YY/T 0722-2016Medical electrical equipment—Dosimetric instrument used for non-invasive measurement of X-ray tube voltage in diagnostic radiology (English PDF)
医用电气设备在诊断放射学中用于X射线管电压非接入式测量的剂量学仪器
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0722-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
July 29, 2016
Implementation date
June 1, 2017
Scope
YY/T 0722-2016 is the English-translated version of 医用电气设备在诊断放射学中用于X射线管电压非接入式测量的剂量学仪器.
YY/T 0722-2016 is the Chinese pharmaceutical industry standard for the non-invasive instruments that measure X-ray tube voltage up to 150 kV in diagnostic radiology.
It sets the performance requirements and the tests that show compliance, describes the calibration method and gives guidance on estimating the uncertainty under the different measurement conditions, so that the practical peak voltage of equipment used in general radiography, mammography, computed tomography and dental radiology can be checked reliably. Issued on 29 July 2016, it has applied since 1 June 2017.
Found what you need?
Document preview — YY/T 0722-2016
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C 43
- Replacing
- YY/T 0722-2009
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- Introduction
- 1 Scope
- 2 normative reference document
- 3 Terms and definitions
- 4.1 Amount to be measured
- 4.2 Limits of performance characteristics
- 4.2.1 Limits
- 4.2.2 Maximum error
- 4.2.2.1 Maximum relative error at voltages above 50 kV
- 4.2.2.2 Maximum inherent error at voltages below 50 kV
- 4.2.3 Super range and below range indication
- 4.2.4 Repeatability
- 4.2.5 Long-term stability
- 4.3 Variance limits due to influence quantity effect
- 4.3.1 Impact
- 4.3.2 Minimum rated rating for use
- 4.3.3 Reference conditions
- 4.3.4 Standard test conditions
- 4.3.5 Variability limits
- 4.4 Performance test procedures
- 4.4.1 Overview
- 4.4.2 Dependence of instrument response on voltage waveform and frequency
- 4.4.3 Dependence of instrument response on anode angle
- 4.4.4 Dependence of Instrument Response on Filtration
- 4.4.5 Dependence of instrument response on dose rate
- 4.4.6 Dependence of instrument response on irradiation time
- 4.4.7 Dependence of instrument response on radiation field size
- 4.4.7.1 Dependence of instrument response on changes in the size of the radiation field within the rated range
- 4.4.7.2 Dependence of instrument response on large radiation field size
- 4.4.8 Dependency of instrument response on focus point - detector distance
- 4.4.9 Dependence of instrument response on radiation incidence angle
- 4.4.10 The dependence of the instrument on the rotation angle of the detector and the X-ray tube axis
- 4.4.10.1 Describes the dependency of the instrument on the preferred alignment direction to the manufacturer
- 4.4.10.2 The manufacturer does not specify the dependency of the instrument on the preferred alignment direction
- 4.4.11 Dependence of Instrument Response on Temperature and Humidity
- 4.4.12 The dependence of the instrument on the operating voltage
- 4.4.13 Dependence of Instrument Response on Electromagnetic Compatibility
- 4.4.13.1 Electrostatic discharge
- 4.4.13.2 RF electromagnetic field
- 4.4.13.3 Conducted interference caused by bursts and radio frequencies
- 4.4.13.4 Voltage sags, short interruptions and voltage changes
- 4.4.13.5 Surge (impact)
- 4.4.14 Additional tungsten filtration (X-ray tube aging)
- 5.1 Requirements for the entire instrument
- 5.2 Overview
- 5.3 Display
- 5.4 Measuring range
- 5.5 connectors and cables
- 6 random file
- 6.1 Overview
- 6.2 Information provided
- 6.3 Instrument Description
- 6.4 Detector
- 6.5 Delay
- 6.6 Measuring window
- 6.7 Data Interface
- 6.8 Transport and storage
- Appendix A (informative) Recommended Performance Standard for Access Dividers
- Appendix B (informative) Additional instructions for practical peak voltages
- Appendix C (informative) Glossary Figure B.1 Example of a double pulse generator waveform
Foreword
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces YY/T 0722-2009 "Medical electrical equipment for diagnostic radiology for X-ray tube voltage Non-access measurement Of the dose instrument. " Compared with YY/T 0722-2009, the main technical changes are as follows.
--- 3.16 "actual peak voltage" to "practical peak voltage";
- Table 2 adds a line to the minimum rated rating, reference conditions, and criteria for additional tungsten filtration (X-ray tube aging)
Test conditions, variation limits (± L), the use of the effective range of the inherent error (E) and other values;
--- Added Article 4.4.14 Additional tungsten filtration (X-ray tube aging);
- Added Table 4, depending on the maximum half-value layer (HVL) of the anode angle.
This standard uses the translation method equivalent to IEC 61676.:2009 "Medical electrical equipment used in diagnostic radiology for X-ray tube voltage Non-access measurement of dosimetry instruments ". For ease of use, this standard has been modified as follows.
In Chapter 2, "normative reference documents", the cited IEC and ISO and other standards, which have been converted into national standards, to Reference to national standards.
Introduction
The X-ray tube voltage measured by an access or non-access instrument, the result is usually a single number representing the tube voltage Value, and whether the tube voltage is a constant voltage, whether it is dependent on the waveform changes with time. The market is used to measure the X-ray tube voltage The non-access instrument often refers to the tube voltage with "average peak voltage", but the "average peak voltage" is not clearly defined and And may be any of the mean values of all voltage peaks. In the case where the amount considered is not clearly defined, it is not possible to establish a measuring X-ray Tube voltage non-access instrument performance requirements of the test procedures, for this purpose, according to the recent literature 1) proposed, known as "practical peak power Pressure "as the basis for the establishment of this standard. Practical peak voltage has a clear definition, and applies to any waveform, the amount of emission X-ray Line spectrum distribution and image attributes, even when the tube voltage waveform is very different, work in the same practical peak voltage X-ray hair The device also produces the same low contrast in X-ray photographs. Details of this concept are given in Appendix B, Appendix B Example of practical peak voltage calculation in the case of "falling load" waveforms.
