YY/T 1533-2017Automatic time-resolved fluorescence immunoassay analyzer (English PDF)
全自动时间分辨荧光免疫分析仪
This is a limited preview
Buy now to download the full PDF (20 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1533-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 28, 2017
Implementation date
April 1, 2018
Scope
YY/T 1533-2017 is the English-translated version of 全自动时间分辨荧光免疫分析仪.
YY/T 1533-2017 is the Chinese pharmaceutical industry standard for the automatic time-resolved fluorescence immunoassay analyser.
It sets the requirements, the test methods, the markings, labels and instructions for use and the rules for packaging, transport and storage, covering the stability and linearity of the fluorescence measurement, the accuracy and repeatability of the result, the carry-over between samples, the temperature control of the incubation, the sample and reagent handling and the alarms and quality control functions of the instrument. Issued on 28 March 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 1533-2017
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Requirements5
- 4 Test method6
- 5 Markings, labels, user instructions11
- 6 Packaging, transportation, storage11
1 Scope
This standard specifies the requirements, test methods, markings, labels, and instructions for use, packaging, transportation, storage of automatic time-resolved fluorescence counter (hereinafter referred to as analyzers).
This standard applies to automatic time-resolved fluorescence counter.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB/T 191 Packaging - Pictorial marking for handling of goods
GB 4793.1 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
GB 4793.9 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 9: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
GB/T 14710 Environmental requirement and test methods for medical electrical equipment
GB/T 18268.1 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
GB/T 18268.26 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 26: Particular requirements - In vitro diagnostic (IVD) medical equipment
GB/T 29791.3 In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 3: In vitro diagnostic instruments for professional use
YY 0648 Safety requirements for electrical equipment for measurement control and laboratory use - Part 2-101: Particular requirements for in vitro If pipetting volume of >= 10 µL and < 50 µL, the relative deviation of loading accuracy is <= ±5%; if pipetting volume of >= 50 µL, the relative deviation of loading accuracy is <= ±3%.
3 Requirements
3.3.2 Precision of sample loading
If pipetting volume of >= 10 µL and < 50 µL, the precision of sample loading CV is <= 5%. If pipetting volume of >= 50 µL, the precision of sample loading CV is < 2%.
3.3.3 Contamination by sample loading needle
Carrying contamination rate is < 0.01%.
3.3.4 Residual amount of lotion
The residual amount of lotion is < 3 µL/hole.
3.4 Temperature accuracy and fluctuation
The temperature accuracy of the incubation chamber is within ±1.5 °C. The temperature fluctuation is within ±1.5 °C.
3.5 Repeatability of clinical test
The in-batch repeatability CV of the kit is <= 15%.
3.6 Environmental test requirements
It shall meet the requirements of the applicable clauses in GB/T 14710.
3.7 Electrical safety requirements
It shall meet the requirements of the applicable clauses in GB 4793.1, GB 4793.9, YY 0648.
3.8 Electromagnetic compatibility requirements
It shall meet the requirements of the applicable clauses of GB/T 18268.1 and GB/T 18268.26.
4 Test method
4.1 Appearance
Visual inspection with normal vision or corrected vision under natural light shall meet the requirements of 3.1. error of 10-10 mol/L europium standard solution.
4.2.5 Stability
After the analyzer is turned on and in a stable working state, take one set of 200 µL 10-10 mol/L europium standard solution to repeat the test 10 times.
Calculate the average value of the measurement results. Then re-start the machine after 4 h and 8 h to repeat the test 10 times; calculate the average value of the measurement result. Take the first measurement result as the reference value; calculate the relative offset (alpha, %) according to formula (2), which shall meet the requirements of 3.2.5. Wherein:
- The average of the measured values in the 4th and 8th hours;
- The average value of the initial measured value.
4.3 Sample loading and cleaning system
4.3.1 Sample loading accuracy
The following methods can be used for verification:
a) Take 12 clean microwell cups; use a calibrated analytical balance (accuracy is ±0.1 mg) to weigh each empty cup; record it as W0i (where i
b) Put the weighed empty cups into the first row of flat-bottomed microplates in turn; add 100 µL of 0.9% saline to the first row of micro-bottomed cups of flat-bottomed microplates; weigh each microwell cup and record it as W1i (where i = 1,..., 12);
c) Calculate the result of (W1i - W0i) and convert it to volume, which is the sample loading volume per hole;
d) Calculate the relative deviation between the actual sample loading volume per hole and the set sample loading volume, which is the sample addition accuracy and shall meet the requirements of 3.3.1.
Note: It can also be verified according to the test method provided by the manufacturer. formula (3) and shall meet the requirements of 3.3.3. Wherein: c - Concentration value of high concentration sample.
4.3.4 Residual volume of lotion
The following methods can be used for verification:
a) Take a clean microplate and use a calibrated ten-thousandth balance to weight it; record it as M1;
b) Put the weighed microplate on any path of plate washing platform; run the plate washing twice; soak it for 2 s;
c) Weigh the microplate after washing the plate and record it as M2;
d) Calculate the average value of the residual volume of each hole; convert it into volume, which is the residual volume of the washing solution and shall meet the requirements of 3.3.4.
4.4 Temperature accuracy and fluctuation
The following methods can be used for verification:
a) Put the temperature sensor probe of the thermometer (accuracy 0.1 °C) into any incubation chamber;
b) Set the temperature of the incubation chamber to be tested to 37 °C and run it continuously for 1 h with continuous shaking;
c) Record the temperature from the 16th minute; record once every 1 minute; record 10 times in a row;
d) Calculate the average value as well as the difference between the maximum and minimum temperatures in the chamber to be tested. The difference between the average value of the chamber temperature and the set temperature value is the temperature accuracy; the half of the difference between the maximum value and the minimum value of the chamber temperature is the temperature fluctuation. The result shall meet the requirements of 3.4.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 20 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 4793.1Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 18268.1Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- GB/T 18268.26Electrical equipment for measurement, control and laboratory use -- EMC requirements -- Part 26: Particular requirements -- In vitro diagnostic (IVD) medical equipment
- GB/T 29791.3In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 3: In vitro diagnostic instruments for professional use
GB 4793.9
Similar standards
How to Download YY/T 1533-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 18268.1-2025 — Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements