YY/T 0654-2017Automatic chemistry analyzer (English PDF)

全自动生化分析仪

Preview PDF

This is a limited preview

Buy now to download the full PDF (18 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0654-2017 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

March 28, 2017

Implementation date

April 1, 2018

Scope

YY/T 0654-2017 is the English-translated version of 全自动生化分析仪.

YY/T 0654-2017 is the Chinese pharmaceutical industry standard for the automatic chemistry analyser.

It gives the terms and definitions, the classification, the requirements, the test methods, the marks and instructions for use and the rules for packaging, transport and storage, covering the accuracy and repeatability of the photometric measurement, the linearity and the stability of the absorbance, the carry-over between samples and reagents, the temperature control of the reaction, the throughput, the sample and reagent handling and the alarms and quality control functions. Issued on 28 March 2017, it has applied since 1 April 2018.

Found what you need?

Document preview — YY/T 0654-2017

National Standard of the People's Republic of China

Replacing
YY/T 0654-2008

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions6
  • 4 Classification6
  • 5 Requirements7
  • 6 Test methods9
  • 7 Marks and instructions for use15
  • 8 Packaging, transport and storage15
  • Annex A (normative) Preparation method of 50g/L sodium nitrite solution17
  • Annex B (informative) Density of pure water at different temperatures at standard atmospheric pressure18

Foreword

This Standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This Standard was revised based on YY/T 0654-2008. Compared with YY/T 0654-2008, in addition to the editorial modifications, the main technical changes are as follows.

- modified the applicable scope as automatic biochemical analyzer for quantitative analysis of various samples by ultraviolet-visible spectrophotometry;

- in normative references, modified GB/T 14710 "The environmental requirements and test methods for medical electrical equipment" to GB/T 14710 "Medical electrical environment requirements and test methods.' - deleted GB/T 2829 "Sampling procedures and tables for periodic inspection by attributes (Apply to inspection of process stability)" from normative references;

- deleted YY 0466 "Medical devices - Symbols to be used with medical device labels labelling and information to be supplied" (ISO 15233:2000, IDT) from normative references;

- modified the sample carryover rate as not exceeding 0.1% and modified its test methods (see 5.8, 6.7);

- adjusted the UREA (urea) test range in the intra-assay precision of clinical projects to 7.0 mmol/L ~ 11.0 mmol/L (see 5.10);

- added the requirements and test methods for applicable provisions of GB 4793.9 and YY 0648 in safety requirements (see 5.13, 6.12);

- added GB/T 18268.1, GB/T 18268.26 electromagnetic compatibility requirements and test methods (see 5.14, 6.13);

- in the absorbance stability test method, the potassium dichromate solution used at a wavelength of 340 nm was changed to a yellow orange G solution (see 6.4);

- in the absorbance repeatability test method, the potassium dichromate solution used at a wavelength of 340 nm was changed to a yellow orange G solution (see 6.5);

- the accuracy and repeatability of the test method is modified by the manufacturer to choose one of two methods (see 6.8);

- the mark and instructions for use shall be changed to meet the requirements of GB/T 29791.3 (see 7);

- Annex B was replaced by the 1990 International Temperature Standard Pure Water Density Table.

1 Scope

This Standard specifies the terms and definitions, classification, requirements, test methods, marks and instructions for use, packaging, transport and storage of automatic chemistry analyzer (hereinafter referred to as the analyzer).

This Standard is applicable to automatic biochemical analyzer for quantitative analysis of various samples by UV-visible spectrophotometry.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 191, Packaging and storage marks

GB 4793.1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements

GB 4793.9, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes

GB/T 14710, Environmental requirement and test methods for medical electrical equipment

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

3.1 Absorbance

The ratio of the transmitted light intensity to the incident light intensity is the transmittance; the common logarithm of the reciprocal of the transmittance is called the absorbance.

3.3 Carry-over

Carrying a test sample reaction by the measurement system to another analyte discontinuity that detects the sample reaction, thereby erroneously affecting the amount of performance of the other test sample.

3.4 Stray light

Light other than the wavelength that deviates from the normal light path and reaches the detector.

