YY/T 1078-2008Direct impedance blood flow recorder (English PDF)
直接式阻抗血流图仪
Open the YY/T 1078-2008 preview as PDF
This is a limited preview
Buy now to download the full PDF (12 pages)
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
April 25, 2008
Implementation date
June 1, 2009
Scope
YY/T 1078-2008 is the English-translated version of 直接式阻抗血流图仪.
YY/T 1078 is the product standard for the direct impedance rheograph, which measures the impedance of a body segment directly rather than through a bridge circuit to derive blood flow from the pulsatile change in blood volume. It is the companion to YY/T 1143, covering the bridge-type instrument, and the two were issued on the same day. The performance of either depends on how the excitation current is applied, how the electrodes are placed, and how the instrument extracts a small periodic signal from a much larger static impedance. For a manufacturer or a distributor of impedance rheography equipment serving Chinese hospitals, this is the standard the type testing is run against.
Document preview — YY/T 1078-2008
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C39
- Replacing
- YY 91078-1999
Issued by: National Medical Products Administration
Contents
- 1 Scope
- 2 Normative references
- 3 Basic parameters
- 5 Test Method 3 6 7 inspection rules 7 signs, packaging, transportation and storage
Foreword
This standard is YY 91078-1999 "Direct ICG instrument" revision. This standard YY 91078-1999 major differences are.
--- Safety requirements for the modification according to GB 9706.1-2007 "Medical Electrical Equipment Part 1. General Requirements for Safety" and YY 0505- 2005 "Medical electrical equipment - Part 1-2. General requirements for safety Collateral standard. Electromagnetic compatibility requirements and tests" execution;
--- Environmental testing requirements modified to press GB/T 14710-1993 "Medical electrical equipment environmental requirements and test methods" execution;
--- Delete the original standard technical requirements for non-technical content 2.1,
--- Update outdated reference standard. From the date of implementation of this standard, YY 91078-1999 repealed. This standard was proposed by the National Standardization Technical Committee on Medical Electrical Medical Electronic Instrument Standardization Technical Committee. This standard by the National Standardization Technical Committee on Medical Electrical Medical Electronic Instrument Standardization Technical Committee. This standard was drafted. Shanghai Medical Device detection. The main drafters of this standard. and Yu. This standard replaces the standards previously issued as follows:
--- ZBC39002-1986;
--- YY 91078-1999. Direct ICG instrument
1 Scope
YY/T 1078 is the product standard for the direct impedance rheograph, which measures the impedance of a body segment directly rather than through a bridge circuit to derive blood flow from the pulsatile change in blood volume. It is the companion to YY/T 1143, covering the bridge-type instrument, and the two were issued on the same day. The performance of either depends on how the excitation current is applied, how the electrodes are placed, and how the instrument extracts a small periodic signal from a much larger static impedance. For a manufacturer or a distributor of impedance rheography equipment serving Chinese hospitals, this is the standard the type testing is run against.
This standard specifies the basic parameters of the instrument, technical requirements, test methods, inspection rules and signs, packaging, transportation and storage requirements. The instrument is mainly used for brain, limbs, internal organs and other flow imaging examination, diagnosis for use of clinical or physiological research. This standard applies to direct blood flow impedance analyzer (hereinafter referred to as the instrument).
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
GB/T 191-2008 Packaging - Pictorial signs
GB 9706.1-2007 Medical electrical equipment - Part 1. General Requirements for Safety ( IEC 60601-1. 1988, IDT)
GB/T 14710-1993 Medical electrical equipment environmental requirements and test methods
YY 0505-2005 Medical electrical equipment - Part 1-2. General requirements for safety Collateral standard. Electromagnetic compatibility requirements and tests ( IEC 60601-1-2.2001, IDT)
3 Basic parameters
3.1 Warm-up time. 2min.
3.2 Continuous working time. >=6h.
3.3 tracings recorded the most significant. >= ± 20mm. 4. Technical Requirements
4.1 Instrument Operating Conditions Ambient temperature. 5 °C ~ 40 °C; Relative humidity. <=80%; Atmospheric pressure. 860hPa ~ 1060hPa; Power supply. AC 220V ± 22V, 50Hz ± 1Hz;
4.2 Input impedance Instrument input impedance is not less than 40kOmega.
4.3 base impedance Z0
4.3.1 Z0 measuring range. 5Omega ~ 100Omega.
4.3.2 Z0 measurement error of ± 4%. 4.4.1 DeltaZ. 0 ~ 0.4Omega;
4.5.1 Maximum sensitivity 4.5.1.1 DeltaZ file. >=20mm/0.1Omega;
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 12 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191-2008Graphical symbols marking for handling and storage of packages
- GB 9706.1-2007Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 14710-1993The environmental requirement and test methods for medical electrical equipment
IEC 60601 · YY 0505-2005
Editions of YY/T 1078
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 1078-2008 | Direct impedance blood flow recorder | current edition | Current |
| YY 91078-1999 | Direct impedance blood flow recorder | previous edition | In force until 2009-06-01 |
This page sells the current edition, YY/T 1078-2008. Earlier editions are listed for reference only.
How to Buy YY/T 1078-2008
- 1Add to cart. Click the "Buy YY/T 1078-2008" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0492-2026 — Implantable leads
YY/T 0517-2026 — Dentistry - Posts
Secure payment via Stripe
Payments accepted
YY/T 1078-2008
$140.00