YY 0285.5-2018 is the Chinese standard "Intravascular catheters - Sterile and single-use catheters - Part 5: Over-needle peripheral catheters". Its scope clause reads: This part of YY 0285 specifies the requirements for over-needle peripheral catheters for insertion into the peripheral vasculature that are supplied sterile and in single-use. Its clauses include terms and definitions.
Found what you need?
Document preview — YY 0285.5-2018
National Standard of the People's Republic of China
- Replacing
- YY 0285.5-2004
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions5
- 4 Requirements7
- hub and needle tube10
- Appendix B (Informative) Opaque catheter seat color11
- Appendix C (Informative) Needle tip geometry12
- Appendix D (Normative) Determination of liquid leakage from vent fittings13
- References15
1 Scope
This part of YY 0285 specifies the requirements for over-needle peripheral catheters for insertion into the peripheral vasculature that are supplied sterile and in single-use.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB/T 1962.1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1. General requirement (GB/T 1962.1-2015, ISO 594-1.1986, IDT)
GB/T 18457 Stainless steel needle tubing for the manufacture of medical devices (GB/T 18457-2015, ISO 9626.1991, MOD)
YY 0285.1 Intravascular catheters - Sterile and single-use catheters - Part 1.
General requirements (YY 0285.1-2017, ISO 10555-1.2013, MOD)
3 Terms and definitions
The terms and definitions as defined in YY 0285.1 and the following terms and definitions apply to this document.
3.1 Over-needle peripheral intravascular catheter
A catheter for introducing in or withdrawing out a fluid or instrument from a peripheral vascular system.
3.2 Needle
Appendix A (Normative)
Determination of connection strength between needle hub and needle tube
A.1 Principle
Apply tension and compression to the needle tube and the needle hub, then check for loose connections between the needle tube and the needle hub.
A.2 Instrument
The tensile tester can apply a force of 20 N with an accuracy of ±1%.
A.3 Test procedure
A.3.1 Condition the needle for 2 h under conditions of relative humidity of 40%
~ 60% and temperature of 22 °C ± 2 °C, immediately perform the test.
A.3.2 Clamp the needle tube and the needle hub on the fixture of the tensile tester, then apply the tension and pressure one time at the speed of 100 mm/min in accordance with the following force values.
- The test needle with a nominal outer diameter less than 0.6 mm. 10 N;
- The test needle with the nominal outer diameter greater than or equal to
0.6 mm. 20 N.
A.3.3 Check the connection between the needle tube and the needle hub, record whether the needle is loose.
A.4 Test report
The test report shall include the following information.
a) Identification of the needle;
b) The outer diameter of the needle, expressed in mm;
c) The applied force (i.e. 10 N or 20 N);
d) Whether the needle in the needle hub loosens.
Appendix D (Normative)
Determination of liquid leakage from vent fittings
D.1 Principle
The catheter is connected to a simulated blood source under hydrostatic pressure. Allow liquid to flow into the needle and measure the time it takes for the liquid to leak through the vent fitting.
D.2 Test solution
D.2.1 Dissolve 9 g of analytically pure sodium chloride in distilled or deionized water to prepare 1 L sodium chloride solution [0.9% (mass concentration)].
D.2.2 Mix 550 mL of sodium chloride solution (D.2.1) and 450 mL of USP grade or higher glycerin to prepare test solution.
Note. To improve the observability, a coloring agent such as a red or blue food coloring agent may be added.
D.3 Instrument
D.3.1 Constant liquid level container, capable of providing 400 mm ± 20 mm static pressure head, equipped with a liquid supply pipeline with an inner diameter of not less than 3 mm, the liquid supply pipeline is equipped with a clip or valve, the end has a puncture membrane such as latex cover). An example of this device is shown in Figure D.1.
D.3.2 Stopwatch, or similar device.
D.4 Procedure
D.4.1 Fill the test solution (D.2) at 23 °C ± 2 °C into a constant liquid level container (D.3.1).
D.4.2 Discharge all air in the supply line and close the clamp or valve.
D.4.3 Insert the needle tip into the membrane to ensure that the needle is maintained at ±5° from the horizontal plane.
D.4.4 Open the clamp or valve to allow liquid to flow into the needle tube and measure the time it takes for the first drop to drop at the vent fitting.
D.5 Test report
YY/T 0285.6-2020 | YY 0285.1-2017 | YY 0285.3-2017 | YY 0285.4-2017 | YY/T 0294.1-2024 | […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 20 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 1962.1Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirement
- GB/T 18457Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- YY 0285.1Intravascular catheters—Sterile and single-use catheters - Part 1: General requirements
Similar standards
Editions of YY 0285.5
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0285.5-2018 | Intravascular catheters - Sterile and single-use catheters - Part 5: Over-needle peripheral catheters | current edition | Current |
| YY 0285.5-2004 | Intravascular catheters - Sterile and single-use catheters - Part 5: Over-needle peripheral catheters | previous edition | Superseded |
This page sells the current edition, YY 0285.5-2018. Earlier editions are listed for reference only.
How to Download YY 0285.5-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 18457-2024 — Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
GB/T 1962.1-2015 — Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirement
YY 0285.1-1999 — Sterile, single-use intravascular catheters - Part 1: General requirements
Secure payment via Stripe
Payments accepted
YY 0285.5-2018
$125.00