YY/T 0285.6-2020Intravascular catheters—Sterile and single-use catheters - Part 6: Subcutaneous implanted ports (English PDF)

血管内导管 一次性使用无菌导管 第6部分:皮下植入式给药装置

Preview PDF

This is a limited preview

Buy now to download the full PDF (19 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0285.6-2020 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

September 27, 2020

Implementation date

September 1, 2021

Scope

YY/T 0285.6-2020 is the English-translated version of 血管内导管 一次性使用无菌导管 第6部分:皮下植入式给药装置.

YY/T 0285.6-2020 is Part 6 of the Chinese standard series on sterile single-use intravascular catheters, and covers subcutaneous implanted ports - the infusion port placed under the skin and connected to a catheter, used for long-term chemotherapy, parenteral nutrition and repeated drug delivery.

A port is punctured hundreds of times over months or years and must neither leak nor shed septum debris into the bloodstream, so the standard sets the performance and safety requirements a manufacturer has to demonstrate.

It covers biocompatibility and bacterial endotoxin, chemical properties of silicone and other components, ethylene oxide residue, distance marks, nominal dimensions, gas tightness, self-sealing after repeated puncture, puncture chips, peak tension, bending and dynamic breaking strength of the connection, flow for ports used and not used for power injection, burst pressure, MRI compatibility and the information supplied by the manufacturer.

It relates to ISO 10555-6:2015 (NEQ); requirements for the non-coring needle are in YY 0881. Issued by the National Medical Products Administration on 27 September 2020, it has applied since 1 September 2021.

Found what you need?

Document preview — YY/T 0285.6-2020

National Standard of the People's Republic of China

ICS
11.040.25
Classification
C 31

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Requirements
  • 4.1 General
  • 4.2 Biochemical properties
  • 4.3 Distance mark
  • 4.4 Nominal dimensions of subcutaneous implanted port
  • 4.5 Physical requirements
  • 4.6 Flow
  • 4.7 Burst pressure for subcutaneous implanted ports and catheters
  • 5 Magnetic resonance imaging (MRI) compatibility
  • 6 Information provided by the manufacturer
  • Annex A (normative) Test method for no gas leakage
  • Annex B (informative) Detection limit of different matrix specimens
  • Annex C (informative) Further characterization test guidelines: needle penetration and extraction
  • Annex D (normative) Test method for no leakage after multiple punctures
  • Annex E (normative) Puncture chips
  • Annex F (normative) Peak tension
  • Annex G (normative) Bending test
  • Bibliography

Foreword

This document was issued on 27 September 2020 by the National Medical Products Administration of the People's Republic of China, and has applied since 1 September 2021.

It is a YY/T standard: a recommended industry standard of China's pharmaceutical and medical device sector.

YY/T 0285, Intravascular catheters - Sterile and single-use catheters, is published in several parts; this is Part 6.

This part makes reference to ISO 10555-6:2015; its degree of correspondence with that standard is not equivalent (NEQ).

1 Scope

This part of YY/T 0285 specifies the performance and safety requirements for sterile and single-use subcutaneous implanted ports. It applies to subcutaneous implanted ports intended for long-term or persistent intravascular use, consisting of a subcutaneously implanted infusion port and a catheter.

Note 1: Subcutaneous implanted ports are also used for non-endovascular indications such as intraperitoneal, intrathecal and epidural access. Ports intended only for non-endovascular use may refer to this part on the basis of a risk assessment, and are encouraged to meet its requirements.

This part does not specify the performance and safety requirements of the non-coring needle used with the port.

Note 2: See YY 0881 for the requirements for the non-coring needle.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 6682, Water for analytical laboratory use - Specification and test methods

GB/T 14233.1-2008, Test methods for infusion, transfusion, injection equipments for medical use - Part 1: Chemical analysis methods

GB 15810, Sterile syringes for single use

YY 0285.1-2017, Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements

YY 0285.3-2017, Intravascular catheters - Sterile and single-use catheters - Part 3: Central venous catheters

YY 0334-2002, General specification for surgical implants made of silicone elastomer

Pharmacopoeia of the People's Republic of China, 2015 Edition

3 Terms and definitions

For the purposes of this document, the terms and definitions given in YY 0285.1-2017 and the following apply.

