YY/T 1155-2019Automatic luminescence immunoassay analyzer (English PDF)

全自动发光免疫分析仪

Preview PDF

This is a limited preview

Buy now to download the full PDF (25 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1155-2019 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

May 31, 2019

Implementation date

June 1, 2020

Scope

YY/T 1155-2019 is the English-translated version of 全自动发光免疫分析仪.

YY/T 1155-2019 is the Chinese pharmaceutical industry standard for the automatic luminescence immunoassay analyser.

It sets the requirements, the test methods, the labels, markings and instructions for use, and the rules for packaging, transport and storage, covering the stability and linearity of the luminescence measurement, the accuracy and repeatability of the result, the carry-over between samples, the temperature control of the incubation, the sample and reagent handling and the alarms and quality control functions of the instrument. Issued on 31 May 2019, it has applied since 1 June 2020.

Found what you need?

Document preview — YY/T 1155-2019

National Standard of the People's Republic of China

Replacing
YY/T 1155-2009

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions6
  • 4 Requirements6
  • 5 Test methods9
  • 6 Labels, markings, and instructions for use15
  • 7 Packaging, transportation, and storage15
  • Appendix A (Informative) Densities of pure water at different temperatures under standard atmospheric pressure References16

Foreword

This standard was drafted according to the rules given in GB/T 1.1-2009.

1 Scope

This standard specifies the requirements, test methods, labels, markings and instructions for use, packaging, transportation, and storage of automatic luminescence immunoassay analyzers.

This standard applies to automatic luminescence immunoassay analyzers (hereinafter referred to as analyzers) used in medical laboratories. Analyzers use luminescent systems and immunoassay methods to quantitatively or qualitatively detect various analytes in human serum, plasma, or other body fluids, including luminescence immunoassay analyzers based on chemiluminescence, electrochemiluminescence, fluorescence, and other principles.

This standard does not apply to luminescence immunoassay analyzers based on image recognition and automatic luminescence immunoassay analyzers for point-of-care testing (POCT).

2 Normative references

The following documents are essential for the application of this document. For any referenced document with a date, only the version with the date applies to this document.

For any referenced document without a date, the latest version (including all amendments) applies to this document.

GB/T 191 Packaging - Pictorial marking for handling of goods

GB 4793.1 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements

GB 4793.6 Safety requirements for electrical for measurement, control, and laboratory use - Part 6: Particular requirements for laboratory equipment for the heating of materials

GB 4793.9 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes

GB/T 14710 Environmental requirement and test methods for medical electrical equipment

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Luminescence immunoassay

A method that combines a luminescent system with an immune response to detect antigens or antibodies.

3.3 Automatic luminescence immunoassay analyzer

A luminescence immunoassay analyzer by which all analytical processes including the injection of samples and reagents, the provision of the immune binding reaction environment, data measurement, result calculation and output are implemented automatically.

4 Requirements

4.1 Accuracy and repeatability of sample addition

The minimum and maximum sample loading volumes and the minimum and maximum reagent loading volumes specified on the instrument shall be tested and shall comply with the requirements of Table 1.

4.3 Light detection device

4.3.1 Instrument noise

The manufacturer's specifications shall not be exceeded.

4.3.2 Linearity of luminescence values

In the range of no less than 3 luminescence value orders of magnitude, the linear correlation coefficient (r) shall be >=0.99.

4.4 Carryover

The carryover rate shall be <=10-5.For analyzers that can only report qualitative test results, negative samples cannot be tested as positive after high-concentration positive samples are tested.

4.5 Intra-batch precision of clinical items

Select at least one clinical item and conduct the intra-batch precision test using the calibrators and kits specified by the manufacturer. The intra-batch precision shall comply with the requirements of the corresponding national standards and industry standards.

4.6 Main functions of the analyzer

The analyzer shall have the following main functions.

4.7 Appearance

The appearance of the analyzer shall meet the following requirements.

4.8 Environmental test requirements

It shall meet the requirements of climate environment group I and mechanical environment group I in GB/T 14710.

4.9 Security requirements

It shall comply with the requirements of applicable clauses in GB 4793.1, GB 4793.6,

GB 4793.9, and YY 0648.

5 Test methods

5.1 Normal working environment conditions

5.1.1 Power supply voltage. 220 V±22 V; 50 Hz±1 Hz;

5.2 Accuracy and repeatability of sample addition

There are two types of determination methods. colorimetry and weighing method. Manufacturers can choose either method.

5.2.1 Weighing method

The weighing method is carried out according to the following steps.

5.2.2 Colorimetry

The colorimetric method is carried out according to the following steps.

5.3 Accuracy and fluctuation of temperature control in the reaction zone

Place the probe of a temperature detector with a resolution of not less than 0.1 °C, or a calibrated special temperature measuring tool with the same accuracy provided by the analyzer manufacturer, at the location specified by the manufacturer.

5.4 Light detection device

5.4.1 Instrument noise

After the analyzer is turned on and in a stable working state, use a blank sample to test for 20 consecutive times, record the luminescence value of each test, calculate the arithmetic mean 𝐼 of the luminescence values, and calculate the instrument noise IB according to formula (5), which shall meet the requirements of 4.3.1.

5.4.2 Linearity of luminescence values

One of the following two methods can be chosen.

5.6 Intra-batch precision of clinical items

Using calibrators and kits for the clinical test item specified by the manufacturer, test the intra-batch precision in accordance with the corresponding national standards, industry standards, or test kit technical requirements, which shall comply with the provisions of 4.5.

5.7 Main functions of the analyzer

The verification is performed according to the instructions; the functions shall meet the requirements of 4.6.

5.8 Appearance

Visual inspection is carried out; the appearance shall comply with the requirements of 4.7.

5.9 Environmental test requirements

The test is carried out according to the method specified in GB/T 14710. The results shall meet the requirements of 4.8.

5.10 Security requirements

The test is carried out according to the methods specified in GB 4793.1, GB 4793.6,

GB 4793.9, and YY 0648, and the results shall meet the requirements of 4.9.

6 Labels, markings, and instructions for use

They shall comply with the requirements of GB/T 29791.3.

7 Packaging, transportation, and storage

7.1 Packaging

The packaging of the analyzer shall comply with the following provisions.

7.2 Transportation

Transportation shall be carried out in accordance with the requirements specified by the manufacturer.

7.3 Storage

Storage shall be carried out in accordance with the manufacturer's instructions.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 25 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 1155

EditionTitleRevisionStatus
YY/T 1155-2019Automatic luminescence immunoassay analyzercurrent editionCurrent
YY/T 1155-2009Automatic luminescence immunoassay analyzerprevious editionSuperseded

This page sells the current edition, YY/T 1155-2019. Earlier editions are listed for reference only.

How to Download YY/T 1155-2019

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
25 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1155-2019

$225.00