YY 0450.1-2020Accessory devices for sterile single-use intravascular catheters - Part 1: Introducers (English PDF)
一次性使用无菌血管内导管辅件 第1部分:导引器械
This is a limited preview
Buy now to download the full PDF (36 pages)
Need quotation for your management?
Get a formal quotation for YY 0450.1-2020 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
February 25, 2020
Implementation date
August 1, 2021
Scope
YY 0450.1-2020 is the English-translated version of 一次性使用无菌血管内导管辅件 第1部分:导引器械.
YY 0450.1-2020 is the Chinese pharmaceutical industry standard for the introducers supplied sterile for single use with the intravascular catheters of YY 0285: the introducer needle, the introducer catheter, the sheath introducer, the guidewire and the dilator.
It is the first part of the YY 0450 series and sets the requirements and the test methods for them, covering the dimensions and the compatibility with the catheter, the surface and the freedom from particles, the strength of the joints and of the guidewire, the tightness of the haemostasis valve, the materials and their biological evaluation and sterility. Issued on 25 February 2020, it applies from 1 August 2021.
Found what you need?
Document preview — YY 0450.1-2020
National Standard of the People's Republic of China
- Replacing
- YY 0450.1-2003
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword5
- 1 Scope8
- 2 Normative References8
- 3 Terms and Definitions9
- 4 General Requirements13
- 4.1 Sterilization13
- 4.2 Biocompatibility13
- 4.3 Surface13
- 4.4 Corrosion Resistance14
- 4.5 Radio-detectability14
- 4.6 Information to be Supplied by the Manufacturer14
- 5 Additional Requirements for Introducer Needles15
- 5.1 General Rules15
- 5.2 Size Designation15
- 5.3 Needle Point15
- 5.4 Hub15
- 5.4.1 Conical fitting15
- 5.4.2 Strength of union of needle tube and needle hub15
- 5.5 Information to be Supplied by the Manufacturer15
- 6 Additional Requirements for Introducer Catheters16
- 6.1 General Rules16
- 6.2 Tip16
- 6.3 Peak Tensile Force16
- 6.4 Hub16
- 6.5 Size Designation17
- 6.6 Information to be Supplied by the Manufacturer17
- 7 Additional Requirements for Sheath Introducers17
- 7.1 General Rules17
- 7.2 Size Designation17
- 7.3 Free of Leakage from Sheath Introducer17
- 7.4 Free of Leakage through Hemostasis Valve18
- 7.5 Hub18
- 7.6 Peak Tensile Force18
- 7.7 Information to be Supplied by the Manufacturer18
- 8 Additional Requirements for Guidewires18
- 8.1 General Rules18
- 8.2 Size Designation18
- 8.3 Safety Wire18
- 8.4 Fracture Test18
- 8.5 Bending Test19
- 8.6 Peak Tensile Force of Guidewire19
- 8.7 Information to be Supplied by the Manufacturer19
- 9 Additional Requirements for Dilators20
- 9.1 General Rules20
- 9.2 Size Designation20
- 9.3 Hub20
- 9.3.1 General rules20
- 9.3.2 Conical fitting20
- 9.3.3 Strength of union between hub and dilator20
- 9.4 Information to be Supplied by the Manufacturer20
- 10 Additional Requirements for Kits Containing Combinations of Devices Specified in this Part21
- 11 Determination of Strength of Union of Needle Hub and Needle21
- Appendix A (informative) Materials and Design22
- Appendix B (normative) Test Method for Corrosion Resistance23
- Appendix C (normative) Determination Method of Peak Tensile Force of Introducer Catheters, Sheath Introducers and Dilators24
- Appendix D (normative) Test Method for Liquid Leakage from Sheath Introducers under Pressure26
- Appendix E (normative) Test Method for Liquid Leakage from Hemostasis Valves of Sheath Introducers28
- Appendix F (normative) Test Method for Fracture of Guidewires29
- Appendix G (normative) Test Method for Guidewire Resistance to Bending Damage31
- Appendix H (normative) Determination Method for Peak Tensile Force of Guidewire33
- Appendix I (normative) Determination of Strength of Union of Needle Hub and Needle35
- Bibliography36
Foreword
Some content of 6.2 in this Part is recommended, and the remaining is mandatory.
YY 0450 Accessory Devices for Sterile Single-use Intravascular Catheters is composed of the following parts.
---Part 1: Introducers;
---Part 2: Obturators for Over-needle Peripheral Catheters;
---Part 3: Inflation Devices for Balloon of Balloon Dilatation Catheters. This Part is Part 1 of YY 0450.
This Part was drafted in accordance with the rules provided in GB/T 1.1-2009.
This Part serves as a replacement of YY 0450.1-2003 Accessory Devices for Sterile Single-use Intravascular Catheters - Part 1: Introducers.
In comparison with YY 0450.1-2003, other than editorial modifications, the main technical changes are as follows.
---Some requirements and test methods of guidewires are modified (see Chapter 8; Chapter 8 of Version 2003);
---The requirements for size designation are modified (see 5.2, 6.5, 7.2, 8.2 and 9.2; see 5.2, 6.5, 7.2, 8.2 and 9.2 of Version 2003);
---Appendix I "Determination Method of Strength of Union of Needle Hub and Needle" is added (see Appendix I);
This Part adopts the re-drafting law to modify the adoption of ISO 11070:2014 Sterile Single-use Intravascular Introducers, Dilators and Guidewires.
In comparison with ISO 11070:2014, there are structural differences. The differences and the causes for these differences are shown below.
