenced documents, only the versions with the indicated dates are applicable to this document; for the undated referenced documents, only the latest version (including all the amendments) is applicable to this document.
YY/T 0163 Characteristics and calibration of hydrophones for medical ultrasonic fields
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Nominal frequency
The ultrasonic working frequency of an ultrasonic transducer or ultrasonic transducer array tuple given by the designer or manufacturer.
3.2 Acoustic-working frequency
The actual ultrasonic operating frequency obtained from the analysis in which the hydrophone is placed at an appropriate position in the sound field and its output signal is analyzed using the zero-crossing frequency method or spectrum analysis method (see Figure 1). The arithmetic-mean acoustic-working frequency is used in this standard. Symbol: fawf
a) ambient temperature;
b) supply voltage;
c) Driving power setting (if any).
4 Test conditions
4.2 Test equipment
4.2.1 Hydrophone
It shall meet the requirements of Class A or Class B needle-shaped (Z-type) or membrane-type (M-type) hydrophones specified in YY/T 0163. The size of the sensitive element is <=1 mm, the frequency response is 0.5 MHz~15 MHz, and the flatness is ±3 dB. For some equipment that cannot provide the synchronized trigger signal required for measurement, an auxiliary hydrophone (Class A or B) may be required to trigger the oscilloscope.
4.2.2 Oscilloscope
The frequency range is not narrower than 0~100 MHz, and the sensitivity is not lower than 5 mV/div.
4.2.3 Spectrum analyzer
The frequency range is not narrower than 0.1 MHz~30 MHz, and the dynamic range is not less than 60 dB.
4.2.4 Hydrophone amplifier
The frequency range is not narrower than 0~15 MHz, and there is an appropriate amplification factor.
4.2.5 Pulse generator/receiver
Pulse generator: the external pulse leading edge (10%~90%) is < 10 ns, the amplitude is >50 V, the trailing edge is < 50 ns, the repetition frequency is 0~10 kHz, and the internal pulse repetition frequency is 100 Hz~10 kHz. Pulse receiver: the bandwidth is not narrower than 1 kHz~35 MHz(-3 dB).
4.2.6 Water tank and positioning device
The size of the tank shall be such that boundary sound reflection does not have a major impact on the test. If necessary, sound-absorbing materials can be laid at appropriate locations in the tank. A positioning adjustment device equipped with a fixed probe and hydrophone can change the relative position and direction of the probe and hydrophone in three dimensions. The minimum adjustment displacement shall be less than half the wavelength of the sound wave.
b) The ultrasonic output power is controlled at an appropriate position.
c) The scanning angle shall be as small as possible.
5 Test methods
5.2.3 After turning on the ultrasonic output, adjust the orientation and angle of the transducer to maximize the signal received by the hydrophone. If necessary, use a hydrophone amplifier to amplify the signal.
5.2.4 If the equipment under test operates in continuous wave or quasi-continuous wave mode, the frequency measurement function of an oscilloscope and a digital frequency meter can be used directly to measure the acoustic-working frequency without using a spectrum analyzer.
5.2.5 If the equipment under test operates in pulse scanning mode, an additional trigger signal synchronized with each sound repetition period shall be provided. If this cannot be provided, an auxiliary hydrophone can be used to provide an oscilloscope synchronization signal.
5.2.6 Calculation of the arithmetic-mean acoustic-working frequency: Measure f1 and f2 on the sound pressure waveform diagram or sound pressure spectrum diagram (as shown in Figure 1), then the arithmetic-mean acoustic-working frequency is calculated according to formula (1). where: f1 and f2 -- f1 and f2 respectively measured on the sound pressure spectrum diagram (as shown in Figure 1), the unit is megahertz (MHz).
5.2.7 The acoustic-working frequency bandwidth is calculated according to formula (3). where: f1 and f2 -- f1 and f2 respectively measured on the sound pressure spectrum diagram (Figure 1), the unit is megahertz (MHz).
5.2.8 If the equipment has multiple nominal frequencies, the above measurements shall be made separately for each nominal frequency value and the corresponding results shall be given.
5.2.9 The spectrum analysis diagram of the signal at the actual working frequency shall be given. In particular, the component sizes of its second harmonic and other specified […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
Editions of YY/T 1142
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 1142-2013 | Methods of measuring the frequency of medical ultrasonic equipment and probe | current edition | Current |
| YY/T 1142-2003 | Methods of measuring the frequency of medical ultrasonic equipment and probe | previous edition | Superseded |
This page sells the current edition, YY/T 1142-2013. Earlier editions are listed for reference only.
How to Download YY/T 1142-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0163-1994 — Characteristics and calibration of hydrophones for medical ultrasonic fields
YY/T 0163-2005 — Characteristics and calibration of hydrophones for medical ultrasonic fields
YY/T 0297-2026 — General requirements for clinical investigation of medical devices