GB/T 3058-2019Determination of arsenic in coal (English PDF)

Also coversGBT3058-2019

Open the GB/T 3058-2019 preview as PDF

Preview — first pages of GB/T 3058-2019

This is a limited preview

Buy now to download the full PDF

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

June 4, 2019

Implementation date

January 1, 2020

Scope

GB/T 3058-2019 (Determination of arsenic in coal) is available as an English-translated PDF.

GB/T 3058-2019 — This standard specifies the method summary, reagents and materials, instruments and equipment, samples, test procedures, calculation and presentation of results, the precision of the method for the determination of arsenic in coal by arsenic molybdenum blue spectrophotometry and hydride generation-atomic absorption method. Arsenic molybdenum blue spectrophotometry is the arbitration method. This standard applies to lignite, bituminous coal, anthracite.

Document preview — GB/T 3058-2019

National Standard of the People's Republic of China

Issued by: SAMR; SAC

Contents

  • Foreword...3
  • 1 Scope...4
  • 2 Normative references...4
  • 3 Arsenic molybdenum blue spectrophotometry...4
  • 4 Hydride generation-atomic absorption method...9
  • 5 Precision...14
  • 6 Test report...14
  • Appendix A (Informative) Recovery rate of arsenic measuring device...15

Foreword

This standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This standard replaces GB/T 3058-2008 "Method for the determination of

arsenic in coal". As compared with GB/T 3058-2008, except for editorial

changes, the main technical changes are as follows.

- In the method summary, CHANGE that "use hydrochloric acid to dissolve

the burnt" into that "use sulfuric acid and hydrochloric acid to dissolve the

burnt" (see 3.1; 3.1 of the 2008 edition);

- MODIFY the requirements for arsenic-free zinc particles (see 3.2.1; 3.2.1

of the 2008 edition);

- ADD the placement position of lead acetate cotton to the diagram of the

arsenic determination device (see Figure 1; Figure 1 of the 2008 edition);

- ADD the calculation method of the recovery rate of the arsenic

determination device by the arsenic molybdenum blue spectrophotometric

method (see Appendix A).

This standard is prepared by using the redrafting method with reference to ISO

11723.2016 "Solid mineral fuels - Determination of arsenic and selenium -

Eschka's mixture and hydride generation method". The degree of consistency

with ISO 11723.2016 is nonequivalent.

This standard was proposed by the China Coal Industry Association.

This standard shall be under the jurisdiction of the National Coal

Standardization Technical Committee (SAC/TC 42).

Drafting organization of this standard. Testing Branch of Coal Science and

Technology Research Institute Co., Ltd.

Drafters of this standard. Chen Huizhu, Fu Kun, Yang Guang, Yang Huayu.

This standard replaces the standard previously issued as follows.

- GB/T 3058-1982, GB/T 3058-1996, GB/T 3058-2008.

Determination of arsenic in coal

1 Scope

This standard specifies the method summary, reagents and materials,

instruments and equipment, samples, test procedures, calculation and

presentation of results, the precision of the method for the determination of

arsenic in coal by arsenic molybdenum blue spectrophotometry and hydride

generation-atomic absorption method. Arsenic molybdenum blue

spectrophotometry is the arbitration method.

This standard applies to lignite, bituminous coal, anthracite.

2 Normative references

The following documents are essential to the application of this document. For

the dated documents, only the versions with the dates indicated are applicable

to this document; for the undated documents, only the latest version (including

all the amendments) is applicable to this standard.

GB/T 483 General rules for analytical and testing methods of coal

GB/T 2304 Chemical reagent - Zinc granular free from arsenic

3 Arsenic molybdenum blue spectrophotometry

3.1 Method summary

The coal sample is mixed with Eschka reagent and burned. The burned material

is dissolved with hydrochloric acid and sulfuric acid. A reducing agent is added

to reduce pentavalent arsenic to trivalent arsenic. Add zinc particles to form and

release arsenic hydride gas. Use iodine solution to absorb and oxidize it to

arsenic acid. Add ammonium molybdate-hydrazine sulfate solution to produce

arsenic molybdenum blue. Then use a spectrophotometer to determine it.

