YY/T 1601-2018Ultrasonic surgical equipment for osseous tissue (English PDF)
超声骨组织手术设备
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1601-2018 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
January 19, 2018
Implementation date
January 1, 2019
Scope
YY/T 1601-2018 is the English-translated version of 超声骨组织手术设备.
YY/T 1601-2018 is the Chinese pharmaceutical industry standard for ultrasonic surgical equipment used on bone, the ultrasonic bone cutting systems.
It sets the technical requirements, the test methods, the inspection rules, the marks and the user's manual, covering the working frequency and the amplitude of the tip, the output power and its stability, the cutting performance and the temperature reached, the handpiece and the tips, the irrigation, the alarms and the electrical safety. Issued on 19 January 2018, it has applied since 1 January 2019.
Found what you need?
Document preview — YY/T 1601-2018
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Technical requirements5
- 5 Test methods7
- 6 Inspection rules9
- 7 Marks and user's manual10
1 Scope
This Standard specifies technical requirements, test methods, inspection rules, marks and user's manual for ultrasonic surgical equipment for osseous tissue (hereinafter referred to as the equipment).
This Standard is applicable to ultrasonic surgical equipment for osseous tissue.
This Standard is not applicable to ultrasonic tooth cleaning device for removing plaque and stones on teeth.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 191, Packaging and storage marks
GB 9706.1-2007, Medical electrical equipment - Part 1: General requirements for safety
GB/T 9969-2008, General Principles for Preparation of Instructions for Use of Industrial Products
GB/T 14710-2009, Environmental requirement and test methods for medical electrical equipment
YY/T 0644-2008, Ultrasonic - surgical systems - Measurement and declaration of the basic output characteristics
3 Terms and definitions
For the purposes of this document, the terms and definitions defined in YY/T 0644-2008 as well as the followings apply.
3.1 Ultrasonic surgical equipment for osseous tissue
Surgical equipment powered by ultrasound, of which the frequency is generally The output sound power of each type of tip of each ultrasonic hand-held component shall be within the range announced by the manufacturer.
4 Technical requirements
4.8 Duty cycle
If the main amplitude of the tip is modulated, its duty cycle or range of duty cycle shall conform to the manufacturer's published value.
4.9 Static (no load) electrical power
The measured static (no-load) electrical power shall be less than the maximum value of the static (no-load) electrical power of each ultrasonic handheld component published by the manufacturer.
4.10 Maximum electric power
The measured maximum electric power shall be less than the maximum electric power value of each ultrasonic handheld component published by the manufacturer.
4.11 Power reserve index
The power reserve index shall conform to the value published by the manufacturer.
4.12 Main sound output area
If applicable, the output area of the main sound of each type of tip shall conform to the manufacturer's published value.
4.13 Output area of secondary transverse vibration sound
If applicable, the secondary transverse vibration sound output area of each type of tip shall conform to the manufacturer's published value.
4.14 Liquid flow
The manufacturer shall publish the adjustment range of liquid flow. The error of liquid flow relative to the nominal value shall not exceed ±10%.
4.15 Noise
It shall not be greater than 90dB (A-weighted).
4.16 Features to be published in user's manual
a) Main amplitude at the tip of the treatment head;
b) Area of the main sound output of the treatment head;
5 Test methods
5.5 Excitation frequency
Measure according to the method specified in 6.3.1 of YY/T 0644-2008. The result shall meet the requirements of 4.5.
5.6 Tip vibration frequency
Measure according to the method specified in 6.4 of YY/T 0644-2008. The result shall meet the requirements of 4.6.
5.7 Exported output sound power
Measure according to the method specified in 6.5.1 of YY/T 0644-2008. The result shall meet the requirements of 4.7.
5.8 Duty cycle
Measure according to the method specified in 6.8 of YY/T 0644-2008. The result shall meet the requirements of 4.8.
5.9 Static (no load) electrical power
Measure according to the method specified in 6.9 of YY/T 0644-2008. The result must not exceed the published value.
5.10 Maximum electric power
Measure according to the method specified in 6.10 of YY/T 0644-2008. The result must not exceed the published value.
NOTE: Simulated tip loads can use bone approximations, such as phenolic plastic and so on.
5.11 Power reserve index
According to the results of 5.9 and 5.10, calculate the power reserve index according to the method specified in 6.13 of YY/T 0644-2008.
5.12 Main sound output area
According to the definition of 3.9 in YY/T 0644-2008, calculate according to the method specified in 6.11 in YY/T 0644-2008.
5.13 Output area of secondary transverse vibration sound
According to the definition of 3.15 in YY/T 0644-2008, calculate according to the method specified in 6.12 in YY/T 0644-2008.
5.14 Liquid flow
c) When requested by the national quality supervision department.
6 Inspection rules
6.3.2 Type inspection items
All requirements specified in this Standard. The number of samples for type inspection is one.
6.3.3 Determination for conformity of type test
In the inspection items, if there are items that do not meet the requirements, it is allowed to repair the nonconforming items. After the adjustment and repair, for the items that may be related to the failure, the retest must all meet the requirements, otherwise it shall be determined as failure.
The inspection items and determination rules for quality supervision and inspection shall be separately stipulated by the quality supervision agency.
7 Marks and user's manual
7.1 Marks
7.1.1 The following marks shall be on the equipment:
a) Manufacturer's name;
b) Product named and model;
c) Power used;
d) Production date or exit-factory number;
e) Other marks required by applicable standards and regulations.
7.1.2 The following marks shall be on the box:
a) Manufacturer's name and address;
b) Product name, model, quantity;
c) Weight, volume (length × width × height);
d) For "fragile items", "upward", "fear of rain" and other words or marks, the text and graphics shall comply with the relevant provisions of GB/T 191- 2008;
e) Other marks required by applicable standards and regulations.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 9706.1-2007Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 9969-2008General principles for preparation of instructions for use of industrial products
- GB/T 14710-2009The environmental requirement and test methods for medical electrical equipment
- YY/T 0644-2008Ultrasonics- surgical systems - Measurement and declaration of the basic output characteristics
Similar standards
How to Download YY/T 1601-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards