YY 0569-2011Class II biological safety cabinets (English PDF)
II级 生物安全柜
This is a limited preview
Buy now to download the full PDF (47 pages)
Need quotation for your management?
Get a formal quotation for YY 0569-2011 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
December 31, 2011
Implementation date
June 1, 2013
Scope
YY 0569-2011 is the English-translated version of II级 生物安全柜.
YY 0569-2011 is the Chinese pharmaceutical industry standard for the class II biological safety cabinet, the ventilated cabinet that protects the operator, the product and the environment by a downflow of filtered air.
It specifies the terms and definitions, the classification, the materials, the structural and performance requirements, the test methods, the inspection rules, the signs, labels and instruction manual and the requirements for packaging, transport and storage, covering the inflow and downflow velocities, the HEPA filter and its integrity, the containment, the noise, the illumination and the vibration, and the electrical safety. Issued on 31 December 2011, it has applied since 1 June 2013.
Found what you need?
Document preview — YY 0569-2011
National Standard of the People's Republic of China
- ICS
- 11.140
- Classification
- C 48
- Replacing
- YY 0569-2005
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 II level security cabinet types and characteristics of
- 5 requires
- 6 Test methods
- 7 Inspection rules
- 8 label, tag
- 9 packaging, transportation and storage
- Appendix A (informative) security cabinet installation recommendations
- Annex B (informative) Recommended disinfection procedures Preparation of Appendix C (normative) of Bacillus subtilis spore suspension of
- Annex D (normative) sprayer calibration and selection
- Appendix E (normative) potassium iodide method
- Annex F (normative) Measurement circular and rectangular duct air volume
- References
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
1 Scope
This standard specifies the terms and definitions grade II biological safety cabinets, classification, materials, structure and performance requirements, test methods, inspection rules, Signs, labels, brochures, packaging, transportation and storage requirements.
This standard applies to grade II biological safety cabinets (hereinafter referred to as safety cabinets).
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 191-2008 Packaging - Pictorial signs
GB 4793.1-2007 measurement, control and laboratory safety requirements for electrical equipment - Part 1. General requirements Disinfection technical specifications (2002 edition) Ministry of Health
3 Terms and Definitions
The following terms and definitions apply to this document.
3.1 Biological Safety Cabinet
biologicalsafetycabinet; BSC Fume suction filter to prevent the operator and environmental exposure to experiment generated during biological aerosols.
Biological safety cabinet is divided into three levels. level I biological safety cabinet, biological safety cabinets and grade II III class biological safety cabinet.
I level security cabinet has a front window opening operation, the operator can operate through the front window opening in the security cabinet operation. Front window opening inward suction operation Security vacuum airflow to protect operators; exhaust air is passed through HEPA filtered exhaust cabinet to protect the environment.
II level security cabinet has a front window opening operation, the operator can operate in a secure cabinet through the front window opening operation, the front window opening inward suction operation Negative pressure air stream to protect the operator's safety; HEPA filtered through downdraft to protect safety cabinet experiment items; airflow The high efficiency filter after discharging the cabinet to protect the environment.
III level security cabinet is closed, does not leak fume hood structure. By connecting people with the cabinet closed glove enforcement action in the security cabinet Make. Downdraft filtered through high efficiency filter into the cabinet to protect the security cabinet test items; exhaust gas flows through the two high efficiency filter Filter and then by incineration or for the protection of the environment through a high efficiency filter.
3.2 Biological factors
biologicalagents All microorganisms and biologically active substances.
3.3 Biohazard
Immediate or potential danger caused by a biological agent.
3.4 Cross-contamination
Accident outside material into the target object.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 47 pages — is available in the English PDF.
Referenced standards
Normative references
Cited by
- GB/T 13477.25-2024Test method for building sealants - Part 25: Determination of resistance to mold
- GB/T 1741-2020Test method for determining the resistance of paints film to mold
- GB/T 37206-2018Test method for antibacterial properties of organic separation membrane
- JY/T 0601-2017Equipment specifications for the major of hydraulic engineering in vocational colleges
- JY/T 0602-2017Equipment specifications for the major of grain and oil storage, transportation and detection technology in vocational schools
- JY/T 0603-2017Equipment specifications for the major of grain and oil storage and detection technology in vocational colleges
Similar standards
Editions of YY 0569
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY 0569-2011 | Class II biological safety cabinets | current edition | Current |
| YY 0569-2005 | Class II biological safety cabinets | previous edition | Superseded |
This page sells the current edition, YY 0569-2011. Earlier editions are listed for reference only.
How to Download YY 0569-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 191-2008 — Packaging - Pictorial marking for handling of goods
YY 0017-2026 — Implants for osteosynthesis - Metallic bone plates
YY 0569-2011
$620.00