YY 1139-2013Diagnostic electrocardiographic devices (English PDF)
心电诊断设备
Open the YY 1139-2013 preview as PDF
This is a limited preview
Buy now to download the full PDF (32 pages)
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY 1139-2013 is the English-translated version of 心电诊断设备.
YY 1139 is the mandatory Chinese standard for diagnostic electrocardiographs, the twelve-lead machines whose output is read as a diagnostic document rather than as a monitoring trace. The distinction matters: a diagnostic ECG has to reproduce the waveform faithfully enough that ST segment deviation of a fraction of a millivolt can be measured, because that measurement decides whether a patient goes to the catheterisation laboratory. The standard sets the requirements such a device has to meet and the test methods that verify them, covering frequency response, gain accuracy, noise, the common mode rejection that suppresses mains interference, and the automated measurement and interpretation where provided. It replaces YY 1139-2000. Being a YY standard without the /T suffix, compliance is compulsory in China.
Document preview — YY 1139-2013
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C39
- Replacing
- YY 1139-2000
Issued by: National Medical Products Administration
Contents
- ForewordI
- 1 Scope1
- 2 Normative references1
- 3 Definitions2
- 4 Requirements5
- 5 Test methods15
- Appendix A (informative) introduced the basic reason of this standard29
Foreword
This special standard for medical electrical equipment performance standards.
This dedicated drafting standards in accordance with GB/T 1.1-2009 given rules.
This standard replaces YY 1139-2000 "single-channel and multi-channel ECG machine", and YY 1139-2000 main technical differences are as follows.
--- Remove the sensitivity and output impedance require an external output, replace manual should indicate the correct external device is connected;
--- Remove the sensitivity and input impedance requires an external DC input signal;
--- Remove the print resolution requirements;
--- Adjust the frequency characteristics of the requirements;
--- Adjust the noise level requirements;
--- Increasing the time and amplitude scale requirements;
--- To increase the release of energy loss in patients with the requirements;
--- Increased requirements pacing pulse display capabilities.
Since the implementation of this standard, YY 1139-2000 "single-channel and multi-channel ECG machine" abolished.
The standard reference to American National Standard ANSI/AAMIEC 11.1991/(R) 2001/(R) 2007 "ECG diagnostic equipment."
Please note that some of the content of this document may involve patents. Release mechanism of the present document does not assume responsibility for the identification of these patents.
This standard by the National Standardization Technical Committee on Medical Electrical Medical Electronic Instrument Standardization Technical Committee (TC10/SC5) proposed.
This standard by the National Standardization Technical Committee on Medical Electrical Medical Electronic Instrument Standardization Technical Committee (TC10/SC5) centralized.
This standard was drafted. Shanghai Medical Device detection, Shenzhen City rationale State precision Instruments Incorporated.
The main drafters of this standard. Yu and Tao Kan, Qiu everywhere.
1 Scope
This standard ECG diagnostic equipment (hereinafter referred to, or devices) 3.9 definition establishes the minimum performance requirements, the device has credited directly
It means recorded. All parts of ECG diagnostic equipment applicable to this standard, they should include ECG obtained from the body surface, amplifies the signal
And adapted to a diagnosis of cardiac electrical activity displayed the signal. This standard ECG electrodes diagnostic equipment from the input to the output is displayed Regulation Set requirements.
NOTE. In the performance criterion specified in this standard is mainly used to verify the manufacturer's design or evaluation.
Chapter 5 of the ruling purpose test method is to provide some means to explicitly establish compliance with this standard. These test methods are not used
To verify the performance of a single device, whether it is used to check the quality of a manufacturer's warranty or routine inspections within the hospital. In addition, according to the ruling again
Defined test method allows the use of other design verification method, as long as those methods equivalent to those ruled to have test methods comparable test Test results.
1.1 For
Apply to this standard includes the following equipment.
a) direct recording ECG;
b) used in other medical devices (eg. patient monitors, defibrillators, stress testing equipment) in the ECG machine, as long as these devices with
ECG diagnostic functions;
c) the patient ECG signal display functions remotely, through their cable, telephone, telemetry or storage medium, as long as these devices have heart
FIG electrical diagnostic functions. These devices comply with the entire output of the system - the relationship between the input functions of performance requirements.
1.2 Not applicable equipment
Within the scope of this standard does not include the equipment.
a) Equipment ECG data collected from the outer surface of the non-human;
b) equipment for the Interpretation and pattern recognition (for example, QRS detection, alarm circuit, heart rate meter, diagnostic algorithms);
c) fetal electrocardiogram monitor;
d) Holter monitoring equipment, including ECG recorders and auxiliary scanning and reading device;
e) diagnostic ECG devices nonpermanent display;
f) vector ECG machine, it is a cycle;
g) ECG device, expected to be used in extreme or uncontrolled environmental conditions of the device in a hospital or clinic outside;
h) with or without a heart rate meter and the alarm of ECG, the monitor is mainly used to detect heart rhythm (ECG monitoring equipment from these
Instrument care covered by the standard).
Note: If the device has both diagnostic functions with monitoring function, the function should be selected must meet the relevant requirements of the standard. meet the diagnostic function ECG diagnostic equipment
The standards, to meet the monitoring function of ECG standards.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein
Member. For undated references, the latest edition (including any amendments) applies to this document.
Remaining clauses in the full document
- 3 Definitions
- 4 Requirements
- 5 Test methods
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 32 pages — is available in the English PDF.
Referenced standards
Normative references
- YY/T 1933Particular specification for MR injector
- YY/T 1994Vascular interventional surgical control system employing robotic technology
- YY/T 1993Reliability testing methods for robotically assisted surgical equipment
- YY/T 1143Bridge type impedance blood flow recorder
YY/T 1712 · YY/T 1519
Editions of YY 1139
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 1139-2013 | Diagnostic electrocardiographic devices | current edition | Current |
| YY 1139-2000 | Diagnostic electrocardiographic devices | previous edition | In force until 2014-10-01 |
This page sells the current edition, YY 1139-2013. Earlier editions are listed for reference only.
How to Buy YY 1139-2013
- 1Add to cart. Click the "Buy YY 1139-2013" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 1143-2008 — Bridge type impedance blood flow recorder
YY/T 1933-2024 — Particular specification for MR injector
YY/T 1993-2025 — Reliability testing methods for robotically assisted surgical equipment
Secure payment via Stripe
Payments accepted
YY 1139-2013
$710.00