YY/T 0328-2015A.V.fistula needle sets for single use (English PDF)
一次性使用动静脉穿刺器
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0328-2015 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 2, 2015
Implementation date
January 1, 2016
Scope
YY/T 0328-2015 is the English-translated version of 一次性使用动静脉穿刺器.
YY/T 0328-2015 is the Chinese pharmaceutical industry standard for the single-use arteriovenous fistula needle set, the puncture device through which a dialysis circuit is connected to the patient's fistula.
It sets the requirements that make the device safe and suitable for use, covering the needle and its point and dimensions, the tubing and the connectors, the flow rate and the resistance, the strength of the joints and the freedom from leakage, the wings and the clamp, the materials and their biological evaluation and sterility. Issued on 2 March 2015, it has applied since 1 January 2016.
Found what you need?
Document preview — YY/T 0328-2015
National Standard of the People's Republic of China
- Replacing
- YY 0328-2002
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction5
- 1 Scope6
- 2 Normative references6
- 3 Structure7
- 4 Material requirements7
- 5 Physical requirements8
- 6 Chemical requirements11
- 7 Biological requirements12
- 8 Type test13
- 9 Markings13
- 10 Packaging15
- Appendix A (Normative) Particle contamination test16
- References17
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the requirements for the A.V. fistula needle sets for single use (hereinafter referred to as puncture devices), to ensure that they are compatible with the blood flow and blood processing systems that they support.
materials, sterile barrier systems and packaging systems 1)
1 The Chinese standard GB/T 19633.1 equivalent to ISO 11607-1:2006 is currently in the approval stage.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB/T 1962.2 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2. Lock fittings
GB/T 14233.1-2008 Test methods for infusion transfusion injection equipment for medical use - Part 1. Chemical analysis methods
GB/T 14233.2 Test methods for infusion, transfusion, injection equipment for medical use - Part 2. Biological test methods
GB/T 16886.1 Biological evaluation of medical devices - Part 1. Evaluation and testing within a risk management process
GB 18457 Stainless steel needle tubing for the manufacture of medical devices
GB 18671-2009 Intravenous needles for single use
YY/T 0466.1 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1. General requirements
ISO 11607-1:2006 Packaging for terminally sterilized medical devices - Part
GB 18671-2009 gives the evaluation method of needle tip puncture performance.
5 Physical requirements
5.6.3 Lubricants
If the needle is coated with a lubricant and observed with the normal or corrected vision, there shall be no visible accumulation of lubricant on the outside surface of the needle.
Note: The suitable lubricant is undiluted polydimethylsiloxane in accordance with national pharmacopoeia. The amount of lubricant per square centimeter of the surface of the needle shall not exceed 0.25 mg.
5.6.4 Needle handle
5.6.4.1 Appearance
The appearance shall meet the following requirements.
a) The edge of the needle handle shall be complete without burrs;
b) In the appropriate position of the needle handle, it shall have the nominal outer diameter marking of the needle, the marking shall be clear.
Note: There shall be embossed buckles on the needle edge to facilitate the fingers to control the angle of the puncture needle.
5.6.4.2 Color scales
The needle handle color should be used to indicate the outer diameter of the puncture needle. It is recommended to use the color specified in Table 1.
5.6.4.3 Needle handle direction
The needle handle shall be in the same direction as the needle bevel (as shown in Figure 1).
Note: Movable needles that can rotate axially around the needle handle are not subject to this requirement, but additional indications of the direction of the bevel of the needle tip must be provided on the needle handle.
5.7 Flow-stop clip
5.7.1 The flow-stop clip on the puncture device should be a locking pin. The flow-stop clip shall be able to effectively open and close the hose. When closed, it shall be able to block the gas that is 50 kPa above atmospheric pressure for 1 min without leakage. chromium, copper, lead, and tin in the puncture test solution shall not exceed 1 µg/mL. The content of cadmium shall not exceed 0.1 µg/mL.
6 Chemical requirements
6.3.2 When tested in accordance with the method of 5.6.1 of GB/T 14233.1-2008, the color of the puncture device test solution shall not exceed the standard control solution which has a mass concentration rho(Pb2+) = 1 µg/mL.
6.4 pH
When tested in accordance with the method of 5.4.2 of GB/T 14233.1-2008, any standard solution required for graying the indicator shall not exceed 1 mL.
6.5 Evaporation residue
When tested in accordance with them method of 5.5 in GB/T 14233.1-2008, the total amount of evaporation residues shall not exceed 5 mg.
6.6 UV absorbance
When tested in accordance with the method of 5.7 in GB/T 14233.1-2008, the absorbance of the puncture test solution in the range of 250 nm ~ 320 nm shall not exceed 0.1.
6.7 Ethylene oxide residues
When tested in accordance with the methods of clause 9 or clause 10 of GB/T 14233.1-2008, the residual ethylene oxide of each puncture device shall not exceed 0.5 mg.
7 Biological requirements
7.1 Biocompatibility
The puncture device shall be evaluated biologically in accordance with the requirements of GB/T 16886.1.
7.2 Asepsis
The puncture device shall undergo a confirmed sterilization process to make the product sterile.
Note 1. Refer to the reference for suitable sterilization methods.
Note 2. GB/T 14233.2 specifies a sterile test method, but this method should not be used for exit-factory inspection. expressed in kPa.
Note: The graphic symbols given in YY/T 0466.1 may be used to satisfy the above requirements.
9 Markings
9.2 Middle packaging
The middle packaging shall have at least the following clearly identified information.
a) Product name, needle size and length;
b) Quantity;
c) "Sterile";
d) Lot number;
e) Year and month of failure;
f) Words or equivalent text of "single-use";
g) Requirements for handling, storage and transportation (if required);
h) The name and address of the manufacturer and/or distributor.
Note: The graphic symbols given in YY/T 0466.1 can be used to satisfy the above requirements.
9.3 Transport packaging
There must be at least the following clearly identifiable signs on the transport packaging.
a) Product name, needle size and length;
b) Quantity;
c) Words of "sterile";
d) Lot number;
e) Year and month of failure;
f) Words or equivalent text of "single-use";
g) Requirements for handling, storage and transportation;
h) The name and address of the manufacturer and/or distributor. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 1962.2Conical fittings with a 6 %(Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- GB/T 14233.1-2008Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods
- GB/T 14233.2Test methods for infusion, transfusion, injection equipment for medical use - Part 2: Biological test methods
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- GB 18671-2009Intravenous needles for single use
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
GB 18457 · ISO 11607-1:2006
Similar standards
Editions of YY/T 0328
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0328-2015 | A.V.fistula needle sets for single use | current edition | Current |
| YY 0328-2002 | A.V.fistula needle sets for single use | previous edition | Superseded |
This page sells the current edition, YY/T 0328-2015. Earlier editions are listed for reference only.
How to Download YY/T 0328-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 18671-2009 — Intravenous needles for single use
GB/T 14233.1-2008 — Test Methods for Infusion, Transfusion, Injection Equipment for Medical Use — Part 1: Chemical Analysis Methods
GB/T 14233.2-2005 — Test methods for infusion, transfusion, injection equipment for medical use-Part 2: Biological test methods
Secure payment via Stripe
Payments accepted
YY/T 0328-2015
$125.00