YY/T 0752-2016Electrical surgical equipment for osseous tissue (English PDF)
电动骨组织手术设备
This is a limited preview
Buy now to download the full PDF (24 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0752-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
January 26, 2016
Implementation date
January 1, 2017
Scope
YY/T 0752-2016 is the English-translated version of 电动骨组织手术设备.
YY/T 0752-2016 is the Chinese pharmaceutical industry standard for electrical surgical equipment for bone tissue, the powered handpieces and consoles that drill, saw and mill bone.
It gives the terms and definitions, the classification of the products, the technical requirements and the test methods, covering the speed and the torque delivered and their accuracy, the run-out and vibration of the tool, the noise and the heat produced, the coupling of the handpiece and the tool, the irrigation, the cleaning and sterilization and the electrical safety. Issued on 26 January 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 0752-2016
National Standard of the People's Republic of China
- Replacing
- YY/T 0752-2009
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions5
- 4 Product classification6
- 5 Technical requirements7
- 6 Test methods9
1 Scope
This Standard specifies terms and definitions, product classification, technical requirements, test methods for electrical surgical equipment for osseous tissue.
This Standard is applicable to electrical surgical equipment for osseous tissue powered by network power supply (hereinafter referred to as the equipment).
This Standard is inapplicable to electrical surgical equipment for osseous tissue powered by internal power supply. This Standard is inapplicable to similar equipment for dentistry.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB 9706.1-2007, Medical electrical equipment - Part 1: General requirements for safety (IEC 60601-1:1988, IDT)
GB/T 14710-2009, Environmental requirement and test methods for medical electrical equipment
YY/T 0149-2006, Medical instruments of stainless steel - Test methods of corrosion resistance (ISO 13402:1995, MOD)
YY 0505-2012, Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standards: Electromagnetic compatibility - Requirements and tests (IEC 60601-1-2:2004, IDT)
YY 1057-2016, General specifications for medical foot switch
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1 Electric surgical equipment for osseous tissue
a medical instrument that is powered by the network power supply to provide mechanical power required for surgical tools to perform osseous tissue surgery
3.2 Main frame
a device that provides mechanical kinetic energy and/or electrical energy for handpiece in electric osseous tissue surgery equipment, and implements real-time monitoring of its output
3.3 Flexible cable
a component that realizes the transmission of mechanical kinetic energy between the main frame and the handpiece
3.4 Cable
a component that realizes the electrical connection between the main frame and the handpiece
3.5 Handpiece
a component that the handpiece is held by the operator and can drive the clamped surgical knife to achieve the purpose of surgery; the handpiece is divided into integrated type or split type; the split type is composed of the (power) handle that can provide power to the machine head, the machine head that clamps the tool, and (or) the speed change direction mechanism and other components
3.6 Cutter
an apparatus that is clamped on a handpiece to perform osseous tissue surgery
3.7 Rotation speed/frequency of work
within the maximum output speed/frequency range of the handpiece output specified by the manufacturer, the corresponding output speed/frequency under a specific workload
3.8 Load of work
the torque and/or force specified by the manufacturer that can sustain a certain operating speed/frequency
4 Product classification
The classification according to the connection mode of the main frame and the handpiece is as follows:
a) Clamp the test rod or cutter to the handpiece. Then fix the handpiece on the measurement platform. Test with 30N of axial tension acting on the test rod or cutter. The test rod or cutter shall not fall off;
b) Clamp the test rod or cutter with the largest specification to the handpiece.
Then fix the handpiece on the measurement platform. Gradually load the test rod or cutter to 1.5 times the rated load. There shall be no relative slippage or permanent deformation between the test rod or cutter and the chuck.
a) When the set rotation speed is <=3000r/min, the equipment runs at low speed and no load. After the speed is stable, use a dial indicator to measure. Record the radial runout of the test rod or tool 20mm from the output end of the handpiece. Measure 3 times. The arithmetic average value shall meet the requirements of 5.2.3.
b) When the set rotation speed is >3000r/min, run at no-load within the recommended speed range. After the speed is stable, use a non-contact measuring instrument to measure. Record the radial runout of the test rod or cutter 20mm away from the output end of the handpiece. Measure 3 times. The arithmetic average value shall meet the requirements of 5.2.3.