4 Classification

4.1 Instrument type

Discrete, mobile.

4.3 Light path form

Front splitting or post splitting.

4.4 Colorimetric container type

Recycling or single use.

5 Requirements

5.1 Normal working environment

5.1.1 Power supply voltage. 220V ± 22V, 50Hz ± 1Hz

5.2 Stray light

Absorbance is not less than 2.3.

5.3 Absorbance linear range

The maximum absorbance of the relative bias within ±5% shall be no less than 2.0.

5.5 Absorbance stability

The change in absorbance shall not be greater than 0.01.

5.6 Absorbance repeatability

Expressed by coefficient of variation, not exceeding 1.5%.

5.7 Temperature accuracy and volatility

The temperature value is within ±0.3°C of the set value, and the fluctuation is not more than ±0.2°C.

5.8 Sample carry-over rate

The sample carry-over rate shall not exceed 0.1%.

5.10 Intra-assay precision of clinical projects

The coefficient of variation (CV) shall meet the requirements of Table 2.

5.11 Appearance requirements

The appearance shall meet the following requirements. […]

5.12 Environmental test requirements

Meet the requirements of the applicable provisions of GB/T 14710.

5.13 Safety requirements

Meet the requirements of the applicable provisions of GB 4793.1, GB 4793.9, YY 0648.

5.14 Electromagnetic compatibility requirements

Meet the requirements of the applicable provisions of GB/T 18268.1 and GB/T 18268.26.

6 Test methods

6.1 Stray light

Using deionized water as a reference, determine a 50 g/L sodium nitrite standard solution at 340 nm (see Annex A for the preparation method); or measure the absorbance of the JB400 cut-off filter at 340 nm using air as a reference, which shall meet the requirements of 5.2.

6.3 Absorbance accuracy

Use deionized water as a reference. Determine the absorbance of the potassium dichromate standard solution of which the absorbance is about 0.5 (with deionized water as blank, allowable deviation is ±5%) and 1.0 (with deionized water as blank, allowable deviation of ±5%) at 340 nm on the analyzer.

6.4 Absorbance stability

Perform the absorbance stability measurement at any wavelength in the 340 nm and 600 nm to 700 nm wavelength range of the analyzer. The 340 nm measurement solution is Orange G standard solution with an absorbance of 0.5 (with deionized water as a blank and an allowable deviation of ±5%).

6.5 Absorbance repeatability

Carry out the absorbance repeatability measurement at 340 nm wavelength of the analyzer. The 340 nm wavelength measurement solution is Orange G standard solution with an absorbance of 1.0 (with deionized water as a blank and an allowable deviation of ±5%).

6.6 Temperature accuracy and fluctuation

Place the probe of the temperature tester with an accuracy of not less than 0.1°C, or the same precision and calibrated special temperature test tool provided by the analyzer manufacturer, at the position specified by the manufacturer. After the temperature is stable, the temperature shall be measured every other analyzer interval or 30s. The measurement time is the analyzer's nominal maximum reaction time or 10min.

6.10 Appearance

Perform visual inspection, which shall comply with the provisions of 5.11.

6.11 Environmental test

Test in accordance with the method specified in GB/T 14710, the results shall comply with the provisions of 5.12.

6.12 Safety requirements

Test according to the methods specified in GB 4793.1, GB 4793.9, YY 0648, the results shall meet the requirements of 5.13.

6.13 Electromagnetic compatibility

Test according to the method specified in GB/T 18268.1 and GB/T 18268.26, the results shall comply with the provisions of 5.14.

7 Marks and instructions for use

Meet the requirements of GB/T 29791.3.

8 Packaging, transport and storage

8.1 Packaging

The analyzer packaging shall meet the following requirements. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 0654

EditionTitleRevisionStatus
YY/T 0654-2017Automatic chemistry analyzercurrent editionCurrent
YY/T 0654-2008Automatic chemistry analyzerprevious editionSuperseded

This page sells the current edition, YY/T 0654-2017. Earlier editions are listed for reference only.

How to Download YY/T 0654-2017

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
18 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0654-2017

$125.00