YY/T 0980.2-2016Biopsy needles for single use - Part 2: Manual biopsy needle (English PDF)
一次性使用活组织检查针 第2部分:手动式
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0980.2-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
January 26, 2016
Implementation date
January 1, 2017
Scope
YY/T 0980.2-2016 is the English-translated version of 一次性使用活组织检查针 第2部分:手动式.
YY/T 0980.2-2016 is the Chinese pharmaceutical industry standard for the manual type of single-use biopsy needle, the needle whose sample is taken by the operator's own hand rather than by a mechanical device.
It is the second part of the YY/T 0980 series and sets the requirements on top of the general part, covering the needle tube and its materials, the point and the sample notch, the strength of the joint to the hub, the rigidity and toughness, the sterility and the marking. Issued on 26 January 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 0980.2-2016
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction4
- 1 Scope5
- 2 Normative references5
- 3 Classification5
- 4 Materials9
- 5 Requirements10
- 6 Packaging and marking12
- 7 Instructions for use12
- 8 Transportation and storage13
- 9 Type inspection rules13
- Appendix A (Informative) Type inspection rules14
- Bibliography15
Foreword
YY/T 0980 Biopsy Needles for Single Use can be divided into 4 parts.
--- Part 1: General Requirements;
--- Part 2: Manual Biopsy Needle;
--- Part 3: Mechanical Assembly Type;
--- Part 4: Mechanical Complete Type. This part belongs to Part 2 of YY/T 0980.
This Part was drafted as per the rules specified in GB/T 1.1-2009.
1 Scope
This Part specifies the requirements of biopsy needles for single use - manual type (hereinafter referred to as biopsy needle).
This Part is applicable to the manual-type biopsy needle for single use that manually finishes collecting the living sample of soft tissue, bone tissue for the test. This Part is also applicable to the manual-type biopsy needle for single use that coordinates with endoscope for use.
This Part is not applicable to the manual-type biopsy needle for repeated use, mechanical assembly type biopsy needle for repeated or single use, or mechanical complete type biopsy needle.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.
GB/T 1220 Stainless Steel Bars
GB/T 2965 Titanium and Titanium Alloy Bars
YY/T 0980.1-2016 Biopsy Needles for Single Use - Part 1: General Requirements
3 Classification
3.1 General
The structures of the biopsy needles can be divided into manual aspiration, manual cutting, and combination of the former two types.
3.2 Structure
3.2.1 The structure of manual aspiration biopsy needle is shown in Figure 1.
3.2.2 The structure of manual cutting biopsy needle is shown in Figure 2.
3.2.3 The structure of manual aspiration and manual cutting combined biopsy
needle is shown in Figure 3.
4 Materials
4.1 The materials of needle tube and needle seat of biopsy needle shall meet the requirements of 5.2 in YY/T 0980.1-2016.
5 Requirements
5.1 General
In addition to meet the requirements of Clause 5 in YY/T 0980.1-2016, the biopsy needle shall also meet the requirements of 5.2 ~ 5.8.
5.2 Needlepoint structure
5.2.1 If the external needle tube of the biopsy needle adopts inclination edge needlepoint, use 3× magnifier to observe; its internal needle tube or internal needle bar needlepoint puncturing with external needle tube shall not protrude out of the edge bevel of the external needle tube.
5.3 Inner hole cleanliness
The inner hole of biopsy needle shall be clean; take the amount (stipulate din Table 1) of mixture between glycerol and alcohol at the rate of 1.1 to inject through the biopsy needle; use the normal or corrected vision to check whether the mixture contains the foreign objects or impurities.
5.4 Internal needle connecting firmness
The internal needle tube or internal needle bar of biopsy needle shall be connected with internal needle seat firmly, after exerting the axial static tension shown in Table 2 on the internal needle tube/bar and internal needle seat, last for 10s, the two of which shall not be separated.
5.6 Scale identification
If the external needle tube of biopsy needle has scale line for prompting the puncture depth (as shown in Figure 4), the scale line shall form a circle on the external surface of the needle tube, and is clearly visible.
5.7 Positioner
5.7.1 If the biopsy needle is equipped with the puncture depth positioner, when exerting 10N axial static pressure on the front end of the positioner, last for 10s, the positioner shall have no axial movement.
5.8 Sheath
The sheath for protecting the needlepoint shall be equipped with out of the needlepoint of the biopsy needle, and the sheath shall not fall off naturally.
6 Packaging and marking
The packaging and marking of the biopsy needle shall meet the requirements of Clause 6 and 7 of YY/T 0980.1-2016.
7 Instructions for use
The instructions for use shall contain the following contents at least.
8 Transportation and storage
8.1 During the transporting and storing process, the biopsy needle shall prevent heavy pressure, direct exposure to sunshine, rain and snow leaching.
9 Type inspection rules
The type inspection rules can refer to Appendix A.
Appendix A (Informative) Type inspection rules
A.1 Inspection items
The inspection items can refer to the requirements of Clause 5.
A.2 Sampling amount
Physical performance shall randomly sample 10 pieces; while the chemical and biological performance shall sample as per the demand.
A.3 Judgement rules
When all inspection items are qualified, the product passes through the type inspection.
During the inspecting process, if the chemical and biological performances are qualified, but the physical performance is disqualified, the sampling amount is allowed to be doubled, and re-inspect the disqualified item. If the re-inspection is still disqualified, then the product doesn't pass through the type inspection.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 0980.2-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 1220-2007 — Stainless steel bars
GB/T 2965-2023 — Titanium and titanium alloy bars
YY/T 0980.1-2016 — Biopsy needles for single use - Part 1: General requirements
Secure payment via Stripe
Payments accepted
YY/T 0980.2-2016
$125.00