YY/T 1444-2016Total protein test reagent (English PDF)
总蛋白测定试剂盒
This is a limited preview
Buy now to download the full PDF (25 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1444-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
January 26, 2016
Implementation date
January 1, 2017
Scope
YY/T 1444-2016 is the English-translated version of 总蛋白测定试剂盒.
YY/T 1444-2016 is the Chinese pharmaceutical industry standard for the total protein test reagent based on the biuret method.
It specifies the requirements, the test methods, the labels, the instructions and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision and the stability.
It applies to kits that determine total protein in human serum or plasma by the biuret reaction read spectrophotometrically. Issued on 26 January 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 1444-2016
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative reference documents
- 3 Request
Foreword
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the requirements, test methods, labels, instructions, packaging, shipping and storage of the total protein assay kit (biuret method) And so on.
This standard is applicable to the use of biuret method for human serum (plasma) in the total protein quantitative detection kit, based on spectrophotometry Including hand reagents and reagents used in semi-automatic, fully automated biochemical analyzers.
This standard does not apply to the total protein assay kit for dry chemical methods.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
YY/T 0466.1 Symbols for medical devices for labeling, labeling and providing information on medical devices - Part 1. General requirements
3 Request
3.1 Appearance
Should specify the normal appearance requirements.
3.2 Capacity
The net content of the liquid reagent shall be not less than the indicated value.
3.3 Reagent Blank Absorbance
Reagent blank absorbance should not be greater than 0.200.
3.4 Analysis of sensitivity
70g/L sample absorbance difference (deltaA) should be not less than 0.150.
3.5 Linearity
The linear range of the kit should cover [30.0,100.0] g/L.
a) the linear correlation coefficient r should be not less than 0.995;
b) The linear deviation should not exceed ± 6.0%.
3.6 Precision
3.6.1 Repeatability
Repeat the test (70.0 ± 10.0) g/L of the sample, the results of the coefficient of variation (CV) should not be greater than 2%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1444-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.