GB/T 39381.1-2020Cardiovascular implants - Vascular device-drug combination products - Part 1: General requirements (English PDF)
Also coversGBT39381.1-2020
Open the GB/T 39381.1-2020 preview as PDF
This is a limited preview
Buy now to download the full PDF
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
November 19, 2020
Implementation date
December 1, 2021
Scope
GB/T 39381.1-2020 (Cardiovascular implants - Vascular device-drug combination products - Part 1: General requirements) is available as an English-translated PDF.
GB/T 39381.1-2020 — This part of GB/T 39381 specifies the requirements for the intended performance, design attributes, materials, design evaluation, manufacturing, sterilization, packaging, and information provided by the manufacturer, etc. of the vascular device-drug combination products (VDDCPs). NOTE 1: Due to variations in the design of combination products covered by this Part and due to the relatively recent development of some of these combination products, acceptable standardized in vitro test results and clinical study results are not always available. As further scientific development and clinical data accumulation, appropriate revision of this Part might be necessary. Delivery systems or parts of the delivery system are included in the scope of this Part, if they comprise an integral component of the vascular device and if they are drug- covered (e.g.: drug-covered balloon catheters and drug-covered guidewires). This Part is also applicable to the non-permanent implantable VDDCPs. Devices whose PMOA is to provide a conduit for delivery of a drug, are excluded from the scope of this Part (e.g.: infusion catheters), unless they contain a drug component that is intended to have an ancillary action to the device part (e.g.: antimicrobial coated central venous catheter). Procedures and devices used prior to and following the introduction of the VDDCP (e.g.: balloon angioplasty devices) are excluded from the scope of this Part if they do not affect the drug-related aspects of the device. This Part includes the requirements for the absorbable components (such as coating) related to the device drug in the VDDCPs. NOTE 2: See also ISO/TS 17137. This Part does not address issues associated with viable or non-viable biological materials such as tissues, cells, or proteins. This Part does not address issues associated with active surgical implants (i.e.: implants that require power not generated by the human body or gravity).
Document preview — GB/T 39381.1-2020
National Standard of the People's Republic of China
Issued by: SAMR; SAC
Contents
- Foreword...3
- Introduction...5
- 1 Scope...6
- 2 Normative References...7
- 3 Terms and Definitions...8
- 4 Intended Performance...12
- 5 Design Attributes...13
- 6 Materials...15
- 7 Design Evaluation...15
- 8 Manufacturing...32
- 9 Sterilization...35
- 10 Packaging...36
- 11 Information Supplied by the Manufacturer...36
- Annex A (Informative) Definitions of Potential Clinical and Technical Events .40
- Bibliography...46
Foreword
GB/T 39381 Cardiovascular Implants - Vascular Device-Drug Combination Products
consists of the following 2 parts:
--- Part 1: General Requirements;
--- Part 2: Local Regulatory Guidance.
This Part is Part 1 of GB/T 39381.
This Part was drafted as per the rules specified in GB/T 1.1-2009.
This Part uses re-drafting method to modify and adopt ISO 12417-1:2015
Cardiovascular Implants and Extracorporeal Systems - Vascular Device-Drug
Combination Products - Part 1: General Requirements.
Compared with ISO 12417-1:2015, the technical differences and causes of this Part
are as follows:
--- For the normative references, this Part adjusts according to the technical
differences, so as to adapt to China's technical conditions. The adjustment is
reflected in Clause 2 "Normative References", which are specially as follows:
Use GB/T 16886.1 that equivalently adopts the international standard to
replace ISO 10993-1;
Use GB/T 16886.2 that equivalently adopts the international standard to
replace ISO 10993-2;
Use GB/T 16886.7 that equivalently adopts the international standard to
replace ISO 10993-7;
Use GB/T 19633.1 that equivalently adopts the international standard to
replace ISO 11607-1;
Use GB/T 19974 that equivalently adopts the international standard to
replace ISO 14937;
Use YY 0285.4 that equivalently adopts the international standard to replace
ISO 10555-4;
Use YY 0450.1 that equivalently adopts the international standard to replace
ISO 11070;
Use YY/T 0466.1 that equivalently adopts the international standard to
replace ISO 15223-1;
Use YY/T 0663.2 that equivalently adopts the international standard to
replace ISO 25539-2.
--- Add the "solvent residue" to supplement the relevant requirements for the design
attributes of the drug-containing part of the device-drug combination products
(see 7.2.4.3.8 of this Edition).
--- Delete the "clinical evaluation" to avoid inconsistencies with Guidelines for
Medical Devices Clinical Evaluation (see 7.3 of ISO12417-1:2015).
This Part made the following editorial modifications:
--- Modify the standard name;
--- Delete the Annex B of ISO 12417-1:2015.
Please note some contents of this document may involve the patents. The issuing
agency of this document shall not assume the responsibility to identify these patents.
This Part was proposed by National Medical Products Administration.
This Part shall be under the jurisdiction of National Technical Committee for
Standardization of Implants for Surgery and Orthopaedic Devices (SAC/TC 110).
Drafting organizations of this Part: Tianjin Medical Device Quality Supervision and
Inspection Centre; Centre for Medical Device Evaluation, NMPA; Shanghai Microport
Medical (Group) Co., Ltd.; Lepu Medical Technology (Beijing) Co., Ltd.; and Medtronic
(Shanghai) Management Co., Ltd.