---Chapter 11 "Determination of Strength of Union of Needle Hub and Needle" is added, which quotes Appendix I; meanwhile, 5.4.2 is modified, which quotes Chapter 11, so as to comply with the requirements of GB/T 1.1 for the sequence of the appendix of volume in this Part. In comparison with ISO 11070:2014, there is no technical difference.
In comparison with ISO 11070:2014, there are technical differences. The differences and the causes for these differences are shown below.
---In terms of Normative References, this Part makes adjustments with technical differences, so as to adapt to the technical conditions of China. The adjustments are intensively reflected in Chapter 2 "Normative References". See the specific adjustments below.
lambda GB/T 1962.1, which equivalently adopts International Standard, is used to replace ISO 594-1;
lambda GB/T 1962.2, which equivalently adopts International Standard, is used to replace ISO 594-2;
lambda GB/T 7408, which equivalently adopts International Standard, is used to replace ISO 8601;
lambda GB 15810, which equivalently adopts International Standard, is used to replace ISO 7886-1;
lambda GB/T 16886.1, which equivalently adopts International Standard, is used to replace ISO 10993-1;
lambda YY/T 0466.1, which equivalently adopts International Standard, is used to replace ISO 15223-1;
lambda Use YY/T 0586-2016 to replace ASTM F640-12 and DIN 13273-7.
---In terms of footnote, this Part makes adjustments with technical differences. See the specific adjustments below.
Footnote 1) and Footnote 2) are modified, so as to adapt to the current international and national technical conditions.
Please be noted that certain content of this document might involve patents. The institution issuing this document does not undertake the responsibility of identifying these patents.
1 Scope
This Part of YY 0450 specifies the requirements for single-use introducer needles, introducer catheters, sheath introducers, guidewires and dilators supplied in the sterile condition and used in conjunction with intravascular catheters specified in YY 0285.
This Standard is applicable to single-use introducer needles, introducer catheters, sheath introducers, guidewires and dilators supplied in the sterile condition and used in conjunction with intravascular catheters specified in YY 0285.
2 Normative References
The following documents are indispensable to the application of this document. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB/T 1962.11) Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain Other Medical Equipment - Part 1: General Requirement (GB/T 1962.1-2015,
ISO 594-1:1986, IDT)
3 Terms and Definitions
The following terms and definitions are applicable to this document.
Distal end refers to one end of a medical device that is inserted into patient's body.
4 General Requirements
When tested in accordance with the appropriate requirements in GB/T 16886.1, the medical device shall be free of biological hazards.
If the intravascular catheter introducer is coated with lubricant, when normal vision or corrected vision is adopted for the inspection, the outer surface of the effective length of the medical device shall not manifest concentrated lubricant droplets.
5 Additional Requirements for Introducer Needles
When inspected under 2.5 times magnification, the needle point shall be sharp and free of feather edges, burrs or hooks.
If a needle hub is provided, the needle hub shall have a 6% (Luer) taper internal conical fitting that complies with GB/T 1962.1 and / or GB/T 1962.2.
6 Additional Requirements for Introducer Catheters
When tested in accordance with the method provided in Appendix C, the peak tensile force of the introducer catheter and the junction between the introducer catheter and the hub shall comply with the stipulations of Table 2.
7 Additional Requirements for Sheath Introducers
If the sheath introducer is equipped with an integral hemostasis valve, when tested in accordance with Appendix E, the hemostasis valve shall not generate any leakage.
When tested in accordance with the method provided in Appendix C, the minimum peak tensile force of the sheath introducer and the junction between the sheath introducer and the hub shall comply with the stipulations of Table 2.
8 Additional Requirements for Guidewires
When tested in accordance with Appendix G, the guidewire shall not manifest fracture, loosening or failure in the following forms. The manufacturer shall provide the following information.
9 Additional Requirements for Dilators
When tested in accordance with the method provided in Appendix C, the peak tensile force of the dilator and the junction between the dilator and the hub shall comply with the requirements of Table 2.
The manufacturer shall provide the nominal sizes of dilators designated in 9.2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 36 pages — is available in the English PDF.
Referenced standards
Normative references
GB/T 1962.11 · ISO 594-1:1986
Cited by
- YY/T 1925-2024Cardiovascular implants and endovascular devices - Neurovascular thrombectomy stent retrievers
- YY/T 1748-2021Neurovascular implants—Intracranial embolization coils
- GB/T 39381.1-2020Cardiovascular implants - Vascular device-drug combination products - Part 1: General requirements
- YY/T 1554-2017Oviduct catheter
- YY/T 0880-2013Breast localization wire and introducer needle for single use
Similar standards
Editions of YY 0450.1
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0450.1-2020 | Accessory devices for sterile single-use intravascular catheters - Part 1: Introducers | current edition | Current |
| YY 0450.1-2003 | Accessory devices for sterile single-use intravascular catheters - Part 1: Introducers | previous edition | Superseded |
This page sells the current edition, YY 0450.1-2020. Earlier editions are listed for reference only.
How to Download YY 0450.1-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY 0450.2-2003 — Accessory devices for sterile single-use intravascular catheters - Part 2: Obturators for over-needle peripheral catheters
YY 0450.3-2007 — Accessory devices for sterile single-use intravascular catheters - Part 3: Inflation devices for balloon dilatation catheters
YY 0450.1-2003 — Accessory devices for sterile single-use intravascular catheters - Part 1: Introducers
Secure payment via Stripe
Payments accepted
YY 0450.1-2020
$405.00