YY 1057-2016General specifications for medical foot switches (English PDF)
医用脚踏开关通用技术条件
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY 1057-2016 in 30 minutes.
Issued by
China Food and Drug Administration
Level / Type
Industry · Mandatory
Issue date
January 26, 2016
Implementation date
January 1, 2018
Scope
YY 1057-2016 is the English-translated version of 医用脚踏开关通用技术条件.
YY 1057-2016 is the Chinese pharmaceutical industry standard giving the general specifications for medical foot switches. It specifies terms and definitions, requirements and test methods for foot switches used as control circuit devices of medical electrical equipment; pneumatic and hydraulic foot switches are excluded, though they may refer to it.
Performance requirements cover the actuating force (not less than 10 N and not more than 50 N, applied over any 625 mm2 of the operating surface), the working resistance of make- and break-contact elements (measured by the Kelvin four-wire method), mechanical strength, mechanical durability (30 000 no-load operating cycles), a bending test of the connecting cable (3 kg load, 90 degree bends held for 1 min) and ingress of liquids to the IPX class declared by the manufacturer under GB 4208-2008; safety follows GB 9706.1 and the particular standards of the equipment concerned.
It replaces YY 91057-1999. Issued on 26 January 2016 by the China Food and Drug Administration, it has been in force since 1 January 2018. Classification: ICS 11.040.40, CCS C 30.
Found what you need?
Document preview — YY 1057-2016
National Standard of the People's Republic of China
- ICS
- 11.040.40
- Classification
- C 30
- Replacing
- YY 91057-1999
Issued by: China Food and Drug Administration
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions5
- 4 Requirements5
- 5 Test methods7
1 Scope
This standard specifies the terms and definitions, requirements and test methods for medical foot switches.
This standard applies to foot switches used as control circuit devices of medical electrical equipment. Pneumatic and hydraulic foot switches are outside its scope.
Note: foot switches outside the scope of this standard may still refer to its relevant provisions.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB 4208-2008, Degrees of protection provided by enclosure (IP code)
GB 9706.1, Medical electrical equipment - Part 1: General requirements for safety
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1 make-contact element (normally open): a contact element that closes a circuit when the switch is actuated.
3.2 break-contact element (normally closed): a contact element that opens a circuit when the switch is actuated.
4.1 Performance requirements
4.1.1 Actuating force: the actuating force of the foot switch shall be not less than 10 N and not more than 50 N.
4.1.2 Working resistance: for foot switches with make-contact (normally open) or break-contact (normally closed) elements, the working resistance with the contact element closed shall not exceed the value specified in the standard.
4.1.4.1 The foot switch shall have sufficient mechanical durability.
4.1.4.2 After the mechanical durability test, foot switches with make-contact or break-contact elements shall still meet the requirements of 4.1.2.
4.1.5 Bending test of the connecting cable: after the test of 5.1.5, the cable entry of the foot switch shall not be damaged and the function shall be unimpaired.
4.2 Safety requirements
Medical foot switches shall comply with GB 9706.1 and with the relevant requirements of the particular standards applicable to the equipment with which they are used.
5.1 Performance tests
5.1.1 Actuating force: apply the test force, perpendicular to the operating surface, over any 625 mm2 area of the operating surface of the foot switch. Measure the minimum force at which the foot switch performs its intended control function. The result shall meet 4.1.1.
5.1.2 Working resistance: for foot switches with make-contact or break-contact elements, with the contact element closed, measure the working resistance at the far end of the foot switch cable by the Kelvin four-wire method. The result shall meet 4.1.2.
5.1.4.1 Operate the foot switch for 30 000 no-load operating cycles (no operating current); the movable part shall reach its maximum operating position in both directions.
5.1.4.4 After the test of 5.1.4.1, place the foot switch upside down and in every other possible abnormal position on the supporting surface. The control setting of the foot switch shall not change unintentionally.
5.1.5 Bending test of the connecting cable: with the axis of the cable entry vertical, hang a 3 kg mass on the cable 1 m from the entry. Tilt the foot switch so that the cable bends through 90 degrees and hold for 1 min; then tilt it the opposite way so that the cable bends through 90 degrees and hold for 1 min. After the test the cable shall meet 4.1.5.
5.1.6 Ingress of liquids: test the foot switch for ingress of water in accordance with Clause 14 of GB 4208-2008, to the IPX class specified by the manufacturer.
5.2 Safety tests
Compliance with 4.2 is checked by the tests of GB 9706.1 and the relevant tests of the particular standards applicable to the equipment with which the foot switch is used.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
GB 4208-2008
Cited by
- YY/T 0776-2023Liver radiofrequency ablation therapy equipment
- YY 0307-2022Laser therapeutic equipment - Nd: YAG laser therapeutic equipment
- YY 1289-2022Laser treatment equipment - Ophthalmic laser photocoagulator
- YY/T 0740-2022Particular specifications for medical X-ray angiography equipment
- YY 0838-2021Microwave ablation and coagulation equipment
- YY/T 0742-2021Particular specifications for gastrointestinal diagnostic X-ray equipment
Editions of YY 1057
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 1057-2016 | General specifications for medical foot switches | current edition | Current |
| YY 91057-1999 | General specifications for medical foot switches | previous edition | Superseded |
This page sells the current edition, YY 1057-2016. Earlier editions are listed for reference only.
How to Download YY 1057-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
YY 0017-2026 — Implants for osteosynthesis - Metallic bone plates
YY 0018-2026 — Implants for osteosynthesis - Metallic bone screws
Secure payment via Stripe
Payments accepted
YY 1057-2016
$55.00