Scope

YY/T 1519-2017 is the English-translated version of 电生理标测导管.

YY/T 1519-2017 is the Chinese pharmaceutical industry standard for the electrophysiology catheter and its connection cable, the catheter passed into the heart to record electrograms and to stimulate.

It gives the terms and definitions, the requirements and the test methods, covering the dimensions and the flexibility of the shaft, the electrodes and their spacing and impedance, the continuity and insulation of the conductors, the connector and the cable, the tensile strength of the assembly, the radiopacity, the materials and their biological evaluation, and sterility. Issued on 28 March 2017, it has applied since 1 April 2018.

Found what you need?

Document preview — YY/T 1519-2017

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions5
  • 4 Requirements5
  • 5 Test methods8

1 Scope

This Standard specifies the terms and definitions, requirements, test methods for electrophysiology catheter and connection cable.

This Standard is applicable to the electrophysiology catheter specified in 3.1 (hereinafter referred to as the catheter) and connection cable (hereinafter referred to as the cable). For electrophysiological catheters with special functions and features, the manufacturer shall supplement the corresponding requirements.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB 9706.1-2007, Medical electrical equipment - Part 1. General requirements for safety

GB/T 14233.1-2008, Test methods for infusion transfusion injection equipment for medical use - Part 1. Chemical analysis methods

GB/T 14233.2-2005, Test methods for infusion, transfusion, injection equipment for medical use - Part 2. Biological test methods

GB/T 16886.4-2003, Biological evaluation of medical devices - Part 4. Selection of tests for interactions with blood

GB/T 16886.5-2003, Biological evaluation of medical devices - Part 5. Test for in vitro cytotoxicity

GB/T 16886.10-2005, Biological evaluation of medical devices - Part 10. Tests for irritation and delayed-type hypersensitivity

GB/T 16886.11-2011, Biological evaluation of medical devices - Part 11. Tests for systemic toxicity

YY 0285.1-2004, Sterile, single-use intravascular catheters - Part 1. General requirements than 5 Mohm.

4 Requirements

4.4 Chemical properties

4.4.1 Reducing substance

The test solution is compared with the same volume and the same batch of blank control solution. The difference between the consumption of potassium permanganate solution [c(KMnO4)=0.002 mol/L] shall not exceed 2.0 mL.

4.4.2 Heavy metal

The color of the test solution shall not exceed the standard control solution with a mass concentration of rho(Pb2+)=1 µg/mL.

4.4.3 pH

The pH of the test solution is compared with the same batch of blank control solution, and the difference between pH shall not exceed 1.5.

4.4.4 Evaporation residue

In 50mL test solution, the total mass of non-volatiles shall not exceed 2 mg.

4.4.5 Ethylene oxide residue (if applicable)

The residual amount of ethylene oxide shall be no more than 10 µg/g.

4.4.6 Ultraviolet absorbance

When the test solution has a wavelength range of 250 nm to 320 nm, the absorbance shall be no more than 0.1.

4.4.7 Corrosion resistance

After the corrosion test, the metal parts of the catheter shall not have corrosion marks.

4.5 Biological performance

4.5.1 Sterile

Sterilized catheter shall be sterile.

4.5.2 Pyrogen

The catheter shall be pyrogen free.

4.5.3 Cytotoxicity

The chemical performance test liquid can be prepared by using only the material in the catheter that is in contact with the human body to prepare the test solution. Adopt the method of No. 2 in Table 1 of GB/T 14233.1-2008. If the finished catheter product is used for verification, the test solution shall be prepared by using the method without destroying the catheter, and immersed according to the method of No. 2 in Table 1 of GB/T 14233.1-2008.

5 Test methods

5.4.1 Reducing substance

When testing according to 5.2.2 of GB/T 14233.1-2008, it shall meet the requirements of 4.4.1.

5.4.2 Heavy metal

When testing according to 5.6.2 of GB/T 14233.1-2008, it shall meet the requirements of 4.4.2.

5.4.3 pH

When testing according to 5.4.1 of GB/T 14233.1-2008, it shall meet the requirements of 4.4.3.

5.4.4 Evaporation residue

When testing according to 5.5 of GB/T 14233.1-2008, it shall meet the requirements of 4.4.4.

5.4.5 Ethylene oxide residue

When testing according to 9 of GB/T 14233.1-2008, it shall meet the requirements of 4.4.5.

5.4.6 Ultraviolet absorbance

When testing according to 5.7 of GB/T 14233.1-2008, it shall meet the requirements of 4.4.6.

5.4.7 Corrosion resistance

When testing according to Annex A of YY 0285.1-2004, it shall meet the requirements of 4.4.7.

5.5 Biological performance

5.5.1 Sterile

According to the method in GB/T 14233.2-2005, the test results shall be sterile.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

How to Download YY/T 1519-2017

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1519-2017

$125.00