GB/T 16292-2010Test method for airborne particles in clean room(zone) of the pharmaceutical industry (English PDF)

医药工业洁净室(区)悬浮粒子的测试方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (16 pages)

Need quotation for your management?

Get a formal quotation for GB/T 16292-2010 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

September 2, 2010

Implementation date

February 1, 2011

Scope

GB/T 16292-2010 is the English-translated version of 医药工业洁净室(区)悬浮粒子的测试方法.

This standard specifies the test method for airborne particle contamination in the cleanrooms of the pharmaceutical industry.

Clause 1 applies it to the verification of airborne particle testing and of the environment in cleanrooms and clean areas in the pharmaceutical industry, in sterile rooms and in local air purification areas including the clean bench, and states that it cannot be used to characterize the physical, chemical, reflective or reproducible properties of airborne particles.

The foreword records that the standard was formulated on the basis of ISO 14644-1 and JGJ 71-90, that it specifies only the test method and does not assess the cleanliness level of the cleanroom, that the filter microscope test method was deleted and the laser particle counter method used instead, and that a method of determining the minimum number of sampling points was added.

Clause 3 defines the cleanroom (zone), localized air purification, unidirectional airflow, non-unidirectional airflow, the airborne particle, given as solid and liquid particles from 0.1 micrometre to 1000 micrometre, and cleanliness.

Clause 4, Test methods, adopts the counting concentration method and covers the duties and training of the testers, the light scattering particle counter and the handling of the instrument, down to a sampling tube that shall not exceed 1.5 m unless otherwise specified.

Clause 5, Test rules, provides for the as-built, at-rest and operational states with no more than two testers present in the first two, for the layout of sampling points according to normative Annex A, for not fewer than two sampling points and five samplings in a small cleanroom or local air purification area, and for the minimum sampling amount at each cleanliness level in Table 2.

The remaining clauses are Calculation of results, which carries the formulas for the average concentration at a sampling point, the mean of the means, the standard error and the 95% upper confidence limit, then Result evaluation and Test report.

Clause 2 cites YY 0033-2000, and informative Annex B gives technical requirements for airborne particles in the cleanroom (zone). It replaces GB/T 16292-1996, was issued on September 2, 2010 and has applied since February 1, 2011.

Found what you need?

Document preview — GB/T 16292-2010

National Standard of the People's Republic of China

ICS
13.040.30
Classification
C 30
Replacing
GB/T 16292-1996

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions4
  • 4 Test methods6
  • 5 Test rules8
  • 6 Calculation of results10
  • 7 Result evaluation12
  • 8 Test report12
  • Annex A (normative) Sampling point layout in cleanroom (zone)14
  • Annex B (informative) Technical requirements for airborne particles in cleanroom (zone)16

Foreword

This Standard was formulated based on ISO 14644-1 "Cleanrooms and associated controlled environments -- Part 1: Classification of air cleanliness by particle concentration" and JGJ 71-90 "Code for Construction and Acceptance of Cleanroom".

This Standard replaces GB/T 16292-1996 "Test method for airborne particles in cleanroom(area)the pharmaceutical industry".

The main differences between this Standard and GB/T 16292-1996 are as follows.

- adjusted the scope of the standard; this Standard only specifies the test method, and does not assess the cleanliness level of the cleanroom (zone);

- deleted the filter microscope test method, used laser particle counter method;

- added the method of determining the minimum number of sampling points;

- added and provided the latest informative annex. Annex A of this Standard is normative. Annex B of this Standard is informative.

1 Scope

This Standard specifies the test method for airborne particle contamination.

This Standard is applicable to the verification of airborne particle testing and environment in cleanroom and clean areas in the pharmaceutical industry, sterile rooms or local air purification areas (including clean bench).

This Standard cannot be used to characterize physical, chemical, or reflective, or reproducible properties of airborne particles.

NOTE: Within the sampling particle size range, the actual particle concentration is unpredictable and significantly changes with time as the sampling amount increases.

2 Normative references

The following standards contain the provisions which, through reference in this Standard, constitute the provisions of this Standard. For dated references, subsequent amendments (excluding corrections) or revisions do not apply to this Standard. However, the parties who enter into agreement based on this Standard are encouraged to investigate whether the latest versions of these documents are applicable. For undated reference documents, the latest versions apply to this Standard.

YY 0033-2000, Good manufacture practice for sterile medical devices

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

3.1 Cleanroom (zone)

A room or area that requires environmental control for dust and microbial contamination. Its building structure, equipment and its use have the function of reducing the involvement, generation and retention of pollution sources in the area. Other relevant parameters such as temperature, humidity, and pressure are also necessary to control.

3.2 Localized air purification

A method in which only the concentration of suspended particles in the air in an indoor working area or a specific partial space reaches a prescribed air cleanliness level.

3.3 Unidirectional airflow

Airflow with uniform wind speed on a section parallel to the flow direction in a single direction and perpendicular to the direction of the airflow. Vertical unidirectional airflow is perpendicular to the horizontal plane, and horizontal unidirectional airflow is parallel to the horizontal plane.

3.4 Non-unidirectional airflow

Airflow with multiple path cycling characteristics or non-parallel airflow directions.

3.5 Airborne particle

Solid and liquid particles with airborne particle size ranging from 0.1µm to 1000µm for air cleanliness fractionation. For a suspended particle count meter, the area or volume of a particle sphere produces a response value, and different response values are equivalent to different particle diameters.

