YY 0285.4-2017Intravascular catheters—Sterile and single-use catheters - Part 4: Balloon dilatation catheters (English PDF)

血管内导管一次性使用无菌导管 第4部分:球囊扩张导管

Preview PDF

This is a limited preview

Buy now to download the full PDF (25 pages)

Need quotation for your management?

Get a formal quotation for YY 0285.4-2017 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

July 17, 2017

Implementation date

January 1, 2019

Scope

YY 0285.4-2017 is the English-translated version of 血管内导管一次性使用无菌导管 第4部分:球囊扩张导管.

YY 0285.4-2017 is the Chinese pharmaceutical industry standard for the balloon dilatation catheter supplied sterile for single use.

It is the fourth part of the YY 0285 series and sets the requirements particular to such catheters, covering the nominal diameter and length of the balloon and the compliance of its diameter with pressure, the rated burst pressure and the fatigue after repeated inflation, the time to inflate and deflate, the strength of the joints and the tightness, the surface and the coating where present, the materials and their biological evaluation, and sterility. Issued on 17 July 2017, it has applied since 1 January 2019.

Found what you need?

Document preview — YY 0285.4-2017

National Standard of the People's Republic of China

Replacing
YY 0285.4-1999

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Definitions5
  • 4 Requirements6
  • Annex A (Normative) Test for balloon rated burst pressure (RBP)8
  • Annex B (Normative) Balloon fatigue test for freedom from leakage and damage on inflation10
  • Annex C (Normative) Test for balloon deflation time13
  • Annex D (Normative) Test for balloon diameter to inflation pressure15
  • Annex E (Informative) Guidance on the selection of balloon materials17
  • Bibliography18

Foreword

YY 0285 "Intravascular catheters - Sterile and single-use catheters" consists of four parts.

- Part 1: General requirements;

- Part 3: Central venous catheters;

- Part 4: Balloon dilatation catheters;

- Part 5: Over-needle peripheral catheters. This Part is Part 4 of YY 0285.

This Part is drafted in accordance with the rules given in GB/T 1.1-2009.

This Part replaces YY 0285.4-1999 "Sterile, single-use intravascular catheters - Part 4: Balloon dilatation catheters". Compared with YY 0285.4-1999, the main technical changes are as follows.

- UPDATE normative references;

- ADD the requirements and test methods for balloon rated burst pressure, balloon deflation time, and relationship between balloon diameter and inflation pressure.

This Part uses the redraft law to modify and adopt ISO 10555-4:2013 "Intravascular catheters - Sterile and single-use catheters - Part 4: Balloon dilatation catheters".

The technical differences between this Part and ISO 10555-1:2013 and their reasons are as follows.

- As for the normative references, this Standard has made adjustments with technical differences, to adapt to the technical conditions of China. The adjustments are mainly reflected in Clause 2 "Normative references". The specific adjustments are as follows.

lambda Replace ISO 594-1 with GB/T 1962.1, which is identical to the international standard;

lambda Replace ISO 594-2 with GB/T 1962.2, which is identical to the international standard;

lambda Replace ISO 10555-1 with YY 0285.1, which is identical to the international standard.

1 Scope

This Part of YY 0285 specifies requirements for balloon dilatation catheters supplied in the sterile condition, and intended for single use.

2 Normative references

The following documents are indispensable for the application of this document.

For the dated references, only the editions with the dates indicated are applicable to this document. For the undated references, the latest edition (including all the amendments) are applicable to this document.

GB/T 1962.1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirement (GB/T 1962.1-2015, ISO 594-1:1986, IDT)

3 Definitions

For the purposes of this document, the terms and definitions given in YY 0285.1 and the following apply.

Intravascular catheter fitted with a balloon near the distal end, which is introduced into an artery or vein to dilate a part or parts of the vascular system.

4 Requirements

Determine the rated burst pressure with an appropriate safety margin when tested in accordance with Annex A. Longitudinal failure is the desirable balloon failure mode.

Annex A (Normative) Test for balloon rated burst pressure (RBP)

The purpose of this test is to determine the RBP of the balloon.

Annex B (Normative)

Balloon fatigue test for freedom from leakage and damage on inflation The catheter is inflated and deflated a number of times to simulate use in vivo.

The catheter in an inflated condition is examined for leakage, rupture or herniation.

Annex C (Normative)

Test for balloon deflation time The purpose of this test is to determine the time required to deflate the balloon from the rate burst pressure (RBP) level. This test provides information that might be clinically useful for treatment planning (e.g. potential occlusion time).

Annex D (Normative)

Test for balloon diameter to inflation pressure The purpose of this test is to determine the relationship between the balloon diameter and the inflation pressure.

Annex E (Informative)

Guidance on the selection of balloon materials The balloon, if it should fail during use, should burst longitudinally and without fragmentation. Consider the severity of other failure modes. Consideration should be given to this guideline in the selection of the balloon material and the manner of securing the balloon material to the shaft.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY 0285.4

EditionTitleRevisionStatus
YY 0285.4-2017Intravascular catheters—Sterile and single-use catheters - Part 4: Balloon dilatation catheterscurrent editionCurrent
YY 0285.4-1999Intravascular catheters—Sterile and single-use catheters - Part 4: Balloon dilatation cathetersprevious editionSuperseded

This page sells the current edition, YY 0285.4-2017. Earlier editions are listed for reference only.

How to Download YY 0285.4-2017

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
25 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0285.4-2017

$235.00