3.2 Reagents and materials

Unless otherwise specified, only use reagents and distilled or deionized water

or equivalent purity water confirmed to be analytically pure in the analysis.

3.2.1 Arsenic-free zinc metal. In accordance with GB/T 2304, screen out zinc

particles with a particle size of not less than 3 mm for use.

3.2.2 Eschka reagent (hereinafter referred to as Aldrin). Commercially available,

or 2 parts by mass of light magnesium oxide and 1 part by mass of anhydrous

sodium carbonate are mixed and ground to a particle size of less than 0.2 mm,

then stored in a closed container.

3.2.3 Hydrochloric acid. The relative density is 1.18.
3.2.4 Hydrochloric acid solution. c(HCl) = 6 mol/L. Mix 1 volume of hydrochloric

acid with 1 volume of water evenly.

3.2.5 Sulfuric acid solution. c(1/2 H2SO4) = 6 mol/L. Measure 167 mL of sulfuric

acid with a relative density of 1.84 and slowly add it to an appropriate amount

of water. Stir it while adding. Then use water to dilute it to 1 L.

3.2.6 Sulfuric acid solution. c (1/2 H2SO4) = 5 mol/L. Measure 139 mL of sulfuric

acid with a relative density of 1.84 and slowly add it to an appropriate amount

of water. Stir it while adding. Then use water to dilute it to 1 L.

3.2.7 Potassium iodide solution. 176.5 g/L. Dissolve 3.0 g of potassium iodide

in 17 mL water and prepare it before use.

3.2.8 Stannous chloride hydrochloric acid solution. 666.7 g/L. Dissolve 8.0 g of

stannous chloride in 12 mL of hydrochloric acid (3.2.3) and prepare it before

use.

3.2.9 Iodine solution. 1.5 g/L. After using a small amount of water to dissolve

9.0 g of potassium iodide and 1.5 g of iodine. Dilute it to 1 L.

3.2.10 Ammonium molybdate solution. 10 g/L. Dissolve 10.0 g of ammonium

molybdate in 1 L sulfuric acid solution (3.2.6).

3.2.11 Hydrazine sulfate solution. 1.2 g/L. Dissolve 1.2 g of hydrazine sulfate in

1 L of water.

3.2.12 Ammonium molybdate-hydrazine sulfate mixed solution. Mix 1 volume

of ammonium molybdate solution (3.2.10) and 1 volume of hydrazine sulfate

solution (3.2.11) evenly. Prepare it before use.

3.2.17 Lead acetate cotton. It is replaced each time of the experiment. The

absorbent cotton is fully soaked in a lead acetate solution which has a mass

concentration of 400 g/L. Take it out and twist it dry. Dry it at 80 °C ~100 °C.

Store it in a desiccator for later use.

3.2.18 Porcelain crucible. The capacity is 30 mL. The enamel on the inner

surface is intact.

3.3 Instrument and equipment

3.5 Test procedure

3.5.1 Drawing of working curve
3.5.1.1 Respectively use the single marking line pipette to take 0.00 mL, 1.00

mL, 2.00 mL, 3.00 mL, 4.00 mL, 5.00 mL of arsenic standard working solution

(3.2.16). Place it into the round flask of the arsenic determination device. First

add 10 mL of sulfuric acid solution (3.2.5). Then add 20 mL of hydrochloric acid

solution (3.2.4). Use water to dilute it to 50 mL.

3.5.1.2 Add 2 mL of potassium iodide solution (3.2.7), 1 mL of stannous chloride

solution (3.2.8). Shake it uniformly. Place it at room temperature for 15 min.

