YY/T 0980.4-2016Biopsy needles for single use - Part 4: Mechanical complete type (English PDF)
一次性使用活组织检查针第4部分机动一体式
This is a limited preview
Buy now to download the full PDF (20 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0980.4-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
July 29, 2016
Implementation date
June 1, 2017
Scope
YY/T 0980.4-2016 is the English-translated version of 一次性使用活组织检查针第4部分机动一体式.
YY/T 0980.4-2016 is the Chinese pharmaceutical industry standard for single-use biopsy needles of the mechanical complete type, in which the needle and the mechanical power device are assembled together at manufacture and the device performs the cutting automatically or semi-automatically to collect the tissue sample.
It is the fourth part of the YY/T 0980 series and sets the requirements for the complete instrument and the test methods for them, on top of the general requirements of part 1, covering the stainless steel and other materials used, the needle point and sample collection space, the firing and cutting mechanism, sterility and biological evaluation, and the marking, packaging and instructions. Issued on 29 July 2016, it has applied since 1 June 2017.
Found what you need?
Document preview — YY/T 0980.4-2016
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction4
- 1 Scope5
- 2 Normative references5
- 3 Structure type5
- 4 Material7
- 6 Packaging and marking9
- 7 Instructions for use9
- 8 Transportation and storage10
- 9 Type inspection rules10
- Annex A (informative) Type inspection rules11
- Bibliography12
Foreword
YY/T 0980 "Biopsy needles for single use" series of standards consist of the following parts.
- Part 1. General requirements;
- Part 2. Manual biopsy needle;
- Part 3. Mechanical assembly type - Part 4. Mechanical complete type.
This Part is Part 4 of YY/T 0980 "Biopsy needles for single use".
This Part is drafted in accordance with the rules given in GB/T 1.1-2009.
Introduction
This Part of YY/T 0980 specifies the requirements for mechanical complete type biopsy needles for single use.
This Part does not provide requirements and methods for sampling performance and ultrasonic development, because currently there is no suitable test methods and simulation test materials.
In view of the fact that the structure type of mechanical complete type biopsy needles and the location, depth, state, texture and method of the puncture are different, this Part does not specify the puncture force. However, manufacturers are encouraged to determine appropriate puncture force by verifying, combining the specific structure and use of the mechanical complete type biopsy needle produced by each of the manufacturers.
In view of the fact that the trigger device of mechanical complete type biopsy needles may be mistakenly triggered in the course of clinical use, and there is no suitable test method for trigger devices, this Part does not specify the trigger device. However, manufacturers are encouraged to determine appropriate trigger device data and test method by verifying.
1 Scope
This Part of YY/T 0980 specifies the requirements for biopsy needles for single use - mechanical complete type (hereinafter referred to as "biopsy needles").
This Part applies to mechanical complete type biopsy needles for single use, of which the needle for single use and the mechanical power device are assembled together at the time of manufacture; the mechanical power device is for automatic or semi- automatic cutting operations to collect biopsy samples.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 1220 Stainless steel bar
GB/T 2965 Titanium and titanium alloy bar
YY/T 0980.1-2016 Biopsy needles for single use - Part 1. General requirements
3 Structure type
3.1 A biopsy needle consists of an outer needle tube, an inner needle bar, a protective cover and a mechanical power device.
NOTE. Biopsy needles can be used with guide needles, cursors, etc. when necessary.
a) product name;
b) product model description;
c) product performance, main structure;
d) scope;
e) use methods and precautions;
f) storage conditions;
g) other instructions and necessary warnings.
8 Transportation and storage
8.1 During transportation and storage process, biopsy needles shall be prevented from stress, sun and rain and snow.
8.2 Biopsy needles shall be stored in a non-corrosive-gas, well ventilated and clean environment.
9 Type inspection rules
See Annex A for type inspection rules.
Bibliography [1] GB 15811-2001 Sterile hypodermic needles for single use [2] YY 0321.2-2009 Single-use needle for anaesthesia [3] YY/T 0313-2014 Medical polymer products - Requirement for package and information supplied by manufacturer [4] YY/T 0466.1-2009 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1. General requirements
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 20 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 0980.4-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 1220-2007 — Stainless steel bars
GB/T 2965-2023 — Titanium and titanium alloy bars
YY/T 0980.1-2016 — Biopsy needles for single use - Part 1: General requirements
Secure payment via Stripe
Payments accepted
YY/T 0980.4-2016
$115.00