Scope
YY/T 0609-2018 is the English-translated version of 医用诊断X射线管组件通用技术条件.
YY/T 0609-2018 is the Chinese pharmaceutical industry standard for medical diagnostic X-ray tube assemblies.
It gives the terms and definitions, the classification, the requirements and the test methods, covering the loading characteristics and the ratings of the anode, the focal spot and its dimensions, the leakage radiation and the filtration, the cooling of the assembly and its housing, the markings and the accompanying documents. Issued on 7 November 2018, it has applied since 1 November 2019.
Found what you need?
Document preview — YY/T 0609-2018
National Standard of the People's Republic of China
- Replacing
- YY/T 0609-2007
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions6
- 4 Classification6
- 5 Requirements6
- 6 Test methods10
1 Scope
This Standard specifies the terms and definitions, classification, requirements and test methods of medical diagnostic X-ray tube assemblies (hereinafter referred to as the X-ray tube assemblies).
This Standard is applicable to medical diagnostic X-ray tube assemblies (including integrated X-ray tube assembly). This product is for use with medical diagnostic X-ray equipment.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 1804-2000, General Tolerances - Tolerances for Linear and Angular Dimensions without Individual Tolerance Indications
GB 9706.1, Medical electrical equipment - Part 1: General requirements for safety
GB 9706.11, Medical electrical equipment - Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
GB 9706.12, Medical electrical equipment - Part 1: General requirements for safety 3. collateral STANDARD: General requirements for radiation protection in diagnostic X-ray equipment
GB/T 10149, Terminology and symbol for medical X-ray equipment
GB/T 10151-2008, Medical diagnostic X-ray equipment - Specifications for high voltage cable plugs and sockets
YY/T 0063, Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Characteristics of focal spots
YY/T 0064-2016, Electrical and loading characteristics of X-ray tube
c) Atmospheric pressure: 700hPa~1060hPa.
5 Requirements
5.1.2 Power conditions
It is powered by a specific power source.
If applicable, the instruction manual shall give the requirements for the power supply.
5.2 Dimensions and wiring
5.2.1 Dimensions
The instruction manual shall give the main dimensions of the X-ray tube assemblies. The unmarked tolerances of the external dimensions shall meet the requirements of Grade V in GB/T 1804-2000.
5.2.2 Wiring
The instruction manual shall give the wiring diagram of the X-ray tube assemblies.
5.3 Marks
The marks on the X-ray tube assemblies shall meet the requirements of "permanently attached" and "clearly identifiable" in GB 9706.1.
5.4 Focal point of the X-ray tube assemblies
5.4.1 The focal point position shall be marked on the outside of the X-ray tube assemblies.
5.4.2 The nominal value of the focal point shall be marked on the outside of the X-ray tube assemblies. The nominal value of focal point shall be the value determined in accordance with YY/T 0063.
5.5 Nominal X-ray tube voltage
5.5.1 It is specified in the instruction manual.
5.5.2 The X-ray tube components shall be marked with the nominal X-ray tube voltage on the outside.
5.6 Dielectric strength of high voltage circuits
The dielectric strength of high voltage circuit of the X-ray tube assemblies shall be able to withstand 1.1 times the nominal X-ray tube voltage. There shall be no abnormal discharge for 3min, such as flashover and breakdown.
- X-ray tube components, in addition to the ones that are only suitable for mammography of which the nominal X-ray tube voltage is not exceeding 50kV, the mass-equivalent filtering of X-ray tube materials in the X-ray tube components must meet the following requirements:
a) Non-removable material in normal use is not less than 0.5mmAl;
b) All fixed additional filters must be removed with tools;
c) In special applications it shall be stated in the instruction manual.
5.12 Leakage radiation of X-ray tube components
When the X-ray tube assemblies, according to continuous anode input power, are loading in nominal X-ray tube voltage, the leakage radiation at 1m from the focal point shall not exceed 0.88mGy/h.
5.13 High-voltage cable connection
Unless otherwise specified in the instruction manual, the basic dimensions of the connection part between the high-voltage cable socket and the plug shall meet the requirements of 4.1.1 in GB/T 10151-2008.
5.14 Expected service life
The instruction manual shall state the expected life of the X-ray tube assemblies.
The expected service life shall be expressed in years or months or loading times or kilowatt-seconds (kW · s). The CT tube components can be expressed in scan seconds (ss) or scan times.
If the expected service life is expressed in years or months, the basis shall be given in the instruction manual.
5.15 Appearance
The X-ray tube components shall be neat, beautiful, flat, smooth and uniform in color. There must be no defects such as scars and cracks.
5.16 Accompanying documents
5.16.1 The accompanying documents shall meet the requirements for accompanying documents for X-ray tube components in GB 9706.11.
5.16.2 The accompanying documents shall also state the speed of the rotating anode, the single load quota, the series load quota.
5.16.3 The CT tube assemblies shall also specify the nominal CT scan power index in the accompanying documents.
The test voltage of the high voltage circuit is loaded from 50% of the nominal X-ray tube voltage. Start timing after it gradually rises to the specified test voltage value within 10s. The X-ray tube assemblies shall have no abnormal discharge during the test.
In the dielectric strength test, the test voltage shall be kept within the range of 100%~105% of the required value. And it shall be as close as possible to the specified test voltage value of 100%.
During the dielectric strength test of high voltage circuits, if a slight corona discharge occurs in a high-voltage circuit but stops when the test voltage drops to 100% of the voltage indicated by the test conditions, then, this discharge phenomenon can be ignored.
6 Test methods
6.7 Nominal anode input power
6.7.1 Nominal photographic anode input power
Check the accompanying documents.
The nominal photographic anode input power can be tested for the compliance with the following tests:
According to the nominal photographic anode input power at the X-ray tube voltage of 100kV and the nominal X-ray tube voltage, it is carried out according to the provisions of YY/T 0064-2016;
The X-ray tube components of which the nominal X-ray tube voltage is below 100kV shall be carried out at the nominal X-ray tube voltage in accordance with the provisions of YY/T 0064-2016; The test duration is not less than 20min.
6.7.2 Nominal CT anode input power
Check the accompanying documents.
The nominal CT anode input power can be tested for the compliance with the following tests:
According to the nominal CT anode input power at the X-ray tube voltage of 120kV and the nominal X-ray tube voltage, it is carried out according to the provisions of YY/T 0064-2016. The test duration is not less than 2h.
6.8 Nominal continuous input power
Check the accompanying documents.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 1804-2000General tolerances - Tolerances for linear and angular dimensions without individual tolerance indications
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB 9706.12Medical electrical equipment Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
- GB/T 10149Terminology and symbol for medical X-ray equipment
- GB/T 10151-2008Medical diagnostic X-ray equipment - Specifications for high voltage cable plugs and sockets
- YY/T 0063Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Focal spot dimensions and related characteristics
GB 9706.11
Similar standards
Editions of YY/T 0609
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0609-2018 | General specifications of medical diagnostic X-ray tube assemblies | current edition | Current |
| YY/T 0609-2007 | General specifications of medical diagnostic X-ray tube assemblies | previous edition | Superseded |
This page sells the current edition, YY/T 0609-2018. Earlier editions are listed for reference only.
How to Download YY/T 0609-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB 9706.12-1997 — Medical electrical equipment Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
GB/T 10149-1988 — Terminology and symbol for medical X-ray equipment
Secure payment via Stripe
Payments accepted
YY/T 0609-2018
$105.00