YY/T 1749-2020External-vibration based ultrasonic elasticity measurement equipment for liver tissue (English PDF)
基于外部振动的肝组织超声弹性测量设备
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1749-2020 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
September 27, 2020
Implementation date
September 1, 2022
Scope
YY/T 1749-2020 is the English-translated version of 基于外部振动的肝组织超声弹性测量设备.
YY/T 1749-2020 is the Chinese pharmaceutical industry standard for equipment that measures the elasticity of liver tissue with ultrasound, in the machines that apply a low-frequency vibration to the skin from outside the body to raise shear waves in the liver and then read them with ultrasonic pulse-echo.
It sets the requirements, the test conditions and the test methods for the depth of ultrasonic penetration, the accuracy and repeatability of the Young's modulus measurement, the functions the device must offer, among them a reminder of whether the pressure applied is right, and what the accompanying documents must state, together with the environmental testing to be done against GB/T 14710 and YY/T 1420.
Tests are run on ultrasonically tissue-mimicking and sono-elasticity phantoms. Issued on 27 September 2020, it has applied since 1 September 2022.
Found what you need?
Document preview — YY/T 1749-2020
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction4
- 1 Scope5
- 2 Normative References5
- 3 Terms and Definitions6
- 4 Requirements7
- 5 Test Conditions9
- 6 Test Methods10
- Appendix A (informative) Classification of Ultrasound Elastography12
- Bibliography14
1 Scope
This Standard specifies the requirements, test conditions and test methods of external- vibration based ultrasonic elasticity measurement equipment for liver tissue (hereinafter referred to as the equipment).
This Standard is applicable to equipment that excites shear waves in liver tissue by externally applying low-frequency vibration and uses the ultrasonic pulse-echo mode to measure the elasticity of liver tissue.
NOTE: the external-vibration based ultrasonic elasticity measurement technology for liver tissue, to which, this Standard is applicable, corresponds to the instantaneous elasticity image in the classification of ultrasound elastography. The classification of the ultrasound elastography is shown in Appendix A.
2 Normative References
The following documents are indispensable to the application of this document. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB 9706.1 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
GB 9706.237 Medical Electrical Equipment - Part 2-37: Particular Requirements for the Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
GB 10152 B Mode Ultrasonic Diagnostic Equipment
GB/T 14710 Environmental Requirement and Test Methods for Medical Electrical Equipment
YY/T 1142 Methods of Measuring the Frequency of Medical Ultrasonic Equipment and Probe
YY/T 1420 Environmental Requirement and Test Methods for Medical Ultrasonic Equipment
3 Terms and Definitions
What is defined in GB 10152, and the following terms and definitions are applicable to this document.
3.1 External-vibration Based Ultrasonic Elasticity Measurement Equipment for Liver Tissue
External-vibration based ultrasonic elasticity measurement equipment for liver tissue refers to the equipment that applies low-frequency vibration with a mechanical device through the skin from outside the body to generate shear wave propagation in the liver tissue, uses the ultrasonic pulse-echo method for measurement, and calculates and displays the elasticity parameters (Young's modulus and / or shear wave velocity, shear modulus) of the liver tissue based on the echo information.
3.2 Ultrasonically Tissue-mimicking Phantom, Ultrasound Tissue
Phantom Ultrasonically tissue-mimicking phantom / ultrasound tissue phantom refers to a B- ultrasonic performance inspection device composed of tissue-mimicking materials and one or more sets of targets embedded in it. [YY/T 0937-2014, Definition 3.1]
3.3 Sono-elasticity Tissue-mimicking Phantom
Sono-elasticity tissue-mimicking phantom refers to a passive device that is composed of ultrasonic elasticity tissue-mimicking materials in different modes and used to detect the performance of ultrasonic elasticity image diagnostic equipment. [YY/T 1521-2017, Definition 3.1]
3.4 Depth of Ultrasonic Penetration
Depth of ultrasonic penetration refers to the distance between the farthest target line and the acoustic window in the longitudinal line-shaped target group that can be clearly imaged (or can be displayed as an echo signal along the ultrasonic beam axis) in the ultrasonically tissue-mimicking phantom.