The introduction of a new amount, the result will lead to the problem. the standard has been prepared to complete, its purpose is to measure the practical peak voltage Of the instruments, but these instruments in the past design is clearly not the case. However, from the current market of non-access instrument exploratory test obtained The initial results can be expected that the future of the new instruments and most of the instruments on the market will be able to fully meet the requirements of this standard, no Overcome the difficulties. For the most important requirements for voltage waveforms and frequency dependence responses, these studies show that by applying the practical peaks The value voltage is used as the measurement amount and can more easily meet the requirements of the standard.
X-ray generator X-ray tube voltage, its correction and adjustment is usually done by the manufacturer through direct access measurement, the use of non- The instrument of the access measurement method can of course also be used to check the calibrated X-ray tube voltage or adjust the X-ray tube voltage, but require that these instruments Uncertainty of voltage measurement with comparable measurements. One of the most important parameters for diagnosing X-ray devices is to apply to X-ray tubes, Because the quality of the image in the diagnostic radiology and the dose accepted by the patient under examination are both associated with the X-ray tube voltage related. The total uncertainty requirement is less than ± 5%, and this value is a guide to the variation limits of the effect of the effect.
1) See Appendix B.
1 Scope
This standard specifies the performance requirements and associated compliance tests for non-access measuring instruments for 150 kV X-ray tube voltages Test. This standard also describes the calibration method and gives a guide to estimating the uncertainty at different measurement conditions during calibration.
Measurements in diagnostic radiology include mammography, computed tomography (CT), dental radiology, and X-ray examination. this The standards do not cover the safety of such instruments. Applicable electrical safety requirements are found in GB 4793.1-2007.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
GB 4793.1-2007 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1. General requirements (IEC 61010 - 1..2001, IDT)
GB/T 17626.2 Electromagnetic compatibility test and measurement technology Electrostatic discharge immunity test (idtIEC 61000-4-2)
GB/T 17626.3 Electromagnetic compatibility test and measurement technology Radio frequency electromagnetic radiation immunity test (idtIEC 61000-4-3)
GB/T 17626.4 Electromagnetic compatibility test and measurement techniques Electrical fast transient burst immunity test (idtIEC 61000-4-4)
GB/T 17626.5 Electromagnetic compatibility test and measurement technology Surge (impact) immunity test (idtIEC 61000-4-5)
GB/T 17626.6 Electromagnetic compatibility test and measurement techniques Radio frequency field sensing conducted disturbance immunity test (idtIEC 61000- 4-6)
GB/T 17626.11 Electromagnetic compatibility test and measurement techniques Immunity test for voltage sags, short interruptions and voltage variations (idt
IEC 61000-4-11)
IEC 60417 (all parts) Equipment graphic symbols (Graphicalsymbolsforuseonequipment)
IEC 60788 Medical Radiology Terminology (Radiation Therapy, Nuclear Medicine and Radiation Dosage Equipment) (Medicalradiology-Termi-nology
IEC 61187.:1993 Electrical and Electronic Measurement Equipment (Electrical and Electronic Measurement Equipment) Documentation)
ISO Guide 99..1993 Metrology Basic and General Terminology Glossary [Internationalvocabularyofbasicandgen-eraltermsinmetrology (ISBN92-67-01075-1)]
ISO 7000. 1989 Equipment Graphic Symbols and Graphs (Graphicalsymbolsforuseonequipment-Index andsynopsis)
3 Terms and definitions
IEC 60788 and ISO Guide99 are defined and the following terms and definitions apply to this document.
3.1 Correction factor
Will be operated under specified conditions to correct the indication value of the instrument to a non-dimensional multiplier of the indicated value of the instrument under a given reference condition.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 4793.1-2007Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
- GB/T 17626.2Electromagnetic compatibility -- Testing and measurement techniques -- Electrostatic discharge immunity test
- GB/T 17626.3Electromagnetic compatibility - Testing and measurement techniques - Part 3: Radiated, radio-frequency, electromagnetic field immunity test
- GB/T 17626.4Electromagnetic compatibility -- Testing and measurement techniques -- Electrical fast transient/burst immunity test
- GB/T 17626.5Electromagnetic compatibility - Testing and measurement techniques - Surge immunity test
- GB/T 17626.6Electromagnetic compatibility -- Testing and measurement techniques -- Immunity to conducted disturbances, induced by radio-frequency fields
IEC 61000-4-11 · IEC 60417 · IEC 60788 · IEC 61187 · ISO Guide 99 · ISO 7000
Similar standards
Editions of YY/T 0722
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0722-2016 | Medical electrical equipment—Dosimetric instrument used for non-invasive measurement of X-ray tube voltage in diagnostic radiology | current edition | Current |
| YY/T 0722-2009 | Medical electrical equipment—Dosimetric instrument used for non-invasive measurement of X-ray tube voltage in diagnostic radiology | previous edition | Superseded |
This page sells the current edition, YY/T 0722-2016. Earlier editions are listed for reference only.
How to Download YY/T 0722-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 17626.11-2023 — Electromagnetic compatibility - Testing and measurement techniques - Part 11: Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase
GB/T 17626.2-2018 — Electromagnetic compatibility -- Testing and measurement techniques -- Electrostatic discharge immunity test
Secure payment via Stripe
Payments accepted
YY/T 0722-2016
$445.00