3.6 Cleanliness

The cleanliness of a unit volume of air in a clean environment containing a statistical amount greater than or equal to a certain particle size of suspended particles.

4 Test methods

4.1 Method summary

The method adopts the counting concentration method, that is, the level of airborne particles cleanness in the cleanroom (zone) is assessed by testing the number of airborne particles that have a particle size greater than or equal to a certain particle, in unit-volume in a clean environment.

4.2 Staff duties and training

Testers in cleanroom (zone) shall be trained in the profession and qualified to perform their duties in cleanroom (zone) testing, including the health knowledge and basic microbiological knowledge involved.

4.3 Instrument

The instrument shall use any of the following. […]

4.3.1 Principle of light scattering particle counter

The airborne particles in the air produce light scattering under the illumination of light. The intensity of the scattered light is proportional to the surface area of the airborne particles.

4.4 Test points

4.4.1 The test instrument must be periodically verified according to the test cycle of the test instrument. It shall use the instrument that passes the test and is within the validity period of use.

4.4.2 When the test instrument does not enter the tested area, if necessary, clean the surface first, or prepare and store in the corresponding cleanroom (protect the instrument with a protective cover or other appropriate cover).

4.4.3 When using paper in level 100 cleanroom, it shall be covered with a transparent, dust-free cover. It cannot use pencil and eraser in level 100 cleanroom.

4.4.4 When using the test instrument, strictly follow the instructions of the instrument.

4.4.4.3 The sampling tube must be clean, and no leakage is allowed.

4.4.4.4 The length of the sampling tube shall be based on the allowable length of the instrument. Unless otherwise specified, the length shall not exceed 1.5m.

4.4.4.5 The particle counter sampling port and the working position of the instrument shall be at the same pressure and temperature to avoid measurement deviation.

5 Test rules

5.1 Test conditions

Pre-test the relevant parameters of the cleanroom (zone) before testing. Such tests shall provide environmental conditions for testing airborne particles, such as. such pre-tests may include.

5.2 Test state

Three states of as-bulit, at-rest and operational-state can be tested.

In as-bulit or at-rest test, there must be no more than 2 testers in the room.

5.3 Test time

5.3.1 In the case of as-bulit or at-rest a test, for a unidirectional flow cleanroom (zone), the test shall begin after the normal operation time of the purified air conditioning system is not less than 10min.

5.3.2 In operational-state testing, it must record the start of production and test time.

5.4 Airborne particle count

5.4.1 Number of sampling points and their layouts

In as-bulit or at-rest test, the number of airborne particle sampling points and their layout stresses shall be uniform and not less than the minimum number of sampling points. See Annex A for sampling point layout rules. In operational-state testing, the number of airborne particle sampling points and their layouts shall be set according to the production and process key operation areas of the product.

5.4.2 Limit of sampling times

For any small cleanroom (zone) or local air purification area, the number of sampling points shall not be less than 2, and the total number of sampling times shall not be less than 5 times. Each sampling point can be sampled more than once, and the sampling times of different sampling points can be different.

5.4.3 Sampling amount

The minimum sampling amount for each cleanliness level is shown in Table 2.

5.4.4 Sampling considerations

5.4.4.1 For a unidirectional flow cleanroom (zone), the sampling nozzle of the particle counter shall face the direction of the airflow. For non-unidirectional flow cleanroom (zone), the sampling nozzle of the particle counter shall be upwards.

5.4.4.2 When arranging sampling points, try to avoid the return air outlet.

5.4.4.3 When sampling, the tester shall be on the leeward side of the sampling port and have as little activity as possible.

5.4.4.4 After sampling, it is advisable to self-clean the particle counter.

5.4.4.5 It shall take measures as many as possible to prevent contamination during the sampling process.

6 Calculation of results

6.1 Average airborne particle concentration A at the sampling point

The average airborne particle concentration A of the sampling point is calculated according to formula (2).

6.2 Mean value of the average M

The mean value of the mean M is calculated according to formula (3).

6.3 Standard error SE

Standard error SE is calculated according to formula (4).

6.4 95% upper confidence limit (UCL)

The 95% upper confidence limit (UCL) is calculated according to formula (5).

7 Result evaluation

The evaluation the cleanliness level of airborne particles shall meet the following two conditions simultaneously.

8 Test report

Test results from each cleanroom (zone) shall be recorded, and reports with consistent or inconsistent reports shall also be submitted. The test report shall contain the following information.

a) name and address of tester, date of test;

b) test reference;

c) plane position of the cleanroom (zone) to be tested (marking the position of the adjacent area if necessary);

d) particle size of airborne particles;

Annex A (normative) Sampling point layout in cleanroom (zone)

A.1 The sampling point of the cleanroom (zone) shall be evenly arranged to avoid the sampling point being too sparse in the local area. The following sampling point layout diagram for multi-point sampling can be used as a reference (see Figure A.1). […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.

Referenced standards

Cited by

Similar standards

Editions of GB/T 16292

EditionTitleRevisionStatus
GB/T 16292-2010Test method for airborne particles in clean room(zone) of the pharmaceutical industrycurrent editionCurrent
GB/T 16292-1996Test method for airborne particles in clean room(zone) of the pharmaceutical industryprevious editionSuperseded

This page sells the current edition, GB/T 16292-2010. Earlier editions are listed for reference only.

How to Download GB/T 16292-2010

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
16 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 16292-2010

$205.00