NOTE: unit: (mm).
3.5 Measurement Accuracy
Measurement accuracy refers to the consistency between the measured value and the nominal value. In this Standard, measurement accuracy is expressed by absolute error or relative error, and the calculation is shown in Formula (1) and Formula (2):
The depth of ultrasonic penetration shall comply with the requirements announced by the manufacturer.
4.4 Measurement Accuracy of Young's Modulus
The measurement accuracy of Young's modulus shall comply with the requirements announced by the manufacturer.
4.5 Repeatability of Measurements of Young's Modulus
The repeatability of measurements of Young's modulus shall comply with the requirements announced by the manufacturer.
4.6 Functions
The equipment shall be equipped with the functions specified by the manufacturer in the accompanying documents, which shall at least include a reminder of whether the pressure is appropriate during operation.
NOTE: this clause does not involve product design parameters or functional items that cannot be verified by intuitive test methods.
4.7 Accompanying Documents
The accompanying documents shall clarify the following contents:
a) Whether it has a directive function when selecting the measurement site;
b) The frequency of the applied external mechanical vibration;
c) The depth range of elasticity inspection of liver tissue.
4.8 Environmental Test
The requirements of environmental test of the equipment shall be implemented by the manufacturer in accordance with the stipulations of the climate environmental test group II and the mechanical environmental test group II in GB/T 14710; the test time, recovery time and test items shall comply with the supplementary stipulations of Table 1. The special requirements and methods of environmental test may be implemented in accordance with the principles and provisions of YY/T 1420. signal) farthest from the probe surface can be resolved. The distance between the target line and the probe surface is the depth of ultrasonic penetration.
6.4 Measurement Accuracy of Young's Modulus
The equipment enters the elasticity measurement mode. Within the measurement range specified by the manufacturer, at least choose two different sono-elasticity tissue-mimicking phantoms. The Young's modulus of one of the sono-elasticity tissue- mimicking phantoms is <= 7 kPa; the Young's modulus of the other sono-elasticity tissue-mimicking phantom is >= 10 kPa. For each sono-elasticity tissue-mimicking phantom, in accordance with the method specified by the manufacturer, measure the Young's modulus of the sono-elasticity tissue-mimicking phantom for 5 times; each time, the probe shall be removed, and then, placed in the same test position. For the sono-elasticity tissue-mimicking phantom whose Young's modulus is <= 7 kPa, in accordance with Formula (1), calculate the absolute error. For the sono-elasticity tissue-mimicking phantom whose Young's modulus is >= 10 kPa, in accordance with Formula (2), calculate the relative error, which shall comply with the requirements announced by the manufacturer.
6.5 Repeatability of Measurements of Young's Modulus
For the test data of each sono-elasticity tissue-mimicking phantom in 6.4, in accordance with Formula (3), respectively calculate the repeatability of measurements of Young's modulus. Take the most unfavorable one, which shall comply with the requirements announced by the manufacturer.
6.6 Functions
In accordance with the provisions of the instruction manual of the equipment, conduct item-by-item inspection on the main functions, so as to verify whether it can normally work.
NOTE: the functional inspection does not include product design parameters or functional items that cannot be verified by intuitive test methods.
6.7 Accompanying Documents
Review the accompanying documents of the equipment, which shall comply with the requirements of 4.7.
6.8 Environmental Test
The environmental test of the equipment shall be conducted in accordance with the methods and procedures specified in GB/T 14710 and YY/T 1420. The test time and conditions shall comply with the supplementary stipulations of Table 1.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB 9706.237Medical electrical equipment—Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- GB 10152B mode ultrasonic diagnostic equipment
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- YY/T 1142Methods of measuring the frequency of medical ultrasonic equipment and probe
- YY/T 1420Environmental requirement and test methods for medical ultrasonic equipment
Similar standards
How to Download YY/T 1749-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 10152-2009 — B mode ultrasonic diagnostic equipment
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB 9706.237-2020 — Medical electrical equipment—Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Secure payment via Stripe
Payments accepted
YY/T 1749-2020
$165.00