3.1 catheter: single-lumen or multi-lumen tubing whose end can be placed at a point in the body

3.2 connection: system that connects the catheter to the subcutaneous implanted port

3.3 effective surface area: area of the septum that can be punctured by the needle

3.4 flushing volume: volume of solution required to completely replace one solution with another in the subcutaneous implanted port and catheter

3.5 non-coring needle: needle that is not prone to producing chips when puncturing the septum

Note: Chips are debris that can be created when the needle pierces the septum.

3.6 outlet tube: cannula connecting the port outlet to the catheter

3.7 priming volume: total volume of the subcutaneous implanted port and catheter that can be filled with solution

3.8 priming volume of the subcutaneous implanted port: total volume of the port that can be filled with solution, consisting of the reservoir and the outlet tube

3.9 priming volume of the catheter: total volume of the effective length of the catheter that can be filled with solution

4.1 General

Unless otherwise specified in this part, subcutaneous implanted ports and catheters shall comply with the requirements of YY 0285.1-2017.

4.2 Biochemical properties

4.2.1.1 General: the subcutaneous implanted port shall be free of biological hazards.

Note 1: See GB/T 16886.1 for biological evaluation.

Note 2: Demonstrating that the implant materials meet the corresponding standards (such as YY 0334, ISO 5832, YY/T 1557 and YY/T 0660) helps to support the biological evaluation required by GB/T 16886.1.

4.2.1.2 Bacterial endotoxin: inject not more than 10 mL of water for bacterial endotoxin test into each port set and extract at (37 ± 1) °C for (1 ± 0.1) h. Test in accordance with the bacterial endotoxin test of the Pharmacopoeia of the People's Republic of China, 2015 Edition. The bacterial endotoxin content shall be less than 0.5 EU/mL.

Note: See YY/T 0618 for the routine testing of bacterial endotoxin.

4.2.2.1 Silicone rubber elastic parts: the silicone rubber elastic parts of the port shall meet the requirements of 5.3 and 5.4 of YY 0334-2002.

4.2.2.2.1 Test solution preparation (components other than silicone rubber elastic parts): extract 0.2 g of sample per 1 mL of water complying with GB/T 6682 at (37 ± 1) °C for 72 h to obtain the test solution. Prepare the blank control solution in the same way.

4.2.2.2.2 pH: when tested by the method of 5.4.1 of GB/T 14233.1-2008, the pH of the test solution shall not differ from that of the blank control solution by more than 1.0.

4.2.2.2.3 Metal ions: when determined by atomic absorption spectrophotometry or an equivalent method (5.9.1 or 5.6.1 of GB/T 14233.1-2008), the total content of barium, chromium, copper, lead and tin in the test solution shall not exceed 1 µg/mL and the content of cadmium shall not exceed 0.1 µg/mL.

4.2.2.2.4 Reducing substances: when tested by the indirect titration method of 5.2.2 of GB/T 14233.1-2008, the difference in the volume of potassium permanganate solution [c(KMnO4) = 0.002 mol/L] consumed by the test solution and the blank control shall not exceed 2.0 mL.

4.2.2.2.5 UV absorbance: when tested by the method of 5.7 of GB/T 14233.1-2008 over 250 nm to 320 nm, the absorbance of the test solution shall not be greater than 0.1.

4.2.2.2.6 Evaporation residue: when tested by the method of 5.5 of GB/T 14233.1-2008, the evaporation residue shall not exceed 2 mg.

4.2.2.3 Residual ethylene oxide: if the port is sterilized with ethylene oxide, when tested by the gas chromatography method of Clause 9 of GB/T 14233.1-2008, the residual ethylene oxide shall not exceed 0.5 mg per port.