Chief drafting staffs of this Part: Duan Qingjiao, Ma Jinzhu, Zhu Lili, Cheng Maobo,
Han Zheng, Zhang Xiangmei, Li Yong, Xu Jiajia, and Chen Huimin.
Cardiovascular Implants - Vascular Device-Drug
Combination Products - Part 1: General Requirements
1 Scope
This part of GB/T 39381 specifies the requirements for the intended performance,
design attributes, materials, design evaluation, manufacturing, sterilization, packaging,
and information provided by the manufacturer, etc. of the vascular device-drug
combination products (VDDCPs).
NOTE 1: Due to variations in the design of combination products covered by this Part and due
to the relatively recent development of some of these combination products, acceptable
standardized in vitro test results and clinical study results are not always available. As further
scientific development and clinical data accumulation, appropriate revision of this Part might be
necessary.
Delivery systems or parts of the delivery system are included in the scope of this Part,
if they comprise an integral component of the vascular device and if they are drug-
covered (e.g.: drug-covered balloon catheters and drug-covered guidewires). This Part
is also applicable to the non-permanent implantable VDDCPs.
Devices whose PMOA is to provide a conduit for delivery of a drug, are excluded from
the scope of this Part (e.g.: infusion catheters), unless they contain a drug component
that is intended to have an ancillary action to the device part (e.g.: antimicrobial coated
central venous catheter).
Procedures and devices used prior to and following the introduction of the VDDCP
(e.g.: balloon angioplasty devices) are excluded from the scope of this Part if they do
not affect the drug-related aspects of the device.
This Part includes the requirements for the absorbable components (such as coating)
related to the device drug in the VDDCPs.
NOTE 2: See also ISO/TS 17137.
This Part does not address issues associated with viable or non-viable biological
materials such as tissues, cells, or proteins.
This Part does not address issues associated with active surgical implants (i.e.:
implants that require power not generated by the human body or gravity).
2 Normative References
The following documents are essential to the application of this document. For the
dated documents, only the versions with the dates indicated are applicable to this
document; for the undated documents, only the latest version (including all the
amendments) is applicable to this document.
GB/T 16886.1 Biological Evaluation of Medical Devices - Part 1: Evaluation and
Testing within a Risk Management Process (GB/T 16886.1-2011, ISO 10993-
1:2009, IDT)
GB/T 16886.2 Biological Evaluation of Medical Devices - Part 2: Animal Welfare
Requirements (GB/T 16886.2-2011, ISO 10993-2:2006, IDT)
GB/T 16886.7 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide
Sterilization Residuals (GB/T 16886.7-2015, ISO 10993-7:2008, IDT)
GB/T 19633.1 Packaging for Terminally Sterilized Medical Devices - Part 1:
Requirements for Materials, Sterile Barrier Systems and Packaging Systems
(GB/T 19633.1-2015, ISO 11607-1:2006, IDT)
GB/T 19974 Sterilization of Health Care Products - General Requirement for
Characterization of a Sterilization Agent and the Development, Validation and
Routine Control of a Sterilization Process for Medical Devices (GB/T 19974-2018,
ISO 14937:2009, IDT)
YY 0285.4 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4:
Balloon Dilatation Catheters (YY 0285.4-2017, ISO 10555-4:2013, MOD)
YY/T 0316-2016 Medical Devices - Application of Risk Management to Medical
Devices (ISO 14971:2007 corrected version, IDT)
YY 0450.1 Accessory Devices for Sterile Single-Use Intravascular Catheters -
Part 1: Introducers (YY 0450.1-2003, ISO 11070:1998, IDT)
YY/T 0466.1 Medical Devices - Symbols to be Used with Medical Device Labels,
Labelling and Information to be Supplied - Part 1: General Requirements (YY/T
0466.1-2016, ISO 15223-1:2012, IDT)
YY/T 0640-2016 Non-Active Surgical Implants - General Requirements (ISO
14630:2012, IDT)
YY/T 0663.2 Cardiovascular Implants Endovascular Devices - Part 2: Vascular
Stems (YY/T 0663.2-2016, ISO 25539-2:2012, MOD)
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- GB/T 16886.1-2011Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
- YY/T 0640-2016Non-active surgical implants - General requirements
ISO 10993 · GB/T 16886.2 · GB/T 16886.2-2011 · GB/T 16886.7 · GB/T 16886.7-2015 · GB/T 19633.1 · GB/T 19633.1-2015 · ISO 11607 · GB/T 19974 · GB/T 19974-2018 · ISO 14937 · YY 0285.4 · YY 0285.4-2017 · ISO 10555 · YY/T 0316-2016 · ISO 14971 · YY 0450.1 · YY 0450.1-2003 · ISO 11070 · ISO 15223 · YY/T 0663.2 · YY/T 0663.2-2016 · ISO 25539
How to Buy GB/T 39381.1-2020
- 1Add to cart. Click the "Buy GB/T 39381.1-2020" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 16886.1-2022 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
GB/T 16886.10-2024 — Biological evaluation of medical devices - Part 10: Tests for skin sensitization
GB/T 16886.11-2021 — Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
Secure payment via Stripe
Payments accepted
GB/T 39381.1-2020
$425.00