Use a dial indicator to measure the relative displacement between the test rod or cutter and the handpiece chuck, which shall meet the requirements of 5.2.4.
After 3min, measure the highest temperature that can touch the surface of the handpiece shell. The test is carried out at an ambient temperature of 20°C ± 2°C.
Conduct visual inspection with a magnifying glass or hand feel inspection, without sharp edges, burrs, sharp corners.
6 Test methods
6.2.2 Perform the loading and unloading test of the test rod or cutter and the handpiece according to the instructions, which shall meet the requirements of 5.2.2.
6.2.3 Clamp the test rod or cutter to the handpiece. Then fix the handpiece on the test platform. Keep the test rod or cutter parallel to the test platform.
6.2.4 Install a test rod or cutter on the handpiece output. Fix the handpiece on the test platform. Apply an axial push-pull force of 10N on the test rod or cutter.
6.2.5 According to the manufacturer's instructions, load the largest specification test rod or cutter at the maximum speed and run the handpiece with no load.
6.2.6 Use ample block comparison method or electrical measurement method for surface roughness. The electrical measurement method shall be used for measurement during arbitration and shall meet the requirements of 5.2.6.
6.3 Flexible cable performance test
6.3.1 Measure the length of the flexible cable, which shall meet the requirements of 5.3.1.
6.3.2 Load and unload the flexible cable according to the method specified in the manual. Apply 30N of axial pulling force on the connecting end of the flexible cable and the main frame AND the connecting end of the flexible cable and the handpiece respectively, which shall meet the requirements of 5.3.2.
6.3.3 Test according to the following steps:
a) After the flexible cable is connected to the main frame, the output end of the flexible cable is fixed on the test platform;
b) Connect the output end of the flexible cable to the test device through the connector. The output speed/frequency of the equipment is set at the highest value and it is turned on and runs without load. Place the flexible cable under its natural curve during normal operation. Then adjust the distance between the test device and the main frame until the curvature radius of one of the bending parts of the flexible cable reaches the minimum radius specified by the standard;
c) Gradually load until the device reaches the overload protection, it shall meet the requirements of 5.3.3.
6.4 Connecting cable test
6.4.1 Measure the cable length, which shall meet the requirements of 5.4.1.
6.4.2 Apply an axial pulling force of 20N on the cable connecting the cable to the main frame and the cable connecting to the handpiece, which shall meet the requirements of 5.4.2.
6.5 Foot controller test
6.5.1 Test according to the requirements specified in YY 1057-2016, which shall meet the requirements of 5.5.1.
6.5.2 Apply 20N of axial tension on the cable that connects the foot controller and the main frame, which shall meet the requirements of 5.5.2.
6.6 Environmental test
Test according to the requirements of 5.6. The equipment can still work normally after the test.
6.7 Safety requirements
Test according to the method specified in GB 9706.1-2007 and YY 0505-2012,
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1-2007Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 14710-2009The environmental requirement and test methods for medical electrical equipment
- YY/T 0149-2006Medical instruments of stainless steel - Test methods of corrosion resistance
- YY 0505-2012Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
- YY 1057-2016General specifications for medical foot switches
Cited by
- YY/T 1901-2023Requirements and test methods for orthopaedic surgical navigation equipment using robotic technology
- YY/T 1629.3-2022Cutting tools for electric osseous tissue surgical equipment - Part 3: Drill
- YY/T 1629.6-2021Cutting tools for electric osseous tissue surgical equipment - Part 6: Rasp
- YY/T 1629.4-2020Cutting tools for electric osseous tissue surgical equipment - Part 4: Milling cutter
- YY/T 1629.5-2020Cutting tools for electric osseous tissue surgical equipment - Part 5: Saw blade
- YY/T 1629.1-2018Cutting tools for electric osseous tissue surgical equipment - Part 1: Burs
Similar standards
Editions of YY/T 0752
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0752-2016 | Electrical surgical equipment for osseous tissue | current edition | Current |
| YY/T 0752-2009 | Electrical surgical equipment for osseous tissue | previous edition | Superseded |
This page sells the current edition, YY/T 0752-2016. Earlier editions are listed for reference only.
How to Download YY/T 0752-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
YY 0505-2012 — Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests