YY/T 0286.6-2020Special infusion sets - Part 6: Infusion sets for single use with graduated flow regulator (English PDF)

专用输液器 第6部分:一次性使用刻度流量调节式输液器

Preview PDF

This is a limited preview

Buy now to download the full PDF (22 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0286.6-2020 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

March 31, 2020

Implementation date

April 1, 2021

Scope

YY/T 0286.6-2020 is the English-translated version of 专用输液器 第6部分:一次性使用刻度流量调节式输液器.

YY/T 0286.6-2020 is the Chinese pharmaceutical industry standard for the single-use gravity infusion set fitted with a graduated flow regulator, the roller or dial marked with numbers that let the nurse set a flow rate.

It is the sixth part of the YY/T 0286 series and sets the requirements that make the set compatible with the other intravenous devices it is used with, together with the accuracy of the flow at the marked settings, the tubing, drip chamber and filter, the connectors and their tightness, sterility and the marking.

It does not apply to sets whose regulator carries no scale numbers. It was issued on 31 March 2020 and has applied since 1 April 2021.

Found what you need?

Document preview — YY/T 0286.6-2020

National Standard of the People's Republic of China

ICS
11.040.20
Classification
C 31
Replacing
YY 0286.6-2009

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 General requirements
  • 5 Materials
  • 6 Physical requirements
  • 7 Chemical requirements
  • 8 Biological requirements
  • 9 Signs
  • 10 Packaging
  • 11 Disposal
  • Appendix A (Normative appendix) Flow control characteristic test method
  • Appendix B (Normative appendix) Infusion set leak test

Foreword

YY/T 0286 "Special infusion set" includes the following parts.

---Part 1: Disposable microporous filter infusion set;

---Part 2: Gravity infusion of disposable burette infusion set;

---Part 3: Disposable light-proof infusion set;

---Part 4: Infusion sets for disposable pressure infusion equipment;

---Part 5: Disposable bottle type and bag type infusion set;

---Part 6: One-time use scale flow regulating infusion set. This part is part 6 of YY/T 0286.

This section was drafted in accordance with the rules given in GB/T 1.1-2009.

This part replaces YY 0286.6-2009 "Special infusion set Part 6: One-time use flow setting fine-tuning infusion set", and Compared with YY 0286.6-2009, the main technical changes except for editorial changes are as follows.

---Added terms and definitions (see Chapter 3);

---Added the schematic diagram of the scale flow regulator (see Figure 2);

---Reliability requirements are modified (see 6.2.1);

---Modified the performance requirements for infusion flow rate (see 6.3);

---Modified the performance requirements of the stop clamp or switch (see 6.4);

---Added the leakage test of the infusion set (see 6.5);

---Modified the single package mark (see 9.2);

---Modified the flow control characteristic test method (see Appendix A);

---Added the leakage test method of the infusion set (see Appendix B);

---Added references.

This part uses the redrafting method to refer to ISO 8536-13:2016 ``Liquid contact single-use scaled flow regulators''.

The degree of consistency with ISO 8536-13:2016 "Liquid contact single-use scaled flow regulator" is not equivalent.

Please note that certain contents of this document may involve patents. The issuing agency of this document is not responsible for identifying these patents.

Introduction

With the continuous development of infusion technology and the increasing clinical requirements, some infusions that can be adapted to special clinical requirements have been produced.

Device. Since the development of products is endless, it is impossible to expect all infusion sets with special requirements to be included in a standard of. Therefore, each part of "Special Infusion Sets" only regulates these special infusion sets for a certain clinical special requirement. Some dedicated input The infusion set may also belong to a variety of special infusion sets, and at the same time implement the applicable part of the "special infusion set".

In some special applications, it is required to control the infusion flow rate stably. The scale flow regulating infusion set is to meet this Specially designed for the requirements of clinical applications. Compared with ordinary flow regulators, the scale flow regulator of this infusion set has the advantage of being able to Keep the flow stable and realize the flow setting according to the scale. Appendix D in GB 8368 is applicable to this part.

1 Scope

This part of YY/T 0286 specifies the single-use infusion set used for gravity infusion and can achieve scale flow adjustment (hereinafter abbreviated Called the infusion set) to ensure compatibility with other intravenous infusion devices.

This section applies to single-use infusion sets that mark the scale numbers on the scale flow adjustment device and are used for gravity infusion.

This section does not apply to infusion sets that are not marked with scale numbers on the scale flow adjustment device.

2 Normative references

The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated references, the latest version (including all amendments) applies to this document.

GB 8368 single-use infusion set gravity infusion type (GB 8368-2018, ISO 8536-4:2010/AMD1.2013, MOD)

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Scale flow regulator

The components that are in contact with the liquid medicine with scale marks and can set the flow rate according to the scale.

3.2 Scale

A series of identifications and corresponding identification numbers related to the position of the observed indicator.

4 General requirements

4.1 The infusion set and scale flow regulator are shown in Figure 1 and Figure 2.

Note: Figure 1 shows an air intake infusion set, it can also be a non air intake infusion set, see GB 8368. Figure 1 and Figure 2 are not unique, as long as the same effect can be achieved As a result, other configurations can also be used.

4.2 The infusion set should have a protective cover to keep the inner cavity of the infusion set sterile before use. The cork piercer or needle of the air intake device should have a protective cover.

5 Materials

The materials for manufacturing the infusion set and its components given in Chapter 4 should meet the requirements of Chapter 6: The components of the infusion set in contact with the liquid The materials should also meet the requirements specified in Chapter 7 and Chapter 8.

6 Physical requirements

6.1 General

The physical requirements of the infusion set should meet the requirements of GB 8368 (except for the requirements of leakage, flow regulator and infusion flow rate).

6.2 Scale flow regulator

6.2.1 Reliability

The infusion set is free to fall on a thick hardwood board at a height of 1m, and the two ends of the scale flow regulator are subjected to 15N for 15s along the pipeline.

The infusion set bears 0.5m and 1.5m hydrostatic pressure respectively, and each lasts for 10s. Check the scale flow regulator, the scale indication is not Changes have occurred and the settings are stable and reliable. After the above test is completed, the infusion set should meet the requirements of 6.2.4 and 6.5.

6.2.2 Continuously adjustable

The scale flow regulator should be able to continuously adjust from zero ("off" position) to smooth flow ("on" position).

6.2.3 Indexing and marking

6.2.3.1 The text on the scale flow regulator should be in Chinese (except for "OPEN" and "OFF"), otherwise, it should be added on the single package Description.

6.2.3.2 The scaled flow regulator should clearly indicate "open" or "OPEN", "close" or "OFF" at the open and closed positions, and clearly indicate the flow rate unit. The flow rate should be scaled in milliliters per hour (mL/h).

6.2.4 Flow control characteristics
6.2.4.1 Flow rate indication error

The error of the flow rate (indicated by relative error) should be clearly indicated on the single package or on the accompanying documents. According to A.3 of Appendix A During the test, the error of the flow rate indication at each test point should conform to the manufacturer's stated value [see 9.2.1c)].

Note: A uniform error can be given for all scales, or different errors can be given for different areas.

6.2.4.2 Flow control stability

When testing in accordance with Appendix A.4, the maximum number of graduations, the minimum number of graduations and the number of intermediate graduations in the range of 20mL/h~125mL/h are appropriate The flow stability at any point should not exceed 10%.

6.3 Infusion flow rate

According to GB 8368, the infusion set with the scale flow regulator set in the "on" state outputs sodium chloride solution [mass concentration is rho(NaCl)= 9g/L] should not be less than 500mL.

Note: For the infusion flow rate of the infusion set with sterile grade liquid filter, please refer to YY 0286.1.

6.4 Stop clamp or switch (if any)

Close the stop clip or switch, immerse the infusion set in (40±1)°C water, and apply a pressure of 50kPa higher than atmospheric pressure from the inside of the infusion set.

Keep the air pressure for 15s, check if there is air leakage at the other end.

6.5 Leak

When tested in accordance with Appendix B, there should be no leakage.

7 Chemical requirements

GB 8368 applies.

8 Biological requirements

GB 8368 applies.

9 Signs

9.1 General

GB 8368 applies.

9.2 Single package mark

9.2.1 In addition to complying with the provisions of GB 8368, the single package should also provide the following information.

a) "The scale indication is only an approximate value and should be corrected by the number of drops" or similar words;

b) The Chinese description of the special text on the scale flow regulator (if any, see 6.2.3.1);

c) The indication error of static head height and flow rate (see 6.2.4.1);

d) Description of exhaust method during adjustment (if necessary);

e) Other necessary information, such as instructions for use, etc.

9.2.2 The unit of measurement on the package should adopt the International System of Units.

10 Packaging

GB 8368 applies.

11 Disposal

Information on the safe and environmentally sound disposal of single-use infusion sets should be provided in the accompanying documents, such as "must be used in accordance with the established "No biohazard disposal procedures for the disposal of blood-contaminated products" or equivalent text.

Appendix A (Normative appendix) Flow control characteristic test method

A.1 Instruments and reagents

A.1.1 Balance, accurate to 0.01g.

A.1.2 The rigid test container (applicable to the air inlet infusion set) as shown in Figure A.1 or the constant as shown in Figure A.2 is filled with the test solution (A.1.5) Liquid level test container (applicable to non-air infusion set).

A.1.3 6 liquid storage containers, equipped with devices to avoid liquid evaporation.

A.1.5 Test solution, sodium chloride solution [mass concentration

A.1.4 Timer

A.2 Test preparation

Take 6 infusion sets and adjust them for at least 2h under the test condition of (23±2)°C. Operate each input according to the instructions provided by the manufacturer Liquid container. The height of the static pressure head is based on the value indicated on the package or attached documents. If there is no indication, it is 1m. For air intake infusion sets, static The height of the indenter is shown in Figure A.1.For the non-air-injection infusion set, use the constant liquid level test container shown in Figure A.2 to test the static head height as As shown in Figure A.2.

A.3 Flow error

A.3.1 Take three infusion sets and connect them to the infusion container (A.1.2), and set the scale flow regulators to low, medium and high. The digital scale makes it in the infusion state.

A.3.2 Make the effluent flow into the liquid storage container (A.1.3) under the specified static pressure head, close the infusion set after 1h, measure by weighing and record for 1h The volume of liquid collected in the liquid storage container, see formula (A.1).

Note: The smallest scale marked with a number refers to the scale marked with a number within the specified range that is closest to 20mL/h, and the largest scale marked with a number refers to the scale within the specified range The scale marked with numbers closest to 125mL/h, the scale marked with numbers in the middle refers to the scale marked with numbers closest to 72.5mL/h.

A.4.2 Connect the three infusion sets to the infusion container (A.1.2), set the scale flow regulators at the three test points, and press Use instructions to operate each infusion set to make it in infusion state.

A.4.3 Under the specified static pressure head, start the test after 15 minutes of stable operation, and continue the test for 6 hours. Use the weighing method to measure and record each small The volume collected at time (Vi).

A.4.4 The relative deviation of the measured volume calculated by formula (A.2) represents the stability of the flow.

Appendix B (Normative appendix) Infusion set leak test

B.1 Before the start of the test, adjust the state of the infusion set at the test temperature. B.2 Follow the instructions provided by the manufacturer.

B.3 Set the scale flow regulator to the "on" position and connect one end of the infusion set to a compressed air source. Close the other end of the infusion set, Immerse in (40±1)°C water, apply pressure 50kPa higher than atmospheric pressure to the inside of the infusion set for 15s, and check whether the infusion set is free No leakage, and repeat the test on the scale with the maximum scale value of 25%, 50% and 75% of the scale flow regulator closest to the number marked.

B.4 Set the scale flow regulator to the "off" position, connect one end of the infusion set to the compressed air source and open the other end. Dip the infusion set Into the water at (40±1)°C, apply a pressure of 50kPa higher than the atmospheric pressure to the inside of the infusion set for 15s. Check the open end of the infusion set Are there any leaks. Finally, close the open end of the pipeline, keep the test sample at a pressure of 50kPa higher than the atmospheric pressure for 15s, and check Check the scale flow regulator for any air leakage.

B.5 Set the scale flow regulator to the "on" position, and fill the infusion set with debubbled distilled water; one end of the infusion set is sealed, and the other end is connected Connect to the vacuum device, apply a pressure of -20kPa to the inside at (40±1)°C for 15s to check whether air enters the infusion set. On the scale The flow regulator is set in the "off" position, and the maximum scale values of 25%, 50% and 75% are closest to the scale marked with numbers and repeat the test. references

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 22 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 0286.6

EditionTitleRevisionStatus
YY/T 0286.6-2020Special infusion sets - Part 6: Infusion sets for single use with graduated flow regulatorcurrent editionCurrent
YY 0286.6-2009Special infusion sets - Part 6: Infusion sets for single use with graduated flow regulatorprevious editionSuperseded

This page sells the current edition, YY/T 0286.6-2020. Earlier editions are listed for reference only.

How to Download YY/T 0286.6-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
22 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0286.6-2020

$210.00