4.3 Distance mark

If the catheter carries distance marks, they shall indicate: a) for a split catheter, the distance from the end of the catheter; b) for a pre-connected catheter, the distance from the proximal end of the catheter.

From the first mark, the distance between marks shall not be greater than 5 cm.

4.5 Physical requirements

4.5.5.2 Self-sealing after puncture: when tested by the method of Annex D, the septum of the port shall show no gas leakage.

4.5.5.3 Puncture chips: when tested in accordance with Annex E, after 1,000 punctures per cm2 of septum or the nominal number of punctures, whichever is greater, the number of chips shall not exceed 30.

4.5.6.1 Peak tension: when tested in accordance with Annex F, the minimum peak tensile force of the connection between port and catheter shall be 5 N; the other parts of the catheter shall meet 4.4 of YY 0285.3-2017.

4.5.6.2 Bending resistance: when tested in accordance with Annex G, the flow rate of the bent catheter shall be not less than 50 % of that of the straight catheter.

4.5.6.3 Dynamic breaking strength: the connection between port and catheter, and each connection of the catheter, shall withstand 100,000 cycles at (1 ± 0.2) Hz, stretched in the flow direction to 10 % elongation or to a maximum force of 5 N, whichever is reached first, without any component breaking or cracking. Split catheters shall also meet this requirement after being connected to the port according to the manufacturer's instructions.

4.6 Flow

4.6.1 General: if a nominal flow rate is marked, when ports not intended and intended for power injection are tested by the methods of Annexes E and G of YY 0285.1-2017, the flow rate of each lumen shall be not less than 80 % of the manufacturer's nominal value for catheters with a nominal outside diameter below 1.0 mm, and not less than 90 % for catheters of 1.0 mm or more. For hydrating catheters, the flow rate shall be measured after hydration.

4.6.2 Ports indicated not to be used for power injection: the test of Annex E of YY 0285.1-2017 is carried out with the non-coring needle connected to the Luer connector, placed through the septum, the port connected to the test catheter, and the in-line pressure sensor at the proximal end of the needle. Results are expressed as the minimum and maximum gravity flow; the needle gauge and length are recorded.

4.6.3 Ports indicated to be used for power injection: the test of Annex G of YY 0285.1-2017 is carried out with the locking device connected to the hub of the power-injection non-coring needle, the needle placed through the septum and the port connected to the test catheter.

4.7 Burst pressure for subcutaneous implanted ports and catheters

4.7.1 General: the test is carried out by the method of Annex F of YY 0285.1-2017, with the locking device accessories connected to the hub of the non-coring needle as justified by a risk-based clinical rationale.

A Annex A (normative) Test method for no gas leakage

A.1 Principle: air pressure is applied to the port and to the connection between port and catheter, and the assembly is checked for gas leakage.

A.2 Apparatus: leak-proof connectors; syringe or constant pressure source; pressure gauge capable of measuring up to 300 kPa; temperature-controlled test chamber that can be filled with liquid.

A.3 Procedure: bring the chamber liquid to (37 ± 2) °C and maintain it; connect a split catheter to the port; cut the catheter leaving up to 10 cm for connection to the gauge and pressure source; connect through a leak-proof connector; immerse the port and allow it to reach thermal equilibrium; raise the pressure to at least 200 kPa and wait 2 min.

A.4 A gas leak is a pressure drop of more than 2.65 kPa within 2 min, or failure to reach 200 kPa.

D Annex D (normative) Test method for no leakage after multiple punctures

D.1 Principle: the effective area of the septum is punctured the appropriate number of times at random positions with the non-coring needle, and the port is then leak tested according to Annex A.

D.3 Procedure: at least 1,000 punctures per cm2 of effective septum area are made, or the higher number claimed by the manufacturer, with the port kept at (37 ± 2) °C. Each needle is used for at most 50 punctures and is replaced if its tip is bent or wobbles after leaving the septum. The leak test of Annex A is then carried out.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 0285.6-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
19 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0285.6-2020

$285.00