YY/T 1422-2016Pregnancy-associated plasma protein A (PAPP-A) in serum test reagent (quantitative labelling immunoassay) (English PDF)
血清妊娠相关血浆蛋白A检测试剂(盒)(定量标记免疫分析法)
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1422-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
July 29, 2016
Implementation date
June 1, 2017
Scope
YY/T 1422-2016 is the English-translated version of 血清妊娠相关血浆蛋白A检测试剂(盒)(定量标记免疫分析法).
YY/T 1422-2016 is the Chinese pharmaceutical industry standard for the kit that measures pregnancy-associated plasma protein A in serum by a labelled immunoassay, the marker used in first trimester screening.
It sets the requirements, the test methods, the identification, labelling and instruction manual and the rules for packaging, transport and storage, covering the appearance, the traceability of the calibrator, accuracy, linearity, the detection limit, precision, the hook effect and stability. Issued on 29 July 2016, it has applied since 1 June 2017.
Found what you need?
Document preview — YY/T 1422-2016
National Standard of the People's Republic of China
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative reference documents
- 3 Request
Foreword
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the requirements for serum pregnancy-related plasma protein A detection reagents (cartridges) (quantitative labeling immunoassay), test methods and Identification, labeling, instruction manual, packaging, transportation and storage.
This standard applies to the serum pregnancy-related plasma protein A quantitative detection reagent (box) (quantitative labeling immunoassay) [hereinafter referred to as PAPP-A reagent (box)]. Quantitative labeling immunoassays include enzyme-linked immunosorbent assay (ELISA), time-resolved fluoroimmunoassay, (Electric) chemiluminescence and so on. This standard does not apply to.
a) various types of colloidal gold marking test paper;
b) Various radioimmunoassay or immunoassay kits labeled with radioisotope such as 125I.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
GB/T 191 packaging and storage icon
GB/T 21415-2008 In Vitro Diagnostic Medical Devices Quantitative Measurement of Biological Samples Metering of Calibration and Control Substance The origin of traceability
YY/T 0466.1 Symbols for medical devices for labeling, labeling and providing information on medical devices - Part 1. General requirements
3 Request
3.1 Appearance
Should be based on the characteristics of the product packaging to specify the appropriate appearance requirements. Generally should be the composition of the components of the kit, traits; inside and outside packaging, labels Clear and so on.
a) kit should be complete components, both inside and outside the packaging should be complete, clear label;
b) liquid reagents no leakage, freeze-dried components were loose body, after the liquid solution evenly (no visible particles, no precipitation).
3.2 Traceability
The source, assignment process and measurement of the PAPP-A calibrator used in accordance with GB/T 21415-2008 and the relevant regulations shall be provided And so on.
3.3 Accuracy
Accuracy should meet one of the following requirements.
a) test with a reference material (CRM) or other recognized reference material that can be used to evaluate the conventional method as a sample.
The relative deviation of the results should not exceed ± 10%;
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB/T 21415-2008In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
Similar standards
How to Download YY/T 1422-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 191-2000 — Packaging-Pictorial marking for handling of goods
GB/T 191-2008 — Packaging - Pictorial marking for handling of goods
GB/T 191-2025 — Graphical symbols marking for handling and storage of packages
YY/T 1422-2016
$115.00