3.5.1.3 Use a pipette to add 3 mL of iodine solution (3.2.9), 1 mL of sodium

bicarbonate solution (3.2.13), 6 mL of water to the absorber of the arsenic

determination device. Insert the absorber into the absorber casing which

contains lead acetate cotton (3.2.17).

abscissa and the corresponding absorbance as the ordinate. Proceed it

simultaneously with the analysis of the coal sample.

3.5.2 Sample determination
3.5.2.1 Weigh 2 g of aldrin agent in a porcelain crucible (accurate to 0.01 g).

Then weigh 0.99 g ~ 1.01 g (accurate to 0.0002 g) of coal sample for general

analysis test which has a particle size of less than 0.2 mm. Use a glass rod to

stir it uniformly. Then use 1 g (accurate to 0.01 g) of Aldrin to evenly cover the

mixed coal sample (no black coal particles can be seen). When the ash content

in the coal sample is greater than 40%, or the arsenic content is greater than

50 µg/g or the total sulfur content is greater than 8%, the sample weight is 0.50

g.

3.5.2.5 Determine the absorbance of the sample blank solution and the sample

solution according to the method specified in 3.5.1.6.Check the corresponding

arsenic mass [micrograms (µg)] from the drawn working curve (3.5.1.7).

3.6 Calculation and expression of results

3.6.1 Result calculation

4 Hydride generation-atomic absorption method

4.1 Method summary

The coal sample is mixed with aldrin and burned. The burned material is

dissolved by hydrochloric acid. The pentavalent arsenic is reduced to trivalent

arsenic by potassium iodide.

4.2 Reagents and materials

Unless otherwise specified, only use reagents and distilled or deionized water

or equivalent purity water confirmed to be analytically pure in the analysis.

4.2.1 Aldrin. Same as 3.2.2.
4.2.2 Hydrochloric acid. Same as 3.2.3.
4.2.3 Sodium hydroxide solution. 5 g/L. Weigh 5.0 g of sodium hydroxide and

dissolve it in 1 L of water.

4.2.10 Nitrogen. Purity above 99.9%.
4.2.11 Porcelain crucible. Same as 3.2.18.
4.2.12 Porcelain evaporating dish. The capacity is 100 mL. The enamel on the

inner surface is intact.

4.3 Instrument and equipment

4.3.1 Atomic absorption spectrophotometer. It has the functions of absorption

peak area integration and peak height measurement.

4.3.2 Hydride generator. It can automatically wash and measure liquid, which

has an accuracy of 0.5%.

4.3.3 Light source. Arsenic hollow cathode lamp or arsenic electrodeless

discharge lamp.

4.3.4 Muffle furnace. Same as 3.3.3.
4.3.5 Analytical balance. Same as 3.3.4.
4.3.6 Balance. The division value is 0.01 g.

4.4 Sample

The sample is the same as 3.4.

4.5 Test procedure

4.5.1 Preparation of sample solution
4.5.2 Preparation of blank solution
4.5.4 Hydride generation-atomic absorption determination
4.5.4.1 Instrument preparation

Install the atomizer of the hydride generator on the burner of the atomic

absorption spectrophotometer according to the instrument instructions, so that

the quartz tube is located directly above the burner slit. Adjust the up and down,

front and rear positions and rotation angle of the burner, to make the axis of the

quartz tube coincide with the main optical axis of the atomic absorption

spectrophotometer. Connect the gas path.

4.5.4.2 Selection of working parameters of atomic absorption

spectrophotometer

The parameter selection is as follows.

4.6 Calculation and expression of results

4.6.1 Result calculation

The arsenic content (mass fraction) in the air-dried coal sample is calculated

according to formula (2).

5 Precision

The precision of the determination of arsenic content in coal is specified in Table

1.

6 Test report

The test report shall include at least the following information.

......

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.

Referenced standards

How to Buy GB/T 3058-2019

  1. 1Add to cart. Click the "Buy GB/T 3058-2019" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 3058-2019

$150.00

$130.00for partners